Weekly archive
Week of September 9, 2024
The FDA cleared 59 510(k)s, granted 1 De Novo request and approved no original PMAs. 24 device recalls were posted, 2 of them Class I. Every 510(k), De Novo and PMA decision dated September 9 to 15, 2024, and every device recall the FDA posted those days.
- 510(k) clearances
- 59median review 141 days
- De Novo grants
- 1
- Original PMA approvals
- 0plus 41 supplements
- Recalls posted
- 2468 products, 2 Class I
Recalls posted
Device recalls the FDA posted September 9 to 15, 2024, Class I first. Class I is the most serious: a reasonable chance the product causes serious harm or death. The class comes from the FDA’s weekly enforcement report, so the newest recalls can show “Class pending” for a few days.
| Recall and class | Firm and reason | Product code |
|---|---|---|
| Z-2973-2024Class I | Smiths Medical ASD Inc.There is a potential for inadvertent tidal volume knob movement from the original setting when set at high (1000-1500 mL) and low (70-150 mL) settings. | BTL2 products |
| Z-3086-2024Class I | Smiths Medical ASD Inc.There is the potential for a disconnection of the pilot balloon from the tracheostomy inflation line within specific lots of the BLUSelect, BLUgriggs and BLUperc products due to a manufacturing defect. | BTO3 products |
| Z-3138-2024Class II | CareFusion 303, Inc.Due to affected devices that may have bent female IUI connector and may not be operable and potentially cause delays in the start of the infusion or monitoring. | FRN5 products |
| Z-3143-2024Class II | SEER MEDICAL PTY LTDDuring CAPA investigations of sight production inconsistencies, it was highlighted that previous Electromagnetic Compatibility (EMC) assessments related to RF radiated emissions may not be reasonably equivalent to the c… | GWQ1 product |
| Z-3144-2024Class II | B. Braun Medical, Inc.Potential for product leakage from the Spin Lock Connector on the distal end of the device due to the migration of excess solvent during the assembly process. | FPA1 product |
| Z-3145-2024Class II | GE Healthcare GmbHThe National Institute of Standards and Technology (NIST) has identified in the National Vulnerability Database (NVD) that NexGen Healthcare Mirth Connect, prior to version 4.4.1, has a potential vulnerability allowing… | LLZ2 products |
| Z-3147-2024Class II | Boston Scientific CorporationThere is the potential for holes in the clear sterile barrier of the device pouch which may impact device sterility. | FFA5 products |
| Z-3152-2024Class II | Medtronic Inc.Medtronic CareLink 2090 programmers with serial number prefixes PKK0 and PKK1 have the potential for autonomous cursor motion when Finger Touchscreen capability is enabled by software version 3.2 or higher. | NVZ1 product |
| Z-1547-2024Class II | Olympus Corporation of the AmericasDue to increase in complaints for the KD-640L and KD-645L Triangle Tip Electrosurgical Knives breaking off during use -issuing reminder to users to utilize the electrosurgical knife in accordance with the Instructions f… | KNS2 products |
| Z-3125-2024Class II | Hitachi America, Ltd., Particle Therapy DivisionDuring positioning within the system's 3D to 3D matching mode while transferring CBCT images, when the Float Image Selection button is pressed, it may display CBCT images that are insufficient in number. | LHN3 products |
| Z-3128-2024Class II | Medtronic Perfusion SystemsDuring manufacturing, DLP Single Stage Venous Cannulae - straight tip (Model 66118) was incorrectly placed into a product labeled as DLP Single Stage Venous Cannulae - right angle metal tip (Model 67312). | DWF1 product |
| Z-3129-2024Class II | Medical Action Industries, Inc. 306The protective pouches were identified to have been improperly labeled with UDI that is associated with an incorrect GMDN (Global Medical Device Nomenclature) code. These blue foam pouches were misused due to misclassif… | FSM1 product |
| Z-3130-2024Class II | Medtronic Xomed, Inc.Product was distributed past expiration date. | EQJ1 product |
| Z-3131-2024Class II | Technomed EuropeProduct has have presence of brown discoloration or substance on the hinge area of the device | ETN1 product |
| Z-3132-2024Class II | Zimmer GmbHMislabeled: Incorrect sleeve in the packaging. The outer packaging is labeled as a standard neck sleeve, however, the product within the box is a -6mm neck sleeve, and vice versa. The neck of the sleeve can be identifie… | LZO2 products |
| Z-3100-2024Class II | Siemens Healthcare Diagnostics, Inc.Positive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the Asia Pacific sample population. The observed bias is not pr… | NIG5 products |
| Z-3105-2024Class II | MEDLINE INDUSTRIES, LP - NorthfieldDue to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction du… | CAI FDE FTN17 products |
| Z-3122-2024Class II | BioFire Diagnostics, LLCSome automated in vitro diagnostic modules and instruments manufactured or serviced between May 2022 and May 2024 were calibrated with internal equipment that may lead modules and instruments to report higher melting te… | NSU OCC2 products |
| Z-3124-2024Class II | Fresenius Medical Care Holdings, Inc.Machines equipped with BPM3 Blood Pressure Modules (Blood pressure monitor TM-2917B with firmware version S005) will not record Blood Pressure measurements and does not alarm if the cuff tubing is kinked close to the cu… | KDI1 product |
| Z-2975-2024Class II | PACIFIC LASERTECH, LLCThese laser products fail to comply with 21 CFR 1040.10(f)(4) Key control which requires, Each laser system classified as a Class IIIb or IV laser product shall incorporate a key-actuated master control. The key shall b… | RDW3 products |
| Z-3089-2024Class II | Ion Beam Applications S.A.Safety Parameters Verification Mechanisms can be deactivated in clinical runtime | LHN1 product |
| Z-3092-2024Class II | MEDLINE INDUSTRIES, LP - NorthfieldPotential for an incomplete seal on the packaging tray. | NLQ NUJ4 products |
| Z-3096-2024Class II | Smith & Nephew Inc.Product packaging process may result in an improper or incomplete seal of the outer foil pouch surrounding the inner Tyvek pouch, compromising product sterility | GDW2 products |
| Z-3098-2024Class II | Zoe Medical IncorporatedThis product labeling does not adequately provide warning about the risks associated with the use of damaged batteries or batteries past the intended use life. | DQA2 products |
Original PMA approvals
Original premarket approvals for Class III devices. The FDA also decided 41 PMA supplements this week, which change devices already approved.
No original PMAs were approved this week.
510(k) clearances
59 510(k) clearances by review panel. The median took 141 calendar days from receipt to decision; the fastest, K242612, 7 days; the longest, K220999, 892 days.
Calendar days include the time a submission waits on the company, so they run longer than the FDA’s 90-day review goal. How to search the 510(k) database
- Radiology13
- Orthopedic9
- Cardiovascular8
- Dental6
- General, Plastic Surgery5
- Physical Medicine4
- Anesthesiology3
- General Hospital3
- Gastroenterology, Urology2
- Obstetrics/Gynecology2
- Clinical Chemistry1
- Microbiology1
- Neurology1
- Panel not stated1
Every clearance
| Number | Device and applicant | Product code | Decision and review time |
|---|---|---|---|
| K240271 | Linshom Continuous Predictive Respiratory Monitoring System (CPRMS)Linshom Medical, Inc. | BZQAnesthesiology | Sep 11, 2024224 days |
| K240929 | Sleep Apnea Notification Feature (SANF)Apple, Inc. | QZWAnesthesiology | Sep 13, 2024162 days |
| K241954 | SonoBlock; SonoBlock IIPAJUNK GmbH Medizintechnologie | BSPAnesthesiology | Sep 11, 202470 days |
| K231982 | Quantum Perfusion Blood Oxygenator VT75-E1, Quantum Perfusion Blood Oxygenator VT75-E2Qura S.R.L | DTZCardiovascular | Sep 12, 2024435 days |
| K234064 | Medline ReNewal Reprocessed Siemens ACUSON AcuNav 8F Ultrasound Catheter (for Siemens Systems) (10135936); Medline ReNewal Reprocessed Siemens ACUSON AcuNav 8F Ultrasound Catheter (for GE Systems) (10135910)Surgical Instrument Service and Savings, Inc. | OWQCardiovascular | Sep 10, 2024263 days |
| K240087 | RT Carbon ECG LeadsIvy Biomedical Systems, Inc. | DSACardiovascular | Sep 13, 2024246 days |
| K240787 | Selectra 3D Lead Delivery System (443624-443629, 451789-451791); Selectra Slitter Tool (383119); Selectra Accessory Kit (375518)Biotronik, Inc. | DQYCardiovascular | Sep 12, 2024174 days |
| K240880 | VitalFlow Set with Balance BiosurfaceMedtronic, Inc. | QJZCardiovascular | Sep 12, 2024164 days |
| K241156 | Medline ReNewal Reprocessed St. Jude Medical Response Electrophysiology Catheter, Supreme Electrophysiology Catheter (401150, 401206, 401207, 401210, 401211, 401212, 401222, 401223, 401226, 401227, 401228, 401260, 401261, 401271, 401305, 401306, 401309, 401310, 401311, 401312, 401317, 401353, 401357, 401381, 401392, 401399, 401400, 401425, 401430, 401433, 401434, 401435, 401436, 401438, 401441, 401442, 401443, 401444, 401445, 401448, 401449, 401450, 401451, 401453, 401466, 401468, 401Surgical Instrument Service and Savings, Inc. | NLHCardiovascular | Sep 14, 2024141 days |
| K241362 | Pounce XL Thrombectomy System (PTS-1011-7F135)Surmodics,Inc. | QEWCardiovascular | Sep 11, 2024120 days |
| K242612 | Option ELITE Vena Cava Filter System (352506070E 352506100E)Argon Medical Devices, Inc. | DTKCardiovascular | Sep 10, 20247 days |
| K220999 | Hipro Glycosylated Hemoglobin (HbA1c) Test SystemShijiazhuang Hipro Biotechnology Co., Ltd. | PDJClinical Chemistry | Sep 12, 2024892 days |
| K232559 | C&B NFH HybridArum Dentistry Co., Ltd. | EBFDental | Sep 11, 2024385 days |
| K240232 | EK D3.3 and Ultra Wide ImplantsHiossen, Inc. | DZEDental | Sep 13, 2024228 days |
| K240803 | Surgikor Fixation One, Abutment Blanks and AbutmentsSurgikor, LLC | DZEDental | Sep 12, 2024171 days |
| K240977 | DenQ Sub SLA Implant SystemDenq | DZEDental | Sep 13, 2024156 days |
| K241493 | Glidewell™ 3DP Denture Base ResinPrismatik Dentalcraft, Inc. | EBIDental | Sep 12, 2024107 days |
| K241722 | Elos Accurate® Hybrid Base™; Elos Accurate® Customized AbutmentElos Medtech Pinol A/S | NHADental | Sep 13, 202491 days |
| K240544 | EpitomeeEpitomee Medical , Ltd. | QFQGastroenterology, Urology | Sep 13, 2024199 days |
| K241638 | 8mm SureForm 30 Curved-Tip Stapler; 8mm SureForm 30 Stapler; 8mm SureForm 30 ReloadsIntuitive Surgical, Inc. | NAYGastroenterology, Urology | Sep 10, 202495 days |
| K240266 | Latex Examination Glove Powder Free (Ocean Blue and Natural White)Sri Trang Gloves (Thailand) Public Company Limited | LYYGeneral Hospital | Sep 9, 2024222 days |
| K241710 | 3M™ Attest™ Super Rapid Readout Biological Indicator 1492V; 3M™ Attest™ Auto-reader 490; 3M™ Attest™ Auto-reader 490H; 3M™ Attest™ Mini Auto-reader 490M3M Company | FRCGeneral Hospital | Sep 12, 202490 days |
| K241745 | Powder Free Nitrile Examination Glove, Non-sterile, Tested for Use with 46 Chemotherapy Drugs, the Opioid Fentanyl citrate, simulated Gastric Acid, Fentanyl in Gastric Acid, the tranquilizer Xylazine, and Xylazine in Fentanyl citrate (Blue)Ksg Medicare Sdn. Bhd. | LZAGeneral Hospital | Sep 11, 202485 days |
| K233732 | CoolSculpting Elite SystemZeltiq Aesthetics, Inc. (Acquired by Allergan Aesthetics | OOKGeneral, Plastic Surgery | Sep 12, 2024296 days |
| K240188 | Vitruvian LiposaberBlack & Black Surgical, Inc. | QPBGeneral, Plastic Surgery | Sep 13, 2024233 days |
| K241205 | Ice Cooling IPL Hair Removal Device (Y6S GR, Y6S GA,Y6S GT, Y6S JU, Y6S JM, Y6S MU, Y6S OG, Y6S CH,Y6S PN,Y6S NK, Y6S PG, Y6S TG, Y6S OE)Shenzhen Ulike Smart Electronics Co., Ltd. | OHTGeneral, Plastic Surgery | Sep 9, 2024132 days |
| K241214 | ProPel SG3™ Surgical GownStandard Textile Co., Inc. | FYAGeneral, Plastic Surgery | Sep 10, 2024133 days |
| K242349 | FoLixLumenis Be, Ltd. | GEXGeneral, Plastic Surgery | Sep 9, 202432 days |
| K234063 | T2Candida 1.1 PanelT2biosystems, Inc. | PIIMicrobiology | Sep 13, 2024266 days |
| K241678 | TENS & EMS Stimulator (MHD-1083)Shenzhen Minghuangda Electronics Co., Ltd. | NUHNeurology | Sep 9, 202490 days |
| K240176 | V-VITFREEZE and V-VITWARMVitromed GmbH | MQLObstetrics/Gynecology | Sep 13, 2024234 days |
| K241004 | Wearable Breast Pump (ABP-1508PRO, ABP-1506, ABP-1306, ABP-1306 PRO, M1, M3, M8, M5, M6, K1, K2, K3, K5, ABP-1306PLUS, ABP-1308, M2)Huizhou Lvb Maternal and Infant Supplies Co., Ltd. | HGXObstetrics/Gynecology | Sep 11, 2024152 days |
| K240233 | Pitkar Spinal Pedicle Screw SystemS.H.Pitkar Orthotools Pvt. , Ltd. | NKBOrthopedic | Sep 9, 2024224 days |
| K240683 | Rx Knee SystemSignature Orthopaedics Pty, Ltd. | JWHOrthopedic | Sep 11, 2024183 days |
| K241010 | HEALIX ADVANCE Knotless anchors; HEALIX ADVANCE Self-Punching anchorsDepuy Mitek | MAIOrthopedic | Sep 11, 2024152 days |
| K241438 | Signus Tetris™ St; Signus Tetris™ R StSignus Medizintechnik GmbH | MAXOrthopedic | Sep 11, 2024113 days |
| K241609 | Tornier Humeral Reconstruction System (Tornier HRS)Stryker Corporation (Tornier, Inc.) | KWSOrthopedic | Sep 10, 202497 days |
| K241695 | Extremity All Suture SystemField Orthopaedics | MBIOrthopedic | Sep 12, 202492 days |
| K241875 | TheRay Collared and Collarless Femoral StemNextstep Arthropedix | MEHOrthopedic | Sep 13, 202477 days |
| K242042 | TriALTIS™ Spine SystemMedos International SARL | NKBOrthopedic | Sep 9, 202459 days |
| K242401 | EXULT Knee Replacement SystemCorentec Co., Ltd. | JWHOrthopedic | Sep 12, 202430 days |
| K233399 | Microdacyn Wound Care SolutionSonoma Pharmaceuticals, Inc. | FROPanel not stated | Sep 13, 2024345 days |
| K240168 | N7-SNuga Medical Co., Ltd. | JFBPhysical Medicine | Sep 12, 2024234 days |
| K240990 | Mobility ScooterZhejiang Innuovo Rehabilitation Devices Co.,Ltd | INIPhysical Medicine | Sep 10, 2024152 days |
| K242136 | Electric wheelchair (C001)Zhejiang Hfizer Medical Equipment Co., Ltd. | ITIPhysical Medicine | Sep 12, 202452 days |
| K242441 | Revian Lyte (10011)Revian, Inc. | OAPPhysical Medicine | Sep 13, 202428 days |
| K240094 | LumiNE US; LumiAugmedit B.V. | QIHRadiology | Sep 10, 2024242 days |
| K240697 | See-Mode Augmented Reporting Tool, Thyroid (SMART-T)See-Mode Technologies Pte, Ltd. | QDQRadiology | Sep 9, 2024179 days |
| K240791 | ADAS 3DAdas3D Medical S.L. | QIHRadiology | Sep 9, 2024171 days |
| K240942 | CINA-CSpineAvicenna.Ai | QASRadiology | Sep 12, 2024160 days |
| K241161 | Ultrasound Imaging System (LK128L, LK128LH, LK128C, LK128M, LK128PA, LK128E, LU700C, LU700L, LU710L, LU710LH, LU710C, LU710M, LU710PA); Ultrasound ImagingLeltek, Inc. | IYNRadiology | Sep 11, 2024138 days |
| K241249 | Orthopantomograph™ OP 3D LX (PAN 3D, PAN CEPH 3D)Palodex Group OY | OASRadiology | Sep 12, 2024132 days |
| K241280 | Constellation (CON-001)Q Bio, Inc. | QIHRadiology | Sep 12, 2024128 days |
| K241380 | FETOLY-HEARTDiagnoly | IYNRadiology | Sep 11, 2024119 days |
| K241432 | MX7/MX7T/Vaus7/Zeus/ME7/Anesus ME7/Anesus ME7T/MX7P/MX7W/MX8/MX8T/Vaus8/ME8 Diagnostic Ultrasound SystemShenzhen Mindray Bio-Medical Electronics Co., Ltd. | IYNRadiology | Sep 12, 2024114 days |
| K241524 | Elucid PlaqueIQElucid Bioimaging, Inc. | LLZRadiology | Sep 12, 2024106 days |
| K241582 | Aplio i900/i800/i700 Diagnostic Ultrasound System, Software V7.0 (TUS-AI900, TUS-AI800, TUS-AI700)Canon Medical Systems Corporation | IYNRadiology | Sep 12, 2024101 days |
| K241681 | Overjet Image Enhancement AssistOverjet, Inc. | QIHRadiology | Sep 9, 202490 days |
| K242394 | Digital Flat Panel X-ray Detector (1717WCE, 1717WCE-HR, 1717WCE-HS, 1717WCE-GF)Rayence Co., Ltd. | MQBRadiology | Sep 9, 202427 days |
Only see the weeks that matter to you
DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and every day it checks openFDA and tells you about new clearances, recalls and MAUDE reports for them. It also puts all of this beside the FDA’s own record pages.
Only your product codes, every Monday
One short email after each openFDA update with just the decisions and recalls in up to three product codes you choose. Free.
Earlier weeks
Each week keeps its own page.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Decisions arrive in weekly updates; check the FDA record before you rely on a figure.
- device/510k · search decision_date:[20240909 TO 20240915]Retrieved Nov 29, 2026, 11:23 PM UTC. FDA data updated Nov 16, 2026.
- device/pma · search decision_date:[20240909 TO 20240915]Retrieved Nov 29, 2026, 11:23 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search event_date_posted:[20240909 TO 20240915]Retrieved Nov 29, 2026, 11:23 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-3138-2024" "Z-3139-2024" "Z-3140-2024" "Z-3141-2024" "Z-3142-2024" "Z-3143-2024" "Z-3144-2024" "Z-3145-2024" "Z-3146-2024" "Z-3147-2024" "Z-3148-2024" "Z-3149-2024" "Z-3150-2024" "Z-3151-2024" "Z-3152-2024" "Z-1547-2024" "Z-1548-2024" "Z-3125-2024" "Z-3126-2024" "Z-3127-2024" "Z-3128-2024" "Z-3129-2024" "Z-3130-2024" "Z-3131-2024" "Z-3132-2024" "Z-3133-2024" "Z-2973-2024" "Z-2974-2024" "Z-3086-2024" "Z-3087-2024" "Z-3088-2024" "Z-3100-2024" "Z-3101-2024" "Z-3102-2024" "Z-3103-2024" "Z-3104-2024" "Z-3105-2024" "Z-3106-2024" "Z-3107-2024" "Z-3108-2024" "Z-3109-2024" "Z-3110-2024" "Z-3111-2024" "Z-3112-2024" "Z-3113-2024" "Z-3114-2024" "Z-3115-2024" "Z-3116-2024" "Z-3117-2024" "Z-3118-2024")Retrieved Nov 29, 2026, 11:23 PM UTC. FDA data updated Nov 18, 2026.
- device/enforcement · search recall_number:("Z-3119-2024" "Z-3120-2024" "Z-3121-2024" "Z-3122-2024" "Z-3123-2024" "Z-3124-2024" "Z-2975-2024" "Z-2976-2024" "Z-2977-2024" "Z-3089-2024" "Z-3092-2024" "Z-3093-2024" "Z-3094-2024" "Z-3095-2024" "Z-3096-2024" "Z-3097-2024" "Z-3098-2024" "Z-3099-2024")Retrieved Nov 29, 2026, 11:23 PM UTC. FDA data updated Nov 18, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.