FDA product code
QJZ: Extracorporeal System For Long-Term Respiratory / Cardiopulmonary Failure (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 870.4100
- Review panel
- Cardiovascular
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- Yes
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Ineligible
- 510(k) and De Novo decisions
- 41 in the last 12 months
- Median review time
- 186 daysthe latest 4 decisions
- Recall records
- 20 in the last 12 months
- Adverse event reports
- 117last 12 months, 10 deaths
510(k) and De Novo decisions
Decisions for product code QJZ per year, by decision date, 4 since 2020. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 1 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 2 |
| 2025 | 0 |
| 2026 (so far) | 1 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K261335 | VitalFlow Set with Balance™ BiosurfaceMedtronic, Inc. | Jul 31, 202699 days |
| K240880 | VitalFlow Set with Balance BiosurfaceMedtronic, Inc. | Sep 12, 2024164 days |
| K232926 | Novalung ultimate kit (US)Fresenius Medical Care Renal Therapies Group, LLC | Apr 15, 2024208 days |
| K191407 | Novalung SystemFresenius Medical Care Renal Therapies Group, LLC | Feb 21, 2020269 days |
Recalls
Recall records for product code QJZ per year, by the date the recall began. Each recalled product is one record; these 2 records belong to 2 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 1 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 1 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-1822-2024Class II | Fresenius Medical Care Holdings, Inc.Issue related to Novalung sensor box, which is a component of the Novalung Console (Part Number F30000163). In certain instances, error messages #206 (yellow)… | Apr 8, 2024 |
| Z-0562-2022Class II | Fresenius Medical Care Holdings, Inc.Error messages 206 (yellow) and 208 (red) technical failure, flow measurement during use of the Novalung System imply that the communication between flow senso… | Dec 7, 2021 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code QJZ. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 17 |
| 2021 | 52 |
| 2022 | 53 |
| 2023 | 104 |
| 2024 | 75 |
| 2025 | 131 |
| 2026 (so far) | 107 |
Last 12 months by event type
- Deaths10
- Injuries12
- Malfunctions95
- Other or not stated0
Most reported device problems, last 12 months
- Insufficient Device Problem Information78
- Incorrect, Inadequate or Imprecise Result or Readings41
- Output Problem40
- Leak/Splash16
- Mechanical Problem13
- Device Alarm System7
In plain English: what the newest reports describe Written by AI
Several reports describe the pump head making an abnormal noise, such as grinding or a loud sound, during support. In some cases the noise appeared when the pump speed was raised above about 7,100 RPM, and staff lowered the speed or changed the circuit as a result.
5 reports: 3012172416-2026-00069, 3012172416-2026-00071, 3012172416-2026-00084, 3012172416-2026-00089, 3012172416-2026-00093
Several reports describe the oxygenator, which adds oxygen to the blood, failing early and needing replacement within hours or a day. In one report oxygenation was poor, and the circuit was changed.
4 reports: 3012172416-2026-00073, 3012172416-2026-00081, 3012172416-2026-00086, 3012172416-2026-00065
Several reports describe kits that had to be exchanged, and blood loss of about 500 to 600 ml occurred during some of those changes. One report describes a vent connection tube on the venous side rupturing and leaking blood about 10 minutes after support started.
4 reports: 3012172416-2026-00068, 3012172416-2026-00070, 3012172416-2026-00086, 3012172416-2026-00089
A report describes a hairline crack in the oxygenator, and a manufacturer report describes a leak from a sensor port found when the returned device was analyzed. In one case a patient's blood clotting problems were thought, in retrospect, by physicians possibly to be related to the device setup, while the manufacturer found only small clots in the oxygenator.
3 reports: 3012172416-2026-00086, 9617601-2026-05627, 9617601-2026-06534
A report describes the pump head failing to rotate during priming. Normal operation returned after a new kit was used, and the manufacturer's check ruled out equipment failure.
Report: 3012172416-2026-00090
Written by AI from the narratives of the 17 newest reports that have one, received Aug 11, 2026 to Oct 29, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 5in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 14registered establishments that list this product code
Other codes under 21 CFR 870.4100
Product codes classified under the same regulation.
Watch QJZ in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get QJZ every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in QJZ. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:QJZRetrieved Nov 28, 2026, 9:38 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:QJZRetrieved Nov 29, 2026, 6:40 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:QJZ · count decision_dateRetrieved Nov 28, 2026, 9:38 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:QJZRetrieved Nov 29, 2026, 6:39 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:QJZ · count event_date_initiatedRetrieved Nov 28, 2026, 9:38 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:QJZ · count res_event_numberRetrieved Nov 28, 2026, 9:38 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-1822-2024" "Z-0562-2022")Retrieved Nov 29, 2026, 6:40 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:QJZ AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 6:39 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:QJZ AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 6:40 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:QJZ AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 1:44 AM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:QJZRetrieved Nov 28, 2026, 9:38 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:QJZRetrieved Nov 28, 2026, 9:38 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:870.4100Retrieved Nov 28, 2026, 9:38 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil