Weekly archive
Week of April 27, 2026
The FDA cleared 69 510(k)s, granted 2 De Novo requests and approved no original PMAs. 25 device recalls were posted, 5 of them Class I. Every 510(k), De Novo and PMA decision dated April 27 to May 3, 2026, and every device recall the FDA posted those days.
Week beforeWeek afterRSS feedThis week as lead lists
- 510(k) clearances
- 69median review 146 days
- De Novo grants
- 2
- Original PMA approvals
- 0plus 47 supplements
- Recalls posted
- 25109 products, 5 Class I
Recalls posted
Device recalls the FDA posted April 27 to May 3, 2026, Class I first. Class I is the most serious: a reasonable chance the product causes serious harm or death. The class comes from the FDA’s weekly enforcement report, so the newest recalls can show “Class pending” for a few days.
| Recall and class | Firm and reason | Product code |
|---|---|---|
| Z-1855-2026Class I | Intuitive Surgical, Inc.Reloads for curved-tipped stapler may potentially produce an incomplete staple line. | NAY1 product |
| Z-1934-2026Class I | Cook IncorporatedMarker bands may be at an increased risk of cracking/breakage. Potential adverse events that may result include increased procedural time, harms associated with device fragmentation/separation, and vessel injury. In a w… | DQO3 products |
| Z-1942-2026Class I | Ventec Life Systems, Inc.Respiratory system intended to provide continuous/intermittent ventilator support, may not have been evaluated under all intended high-pressure conditions during production testing, so an oxygen leak condition could occ… | NOU1 product |
| Z-1970-2026Class I | AVID Medical, Inc.Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold. | LRO OES OEZ10 products |
| Z-1947-2026Class I | North American Rescue LLC.Kits contain the McKesson TRUE Metrix PRO Professional Monitoring Blood Glucose System that has been recalled for an issue with the software design of the E-5 Error Code where the meter displays an E-5 error code for a… | LRR1 product |
| Z-2046-2026Class II | Intuitive Surgical, Inc.Due to manipulator containing screws with in the arm sub-assembly that my be susceptible to breaking. | NAY1 product |
| Z-2048-2026Class II | Boston Scientific CorporationFor some patients upgraded from a LUX-Dx M301 device to a new Model M302 or M312, the new ICM device is not collecting PVC Burden data (both M302/M312) or monitoring for Bradyarrhythmia and Pause (M312 only). In additio… | MXD2 products |
| Z-2050-2026Class II | Centinel Spine, Inc.Product labeling mix up. The prodisc C SK U.S. Implant Extra Large 6mm product was labeled as a 5mm and prodisc C SK U.S. Implant Extra Large 5mm product was labeled as a 6mm. | MJO2 products |
| Z-2052-2026Class II | Spacelabs Healthcare, Inc.Due to a manufacturing issue (i.e. malfunctioning of electrical Leakage tester). | DSI1 product |
| Z-2062-2026Class II | Erbe Medical, LLCDevices providing sterile water/CO2/air to endoscopes were manufactured with a distal irrigation segment connector configuration that may be more susceptible to unintended water flow from distal tip under unsupported th… | FEQ OCX3 products |
| Z-2008-2026Class II | Medline Industries, LPDuring an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin. | CAI FDE FMH31 products |
| Z-2039-2026Class II | Medline Industries, LPMedline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing Cardinal Health Webcol Large Alcohol Prep Pads. Cardinal Health has deemed Webcol Large Alcohol Prep Pads to be non-s… | CBT LRO LRP7 products |
| Z-1982-2026Class II | Becton Dickinson & Co.BD identified an issue affecting certain lots of BD GasPak EZ CO2 and BD GasPak EZ Campy Gas Generating Pouch Systems, where gas generating sachets may produce carbon dioxide levels below specification. | KZJ2 products |
| Z-1984-2026Class II | Immuno-Mycologics, IncNeutralization Buffer may contain contaminants | PPM1 product |
| Z-1985-2026Class II | DEPUY (IRELAND)The external sterile packaging was found adhered to the internal sterile packaging, potentially compromising sterility. | KRO9 products |
| Z-1994-2026Class II | Straumann USA LLCIncludes an incorrect screw seat interface. | NHA2 products |
| Z-1996-2026Class II | Beta Bionics, Inc.insulin pump software versions 1.4.2, and 1.4.3 issue with Dexcom Continuous Glucose Monitoring (CGM) G7 sensor may result in delayed CGM reading then switch to non-CGM dosing, leading to only basal/requested meal doses… | QFG1 product |
| Z-1999-2026Class II | Abbott Point Of Care Inc.Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cartridge lots in the U.S. marketplace may report higher than e… | CHL3 products |
| Z-2002-2026Class II | Merit Medical Systems, Inc.catheter may experience resistance when being advanced over the guidewire | OEQ3 products |
| Z-2005-2026Class II | Siemens Healthcare Diagnostics IncUse of the affected product can result in discrepant low pH and measured Total Carbon Dioxide (mTCO2) results and corresponding calculated values in arterial, venous, and capillary patient samples. An erroneously low pH… | CHL1 product |
| Z-2006-2026Class II | Elekta, Inc.Locking mechanisms may not properly secure to the Arc system, resulting in shifting of coordinates during procedures. | HAW1 product |
| Z-1980-2026Class II | Siemens Healthcare Diagnostics, Inc.Siemens Healthcare Diagnostics is recalling certain lots of Dimension Creatinine Flex reagent cartridge (Dimension CRE2) because a potential imprecision in the Dimension CRE2 quality control (QC) and patient sample resu… | CGX1 product |
| Z-1950-2026Class II | Advanced Bionics, LLCThe non-tamperproof battery door supplied with the sound processor fails to conform to IEC 60601-1-11:2015 by not requiring the use of a tool to replace the button/coin cell battery. | MCM1 product |
| Z-1951-2026Class II | Philips North AmericaThe potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE… | LNH19 products |
| Z-1997-2026Class III | Heraeus Medical GmbH (Dental Division)Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the… | JDZ2 products |
De Novo grants
New device types the FDA classified through the De Novo pathway. Each one creates a product code and becomes a possible predicate for later 510(k)s.
Original PMA approvals
Original premarket approvals for Class III devices. The FDA also decided 47 PMA supplements this week, which change devices already approved.
No original PMAs were approved this week.
510(k) clearances
69 510(k) clearances by review panel. The median took 146 calendar days from receipt to decision; the fastest, K261404, 1 day; the longest, K250680, 420 days.
Calendar days include the time a submission waits on the company, so they run longer than the FDA’s 90-day review goal. How to search the 510(k) database
- Orthopedic13
- Cardiovascular8
- Anesthesiology6
- Dental6
- Radiology5
- Gastroenterology, Urology4
- General, Plastic Surgery4
- Obstetrics/Gynecology4
- Clinical Chemistry3
- Microbiology3
- Ophthalmic3
- Physical Medicine3
- General Hospital2
- Neurology2
- Clinical Toxicology1
- Hematology1
- Panel not stated1
Every clearance
| Number | Device and applicant | Product code | Decision and review time |
|---|---|---|---|
| K251583 | Ambu® Virobac II® Exhalation FilterAmbu A/S | CAHAnesthesiology | Apr 30, 2026342 days |
| K252628 | CASSIEWesper, Inc. | MNRAnesthesiology | Apr 27, 2026250 days |
| K252849 | AffloVest® Mobile Airway Clearance TherapyTactile Medical | BYIAnesthesiology | Apr 30, 2026234 days |
| K253030 | Capnostream™35 Portable Respiratory Monitor (PM35MN)Covidien, LLC | CCKAnesthesiology | May 1, 2026224 days |
| K253166 | F6S Full Face Mask (F6S); P6S Nasal Pillows Interface (P6S)BMC Medical Co., Ltd. | BZDAnesthesiology | Apr 29, 2026215 days |
| K260910 | EndoTool IV Cloud 1.0Glooko, Inc. | NDCAnesthesiology | May 1, 202643 days |
| K252445 | Fusion Bioline Vascular GraftMaquet Cardiovascular, LLC | DSYCardiovascular | Apr 27, 2026266 days |
| K252501 | Arm-type Fully Automatic Digital Blood Pressure Monitor (DBP-62E3B, DBP-62E2B, DBP-62E1B, DBP-61E3, DBP-61E2, DBP-61E1, DBP-61D7G, DBP-6281L, DBP-6282L, DBP-6285L)Joytech Healthcare Co. , Ltd. | DXNCardiovascular | Apr 28, 2026263 days |
| K253436 | c-med0 alphaCosinuss GmbH | MWICardiovascular | Apr 30, 2026211 days |
| K253579 | Aristotle 14 Guidewire; Aristotle 18 Guidewire; Aristotle 24 Guidewire; Zoom Wire 14 GuidewireScientia Vascular, Inc. | MOFCardiovascular | Apr 29, 2026163 days |
| K253887 | Nasal Alar SpO2 Sensor (989803205381, 989803205391); Nasal Alar Sp02 Sensor (989803205401)Philips Medizin Systeme | DQACardiovascular | May 1, 2026148 days |
| K260260 | Automatic Digital Blood Pressure Monitor (LD-587, LD-528, LD-562, LD-581, LD-521, LD-527, LD-572 & LD-569)Honsun (Nantong) Co., Ltd. | DXNCardiovascular | Apr 28, 202690 days |
| K260606 | Echo Large Bore Introducer SheathEcho Medical, LLC | DYBCardiovascular | Apr 29, 202664 days |
| K261074 | BIOMONITOR IV (471155)Biotronik, Inc. | MXDCardiovascular | May 1, 202630 days |
| K252475 | ABL90 FLEX PLUS SystemRadiometer Medicals Aps | JGSClinical Chemistry | May 1, 2026267 days |
| K252488 | ABL90 FLEX PLUS SystemRadiometer Medicals Aps | CGAClinical Chemistry | May 1, 2026266 days |
| K253976 | iLet ACE PumpBeta Bionics, Inc. | QFGClinical Chemistry | Apr 30, 2026139 days |
| K260065 | SAFElife T-Dip Multi-Drug Urine Test Panel; SAFElife T-Dip Multi-Drug Urine Test Panel DxGuangzhou Wondfo Biotech Co., Ltd. | NFTClinical Toxicology | Apr 28, 2026109 days |
| K252073 | VOCO Profluorid Varnish + BioMinVoco GmbH | LBHDental | May 1, 2026303 days |
| K252464 | Surgical Microscope (OMS2360, OMS2380)Zumax Medical Co., Ltd. | NBLDental | Apr 28, 2026265 days |
| K260819 | Any-Com SealMediclus Co., Ltd. | EBCDental | May 1, 202649 days |
| K261356 | Dura-CrownHangzhou SHINING3D Dental Technology Co., Ltd. | EBFDental | Apr 27, 20263 days |
| K261357 | Grandio disc multicolorVoco GmbH | EBFDental | Apr 27, 20263 days |
| K261404 | UltraEZUltradent Products, Inc. | LBHDental | Apr 30, 20261 day |
| K252831 | Pelvic Floor Treatment Device (DLR-8920, DLR-8921, DLR-8922, DLR-8923, DLR-8924, DLR-8925, DLR-8926, DLR-8927)Dolanvy (Suzhou) Medical Technology Co., Ltd. | KPIGastroenterology, Urology | Apr 28, 2026235 days |
| K252858 | 4K UHD Endoscopy Camera System (KMS-4K-2088, KMS-4K-2188, KMS-4K-2288)Kms Medical Technology Co., Ltd. | GCJGastroenterology, Urology | Apr 30, 2026233 days |
| K260314 | Ancora-SBAspero Medical, Inc. | FDAGastroenterology, Urology | May 1, 202691 days |
| K261086 | Break WaveSonomotion, Inc. | LNSGastroenterology, Urology | Apr 29, 202628 days |
| K250680 | Bayesian Health Sepsis Flagging DeviceBayesian Health, Inc. | SAKGeneral Hospital | Apr 30, 2026420 days |
| K252471 | Paptizer 360LiViliti Health Products Corporation | QXQGeneral Hospital | Apr 29, 2026266 days |
| K253243 | Cardinal Health™ Poly Reinforced Surgical GownCardinal Health 200, LLC | FYAGeneral, Plastic Surgery | Apr 28, 2026211 days |
| K254234 | Phototherapy System (DUV-COMBO)Choyang Medics Co., Ltd. | FTCGeneral, Plastic Surgery | May 1, 2026123 days |
| K261045 | Verifine Safety LancetsPromisemed Hangzhou Meditech Co., Ltd. | FMKGeneral, Plastic Surgery | May 1, 202631 days |
| K261094 | YellowStarAsclepion Laser Technologies GmbH | GEXGeneral, Plastic Surgery | Apr 29, 202627 days |
| K253658 | STA Satellite Max®Diagnostica Stago, Inc. | JPAHematology | Apr 27, 2026158 days |
| K252424 | Anti-HCV NextAbbott Laboratories | MZOMicrobiology | Apr 28, 2026270 days |
| K254139 | BioCode® Respiratory Pathogen Panel (RPP)Applied BioCode, Inc. | OCCMicrobiology | May 1, 2026130 days |
| K260342 | AllTest Strep A Rapid TestHangzhou AllTest Biotech Co., Ltd. | GTYMicrobiology | Apr 30, 202687 days |
| K252439 | Tutamen Self Adhesive ElectrodesDongguan Tutamen Metalwork Co., Ltd. | GXYNeurology | May 1, 2026270 days |
| K253663 | Zeta Navigation SystemZeta Surgical, Inc. | HAWNeurology | May 1, 2026162 days |
| K252748 | Pelvic Floor Exerciser (PF01)Hunan Accurate Bio-Medical Technology Co., Ltd. | HIRObstetrics/Gynecology | Apr 27, 2026241 days |
| K253914 | Momcozy Wearable Breast Pump (Model: BP334, BP334-A, BP334-B, BP334-C, BP334-D, BP434)Shenzhen Root Innovation Technology Co., Ltd. | HGXObstetrics/Gynecology | Apr 29, 2026142 days |
| K260194 | Vacuum Pump (Model: AD-VPUM-500)Sheng Sheng Yi (Beijing) Technology Company Limited | MQGObstetrics/Gynecology | Apr 28, 202696 days |
| K260248 | Ultra-Fast WarmKitazato Corporation | MQLObstetrics/Gynecology | Apr 28, 202691 days |
| K252409 | TearCare MGX SystemSight Sciences, Inc. | ORZOphthalmic | Apr 27, 2026269 days |
| K252455 | FSYX Ocular Pressure Adjusting PumpBalance Ophthalmics, Inc. | QQJOphthalmic | Apr 30, 2026268 days |
| K253885 | PUSCON WATER BALANCE-1DAY TINT (omafilcon A) SOFT (HYDROPHILIC) CONTACT LENS; PUSCON WATER BALANCE-1DAY COLOR (omafilcon A) SOFT (HYDROPHILIC) CONTACT LENSESGEO MEDICAL CO., LTD. | LPLOphthalmic | Apr 29, 2026146 days |
| K252542 | LEO Spinal SystemZheJiang Decans Medical Devices Co., Ltd. | NKBOrthopedic | Apr 29, 2026260 days |
| K252635 | ArthroTAK Tendon Anchor KitArthroTAK, LLC | MBIOrthopedic | May 1, 2026254 days |
| K253326 | OxiplexFziomed, Inc. | QVLOrthopedic | May 1, 2026213 days |
| K253401 | SCRIPT™ Implant SystemGlobus Medical, Inc. | OVDOrthopedic | Apr 29, 2026211 days |
| K253940 | M.U.S.T. Pedicle Screw System - ExtensionMedacta International S.A. | NKBOrthopedic | Apr 28, 2026140 days |
| K253966 | PERLA® TL Posterior Thoraco-lumbar Fixation SystemSpineart SA | NKBOrthopedic | May 1, 2026141 days |
| K254077 | OSSIOfiber® Threaded Trimmable Fixation NailOSSIO , Ltd. | HWCOrthopedic | May 1, 2026134 days |
| K254154 | FX V135 EASYTECH® Shoulder SystemFX Shoulder Solutions, Inc. | PKCOrthopedic | May 1, 2026130 days |
| K260182 | Avenir® Müller Stem; Avenir Complete™ Hip SystemZimmer, Inc. | LZOOrthopedic | Apr 27, 202696 days |
| K260323 | OrthoPediatrics® Locking Cannulated Blade Plate SystemOrthoPediatrics Corp. | HRSOrthopedic | Apr 30, 202690 days |
| K260660 | SABER-C SystemElevation Spine | OVEOrthopedic | Apr 27, 202656 days |
| K261032 | NovoKnee (SteriKnee)NovoSource | JWHOrthopedic | Apr 27, 202628 days |
| K261067 | BEE PLIF CageNGMedical GmbH | MAXOrthopedic | Apr 30, 202629 days |
| K253017 | BloodSTOP iX Trauma Matrix (TM-iX-20)Lifescience Plus, Inc. | QSYPanel not stated | Apr 30, 2026223 days |
| K253064 | Air Pressure Therapy System (VU-IPC8M, VU-IPC6M, VU-IPC4X)Xiamen Weiyou Intelligent Technology Co., Ltd. | IRPPhysical Medicine | Apr 30, 2026220 days |
| K260076 | SAM 3.0 Multi-Hour Continuous Ultrasound DeviceZetrOZ Systems, LLC | PFWPhysical Medicine | Apr 28, 2026106 days |
| K260121 | Manual Wheelchair (7101L, 7102LHD)Foshan Nanhai Hongchen Medical Equipment Co., Ltd. | IORPhysical Medicine | Apr 28, 2026103 days |
| K252503 | Intelligent NRCanon, Inc. | OWBRadiology | Apr 30, 2026265 days |
| K253682 | DeepHealth ProstateAIQuantib B.V. | QDQRadiology | Apr 29, 2026159 days |
| K254184 | syngo Application SoftwareSiemens Medical Solutions USA, Inc. | QIHRadiology | Apr 29, 2026127 days |
| K254189 | Synapse 3D Base Tools (V7.2)Fujifilm Corporation | QIHRadiology | Apr 30, 2026128 days |
| K260589 | ES-SeriesE-Scopics | IYNRadiology | Apr 29, 202668 days |
Only see the weeks that matter to you
DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and every day it checks openFDA and tells you about new clearances, recalls and MAUDE reports for them. It also puts all of this beside the FDA’s own record pages.
Only your product codes, every Monday
One short email after each openFDA update with just the decisions and recalls in up to three product codes you choose. Free.
Earlier weeks
Each week keeps its own page.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Decisions arrive in weekly updates; check the FDA record before you rely on a figure.
- device/510k · search decision_date:[20260427 TO 20260503]Retrieved Nov 29, 2026, 9:23 PM UTC. FDA data updated Nov 16, 2026.
- device/pma · search decision_date:[20260427 TO 20260503]Retrieved Nov 29, 2026, 9:23 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search event_date_posted:[20260427 TO 20260503]Retrieved Nov 29, 2026, 9:23 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-1855-2026" "Z-1934-2026" "Z-1935-2026" "Z-1936-2026" "Z-1942-2026" "Z-1970-2026" "Z-1971-2026" "Z-1972-2026" "Z-1973-2026" "Z-1974-2026" "Z-1975-2026" "Z-1976-2026" "Z-1977-2026" "Z-1978-2026" "Z-1979-2026" "Z-2046-2026" "Z-2048-2026" "Z-2049-2026" "Z-2050-2026" "Z-2051-2026" "Z-2052-2026" "Z-2062-2026" "Z-2063-2026" "Z-2064-2026" "Z-2008-2026" "Z-2009-2026" "Z-2010-2026" "Z-2011-2026" "Z-2012-2026" "Z-2013-2026" "Z-2014-2026" "Z-2015-2026" "Z-2016-2026" "Z-2017-2026" "Z-2018-2026" "Z-2019-2026" "Z-2020-2026" "Z-2021-2026" "Z-2022-2026" "Z-2023-2026" "Z-2024-2026" "Z-2025-2026" "Z-2026-2026" "Z-2027-2026" "Z-2028-2026" "Z-2029-2026" "Z-2030-2026" "Z-2031-2026" "Z-2032-2026" "Z-2033-2026")Retrieved Nov 29, 2026, 9:23 PM UTC. FDA data updated Nov 18, 2026.
- device/enforcement · search recall_number:("Z-2034-2026" "Z-2035-2026" "Z-2036-2026" "Z-2037-2026" "Z-2038-2026" "Z-2039-2026" "Z-2040-2026" "Z-2041-2026" "Z-2042-2026" "Z-2043-2026" "Z-2044-2026" "Z-2045-2026" "Z-1947-2026" "Z-1982-2026" "Z-1983-2026" "Z-1984-2026" "Z-1985-2026" "Z-1986-2026" "Z-1987-2026" "Z-1988-2026" "Z-1989-2026" "Z-1990-2026" "Z-1991-2026" "Z-1992-2026" "Z-1993-2026" "Z-1994-2026" "Z-1995-2026" "Z-1996-2026" "Z-1997-2026" "Z-1998-2026" "Z-1999-2026" "Z-2000-2026" "Z-2001-2026" "Z-2002-2026" "Z-2003-2026" "Z-2004-2026" "Z-2005-2026" "Z-2006-2026" "Z-1980-2026" "Z-1950-2026" "Z-1951-2026" "Z-1952-2026" "Z-1953-2026" "Z-1954-2026" "Z-1955-2026" "Z-1956-2026" "Z-1957-2026" "Z-1958-2026" "Z-1959-2026" "Z-1960-2026")Retrieved Nov 29, 2026, 9:23 PM UTC. FDA data updated Nov 18, 2026.
- device/enforcement · search recall_number:("Z-1961-2026" "Z-1962-2026" "Z-1963-2026" "Z-1964-2026" "Z-1965-2026" "Z-1966-2026" "Z-1967-2026" "Z-1968-2026" "Z-1969-2026")Retrieved Nov 29, 2026, 9:23 PM UTC. FDA data updated Nov 18, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil