Product codes

FDA product code

CAH: Filter, Bacterial, Breathing-Circuit (FDA product code)

Follow CAH: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 868.5260
Review panel
General Hospital
Premarket route
510(k)
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
2022 in the last 12 months
Median review time
237 daysthe latest 25 decisions
Recall records
280 in the last 12 months
Adverse event reports
95last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code CAH per year, by decision date, 202 since 1976. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20171
20180
20192
20202
20212
20223
20233
20242
20250
2026 (so far)2

Latest decisions

NumberDevice and applicantDecision and review time
K252643 Iso-Gard Filter Straight; Humid-Vent Filter Compact AngledTeleflex Medical Sdn. Bhd. May 15, 2026267 days
K251583 Ambu® Virobac II® Exhalation FilterAmbu A/S Apr 30, 2026342 days
K241339 Besmed Bacterial Filter and HMEFBesmed Health Business Corp Nov 14, 2024185 days
K222917 Breathing circuit bacterial/viral filterShaoxing Haitech Medical Products Co., Ltd. Feb 15, 2024507 days
K230483 Passy Muir Tracheostomy Viral & Bacterial Airway Protection Filter (PM-APF15)Passy-Muir, Inc. Sep 16, 2023205 days
K223258 TNI Clear-Guard™ 3 angled breathing filter (1545020)Intersurgical , Ltd. Jun 15, 2023237 days
K221472 Filter and HME/FilterNingbo Huakun Medical Equipment Co., Ltd. Feb 8, 2023264 days
K221836 Filter CareStar Plus, Filter SafeStar Plus, Filter/HME TwinStar PlusDrägerwerk AG & Co. KGaA Dec 7, 2022167 days
K221250 EasyOne FilterNdd Medizintechnik AG Nov 22, 2022204 days
K210352 GGM Breathing Circuit Bacterial FilterGreat Group Medical Co, Ltd. Mar 27, 2022412 days

Recalls

Recall records for product code CAH per year, by the date the recall began. Each recalled product is one record; these 28 records belong to 16 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20171
20180
20190
20200
20214
20227
20232
20241
20257
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-0449-2026Class I Medline Industries, LPFirm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypo… Oct 10, 2025
Z-2158-2025Class I Draeger, Inc.Reports of slow increase of CO2-curve when gas sample is taken from the gas sampling port of filters. Jun 16, 2025
Z-2159-2025Class I Draeger, Inc.Reports of slow increase of CO2-curve when gas sample is taken from the gas sampling port of filters. Jun 16, 2025
Z-2160-2025Class I Draeger, Inc.Reports of slow increase of CO2-curve when gas sample is taken from the gas sampling port of filters. Jun 16, 2025
Z-2161-2025Class I Draeger, Inc.Reports of slow increase of CO2-curve when gas sample is taken from the gas sampling port of filters. Jun 16, 2025

Adverse event reports (MAUDE)

Medical device reports received per year that name product code CAH. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
201729
201833
201930
202033
202156
202267
202340
202451
202579
2026 (so far)85

Last 12 months by event type

  • Deaths0
  • Injuries14
  • Malfunctions79
  • Other or not stated2

Most reported device problems, last 12 months

  1. Gas/Air Leak28
  2. Connection Problem24
  3. Loose or Intermittent Connection21
  4. Use of Device Problem11
  5. Crack7
  6. Material Split, Cut or Torn7

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
2,889in the FDA’s Global Unique Device Identification Database
Establishment listings
305registered establishments that list this product code

Watch CAH in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get CAH every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in CAH. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.