Product codes

FDA product code

EQJ: Bur, Ear, Nose And Throat (FDA product code)

Follow EQJ: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class I
Regulation
21 CFR 874.4140
Review panel
Ear, Nose and Throat
Premarket route
510(k) exempt
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
180 in the last 12 months
Median review time
48 daysthe latest 18 decisions
Recall records
40 in the last 12 months
Adverse event reports
212last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code EQJ per year, by decision date, 18 since 1979. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20171
20180
20190
20200
20210
20220
20230
20240
20250
2026 (so far)0

Latest decisions

NumberDevice and applicantDecision and review time
K161514 Precision Thin Reciprocating Blade, 0.010in.Stryker Corporation Jan 17, 2017229 days
K041985 XPS CURVED BURMedtronic Xomed, Inc. Aug 12, 200420 days
K974232 ESSENTIAL SHAVER SYSTEMGyrus Ent LLC Jan 12, 199861 days
K960853 ACE SYSTEMXomed, Inc. Apr 3, 199633 days
K955704 TREBAY MICRODEBRIDER SYSTEMTrebay Medical Corp. Jan 30, 199646 days
K955294 SOF - ABRASION BURTrebay Medical Corp. Dec 22, 199535 days
K954912 TREBAY IRRIGATED CUTTER BLADE (MODIFICATION)Trebay Medical Corp. Dec 6, 199544 days
K953370 KARL STORZ ENT SHAVERKARL STORZ Endoscopy-America, Inc. Sep 6, 199550 days
K950337 KARL STORZ STAMMBERGER-SACHSE INTRANASAL-FRILLKARL STORZ Endoscopy-America, Inc. Aug 23, 1995208 days
K952609 REDI DRILL DRILL SYSTEMTrebay Medical Corp. Jul 31, 199554 days

Recalls

Recall records for product code EQJ per year, by the date the recall began. Each recalled product is one record; these 4 records belong to 4 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20171
20180
20190
20200
20211
20220
20230
20241
20250
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-3130-2024Class II Medtronic Xomed, Inc.Product was distributed past expiration date. Jul 5, 2024
Z-0156-2022Class II Medtronic Xomed, Inc.There was an increase in complaints related to tip breaks and wobble/vibration. Aug 31, 2021
Z-1337-2018Class II Gyrus ACMI, Inc."Chattering" when cutting burrs are used in bone. Aug 3, 2017
Recall Medtronic Xomed, Inc.Product packaging pouches may be open thereby compomising the sterile barrier. Jan 25, 2002

Adverse event reports (MAUDE)

Medical device reports received per year that name product code EQJ. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
2017118
2018204
2019129
2020100
2021131
2022229
2023261
2024368
2025445
2026 (so far)183

Last 12 months by event type

  • Deaths0
  • Injuries4
  • Malfunctions208
  • Other or not stated0

Most reported device problems, last 12 months

  1. Break104
  2. Vibration67
  3. Material Fragmentation30
  4. Physical Resistance/Sticking12
  5. Mechanical Jam11
  6. Material Protrusion/Extrusion4

In plain English: what the newest reports describe Written by AI

  • In seven reports, a bur or blade broke inside the handpiece or drill, and a piece stayed lodged in the device. Some were found before surgery and some during it. The reports state no patient was involved.

    7 reports: 1045254-2026-01428, 1045254-2026-01399, 1045254-2026-01285, 1045254-2026-01280, 1045254-2026-01416, 1045254-2026-01266, 1045254-2026-01281

  • In three reports, a bur or blade broke during use, in one case within five minutes. One report describes a shaft and tip that broke and jammed after one use at a high speed setting. One procedure was finished with backup product, and the reports state no patient impact.

    3 reports: 1045254-2026-01398, 1045254-2026-01377, 1045254-2026-01275

  • In three reports, the blade would not cut or would not turn. One report says the inner blade was bent. One procedure was finished with backup product, and one report says no patient was injured.

    3 reports: 1045254-2026-01309, 1045254-2026-01305, 1045254-2026-01314

  • In three reports, the blade or attachment wobbled or rattled. One report describes a rattling blade during an endoscopic septoplasty, which was finished with a replacement device. The reports state no patient impact.

    3 reports: 1045254-2026-01311, 1045254-2026-01323, 1045254-2026-01277

  • One report describes a bur that broke into multiple pieces inside a patient's ear during cholesteatoma surgery, a growth in the ear. The report says the bur had been re-sterilized 21 times, and the surgeons removed all the pieces. Nerve monitoring showed no reaction, and the patient was watched for 48 hours and discharged.

    Report: 1045254-2026-01337

Written by AI from the narratives of the 20 newest reports that have one, received Sep 11, 2026 to Oct 30, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
2,297in the FDA’s Global Unique Device Identification Database
Establishment listings
128registered establishments that list this product code

Product codes classified under the same regulation.

Watch EQJ in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get EQJ every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in EQJ. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.