NVZ: Pulse Generator, Permanent, Implantable (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class III
- Regulation
- None listed
- Review panel
- Cardiovascular
- Premarket route
- PMA
- Implant
- Yes
- Life-sustaining or supporting
- Yes
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Ineligible
- PMA decisions
- 1,98216 original PMAs, the rest supplements
- Median review time
- 26 daysthe latest 25 decisions
- Recall records
- 433 in the last 12 months
- Adverse event reports
- 5,139last 12 months, 111 deaths
PMA decisions
Original PMAs and supplements for product code NVZ per year, by decision date, 1,982 since 1989. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 61 |
| 2018 | 59 |
| 2019 | 60 |
| 2020 | 73 |
| 2021 | 40 |
| 2022 | 44 |
| 2023 | 39 |
| 2024 | 42 |
| 2025 | 39 |
| 2026 (so far) | 26 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| P980035 S850 | Adapta, Versa, Sensia IPG ADDR01, ADDR03, ADDRL1, ADDRS1, ADSR06 Adapta, Versa, Sensia IPG SEDRL1 Adapta, Versa, SensiaMedtronic, Inc. | Nov 11, 202622 days |
| P980035 S849 | Astra S DR MRI IPG; Astra S SR MRI IPG; Astra XT DR MRI IPG; Astra XT SR MRI IPG; Azure S DR MRI IPG; Azure S SR MRI IPGMedtronic, Inc. | Oct 29, 202614 days |
| P980035 S848 | Astra S DR MRI IPG (X3DR01); Astra S SR MRI IPG (X3SR01); Astra XT DR MRI IPG (X1DR01); Astra XT SR MRI IPG (X1SR01); AzMedtronic, Inc. | Oct 23, 202629 days |
| P980035 S847 | Astra S DR MRI IPG, Astra S SR MRI IPG, Astra XT SR MRI IPG, Azure S DR MRI IPG, Azure S SR MRI IPG, Azure XT DR MRI IPGMedtronic, Inc. | Oct 5, 202626 days |
| P980035 S846 | Adapta, Versa, Sensia IPG (ADDR03, ADDRL1, ADDRS1, ADSR01, ADSR03, SED01, SES01, SESR01), Adapta, Versa, Sensia IPG (SEDMedtronic, Inc. | Jul 31, 202610 days |
| P980035 S844 | Adapta, Versa, Sensia IPG (ADDR03, ADDR06, ADSR01, SEDR01, SESR01, SEDRL1); Adapta, Versa, SMedtronic, Inc. | Jul 31, 202624 days |
| P890003 S487 | CareLink SmartSync Device Manager Base (24970A)Medtronic, Inc. | Jul 31, 202624 days |
| P980035 S845 | Adapta, Versa, Sensia IPG (ADDR06, ADDRS1, ADSR01, ADSR03, ADSR06, SEDR01, SESR01, SEDRL1)Medtronic, Inc. | Jul 31, 202624 days |
| P980035 S843 | Astra S DR MRI IPG, Astra S SR MRI IPG, Astra XT DR MRI IPG, Astra XT SR MRI IPG, Azure S DR MRI IPG, Azure S SR MRI IPGMedtronic, Inc. | Jul 16, 202630 days |
| P980035 S842 | Astra S DR MRI IPG (X3DR01); Astra S SR MRI IPG (X3SR01); Astra XT DR MRI IPG (X1DR01); Astra XT SR MRI IPG (X1SR01); AzMedtronic, Inc. | Jul 1, 202629 days |
Recalls
Recall records for product code NVZ per year, by the date the recall began. Each recalled product is one record; these 43 records belong to 20 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 2 |
| 2018 | 0 |
| 2019 | 7 |
| 2020 | 9 |
| 2021 | 3 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 3 |
| 2025 | 3 |
| 2026 (so far) | 3 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-0302-2027Class II | Medtronic, Inc.Software problem has the potential for an inaccurate displayed longevity estimate. | Sep 24, 2026 |
| Z-0303-2027Class II | Medtronic, Inc.Software problem has the potential for an inaccurate displayed longevity estimate. | Sep 24, 2026 |
| Z-0304-2027Class II | Medtronic, Inc.Software problem has the potential for an inaccurate displayed longevity estimate. | Sep 24, 2026 |
| Z-2168-2025Class II | Medtronic, Inc.In prior SmartSync application versions, the Abort button stopped the test that was selected. During an induction test, there was a limited window of time for… | Jun 18, 2025 |
| Z-2169-2025Class II | Medtronic, Inc.In prior SmartSync application versions, the Abort button stopped the test that was selected. During an induction test, there was a limited window of time for… | Jun 18, 2025 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code NVZ. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 3,164 |
| 2018 | 5,878 |
| 2019 | 8,188 |
| 2020 | 8,337 |
| 2021 | 8,539 |
| 2022 | 7,791 |
| 2023 | 8,463 |
| 2024 | 6,496 |
| 2025 | 6,059 |
| 2026 (so far) | 4,707 |
Last 12 months by event type
- Deaths111
- Injuries2,660
- Malfunctions2,355
- Other or not stated13
Most reported device problems, last 12 months
- Adverse Event Without Identified Device or Use Problem1,520
- Pacing Problem670
- Premature Discharge of Battery594
- Appropriate Device Problem Term/Code Not Available430
- Mechanical Problem279
- Failure to Interrogate201
In plain English: what the newest reports describe Written by AI
Several reports describe infection at the device pocket, the space under the skin where the pulse generator sits. In some cases the skin eroded or was red, painful or discharging pus, and the pacemaker was removed; one report mentions a recent joint replacement, and one report notes the patient had a fever and a bloodstream infection (bacteremia) caused by MRSA, a drug-resistant bacterium.
7 reports: 2017865-2026-22930, 3008973940-2026-08050, 3008973940-2026-08054, 3004209178-2026-18542, 3008973940-2026-08071, 3008973940-2026-08014, 3008973940-2026-08023
Several reports describe the lead's screw or setscrew, which holds the lead in the pulse generator, failing during implant. The lead could not be secured or released, so the generator was set aside and replaced; in one report the lead could not be removed and had to be cut.
5 reports: 9614453-2026-04438, 3008973940-2026-08015, 3008973940-2026-08019, 9614453-2026-04440, 2182208-2026-62352
Two reports describe the right atrial lead sensing heartbeats incorrectly, including signals arriving during its blanking period (a short window when the device ignores signals) and far-field signals from the ventricle. This led to pacing at times when it was not needed, and the lead remains in use.
2 reports: 3008973940-2026-08055, 3008973940-2026-08059
Several reports describe false impedance warnings, meaning the device signaled a lead connection problem that was not confirmed, along with noise and oversensing on the right ventricular lead in one case. The generator and lead remain in use, and no patient harm was reported.
2 reports: 3008973940-2026-08066, 3008973940-2026-08010
One report describes strong chest pain and abdominal pain during ventricular pacing, with abdominal muscle spasms and diaphragm twitching seen once. The patient was hospitalized, and the generator remains in use.
Report: 9614453-2026-04439
Written by AI from the narratives of the 20 newest reports that have one, received Oct 29, 2026 to Oct 30, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 341in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 91registered establishments that list this product code
Watch NVZ in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get NVZ every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in NVZ. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:NVZRetrieved Nov 29, 2026, 5:41 AM UTC. FDA data updated Nov 16, 2026.
- device/pma · search product_code:NVZRetrieved Nov 29, 2026, 5:41 AM UTC. FDA data updated Nov 16, 2026.
- device/pma · search product_code:NVZ · count decision_dateRetrieved Nov 29, 2026, 5:41 AM UTC. FDA data updated Nov 16, 2026.
- device/pma · search product_code:NVZ · count pma_numberRetrieved Nov 29, 2026, 5:41 AM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:NVZRetrieved Nov 29, 2026, 5:41 AM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:NVZ · count event_date_initiatedRetrieved Nov 29, 2026, 5:41 AM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:NVZ · count res_event_numberRetrieved Nov 29, 2026, 5:41 AM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-0302-2027" "Z-0303-2027" "Z-0304-2027" "Z-2168-2025" "Z-2169-2025")Retrieved Nov 29, 2026, 5:41 AM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:NVZ AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 5:41 AM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:NVZ AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 5:41 AM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:NVZ AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 5:41 AM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:NVZRetrieved Nov 29, 2026, 5:41 AM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:NVZRetrieved Nov 29, 2026, 5:41 AM UTC. FDA data updated Nov 16, 2026.