Product codes

FDA product code

OWQ: Reprocessed Intravascular Ultrasound Catheter (FDA product code)

Follow OWQ: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 870.1200
Review panel
Cardiovascular
Premarket route
510(k)
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
311 in the last 12 months
Median review time
146 daysthe latest 25 decisions
Recall records
103 in the last 12 months
Adverse event reports
107last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code OWQ per year, by decision date, 31 since 2002. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20172
20181
20193
20201
20210
20222
20238
20241
20253
2026 (so far)1

Latest decisions

NumberDevice and applicantDecision and review time
K254149 Reprocessed AcuNav CRYSTAL™ Ultrasound Catheter 9F x 90 CM, For Use on Siemens System (R11510825); Reprocessed AcuNav CRYSTAL™ Ultrasound Catheter 9F x 90 CM, For Use on GE System (R11657484); Reprocessed SOUNDSTAR CRYSTAL™ Ultrasound Catheter 9F x 90 CM, For Use on Siemens System (R11510826); Reprocessed SOUNDSTAR CRYSTAL™ Ultrasound Catheter 9F x 90 CM, For Use on GE System (R11657485)Sterilmed, Inc. Jun 29, 2026189 days
K250545 Medline ReNewal Reprocessed Biosense Webster SOUNDSTAR eco Diagnostic Ultrasound Catheter (8F, 8FG, 10F, 10FG)Surgical Instrument Service and Savings, Inc. Jun 5, 2025100 days
K250592 NES Reprocessed Visions PV .018 Digital IVUS Catheter (R-86700)Northeast Scientific, Inc. May 12, 202574 days
K243101 Medline ReNewal Reprocessed Siemens ACUSON AcuNav 10F Ultrasound Catheter (for Siemens Systems) (08255790 ); Medline ReNewal Reprocessed Siemens ACUSON AcuNav 10F Ultrasound Catheter (for GE Systems) (10043342)Surgical Instrument Service and Savings, Inc. Mar 10, 2025161 days
K234064 Medline ReNewal Reprocessed Siemens ACUSON AcuNav 8F Ultrasound Catheter (for Siemens Systems) (10135936); Medline ReNewal Reprocessed Siemens ACUSON AcuNav 8F Ultrasound Catheter (for GE Systems) (10135910)Surgical Instrument Service and Savings, Inc. Sep 10, 2024263 days
K232130 Medline ReNewal Reprocessed ViewFlex Xtra ICE CatheterSurgical Instrument Services and Savings, Inc. Dec 15, 2023151 days
K232584 Vein360 Reprocessed Visions PV.014P Rx Digital IVUS Catheter, Vein360 Reprocessed Eagle Eye Platinum Rx Digital IVUS Catheter, Vein360 Reprocessed Eagle Eye Platinum ST Rx Digital IVUS CatheterVein 360, LLC Oct 24, 202360 days
K230928 Vein360 Reprocessed Visions PV.018 Digital IVUS CatheterVein 360, LLC Aug 25, 2023144 days
K231621 Reprocessed ViewFlex Xtra ICE Catheter (D087031)Stryker Sustainability Solutions Jul 25, 202353 days
K231588 ViewFlex™ Eco Reprocessed ICE CatheterABBOTT MEDICAL Jun 30, 202329 days

Recalls

Recall records for product code OWQ per year, by the date the recall began. Each recalled product is one record; these 10 records belong to 6 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20170
20180
20190
20201
20211
20220
20230
20240
20257
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-1323-2026Class I Medline Industries, LPThese lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, s… Dec 22, 2025
Z-1324-2026Class I Medline Industries, LPThese lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, s… Dec 22, 2025
Z-1325-2026Class I Medline Industries, LPThese lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, s… Dec 22, 2025
Z-0573-2026Class II Sterilmed, Inc.Affected catheters were reprocessed beyond validated cycles. This may compromise device sterility or mechanical integrity of the device and can lead to severe… Oct 8, 2025
Z-0574-2026Class II Sterilmed, Inc.Affected catheters were reprocessed beyond validated cycles. This may compromise device sterility or mechanical integrity of the device and can lead to severe… Oct 8, 2025

Adverse event reports (MAUDE)

Medical device reports received per year that name product code OWQ. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
20172
20182
20199
202029
202138
202224
202329
202437
202542
2026 (so far)103

Last 12 months by event type

  • Deaths0
  • Injuries4
  • Malfunctions103
  • Other or not stated0

Most reported device problems, last 12 months

  1. Electrical /Electronic Property Problem17
  2. Poor Quality Image17
  3. Communication or Transmission Problem9
  4. Image Display Error/Artifact8
  5. Material Twisted/Bent8
  6. Mechanical Problem8

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
101in the FDA’s Global Unique Device Identification Database
Establishment listings
70registered establishments that list this product code

Watch OWQ in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get OWQ every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in OWQ. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.