Weekly archive
Week of September 16, 2024
The FDA cleared 57 510(k)s, granted no De Novo requests and approved no original PMAs. 29 device recalls were posted, 4 of them Class I. Every 510(k), De Novo and PMA decision dated September 16 to 22, 2024, and every device recall the FDA posted those days.
- 510(k) clearances
- 57median review 170 days
- De Novo grants
- 0
- Original PMA approvals
- 0plus 38 supplements
- Recalls posted
- 2964 products, 4 Class I
Recalls posted
Device recalls the FDA posted September 16 to 22, 2024, Class I first. Class I is the most serious: a reasonable chance the product causes serious harm or death. The class comes from the FDA’s weekly enforcement report, so the newest recalls can show “Class pending” for a few days.
| Recall and class | Firm and reason | Product code |
|---|---|---|
| Z-3090-2024Class I | Smiths Medical ASD Inc.Smiths Medical became aware of an issue related to a potential for the patient outlet connector to loosen/detach from the paraPAC Plus P300 and P310 ventilators impacting the active ventilation function. | BTL2 products |
| Z-3134-2024Class I | Datascope Corp.Firm has developed a software correction (Revision D.00) outside the US to mitigate product problems that pose a risk of hemodynamic instability and failure of transmission of clinical data in Cardiosave Hybrid and Card… | DSP2 products |
| Z-3136-2024Class I | Datascope Corp.Firm has developed a software correction (Revision D.01) for outside the US to mitigate product problems that pose a risk of hemodynamic instability and failure of transmission of clinical data in Cardiosave Hybrid and… | DSP2 products |
| Z-3082-2024Class I | Baxter Healthcare CorporationBaxter Healthcare Corporation has received increased customer reports of particulate matter in the Automated Compounding Device Inlets (disposable inlet), Product Codes H938173, H938174, H938175, H938176, used with the… | LHI4 products |
| Z-3080-2024Class II | Wipro GE Healthcare Private Ltd.GE HealthCare has become aware that complete electrical safety testing was not conducted during manufacturing of certain Giraffe and Panda iRes Warmers. There are multiple electrical safety protections included in the p… | FMT2 products |
| Z-3187-2024Class II | Roche Diagnostics Operations, Inc.Creatine Kinase (CK) Assay Abnormal calibrations on the cobas c 311/501/502 and COBAS INTEGRA 400 plus Analyzers may lead to erroneous results and misdiagnosis. | JHS1 product |
| Z-3188-2024Class II | CooperSurgical, Inc.During the manufacturing process the devices were laser-marked with the incorrect part number. | HHM2 products |
| Z-3190-2024Class II | Merit Medical Systems, Inc.7F sheath introducers labeled as 7.5F | DRE OEQ2 products |
| Z-3192-2024Class II | Argon Medical Devices, IncHoles have been identified in biopsy instrument packaging trays compromising the sterile barrier, which may lead to cross contamination, or the introduction of micro-organisms into the access site, which could lead to a… | KNW1 product |
| Z-3193-2024Class II | Philips North America LlcIntegrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May lead… | LNH13 products |
| Z-3206-2024Class II | Bard Peripheral Vascular IncDue to possible packaging damage (indentations/cuts) cause by manufacturing process. As a result of the packaging damage, may cause the sterile barrier to be compromised. | FCG1 product |
| Z-3171-2024Class II | Smith & Nephew, Inc.Smith & Nephew, Inc., has initiated a field action to voluntarily remove two batches of K-Wires due to a packaging error. A complaint was received indicating that a package contained K-WIRE .062X9 2PT DM wires instead o… | HTY2 products |
| Z-3173-2024Class II | B Braun Medical IncIncomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion during prograde rinseback procedures. If the patient connec… | FJK7 products |
| Z-3180-2024Class II | Beckman Coulter, Inc.Beckman Coulter is recalling the DxI 9000 Access Immunoassay Analyzer due to undersized Pneumatics Tubing in the reagent storage module causing air leaks, compressor failures or sample processing errors. Four complaints… | JJE1 product |
| Z-3181-2024Class II | Safeguard US Operating LLCIncorrect expiration date. | IMD3 products |
| Z-3184-2024Class II | Cypress Medical Products LLCDue to a classification error in our inventory management system, these medical devices were shipped to customers that did not have the requisite license to purchase such devices. | GTY GTZ2 products |
| Z-3186-2024Class II | Unomedical A/SSpecific VariSoftTM Infusion Sets were packaged using incorrect secondary packaging (boxes) branded "Trusteel Infusion Set". | FPA1 product |
| Z-3165-2024Class II | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Potential Detachment of the Cable Hose Carrier (due to broken plastic and/or bolts) which May Result in Parts Falling and/or Part of the Cable Hose Dropping, and result in injury. | KPR OWB4 products |
| Z-3169-2024Class II | Siemens Healthcare Diagnostics, Inc.Atellica CH Microalbumin_2 (ALB_2) and the ADVIA Chemistry Microalbumin_2 (ALB_2) are not meeting the High-Dose Hook Effect claim/Prozone Effect claim as stated in the Instructions for Use (IFU) on the Atellica CH, Atel… | JIR2 products |
| Z-3157-2024Class II | MICROVENTION INC.Due to Retrieval Device product labels stating the incorrect expiration dates that (5 years instead of 3 years) which extends beyond their true expiration date. | NRY1 product |
| Z-3159-2024Class II | ROi CPS LLCMedical convenience kits were distributed with Merit syringes (manufactured by Jiangsu Shenli Medical Production) which were subsequently recalled. | OEQ1 product |
| Z-3160-2024Class II | GE HealthcareFootprint Client (ZFP) versions v6.0 SP9.x and SP10.x where the latest addended report is not shown by default to the user. The issue occurs when the addendum is created on the same day as the original report but at a d… | LLZ1 product |
| Z-3162-2024Class II | Btt Health GmbhSome power supply units in Austria and Germany had a voltage drop in the millisecond range, which led to a cancellation of the treatment and subsequent restart of the medical device. The treatment is only restarted by a… | BXB1 product |
| Z-3163-2024Class II | Innovasis, IncDue to interfacing issues between the implant trial and inserter of the lumbar intervertebral fusion device. | MAX1 product |
| Z-3164-2024Class II | UNIMAX MEDICAL SYSTEMS INCThe tube may fall into the surgical site during the grasping process. | GCJ1 product |
| Z-3154-2024Class II | Cypress Medical Products LLCThe syringes that were inadvertently shipped from the Distribution Center. This product was on quarantine because FDA has stated that these syringes lack FDA clearance and were inadvertently shipped due to an internal o… | FMF1 product |
| Z-3155-2024Class II | Tornier, IncOne lot of Perform Humeral Nucleus may exceed the specification for bacterial endotoxin load (max 20 EU/device). | PKC1 product |
| Z-3156-2024Class II | Mazor Robotics LtdSoftware anomalies which can result in two separate issues: A lagging screen display during a navigated procedure; or the potential for a mismatch between the navigated screw and the screw information presented on the s… | OLO1 product |
| Z-3161-2024Class III | Carbon Medical Technologies, Inc.This recall has been initiated due to a failure to provide an MRI Insert that is to be provided to the patient. | IYE1 product |
De Novo grants
New device types the FDA classified through the De Novo pathway. Each one creates a product code and becomes a possible predicate for later 510(k)s.
No De Novo decisions this week.
Original PMA approvals
Original premarket approvals for Class III devices. The FDA also decided 38 PMA supplements this week, which change devices already approved.
No original PMAs were approved this week.
510(k) clearances
57 510(k) clearances by review panel. The median took 170 calendar days from receipt to decision; the fastest, K242650, 16 days; the longest, K231383, 495 days.
Calendar days include the time a submission waits on the company, so they run longer than the FDA’s 90-day review goal. How to search the 510(k) database
- Orthopedic9
- Dental7
- General Hospital7
- Radiology7
- Cardiovascular5
- Physical Medicine5
- General, Plastic Surgery4
- Neurology4
- Anesthesiology2
- Gastroenterology, Urology2
- Panel not stated2
- Clinical Chemistry1
- Ear, Nose, Throat1
- Obstetrics/Gynecology1
Every clearance
| Number | Device and applicant | Product code | Decision and review time |
|---|---|---|---|
| K240293 | STSTechbopulm, Ltd. | BZCAnesthesiology | Sep 17, 2024229 days |
| K241465 | Midas Flowmeter; eAVSParker Hannifin Corporation | BZRAnesthesiology | Sep 19, 2024119 days |
| K232670 | HiCardi+ H100MEZOO Co., Ltd. | DSHCardiovascular | Sep 20, 2024385 days |
| K240901 | StethophoneSparrow Acoustics, Inc. | DQDCardiovascular | Sep 19, 2024170 days |
| K241230 | Super Sheath Introducer Sheath S-3N5/1005 S-3N7/1006 S-3W5/1007 S-3W7/1008Togo Medikit Co., Ltd. | DYBCardiovascular | Sep 17, 2024138 days |
| K241784 | Arrow® Nitinol WireTeleflex Medical | DQXCardiovascular | Sep 18, 202490 days |
| K242504 | Millipede 088 Access Catheter; Millipede 070 Aspiration Catheter; Perfuze Aspiration Tube SetPerfuze, Ltd. | NRYCardiovascular | Sep 20, 202429 days |
| K241335 | Eversense 365 Continuous Glucose Monitoring (CGM) SystemSenseonics, Incorporated | SBAClinical Chemistry | Sep 16, 2024126 days |
| K233064 | Copra SupremeWhitepeaks Dental Solutions GmbH | EIHDental | Sep 18, 2024358 days |
| K233721 | KLS Martin Drill-Free MMF ScrewKLS-Martin L.P. | DZLDental | Sep 17, 2024301 days |
| K234089 | DNA ApplianceVivos Therapeutics | LRKDental | Sep 16, 2024269 days |
| K234142 | TiGEN Abutment, PMMA Abutment and Scan Healing AbutmentMegagen Implant Co., Ltd. | NHADental | Sep 18, 2024264 days |
| K240183 | Star E900 Handpiece SeriesDentalez, Inc., Stardental Division | EGSDental | Sep 19, 2024240 days |
| K240660 | Signum composite flow; Signum universal bond; Signum metal bond; Signum composite; Signum matrix; Signum liquid; Signum cre-activeKulzer, LLC | EBFDental | Sep 17, 2024193 days |
| K241880 | Air Powered Tooth Polishing System (Perio-Mate, Perio Mate Nozzle Tip)Nakanishi, Inc. | EFBDental | Sep 20, 202484 days |
| K234096 | MonoStereoMedicaltek Co., Ltd. | EOBEar, Nose, Throat | Sep 19, 2024268 days |
| K240274 | PADLOCK CLIP EFTR Kit (00713229)STERIS Corporation | PKLGastroenterology, Urology | Sep 17, 2024230 days |
| K241424 | InnoCare Specialty Foley CatheterInnocare Urologics, LLC | EZLGastroenterology, Urology | Sep 18, 2024121 days |
| K231383 | Terragene® Bionova® Self-Contained Biological Indicator (BT95), Terragene® Bionova® Self-Contained Biological Indicator (BT96), Terragene® Chemdye® Chemical Process Indicator (CD42), Terragene® Cintape® Chemical Process Indicator Tape (CT40)Terragene S.A. | FRCGeneral Hospital | Sep 18, 2024495 days |
| K231893 | LOWTEM Crystal 120Lowtem Co., Ltd. | MLRGeneral Hospital | Sep 18, 2024448 days |
| K234050 | HALYARD® ONE-STEP® Sterilization Wrap, HALYARD® QUICK CHECK® Sterilization Wrap, HALYARD® SEQUENTIAL Sterilization Wrap and HALYARD® SMART-FOLD® Sterilization WrapO&M Halyard, Inc. | FRGGeneral Hospital | Sep 16, 2024270 days |
| K234129 | Ecomed Disposable Administration SetsEcomed Solutions, LLC | FPAGeneral Hospital | Sep 20, 2024267 days |
| K234132 | Medline Reusable Sterilization WrappersMedline Industries, LP | FRGGeneral Hospital | Sep 17, 2024264 days |
| K241123 | Perpak™ Sterilization Tyvek PouchKm Corp. | FRGGeneral Hospital | Sep 18, 2024148 days |
| K241821 | Luer Lock Syringe with Safety Needle; Luer Lock Syringe with Exchangeable Needle; Luer Lock Syringe with Blunt Fill NeedleSol-Millennium Medical, Inc. | FMFGeneral Hospital | Sep 20, 202488 days |
| K240604 | MTONE EVO (MTONE EVO)M&T S.R.L. | GEXGeneral, Plastic Surgery | Sep 20, 2024200 days |
| K241762 | OneMarkView Point Medical | NEUGeneral, Plastic Surgery | Sep 16, 202488 days |
| K241771 | Liftendo; Liftendo 2.0Medical San Indústria DE Equipamentos Médicos Ltda. | GEXGeneral, Plastic Surgery | Sep 16, 202488 days |
| K241825 | Microwave Ablation AntennasCanyon Medical, Inc. | NEYGeneral, Plastic Surgery | Sep 20, 202488 days |
| K232938 | Traus SSG30 Surgical SystemSaeshin Precision Co., Ltd. | HBCNeurology | Sep 16, 2024362 days |
| K233153 | Traus SSG10 Surgical SystemSaeshin Precision Co., Ltd. | HBCNeurology | Sep 20, 2024359 days |
| K240420 | NeuroField Analysis SuiteNeurofield, Inc. | OLUNeurology | Sep 20, 2024220 days |
| K241893 | CROSSNAV Navigation Enabled Instruments and UNAS Navigation ArraysMedos International SARL | OLONeurology | Sep 16, 202480 days |
| K242162 | Coconut HybridGood Clean Love, Inc. | NUCObstetrics/Gynecology | Sep 19, 202457 days |
| K240061 | TRIGEN MAX Tibial Nail System; INTERTAN MAX Hip Fracture Nail SystemSmith & Nephew, Inc. | JDSOrthopedic | Sep 20, 2024255 days |
| K240250 | 3D Printed PEEK Interbody SystemNvision Biomedical Technologies, Inc. | MAXOrthopedic | Sep 17, 2024231 days |
| K241332 | aprevo® anterior and lateral lumbar interbody fusion device; aprevo® anterior lumbar interbody fusion devices with interfixation; aprevo® transforaminal lumbar interbody fusion deviceCarlsmed, Inc. | MAXOrthopedic | Sep 20, 2024133 days |
| K241709 | Pectus Blu Support Bar SystemBiomet Microfixation | HRSOrthopedic | Sep 19, 202498 days |
| K241769 | Response Ortho Metaphyseal Hinge Fixator SystemResponse Ortho Solutions, LLC | KTTOrthopedic | Sep 18, 202490 days |
| K241775 | XeliteMed VertehighFix High Viscosity Spinal Bone CementXelite Biomed , Ltd. | NDNOrthopedic | Sep 18, 202490 days |
| K241942 | TriMed Bridge PlatesTriMed, Inc. | HRSOrthopedic | Sep 17, 202477 days |
| K242147 | Calibrate LTX Interbody System; Calibrate NanoTec LTX Interbody SystemAlphatec Spine, Inc. | MAXOrthopedic | Sep 20, 202459 days |
| K242509 | HAnano InterFuse(R) Modular InterbodyNext Orthosurgical | MAXOrthopedic | Sep 19, 202428 days |
| K240722 | Derma-R Cream (Derma-R Cream)Raya Pharmaceuticals, LLC | FROPanel not stated | Sep 20, 2024186 days |
| K242650 | zLOCK Lumbar Facet Fixation SystemZygofix , Ltd. | MRWPanel not stated | Sep 20, 202416 days |
| K240373 | Air compression Leg Massager (MM0520, MM0521, MM0522)Zhangzhou Easepal Medical Science and Technology Co., Ltd. | IRPPhysical Medicine | Sep 17, 2024223 days |
| K240638 | Power Wheelchair (YH-E7007)Ningbo Youhuan Automation Technology Co., Ltd. | ITIPhysical Medicine | Sep 16, 2024194 days |
| K240689 | Robin's EggTelebrands Corporation | OAPPhysical Medicine | Sep 17, 2024188 days |
| K241379 | Powered Wheelchair (NXN20-205, NXN20-205M)Zhejiang Nysin Medical Co., Ltd. | ITIPhysical Medicine | Sep 16, 2024124 days |
| K242186 | Neo Sculptor (OF-NEO001)Ofan Intelligent Technology (Guangzhou) Co. , Ltd. | NGXPhysical Medicine | Sep 17, 202454 days |
| K234108 | ERI Portable X-ray System (CVX-air, CVX-lite, CVX-E)Energy Resources International Co., Ltd. | IZLRadiology | Sep 20, 2024268 days |
| K234153 | EdgeLife Handheld Wireless Ultrasound System (E8200, E8220)Edgelife Technologies, LLC | IYNRadiology | Sep 18, 2024264 days |
| K240036 | PIUR tUS InfinityPiur Imaging GmbH | QIHRadiology | Sep 20, 2024259 days |
| K241201 | TEX20/TEX20 Pro/TEX20S/TEX20T/TEX20 Exp/TEX20 Elite/TEX10/TEX10 Pro/TEX10S/TEX10T/TEX10 Exp/TEX10 Elite/TE X/TE X Lite/Anesus TEX20 Pro/Eagus TEX20/Eagus TEX20T/Eagus TEX21/Eagus TEX21T/Eros/TE X Plus/Anesus TEX10 Pro/TEX9U/TEX9M/Eagus TEX11/Eagus TEX11T/EagusTEX10/Eagus TEX10T/Eagus TEX/Ares Diagnostic Ultrasound SystemShenzhen Mindray Bio-Medical Electronics Co., Ltd. | IYNRadiology | Sep 16, 2024139 days |
| K242167 | Sterile and Non-Sterile Ultrasonic Coupling AgentAnhui Deepblue Medical Technology Co., Ltd. | MUIRadiology | Sep 19, 202457 days |
| K242223 | SuperSonic HepaVuSupersonic Imagine | IYNRadiology | Sep 20, 202453 days |
| K242478 | GF85 (models GF85-3P, GF85-SP)Samsung Electronics Co., Ltd. | KPRRadiology | Sep 19, 202429 days |
Only see the weeks that matter to you
DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and every day it checks openFDA and tells you about new clearances, recalls and MAUDE reports for them. It also puts all of this beside the FDA’s own record pages.
Only your product codes, every Monday
One short email after each openFDA update with just the decisions and recalls in up to three product codes you choose. Free.
Earlier weeks
Each week keeps its own page.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Decisions arrive in weekly updates; check the FDA record before you rely on a figure.
- device/510k · search decision_date:[20240916 TO 20240922]Retrieved Nov 29, 2026, 11:23 PM UTC. FDA data updated Nov 16, 2026.
- device/pma · search decision_date:[20240916 TO 20240922]Retrieved Nov 29, 2026, 11:23 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search event_date_posted:[20240916 TO 20240922]Retrieved Nov 29, 2026, 11:23 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-3080-2024" "Z-3081-2024" "Z-3090-2024" "Z-3091-2024" "Z-3134-2024" "Z-3135-2024" "Z-3136-2024" "Z-3137-2024" "Z-3187-2024" "Z-3188-2024" "Z-3189-2024" "Z-3190-2024" "Z-3191-2024" "Z-3192-2024" "Z-3193-2024" "Z-3194-2024" "Z-3195-2024" "Z-3196-2024" "Z-3197-2024" "Z-3198-2024" "Z-3199-2024" "Z-3200-2024" "Z-3201-2024" "Z-3202-2024" "Z-3203-2024" "Z-3204-2024" "Z-3205-2024" "Z-3206-2024" "Z-3082-2024" "Z-3083-2024" "Z-3084-2024" "Z-3085-2024" "Z-3171-2024" "Z-3172-2024" "Z-3173-2024" "Z-3174-2024" "Z-3175-2024" "Z-3176-2024" "Z-3177-2024" "Z-3178-2024" "Z-3179-2024" "Z-3180-2024" "Z-3181-2024" "Z-3182-2024" "Z-3183-2024" "Z-3184-2024" "Z-3185-2024" "Z-3186-2024" "Z-3165-2024" "Z-3166-2024")Retrieved Nov 29, 2026, 11:23 PM UTC. FDA data updated Nov 18, 2026.
- device/enforcement · search recall_number:("Z-3167-2024" "Z-3168-2024" "Z-3169-2024" "Z-3170-2024" "Z-3157-2024" "Z-3159-2024" "Z-3160-2024" "Z-3161-2024" "Z-3162-2024" "Z-3163-2024" "Z-3164-2024" "Z-3154-2024" "Z-3155-2024" "Z-3156-2024")Retrieved Nov 29, 2026, 11:23 PM UTC. FDA data updated Nov 18, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil