FDA product code
NIG: System, Test, Carbohydrate Antigen (Ca19-9), For Monitoring And Management Of Pancreatic Cancer (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 866.6010
- Review panel
- Immunology
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 121 in the last 12 months
- Median review time
- 87 daysthe latest 12 decisions
- Recall records
- 190 in the last 12 months
- Adverse event reports
- 102last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code NIG per year, by decision date, 12 since 2002. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 1 |
| 2020 | 1 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 1 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 1 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K253528 | Atellica IM CA 19-9 IISiemens Healthcare Diagnostics, Inc. | Aug 10, 2026270 days |
| K231525 | VITROS Immunodiagnostic Products CA 19-9TM Reagent PackOrtho Clinical Diagnostics | Aug 9, 202375 days |
| K200997 | Lumipulse G CA19-9-NFujirebio Diagnostics,Inc. | May 14, 202028 days |
| K191973 | Lumpipulse G CA19-9-NFujirebio Dianostics, Inc. | Oct 22, 201990 days |
| K100375 | DIMENSION VISTA LOCI CA 19-9 FLEX REAGENT CARTRIDGE, AND DIMENSION LOCI7 CALIBRATOR WITH MODEL(S) K6457, AND KC604Siemens Healthcare Diagnostics, Inc. | Apr 6, 2011418 days |
| K052889 | VITROS CA 19-9Fujirebio Diagnostics,Inc. | Dec 20, 200568 days |
| K052000 | ARCHITECT CA 19-9XR ASSAYFujirebio Diagnostics,Inc. | Oct 25, 200592 days |
| K033038 | GI MONITOR AND GI MONITOR CALIBRATORS ON THE ACCESS IMMUNOASSAY SYSTEMS, MODEL 387687, 387688Beckman Coulter, Inc. | Dec 29, 200391 days |
| K031393 | BAYER ADVIA CENTAUR CA 19-9 ASSAYBayer Healthcare, LLC | Jun 24, 200353 days |
| K023239 | AIA-PACK CA 19-9Tosoh Corp. | Dec 23, 200287 days |
Recalls
Recall records for product code NIG per year, by the date the recall began. Each recalled product is one record; these 19 records belong to 14 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 1 |
| 2018 | 1 |
| 2019 | 1 |
| 2020 | 1 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 5 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-3100-2024Class II | Siemens Healthcare Diagnostics, Inc.Positive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the… | Jul 26, 2024 |
| Z-3101-2024Class II | Siemens Healthcare Diagnostics, Inc.Positive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the… | Jul 26, 2024 |
| Z-3102-2024Class II | Siemens Healthcare Diagnostics, Inc.Positive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the… | Jul 26, 2024 |
| Z-3103-2024Class II | Siemens Healthcare Diagnostics, Inc.Positive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the… | Jul 26, 2024 |
| Z-3104-2024Class II | Siemens Healthcare Diagnostics, Inc.Positive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the… | Jul 26, 2024 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code NIG. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 98 |
| 2018 | 58 |
| 2019 | 135 |
| 2020 | 93 |
| 2021 | 69 |
| 2022 | 83 |
| 2023 | 129 |
| 2024 | 121 |
| 2025 | 100 |
| 2026 (so far) | 93 |
Last 12 months by event type
- Deaths0
- Injuries1
- Malfunctions101
- Other or not stated0
Most reported device problems, last 12 months
- High Test Results84
- Non Reproducible Results28
- Low Test Results17
- Imprecise or Erratic Results8
- Incorrect, Inadequate or Imprecise Result or Readings4
- False Positive Result2
In plain English: what the newest reports describe Written by AI
Several reports describe falsely high CA 19-9 results (a tumor marker blood test used to monitor pancreatic cancer) from the manufacturer's analyzer, where a repeat test or a test on another analyzer gave a much lower value, or a value that matched the patient's history. In the more extreme cases the first result was above 1200 U/mL, while the repeat was far lower.
8 reports: 3002809144-2026-00258, 3002809144-2026-00257, 3002809144-2026-00256, 3002809144-2026-00247, 3002809144-2026-00242, 3002809144-2026-00195, 3002809144-2026-00190, 3002809144-2026-00219
Some reports describe a first result that was far lower than the repeats or the patient's prior results, and the repeat matched the patient's history. In these reports the first result was reported as the questioned value and the repeat was judged correct.
6 reports: 1823260-2026-04406, 1823260-2026-03948, 1823260-2026-02882, 1823260-2026-02812, 1823260-2026-02793, 3002809144-2026-00218
Several reports describe the test giving different results when the same sample was run again or diluted. The reports show the first result, the repeat and the diluted results often disagreed by large amounts. In one report, a 1:5 dilution gave a value of 58000 U/mL, far above the repeat values of about 13000 U/mL from a reference lab.
4 reports: 3002809144-2026-00190, 3002809144-2026-00195, 3002809144-2026-00191, 3002809144-2026-00199
One report on a different laboratory system describes a series of CA 19-9 results where the first result was flagged, the sample was re-centrifuged and retested, and the flagged results were still low. The lab stated the result did not match the patient's clinical diagnosis.
Report: 1823260-2026-02952
In one report, a sample tested on a cobas e 801 gave a first result of 33.9 kU/L that the lab repeated five more times with similar results. The lab did not release the first result because it did not match the patient's clinical diagnosis.
Report: 1823260-2026-04314
Written by AI from the narratives of the 20 newest reports that have one, received Jul 10, 2026 to Oct 23, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 30in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 35registered establishments that list this product code
Other codes under 21 CFR 866.6010
Product codes classified under the same regulation.
Watch NIG in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get NIG every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in NIG. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:NIGRetrieved Nov 28, 2026, 11:37 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:NIGRetrieved Nov 29, 2026, 7:29 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:NIG · count decision_dateRetrieved Nov 28, 2026, 11:37 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:NIGRetrieved Nov 29, 2026, 7:29 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:NIG · count event_date_initiatedRetrieved Nov 28, 2026, 11:37 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:NIG · count res_event_numberRetrieved Nov 28, 2026, 11:37 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-3100-2024" "Z-3101-2024" "Z-3102-2024" "Z-3103-2024" "Z-3104-2024")Retrieved Nov 29, 2026, 7:29 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:NIG AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 7:29 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:NIG AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 7:29 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:NIG AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 7:29 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:NIGRetrieved Nov 28, 2026, 11:37 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:NIGRetrieved Nov 28, 2026, 11:37 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:866.6010Retrieved Nov 29, 2026, 7:53 AM UTC. FDA data updated Nov 16, 2026.