FDA product code

NVZ: Pulse Generator, Permanent, Implantable (FDA product code)

Follow NVZ: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class III
Regulation
None listed
Review panel
Cardiovascular
Premarket route
PMA
Implant
Yes
Life-sustaining or supporting
Yes
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Ineligible
PMA decisions
1,98216 original PMAs, the rest supplements
Median review time
26 daysthe latest 25 decisions
Recall records
433 in the last 12 months
Adverse event reports
5,139last 12 months, 111 deaths

PMA decisions

Original PMAs and supplements for product code NVZ per year, by decision date, 1,982 since 1989. Review time is the days from receipt to decision.

All of them in the FDA’s PMA database

PMA decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
201761
201859
201960
202073
202140
202244
202339
202442
202539
2026 (so far)26

Latest decisions

NumberDevice and applicantDecision and review time
P980035 S850 Adapta, Versa, Sensia IPG ADDR01, ADDR03, ADDRL1, ADDRS1, ADSR06 Adapta, Versa, Sensia IPG SEDRL1 Adapta, Versa, SensiaMedtronic, Inc. Nov 11, 202622 days
P980035 S849 Astra S DR MRI IPG; Astra S SR MRI IPG; Astra XT DR MRI IPG; Astra XT SR MRI IPG; Azure S DR MRI IPG; Azure S SR MRI IPGMedtronic, Inc. Oct 29, 202614 days
P980035 S848 Astra S DR MRI IPG (X3DR01); Astra S SR MRI IPG (X3SR01); Astra XT DR MRI IPG (X1DR01); Astra XT SR MRI IPG (X1SR01); AzMedtronic, Inc. Oct 23, 202629 days
P980035 S847 Astra S DR MRI IPG, Astra S SR MRI IPG, Astra XT SR MRI IPG, Azure S DR MRI IPG, Azure S SR MRI IPG, Azure XT DR MRI IPGMedtronic, Inc. Oct 5, 202626 days
P980035 S846 Adapta, Versa, Sensia IPG (ADDR03, ADDRL1, ADDRS1, ADSR01, ADSR03, SED01, SES01, SESR01), Adapta, Versa, Sensia IPG (SEDMedtronic, Inc. Jul 31, 202610 days
P980035 S844 Adapta, Versa, Sensia IPG (ADDR03, ADDR06, ADSR01, SEDR01, SESR01, SEDRL1); Adapta, Versa, SMedtronic, Inc. Jul 31, 202624 days
P890003 S487 CareLink SmartSync Device Manager Base (24970A)Medtronic, Inc. Jul 31, 202624 days
P980035 S845 Adapta, Versa, Sensia IPG (ADDR06, ADDRS1, ADSR01, ADSR03, ADSR06, SEDR01, SESR01, SEDRL1)Medtronic, Inc. Jul 31, 202624 days
P980035 S843 Astra S DR MRI IPG, Astra S SR MRI IPG, Astra XT DR MRI IPG, Astra XT SR MRI IPG, Azure S DR MRI IPG, Azure S SR MRI IPGMedtronic, Inc. Jul 16, 202630 days
P980035 S842 Astra S DR MRI IPG (X3DR01); Astra S SR MRI IPG (X3SR01); Astra XT DR MRI IPG (X1DR01); Astra XT SR MRI IPG (X1SR01); AzMedtronic, Inc. Jul 1, 202629 days

Recalls

Recall records for product code NVZ per year, by the date the recall began. Each recalled product is one record; these 43 records belong to 20 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20172
20180
20197
20209
20213
20220
20230
20243
20253
2026 (so far)3

Latest recalls

Recall and classFirm and reasonBegan
Z-0302-2027Class II Medtronic, Inc.Software problem has the potential for an inaccurate displayed longevity estimate. Sep 24, 2026
Z-0303-2027Class II Medtronic, Inc.Software problem has the potential for an inaccurate displayed longevity estimate. Sep 24, 2026
Z-0304-2027Class II Medtronic, Inc.Software problem has the potential for an inaccurate displayed longevity estimate. Sep 24, 2026
Z-2168-2025Class II Medtronic, Inc.In prior SmartSync application versions, the Abort button stopped the test that was selected. During an induction test, there was a limited window of time for… Jun 18, 2025
Z-2169-2025Class II Medtronic, Inc.In prior SmartSync application versions, the Abort button stopped the test that was selected. During an induction test, there was a limited window of time for… Jun 18, 2025

Adverse event reports (MAUDE)

Medical device reports received per year that name product code NVZ. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
20173,164
20185,878
20198,188
20208,337
20218,539
20227,791
20238,463
20246,496
20256,059
2026 (so far)4,707

Last 12 months by event type

  • Deaths111
  • Injuries2,660
  • Malfunctions2,355
  • Other or not stated13

Most reported device problems, last 12 months

  1. Adverse Event Without Identified Device or Use Problem1,520
  2. Pacing Problem670
  3. Premature Discharge of Battery594
  4. Appropriate Device Problem Term/Code Not Available430
  5. Mechanical Problem279
  6. Failure to Interrogate201

In plain English: what the newest reports describe Written by AI

  • Several reports describe infection at the device pocket, the space under the skin where the pulse generator sits. In some cases the skin eroded or was red, painful or discharging pus, and the pacemaker was removed; one report mentions a recent joint replacement, and one report notes the patient had a fever and a bloodstream infection (bacteremia) caused by MRSA, a drug-resistant bacterium.

    7 reports: 2017865-2026-22930, 3008973940-2026-08050, 3008973940-2026-08054, 3004209178-2026-18542, 3008973940-2026-08071, 3008973940-2026-08014, 3008973940-2026-08023

  • Several reports describe the lead's screw or setscrew, which holds the lead in the pulse generator, failing during implant. The lead could not be secured or released, so the generator was set aside and replaced; in one report the lead could not be removed and had to be cut.

    5 reports: 9614453-2026-04438, 3008973940-2026-08015, 3008973940-2026-08019, 9614453-2026-04440, 2182208-2026-62352

  • Two reports describe the right atrial lead sensing heartbeats incorrectly, including signals arriving during its blanking period (a short window when the device ignores signals) and far-field signals from the ventricle. This led to pacing at times when it was not needed, and the lead remains in use.

    2 reports: 3008973940-2026-08055, 3008973940-2026-08059

  • Several reports describe false impedance warnings, meaning the device signaled a lead connection problem that was not confirmed, along with noise and oversensing on the right ventricular lead in one case. The generator and lead remain in use, and no patient harm was reported.

    2 reports: 3008973940-2026-08066, 3008973940-2026-08010

  • One report describes strong chest pain and abdominal pain during ventricular pacing, with abdominal muscle spasms and diaphragm twitching seen once. The patient was hospitalized, and the generator remains in use.

    Report: 9614453-2026-04439

Written by AI from the narratives of the 20 newest reports that have one, received Oct 29, 2026 to Oct 30, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
341in the FDA’s Global Unique Device Identification Database
Establishment listings
91registered establishments that list this product code

Watch NVZ in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get NVZ every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in NVZ. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.