FDA device news

Weekly archive

Week of August 12, 2024

The FDA cleared 60 510(k)s, granted 2 De Novo requests and approved no original PMAs. 22 device recalls were posted, 1 of them Class I. Every 510(k), De Novo and PMA decision dated August 12 to 18, 2024, and every device recall the FDA posted those days.

510(k) clearances
60median review 92 days
De Novo grants
2
Original PMA approvals
0plus 39 supplements
Recalls posted
2288 products, 1 Class I

Recalls posted

Device recalls the FDA posted August 12 to 18, 2024, Class I first. Class I is the most serious: a reasonable chance the product causes serious harm or death. The class comes from the FDA’s weekly enforcement report, so the newest recalls can show “Class pending” for a few days.

Recall and classFirm and reasonProduct code
Z-2543-2024Class I Defibtech, LLCProblem in the device's motor may cause the device to stop compressions. This may lead to a delay in therapy if the device fails to operate, and could lead to patient injury or death. DRM1 product
Z-2581-2024Class II Baxter Healthcare CorporationThe Spectrum WBM may fail to auto-document infusion status information back to the hospital's Electronic Medical Record (EMR) system. This software issue only impacts customers where the Spectrum V8 or Spectrum IQ infus… FRN2 products
Z-2638-2024Class II Argon Medical Devices, IncPeripherally Inserted Central Catheter (PICC) is labeled 26G/1.9Fr x 30cm but package contains incorrect 26G/1.9Fr x 50cm PICC, which could lead to procedural delay. LJS1 product
Z-2639-2024Class II Microtek Medical Inc.During sterile barrier testing performed on the decanter product line, the samples were identified to have pin holes and tears in the packaging film. KPE5 products
Z-2644-2024Class II Ventana Medical Systems, Inc.Their is a potential for high background and off-target staining when using the monoclonal primary antibody which can result in false positive results. NJT1 product
Z-2645-2024Class II Alcon Research LLCAlcon has detected an increase in complaint reports related to sharpness for specific ClearCut Dual Bevel Sideport and A-OK Comeal/Scleral V-Lance Ophthalmic knives. HNN HQC LRO36 products
Z-2681-2024Class II Abbott Laboratories1)If Complete Blood Count with differential and reticulocyte(CBC+Diff+Retic) sample is run, then immediately following on same rack CBC+Diff sample is run, then falsely low red blood cell count may occur, generating fal… GKZ1 product
Z-2682-2024Class II BioFire Diagnostics, LLCPneumonia panels have an increased risk of control failures due to an issue in the manufacturing process, which could lead to delayed results for the customer. QDS2 products
Z-2684-2024Class II Ortho-Clinical Diagnostics, Inc.The reason for this recall is QuidelOrtho has determined that during normal use the VITROS XT ALB-TP Slides may shed dust and debris as they travel through the MicroSlide Subsystem of the VITROS XT 3400 or VITROS XT 760… CIX1 product
Z-2685-2024Class II ROi CPS LLCDecanters may have pin holes and tears in sterile packaging that cannot be identified visually. KPE1 product
Z-2686-2024Class II Becton Dickinson & Co.May intermittently produce falsely resistant results for pyrazinamide (PZA) during susceptibility testing of Mycobacterium tuberculosis isolates. MJA1 product
Z-2584-2024Class II CepheidSpecimen collection device may leak after the patient sample swabs have been inserted into the test tubes, which could result in biohazard exposure, cross contamination causing false positive results for other specimens… JSM1 product
Z-2585-2024Class II Becton Dickinson & Co.There is a potential risk that the image colony count does not adequately reflect the actual plate colony count and may lead to inaccurate counts that would impact patient diagnosis and subsequent care. KZB1 product
Z-2586-2024Class II EYE COMFORT CARE LLCDue to a change in the safety clips for the heated eye masks, the wires to the safety clips may detach as a result of tugging and pulling and may cause a short in the heating element of the mask and could potentially ca… IRT1 product
Z-2616-2024Class II Biointellisense Inc.Software anomaly causing the skin temperature measurement accuracy to be slightly out of specification. DRG1 product
Z-2617-2024Class II Boston Scientific Neuromodulation CorporationSpinal Cord Stimulation (SCS) Implantable Pulse Generator (IPG) may experience routine system check during IPG charging, which may cause device reset, which may lead to transient loss of stimulation; patients may experi… LGW1 product
Z-2618-2024Class II DKK Dai-Ichi Shomei Co., Ltd.The weld on the light head frame can crack which could lead to problems such as the light head frame sagging, the plastic front face housing cracking, or the light head detaching. FSY4 products
Z-2583-2024Class II Thoratec LLCLeft ventricular assist system controller UI membrane/screen may lift along the edge of the controller housing, and if controller is exposed to fluid, ingress may occur, which may damage printed circuit board assemblies… DSQ1 product
Z-2413-2024Class II Palodex Group OyIt was noticed in production on 28.10.2021, that the final testing database gives a pass result for Pan ADC repeatability test even when the test result is not within the set acceptance criteria (repeatability within 4%… OAS1 product
Z-2557-2024Class II Merit Medical Systems, Inc.Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health. DYB JOL NWU23 products
Z-2580-2024Class II Ortho-Clinical Diagnostics, Inc.It was confirmed that fibrinogen in patient plasma samples precipitates out of solution upon the addition of folate stabilizer reagent as part of the pre-treatment process causing the increase in "TM5-4MB" condition cod… CGN1 product
Z-2622-2024Class III DRG International, Inc.Invalid Controls. The Low Control is out of range; too high which causes invalid runs that would result in repeating the runs and delays for reporting results to patients. CDZ1 product

De Novo grants

New device types the FDA classified through the De Novo pathway. Each one creates a product code and becomes a possible predicate for later 510(k)s.

NumberDevice and applicantProduct codeDecision and review time
DEN230090 First To Know Syphilis TestNowdiagnostics SBZ Aug 16, 2024233 days
DEN210037 SoClean 3+ Bacterial Reduction DeviceSoclean, Inc. QXQ Aug 12, 20241,062 days

Original PMA approvals

Original premarket approvals for Class III devices. The FDA also decided 39 PMA supplements this week, which change devices already approved.

No original PMAs were approved this week.

510(k) clearances

60 510(k) clearances by review panel. The median took 92 calendar days from receipt to decision; the fastest, K242239, 17 days; the longest, K222126, 760 days.

Calendar days include the time a submission waits on the company, so they run longer than the FDA’s 90-day review goal. How to search the 510(k) database

  • Radiology9
  • General, Plastic Surgery7
  • Cardiovascular6
  • Dental5
  • Gastroenterology, Urology5
  • Obstetrics/Gynecology5
  • Orthopedic5
  • Anesthesiology3
  • Ophthalmic3
  • Clinical Chemistry2
  • Clinical Toxicology2
  • Neurology2
  • Physical Medicine2
  • General Hospital1
  • Hematology1
  • Microbiology1
  • Panel not stated1

Every clearance

NumberDevice and applicantProduct codeDecision and review time
K232836 Family of DIMAR DMAX NIV masksDimar S.P.A. CBKAnesthesiology Aug 14, 2024335 days
K240807 VM-2000Ventis Medical, Inc. CBKAnesthesiology Aug 13, 2024141 days
K241236 XTRA Collection sets; XTRA Sequestration set XSorin Group Italia S.R.L. CACAnesthesiology Aug 16, 2024106 days
K231979 Mecun SpO2 sensorShenzhen Mecun Medical Supply Co., Ltd. DQACardiovascular Aug 16, 2024408 days
K233499 Sapphire NC ULTRA Coronary Dilatation CatheterOrbusNeich Medical (Shenzhen) Co., Ltd. LOXCardiovascular Aug 16, 2024290 days
K233505 Sapphire ULTRA Coronary Dilatation CatheterOrbusNeich Medical (Shenzhen) Co., Ltd. LOXCardiovascular Aug 15, 2024289 days
K233755 HeartKey® RhythmB-Secur Limited DQKCardiovascular Aug 16, 2024268 days
K240957 Catapult Guide SheathContract Medical International, GmbH DYBCardiovascular Aug 14, 2024128 days
K241510 CROSSLEAD TrackerAsahi Intecc Co., Ltd. DQXCardiovascular Aug 14, 202477 days
K222126 VivaChek™ Fad Blood Glucose Monitoring System, VivaChek™ Fad Smart Blood Glucose Monitoring System, VivaChek™ Fad Sync Blood Glucose Monitoring SystemVivachek Biotech (Hangzhou) Co., Ltd. NBWClinical Chemistry Aug 16, 2024760 days
K241978 Shinetell Digital Pregnancy TestHangzhou AllTest Biotech Co., Ltd. LCXClinical Chemistry Aug 13, 202439 days
K241969 Hightop® Home Use Fentanyl/Norfentanyl Urine Rapid Test Panel; Hightop® Fentanyl/Norfentanyl Urine Rapid Test PanelQingdao Hightop Biotech Co., Ltd. NGLClinical Toxicology Aug 14, 202440 days
K242077 PocguideTM Multi-Drug Test Cup OTC, PocguideTM Multi-Drug Test CupAicheck Biotech, Inc. NFTClinical Toxicology Aug 14, 202429 days
K233896 JDentalCare Dental Implant System: JDEvolution; JDEvolution plus; JDIcon; JDIcon plusJdentalcare Srl DZEDental Aug 16, 2024249 days
K240133 Xenograft Bovine Bone ParticulateCollagen Solutions, LLC NPMDental Aug 16, 2024211 days
K240573 iTero Lumina™ ProAlign Technology , Ltd. NTKDental Aug 16, 2024168 days
K240837 TOV Dental Implant SystemRdj Tov Implant, Ltd. DZEDental Aug 15, 2024141 days
K241314 KLS Martin Oral-Max Implants MR Conditional (bundled)KLS-Martin L.P. JEYDental Aug 16, 202498 days
K233700 Avicenna RoboflexElmed Elektronik VE Medikal Sanayi VE Ticaret Anonim Sirketi FGBGastroenterology, Urology Aug 16, 2024273 days
K241401 Connected OR Hub with Device and Voice Control; SDC4K Information Management System with Device and Voice ControlStryker Endoscopy GCJGastroenterology, Urology Aug 15, 202490 days
K241402 Medical Monitor (32HR734S)Lg Electronics.Inc GCJGastroenterology, Urology Aug 12, 202487 days
K242061 Reusable Interface Cable 1.5 m (PRD-IFC-002)Creo Medical, Ltd. KNSGastroenterology, Urology Aug 15, 202431 days
K242108 Ambu® aScope™ 5 Uretero (Standard Deflection, Reverse Deflection); Ambu® aView™ 2 AdvanceAmbu A/S FGBGastroenterology, Urology Aug 16, 202429 days
K240961 Disposable Pen Injector AssemblyWuxi Nest Biotechnology Co., Ltd. FMFGeneral Hospital Aug 15, 2024129 days
K240868 Ultravision2™ IonPencilAlesi Surgical Limited PQMGeneral, Plastic Surgery Aug 12, 2024136 days
K241522 NPseal (20 and 25)Guard Medical, Inc. OKOGeneral, Plastic Surgery Aug 15, 202478 days
K241547 Diode Laser Therapy Device (DF-DIODE LASER-S1)Weifang Dragonfly Electronics Technology Co., Ltd. GEXGeneral, Plastic Surgery Aug 13, 202474 days
K241656 Eneka SelectifTermosalud GEXGeneral, Plastic Surgery Aug 12, 202463 days
K241699 SUPER VELOCEIlooda Co,., Ltd. GEXGeneral, Plastic Surgery Aug 13, 202461 days
K241750 Medifun Safety Lancet ( MSL1 series)Medifun Corporation FMKGeneral, Plastic Surgery Aug 15, 202458 days
K241810 MICROLET®NEXT LancetAscensia Diabetes Care U.S., Inc. QRKGeneral, Plastic Surgery Aug 15, 202455 days
K242127 ACL TOP Family 50 Series (ACL TOP 750; ACL TOP 750 CTS; ACL TOP 750 LAS; ACL TOP 550 CTS; ACL TOP 350 CTS)Instrumentation Laboratory (IL) Co. GKPHematology Aug 16, 202428 days
K233367 iDart Lyme IgG ImmunoBlot KitId-Fish Technology, Inc. LSRMicrobiology Aug 12, 2024315 days
K241365 Pixie Pulse (SM9068)Pixie Cup, LLC NUHNeurology Aug 12, 202490 days
K242057 COOLIEF* Radiofrequency Generator (CRG-BASIC, CRG-BASIC-R, CRG-ADVANCED, CRG-ADVANCED-R)Avanos Medical, Inc. GXDNeurology Aug 14, 202430 days
K240081 Human LubricantJiangxi Wulinxiang Biological Technology Co., Ltd. NUCObstetrics/Gynecology Aug 16, 2024218 days
K240307 VitaVitro Embryo Transfer Catheter (Models ET-S, ET-SI, ET-A, ET-I, ET-IH)Shenzhen Vitavitro Biotech Co., Ltd. MQFObstetrics/Gynecology Aug 16, 2024196 days
K240595 Utepreva Endometrial Sampler (UP01)Utepreva, LLC HFEObstetrics/Gynecology Aug 14, 2024163 days
K240978 SeleneView® Single-Use Digital Hysteroscope (HC29, HC29-G, HC35, HC35-G, HO35, HO35-G, HO42, HO42-G)MacroLux Medical Technology Co., Ltd. HIHObstetrics/Gynecology Aug 16, 2024128 days
K241748 Intimate Rose Vaginal DilatorsPlus EV Holdings Dba Intimate Rose HDXObstetrics/Gynecology Aug 14, 202457 days
K223700 BOSS™Intelon Optics, Inc. MXKOphthalmic Aug 16, 2024613 days
K241571 BostonSight® Specialty LensesBoston Foundation For Sight, Inc. D/B/A Bostonsight HQDOphthalmic Aug 14, 202475 days
K242059 Flexlens Silicone Hydrogel Daily Wear Soft (Hydrophilic) Contact Lens (efrofilcon A)X-Cel Specialty Contacts, A Walman Optical Company LPLOphthalmic Aug 14, 202430 days
K241328 aprevo® Anterior and Lateral Lumbar Interbody Fusion device (ALIF/LLIF); aprevo® Transforaminal Lumbar Interbody Fusion device (TLIF); aprevo® Anterior Lumbar Interbody Fusion device with Interfixation (ALIF-X)Carlsmed, Inc. MAXOrthopedic Aug 12, 202494 days
K241387 CoLink NeoFuse MIS Plating System (see attached parts list)In2bones USA HRSOrthopedic Aug 13, 202490 days
K241754 NCB® Polyaxial Locking Plate System; NCB® Periprosthetic Femur Plate System; NCB® Cable Button; NCB® Straight Narrow Shaft PlateZimmer Switzerland Manufacturing GmbH HRSOrthopedic Aug 15, 202458 days
K241813 SI-TECHNOLOGY® SI-DESIS® X™ Sacroiliac Joint Fusion SystemSi-Technology, LLC OUROrthopedic Aug 13, 202453 days
K241846 E3D™-C Interbody SystemEvolution Spine OVEOrthopedic Aug 12, 202447 days
K240713 TRAUMAGEL®Cresilon, Inc. QSYPanel not stated Aug 14, 2024152 days
K240234 BTL-899MSBTL Industries, Inc. IPFPhysical Medicine Aug 17, 2024201 days
K241686 Power wheelchair (N6101)Zhejiang Zhonglei Medical Technology Co. , Ltd. ITIPhysical Medicine Aug 13, 202462 days
K233965 UltraDrape UGPIV Barrier and Securement (34-15)Parker Laboratories, Inc. ITXRadiology Aug 15, 2024244 days
K240740 qCT LN QuantQure.Ai Technologies QIHRadiology Aug 16, 2024151 days
K241094 BullsaiTuring Medical Technologies, Inc. LLZRadiology Aug 13, 2024113 days
K241211 CoLumboSmart Soft Healthcare QIHRadiology Aug 15, 2024107 days
K241429 ECHELON Synergy MRI SystemFujifilm Healthcare Corporation LNHRadiology Aug 13, 202484 days
K241596 uMI PanvivoShanghai United Imaging Healthcare Co., Ltd. KPSRadiology Aug 15, 202473 days
K242054 OptimMRI (v2)Rebrain, Sas QIHRadiology Aug 12, 202428 days
K242096 dS FootAnkle Coil 8ch 1.5TPhilips Medical Systems Nederland B.V. MOSRadiology Aug 12, 202425 days
K242239 HF-OCT Imaging System with Vis-RX Micro-Imaging Catheter (G10 (HF-OCT Console), G30 (Vis-Rx Micro-Imaging Catheter))Gentuity, LLC NQQRadiology Aug 16, 202417 days

Only see the weeks that matter to you

DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and every day it checks openFDA and tells you about new clearances, recalls and MAUDE reports for them. It also puts all of this beside the FDA’s own record pages.

Add to Chrome — free Add to Edge Or follow this page by RSS

Only your product codes, every Monday

One short email after each openFDA update with just the decisions and recalls in up to three product codes you choose. Free.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Decisions arrive in weekly updates; check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.