Product codes

FDA product code

DSA: Cable, Transducer And Electrode, Patient, (Including Connector) (FDA product code)

Follow DSA: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 870.2900
Review panel
Cardiovascular
Premarket route
510(k)
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
1063 in the last 12 months
Median review time
132 daysthe latest 25 decisions
Recall records
120 in the last 12 months
Adverse event reports
93last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code DSA per year, by decision date, 106 since 1977. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20175
20183
20194
20206
20214
20222
20233
20243
20254
2026 (so far)2

Latest decisions

NumberDevice and applicantDecision and review time
K262816 EXG PulseLinQ Cables (EXGPL)Cb Innovations, LLC Oct 30, 202684 days
K254114 Baylis Connector CableBaylis Medical Technologies, Inc. Jan 18, 202630 days
K252827 IBP cable Becton Dickinson (2606488); IBP cable Edwards (2606489); IBP cable Abbott, Medex (2606490); IBP cable Utah (2606491); IBP Adapter Cable Draeger 7pin (2607558); IBP Adapter Cable Draeger 10pin (2607559)Drägerwerk AG & Co. KGaA Dec 1, 202587 days
K251669 IBP transducer adapter cable, Argon/BD/Ohmeda, 2m; IBP transducer adapter cable, Draeger 10-pin, 2m; IBP transducer adapter cable, Draeger 7-pin, 2m; IBP transd. adapter cbl, Abbott/Medex, Smiths/TranStar 2m; IBP transducer adapter cable, Baxter/Edwards, 0.3m; IBP transducer adapter cable, Baxter/Edwards, 2m; IBP transducer adapter cable, Argon/BD/Ohmeda, 0.3m; IBP transducer adapter cable, Draeger 10-pin, 0.3m; IBP transducer adapter cable, Draeger 7-pin, 0.3m; IBP transducer adapterShanghai Draeger Medical Instrument Co., Ltd. Sep 30, 2025123 days
K242986 SpO2 Extension CableBeijing Rongrui-Century Science & Technology Co., Ltd. May 2, 2025218 days
K243545 Philips Reusable ECG Lead Sets and Trunk Cables : Model Number,Model Name: 989803170171, OR 3-Lead ECG Trunk Cable, AAMI/IEC; 989803170181, OR 5-Lead ECG Trunk Cable, AAMI/IEC; M1667A, 6 Lead ECG Trunk, AAMI/IEC 2.7m; M1668A,5 Lead ECG Trunk, AAMI/IEC 2.7m; M1669A,3 Lead ECG Trunk, AAMI/IEC 2.7m; M1949A,5+5 ECG Trunk cable AAMI, IEC 2.7m; M1663A,10 Lead ECG Trunk AAMI/IEC 2m; M1665A, 6+4 Lead ECG Trunk AAMI/IEC 2.7m; M1671A, 3 Leadset, Grabber, AAMI, ICU; M1673A, 3 Leadset, SnPhilips Medizin Systeme Böblingen GmbH Jan 17, 202563 days
K240087 RT Carbon ECG LeadsIvy Biomedical Systems, Inc. Sep 13, 2024246 days
K240377 LIFEPAK® ECG CablesStryker Physio-Control May 7, 202490 days
K240156 LIFEPAK® Invasive Pressure Adapter Cable (Interfaces with ICU Medical Transpack IV Invasive Pressure Transducers (or equivalent)) (11230-000021); LIFEPAK® Invasive Pressure Adapter Cable (interfaces with Edwards Lifesciences TruWave Invasive Pressure Transducers (or equivalent)) (11230-000024)Physio-Control, Inc. Apr 30, 2024102 days
K222712 Fogg System Patient Monitoring CablesFogg System Company, Inc. Nov 22, 2023440 days

Recalls

Recall records for product code DSA per year, by the date the recall began. Each recalled product is one record; these 12 records belong to 8 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20170
20180
20197
20200
20210
20221
20230
20240
20251
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-1102-2025Class II Physio-Control, Inc.Product shipped with incorrect IFU. IFU does not include proper cleaning/disinfection instructions Jan 21, 2025
Z-0917-2022Class III Volcano CorpMulti-Modality Touch Screen Modules (MM-TSM), used with integrated IVUS systems, may crack, or become damaged, which may affect their usability. They could bec… Feb 3, 2022
Z-2553-2019Class II GN OtometricsThe electrode lead connector has a protruding conductor beyond insulation of approximately 0.1 mm. Aug 1, 2019
Z-2495-2019Class II Philips North America, LLCThe thickness of the connector to the trunk cable on affected limb lead sets may be greater than specified. which may prevent or impede the performance of I 2-… Jul 19, 2019
Z-1875-2019Class III GE Healthcare, LLCECG trunk cables and leadwires may reduce the amount of energy reaching the patient during defibrillation, potentially limiting success of defibrillating the p… May 10, 2019

Adverse event reports (MAUDE)

Medical device reports received per year that name product code DSA. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
20175
201816
201911
202031
20217
202249
202329
202440
202568
2026 (so far)90

Last 12 months by event type

  • Deaths0
  • Injuries1
  • Malfunctions92
  • Other or not stated0

Most reported device problems, last 12 months

  1. Material Integrity Problem70
  2. Incorrect Measurement4
  3. Loose or Intermittent Connection4
  4. Communication or Transmission Problem2
  5. Connection Problem2
  6. Device Sensing Problem2

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
12,366in the FDA’s Global Unique Device Identification Database
Establishment listings
310registered establishments that list this product code

Watch DSA in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get DSA every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in DSA. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.