FDA product code
OHT: Light Based Over-The-Counter Hair Removal (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 878.4810
- Review panel
- General and Plastic Surgery
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 16915 in the last 12 months
- Median review time
- 84 daysthe latest 25 decisions
- Recall records
- 00 in the last 12 months
- Adverse event reports
- 14last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code OHT per year, by decision date, 169 since 2006. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 6 |
| 2018 | 8 |
| 2019 | 7 |
| 2020 | 1 |
| 2021 | 12 |
| 2022 | 25 |
| 2023 | 32 |
| 2024 | 26 |
| 2025 | 25 |
| 2026 (so far) | 13 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K261296 | IPL Hair Removal Device (Model: T32A)Shenzhen Mlay Intelligent Technology Co., Ltd. | Nov 12, 2026206 days |
| K261918 | IPL Hair Removal Device (IN-HR001, IN-HR005L)ShenZhen Qiyue Trading Co., Ltd. | Nov 3, 2026147 days |
| K262352 | IPL Hair Removal Device (AHR01, R3C16-G Pro1, AHR02, AHR03, R4820, R6715, L3B02, JR20, R6312, R6202)Shenzhen Jianchao Intelligent Technology Co., Ltd. | Oct 9, 202691 days |
| K261975 | Braun Aura 7 (6041)CyDen Limited | Sep 30, 2026110 days |
| K261977 | Braun Aura 9 (6042)Cyden Limited | Sep 25, 2026105 days |
| K262524 | Sapphire Cooling IPL Hair Removal Device (Ice 2.0, Ice 2.1, Ice 2.2, Ice 2.0B, Ice 2.1B, Ice 2.2B)ONE BEAUTY Technology Co., Ltd. | Sep 18, 202658 days |
| K261705 | Hair Removal Device (R3C16-G Pro, R3C16-V Pro)Shenzhen Jianchao Intelligent Technology Co., Ltd. | Aug 10, 202680 days |
| K261325 | Ice Cooling IPL Hair Removal Device (UI06S PR, UI06S PN, UI06S WH, UI06S PRU, UI06S PNU, UI06S WHU)Shenzhen Ulike Smart Electronics Co., Ltd. | May 21, 202629 days |
| K260594 | IPL Hair Removal Device (FL-B505AG, FL-B505AP, FL-B505AM, FL-B507AG, FL-B507AP, FL-B507AM, FL-B508AG, FL-B508AP, FL-B508AM, FL-B509AG, FL-B509AP, FL-B509AM, FL-B510G, FL-B510P, FL-B510M, FL-B511G, FL-B511P, FL-B511M, FL-B512AG, FL-B512AP, FL-B512AM, FL-B513AG, FL-B513AP, FL-B513AM)Shenzhen Xiazhifeng Electronic Co., Ltd. | May 20, 202686 days |
| K260742 | Ice Cooling IPL Hair Removal Device (UI20S DB, UI20S RE, UI20S PW, UI20S GP, UI20S GR, UI20S BK)Shenzhen Ulike Smart Electronics Co., Ltd. | May 5, 202660 days |
Recalls
Recall records for product code OHT per year, by the date the recall began. Each recalled product is one record.
No recalls on record for this product code.
Adverse event reports (MAUDE)
Medical device reports received per year that name product code OHT. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 2 |
| 2020 | 3 |
| 2021 | 1 |
| 2022 | 2 |
| 2023 | 6 |
| 2024 | 0 |
| 2025 | 6 |
| 2026 (so far) | 14 |
Last 12 months by event type
- Deaths0
- Injuries8
- Malfunctions3
- Other or not stated3
Most reported device problems, last 12 months
- Overheating of Device4
- Patient Device Interaction Problem3
- Energy Output Problem1
- Explosion1
- Fire1
- Incorrect, Inadequate or Imprecise Result or Readings1
In plain English: what the newest reports describe Written by AI
Several reports describe the device overheating during use, with the glass or treatment window shattering, smoking, or catching fire, and causing burns to the user's leg. One report says the burns left permanent scarring and that a fire extinguisher was needed. The manufacturer's testing, as described in the reports, found the device exceeded temperature limits and failed electrical insulation and breakdown tests.
5 reports: MW6004289, MW5187850, MW6000812, 3029992391-2026-00001, MW5185280
Some reports describe burns, blistering, redness, or bruising from the light pulses, including second-degree burns in one report. One user said the pulses felt unusually painful and left a barcode-like mark that lasted about a year. Users said the pulses were stronger than expected or the device fired without a sign of proper skin contact.
5 reports: MW6004132, MW6001934, MW6001579, MW6000812, 3023339103-2026-00001
One report says the device delivered more energy than the approval paperwork allowed, with a stated maximum of 19.8 joules compared with 13 joules in the clearance. The report says the actual output could be even higher, though this is the reporter's claim and was not confirmed by testing in the report.
2 reports: MW5184906, MW5184093
One report describes the device giving off light when it was not touching the skin, and light leaking from the side of the treatment window. One report says this happened repeatedly despite correct use.
Report: 3023339103-2026-00001
A report describes hair growing back darker, thicker, and coarser after repeated professional laser sessions, which the user later linked to a condition called paradoxical hypertrichosis. The user said the clinic did not warn about this possibility.
Report: MW5187951
Written by AI from the narratives of the 15 newest reports that have one, received Sep 29, 2025 to Oct 21, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 586in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 228registered establishments that list this product code
Other codes under 21 CFR 878.4810
Product codes classified under the same regulation.
Watch OHT in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get OHT every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in OHT. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:OHTRetrieved Nov 29, 2026, 4:13 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:OHTRetrieved Nov 29, 2026, 5:41 AM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:OHT · count decision_dateRetrieved Nov 29, 2026, 4:13 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:OHTRetrieved Nov 29, 2026, 5:41 AM UTC.
- device/recall · search product_code:OHT · count event_date_initiatedRetrieved Nov 29, 2026, 4:13 PM UTC.
- device/recall · search product_code:OHT · count res_event_numberRetrieved Nov 29, 2026, 4:13 PM UTC.
- device/event · search device.device_report_product_code:OHT AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 5:41 AM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:OHT AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 5:41 AM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:OHT AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 5:41 AM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:OHTRetrieved Nov 29, 2026, 4:13 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:OHTRetrieved Nov 29, 2026, 4:13 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:878.4810Retrieved Nov 29, 2026, 10:47 AM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.