FDA device news

Weekly archive

Week of August 5, 2024

The FDA cleared 48 510(k)s, granted no De Novo requests and approved no original PMAs. 21 device recalls were posted, 2 of them Class I. Every 510(k), De Novo and PMA decision dated August 5 to 11, 2024, and every device recall the FDA posted those days.

510(k) clearances
48median review 221 days
De Novo grants
0
Original PMA approvals
0plus 26 supplements
Recalls posted
2147 products, 2 Class I

Recalls posted

Device recalls the FDA posted August 5 to 11, 2024, Class I first. Class I is the most serious: a reasonable chance the product causes serious harm or death. The class comes from the FDA’s weekly enforcement report, so the newest recalls can show “Class pending” for a few days.

Recall and classFirm and reasonProduct code
Z-2447-2024Class I ICU Medical IncDue to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the batter… FRN3 products
Z-2442-2024Class I MEDLINE INDUSTRIES, LP - NorthfieldMedical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health. OES OFT OGJ3 products
Z-2544-2024Class II Beckman Coulter, Inc.The reason for this recall is Beckman Coulter has determined that when a DxI 9000 Access Immunoassay Analyzer is connected to a Laboratory Automation System (LAS); is running with the software versions 1.17 and below; a… JJE1 product
Z-2545-2024Class II Beckman Coulter, Inc.Beckman Coulter, Inc. has identified a rise in reagent carousel motion errors in Access 2 Immunoassay System Analyzers that appear to be caused by misaligned index wheels housed within the reagent carousel module. Custo… JJE1 product
Z-2531-2024Class II Artivion, IncGrafts were found to have been approved, released, shipped and implanted with an invalid test results. DXZ OHA2 products
Z-2533-2024Class II Luminex CorporationPotential for defective test cartridges which may result in false positive results for the Staphylococcus lugdunensis and Enterococcus faecium targets. PAM1 product
Z-2534-2024Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Potential Loss of Imaging (X-ray) Functionality and/or Longer Time to Perform Cold Restart OWB1 product
Z-2536-2024Class II Steris CorporationIFU for hookup were incorrectly reflected within hookup IFUs and online reference system, Hookup Lookup (HULU). Use of an incorrect hookup connection and/or parameter set may prevent an endoscope from being properly dis… FEB4 products
Z-2540-2024Class II Olympus Corporation of the AmericasFractures and breakages in packaging trays and Tyvek covers, which may result in a sterility breach GEI3 products
Z-2515-2024Class II MEDLINE INDUSTRIES, LP - NorthfieldMedical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health. LRO LRP2 products
Z-2517-2024Class II American Contract Systems, Inc.Product released with insufficient aeration time impacting the kits sterilized on the same cycle or in the same aeration chamber, which may have been compromised in sterility FCH FDE LRO8 products
Z-2525-2024Class II Siemens Medical Solutions USA, IncThe support arm may unintentionally lower resulting in injury to persons when they are under the display ceiling/wall suspension when positioning the displays OWB4 products
Z-2529-2024Class II LEICA BIOSYSTEMS NUSSLOCH GMBHAn issue with safe usage of the device was identified whereby toxic smoke and internal fire developed resulting from a problem associated with the sealing of the paraffin tank. IDW1 product
Z-2530-2024Class II Baxter Healthcare CorporationImproperly performed testing prior to release FRN1 product
Z-2512-2024Class II Baxter Healthcare CorporationBaxter Healthcare Corporation is issuing an Urgent Medical Device Correction for the PST 500 U Precision Surgical Table due to a potential issue with loose spring pins in the spindle drives. This can cause the tabletop… GDC1 product
Z-2513-2024Class II W.L. Gore & Associates, Inc.Devices were labeled with an expiration date of four years and ten days rather than the validated two years. FTL1 product
Z-2514-2024Class II BioFire Diagnostics, LLCDue to increase of false positive Vibrio/Vibrio Cholerae results when using a certain transport medium. PCH1 product
Z-2504-2024Class II Olympus Corporation of the AmericasDifficulties in pairing the wireless footswitch with the Soltive Laser, potentially delaying the surgical procedure. GEX3 products
Z-2507-2024Class II Integra LifeSciences Corp.Specific lots of TruDi NAV Suction Instruments 0 Degrees were incorrectly calibrated which when used with the TruDi Navigation System, may result in a discrepancy between the actual tip location and the visualized locat… PGW1 product
Z-2508-2024Class II Philips North America LlcPhilips lntelliVue Patient Monitors MX400/450/500/550 Shipped or Upgraded with Incorrect Software Options. Changes were made to reflect some options becoming standard capabilities for software-version N.x. In the proces… MHX4 products
Z-2535-2024Class III LINEAGE BIOMEDICAL, INCIncorrect GTIN number on carton. FMF1 product

De Novo grants

New device types the FDA classified through the De Novo pathway. Each one creates a product code and becomes a possible predicate for later 510(k)s.

No De Novo decisions this week.

Original PMA approvals

Original premarket approvals for Class III devices. The FDA also decided 26 PMA supplements this week, which change devices already approved.

No original PMAs were approved this week.

510(k) clearances

48 510(k) clearances by review panel. The median took 221 calendar days from receipt to decision; the fastest, K242009, 29 days; the longest, K212969, 1,056 days.

Calendar days include the time a submission waits on the company, so they run longer than the FDA’s 90-day review goal. How to search the 510(k) database

  • Radiology8
  • General, Plastic Surgery7
  • Cardiovascular5
  • Gastroenterology, Urology5
  • Neurology4
  • Anesthesiology3
  • Dental3
  • Physical Medicine3
  • General Hospital2
  • Microbiology2
  • Clinical Chemistry1
  • Immunology1
  • Obstetrics/Gynecology1
  • Ophthalmic1
  • Orthopedic1
  • Panel not stated1

Every clearance

NumberDevice and applicantProduct codeDecision and review time
K233643 F&P Airvo 3 NIV (PT311US)Fisher & Paykel Healthcare, Ltd. MNTAnesthesiology Aug 9, 2024270 days
K234053 F&P Optiflow Flow DiverterFisher & Paykel Healthcare, Ltd. CBPAnesthesiology Aug 9, 2024231 days
K234058 F&P Optiflow Filtered Nasal Interface with Anesthesia Mask Compatibility (AA041J)Fisher & Paykel Healthcare, Ltd. BTTAnesthesiology Aug 8, 2024230 days
K233433 Sphere-9Dx Diagnostic Catheter (AFR-00009)Medtronic MTDCardiovascular Aug 7, 2024300 days
K233702 SMARxT Tubing and ConnectorsLivaNova USA, Inc. DWECardiovascular Aug 8, 2024265 days
K240229 Masimo W1Masimo Corporation DPSCardiovascular Aug 8, 2024192 days
K241708 Abiomed 14Fr Low Profile Introducer KitAbiomed, Inc. DYBCardiovascular Aug 6, 202454 days
K242016 EnSite™ X EP SystemABBOTT MEDICAL DQKCardiovascular Aug 9, 202430 days
K232404 CHOLESTEROL; HDL-cholesterol; LDL-cholesterol; TRIGLYCERIDESMedicon Hellas S.A CHHClinical Chemistry Aug 9, 2024365 days
K222688 The TrustGenoss Co., Ltd. EKXDental Aug 9, 2024703 days
K233356 Illusion Aligner Pro: Illusion Aligner FLXLaxmi Dental Export Pvt., Ltd. NXCDental Aug 8, 2024314 days
K233844 OdneCleanOdne AG ELCDental Aug 8, 2024248 days
K233312 SOLOASSIST IID / DEXTER ENDOSCOPE ARM (212499); ARTip solo (202426) + ARTip solo voice (202445); SOLOASSIST II (141364) + VOICE CONTROL (171894Aktormed GmbH QZBGastroenterology, Urology Aug 8, 2024314 days
K233653 Midfield Light Pipe, 23ga (3269.M06-00); Midfield Light Pipe, 25ga (3269.M05-00); Midfield Light Pipe, 27ga (3269.M04-00)Peregrine Surgical, LLC MPAGastroenterology, Urology Aug 5, 2024265 days
K240664 ALARA BMA Neuro Access Kit (ALARA BMAN and ALARA BMAC Kit)SurGenTec, LLC KNWGastroenterology, Urology Aug 9, 2024154 days
K241360 Universal Seal (5-12 mm)Intuitive Surgical, Inc. GCJGastroenterology, Urology Aug 5, 202483 days
K241530 Stellar Imaging SystemGuangdong Optomedic Technologies, Inc. GCJGastroenterology, Urology Aug 7, 202469 days
K240330 HALYARD* SMART-FOLD* Sterilization Wrap (H450, H650)O&M Halyard, Inc. FRGGeneral Hospital Aug 7, 2024184 days
K241528 Nitrile Powder Free Examination Gloves (Blue), Non-Sterile Low Dermatitis PotentialComfort Rubber Gloves Industries Sdn. Bhd. LZAGeneral Hospital Aug 6, 202468 days
K212969 LiposaverLhbiomed Co., Ltd. QPBGeneral, Plastic Surgery Aug 7, 20241,056 days
K231950 MEDIDIODEMedicreations, LLC GEXGeneral, Plastic Surgery Aug 9, 2024406 days
K240127 PlumeSafe X5 Smoke Management SystemConmed Corporation FYDGeneral, Plastic Surgery Aug 5, 2024201 days
K240782 CellFX Percutaneous Electrode System (SYS3000); CellFX Percutaneous Electrode, 13G (CPE013); CellFX Percutaneous Electrode, 18G (CPE018)Pulse Biosciences, Inc. GEIGeneral, Plastic Surgery Aug 5, 2024137 days
K240932 HybridTherm SystemErbe Elektromedizin GmbH GEIGeneral, Plastic Surgery Aug 5, 2024122 days
K241376 HyperSuture All Blue Extension LineThreadstone, LLC GATGeneral, Plastic Surgery Aug 9, 202486 days
K241635 da Vinci E-200 Electrosurgical GeneratorIntuitive Surgical GEIGeneral, Plastic Surgery Aug 5, 202460 days
K240469 Access TPO AntibodyBeckman Coulter, Inc. JZOImmunology Aug 9, 2024171 days
K233605 ADVIA Centaur EBV-EBNA IgGBiokit, S.A. LLMMicrobiology Aug 7, 2024272 days
K233606 ADVIA Centaur EBV-VCA IgMBiokit, S.A. LSEMicrobiology Aug 7, 2024272 days
K233315 HipXpert and HipInsight SystemSurgical Planning Associates, Inc. OSFNeurology Aug 8, 2024314 days
K233577 Sleepio®Big Health, Inc. QVONeurology Aug 5, 2024272 days
K241644 SPINEART Navigation Instrument SystemSpineart SA OLONeurology Aug 7, 202461 days
K242009 360CAS KneeKico Knee Innovation Company Pty, Ltd. OLONeurology Aug 8, 202429 days
K240605 Ultra RapidWarm™ BlastVitrolife Sweden AB MQLObstetrics/Gynecology Aug 7, 2024156 days
K233720 LuminopiaLuminopia, Inc. QQUOphthalmic Aug 8, 2024261 days
K233992 Precision AI Surgical Planning System (PAI-SPS)Precision AI Pty, Ltd. QHEOrthopedic Aug 5, 2024231 days
K231775 GR Resin System MSIPro3Dure Medical MQCPanel not stated Aug 8, 2024419 days
K240079 Electric Wheelchair (Panda 10, Panda 12)Suzhou Ideal Medical Co., Ltd. ITIPhysical Medicine Aug 9, 2024211 days
K240255 Electric Wheelchair (JLE-W01A-2023)Zhejiang Jianrui Technology Company Limited ITIPhysical Medicine Aug 7, 2024189 days
K241685 SofWave SystemSofwave Medical, Ltd. NGXPhysical Medicine Aug 7, 202457 days
K233697 SonoMax Series Digital Color Doppler Ultrasound SystemCHISON Medical Technologies Co., Ltd. IYNRadiology Aug 9, 2024266 days
K233761 Orthopaedic Surgery Planning Software (AIJOINT)Longwood Valley Medical , Ltd. LLZRadiology Aug 5, 2024255 days
K233822 i2ContourMrimath, LLC QIHRadiology Aug 8, 2024251 days
K233920 EOSedgeEos Imaging KPRRadiology Aug 6, 2024237 days
K240724 SynthVISION 1.0.0Synthesis Health Intelligence, Inc. LLZRadiology Aug 8, 2024143 days
K240796 myAblation Guide (VB80A)Siemens Healthcare GmbH QTZRadiology Aug 6, 2024137 days
K240953 AI Platform 2.0 (AIP002)Exo Imaging QIHRadiology Aug 5, 2024119 days
K241665 Omni LegendGe Medical Systems Israel, Functional Imaging KPSRadiology Aug 8, 202459 days

Only see the weeks that matter to you

DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and every day it checks openFDA and tells you about new clearances, recalls and MAUDE reports for them. It also puts all of this beside the FDA’s own record pages.

Add to Chrome — free Add to Edge Or follow this page by RSS

Only your product codes, every Monday

One short email after each openFDA update with just the decisions and recalls in up to three product codes you choose. Free.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Decisions arrive in weekly updates; check the FDA record before you rely on a figure.