Product codes

FDA product code

QPB: System, Suction, Lipoplasty For Removal (FDA product code)

Follow QPB: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 878.5040
Review panel
General and Plastic Surgery
Premarket route
510(k)
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Ineligible
510(k) and De Novo decisions
605 in the last 12 months
Median review time
165 daysthe latest 25 decisions
Recall records
10 in the last 12 months
Adverse event reports
6last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code QPB per year, by decision date, 60 since 1998. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20173
20180
20192
20202
20210
20222
20230
20244
20252
2026 (so far)5

Latest decisions

NumberDevice and applicantDecision and review time
K254135 LipoSurg ARTISTNOUVAG AG Sep 5, 2026257 days
K261710 AYON™ Power Liposuction Cannulas and AYON™ Power Liposuction Cannula AdaptorsApyx Medical Corporation Aug 21, 202691 days
K251944 Lipocosm Harmonic SystemLipocosm, LLC Jun 8, 2026349 days
K253396 AYON Body Contouring System™ (AYON SYSTEM)Apyx Medical Corporation May 8, 2026220 days
K252495 LSSA (NP-200)Newpong Co., Ltd. Apr 3, 2026238 days
K244050 AYON Body Contouring System (AYON SYSTEM)Apyx Medical Corporation May 12, 2025132 days
K244026 PAL 7020 ConsoleMicroaire Surgical Instruments, LLC Apr 25, 2025116 days
K242804 PAL Aspiration SystemMicroaire Surgical Instruments, LLC Dec 30, 2024104 days
K241835 AirSculpt Body Contouring System (AIRS-1-SYS)Cellmyx Nov 27, 2024155 days
K240188 Vitruvian LiposaberBlack & Black Surgical, Inc. Sep 13, 2024233 days

Recalls

Recall records for product code QPB per year, by the date the recall began. Each recalled product is one record; these 1 records belong to 1 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20170
20180
20190
20200
20210
20220
20230
20241
20250
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-1251-2025Class II HUMAN MED AGThe cannula tubes do not comply with the specification. The water jet emerging from the nozzle of the nonconforming cannulas may emerge as a spot jet rather th… Oct 18, 2024

Adverse event reports (MAUDE)

Medical device reports received per year that name product code QPB. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
20170
20180
20190
20200
20210
20225
202328
202416
202516
2026 (so far)6

Last 12 months by event type

  • Deaths0
  • Injuries4
  • Malfunctions2
  • Other or not stated0

Most reported device problems, last 12 months

  1. Adverse Event Without Identified Device or Use Problem3
  2. Overheating of Device2

In plain English: what the newest reports describe Written by AI

  • Several reports describe burns, bruising and blisters on the abdomen, flank, neck or leg after ultrasound liposuction with the device. One patient reported skin death across half the abdomen that needed a skin graft after months of wound vacuum care.

    6 reports: 3011423170-2026-00338, 3011423170-2026-00339, 3011423170-2026-00334, 3011423170-2025-00282, 3011423170-2025-00247, MW5171525

  • Several reports describe the suction pump, foot pedal or footswitch failing during surgery. Staff faced delays of over an hour with the patient under general anesthesia, and some procedures were cancelled or finished with another device.

    6 reports: 3011423170-2025-00274, 3011423170-2025-00244, 3011423170-2025-00254, 3011423170-2025-00267, 3011423170-2025-00270, 3011423170-2025-00276

  • Two reports describe suction cannulas, the tubes used for suction, found bent before surgery, one still in its sealed package. In one report, staff had to open a second bent cannula before finding a usable one.

    2 reports: 23423510, 23424549

  • Two reports describe cords causing burns. In one, the cable attached to the handpiece touched a patient's shoulder and caused burns that were found after the procedure. In the other, a cord was said to overheat and burn a patient's foot.

    2 reports: 2020601-2026-00003, 25665586

  • One report describes a patient who died after fat transfer liposuction. The reporter stated the cause was a pulmonary embolism, a fat clot that blocked blood flow to the lungs, which occurred as the liposuction was concluded.

    Report: 3011423170-2025-00275

Written by AI from the narratives of the 20 newest reports that have one, received Jan 21, 2025 to Sep 3, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
640in the FDA’s Global Unique Device Identification Database
Establishment listings
83registered establishments that list this product code

Watch QPB in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get QPB every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in QPB. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.