FDA product code
FEQ: Pump, Air, Non-Manual, For Endoscope (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 876.1500
- Review panel
- Gastroenterology and Urology
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 210 in the last 12 months
- Median review time
- 50 daysthe latest 21 decisions
- Recall records
- 51 in the last 12 months
- Adverse event reports
- 282last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code FEQ per year, by decision date, 21 since 2000. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 1 |
| 2018 | 2 |
| 2019 | 1 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 1 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K220395 | EndoGator Endoscopy Irrigation TubingSTERIS Corporation | Mar 11, 202228 days |
| K191559 | AquaShield System CO2, AquaShield System CO2-PENTAX, AquaShield System CO2-FUJIFILMUnited States Endoscopy Group, Inc. | Aug 1, 201949 days |
| K182258 | AquaShield System CO2; AquaShield System CO2 - PENTAX; AquaShield System CO2 - FujifilmUnited States Endoscopy Group, Inc. | Sep 21, 201831 days |
| K173650 | AquaShield System CO2- FujifilmUnited States Endoscopy Group, Inc. | Jan 5, 201838 days |
| K163424 | AquaShield System - PENTAX, AquaShield System CO2 - PENTAX, AquaShield System CO2 - FUJINONUnited State Endoscopy Group, Inc. | Jan 4, 201729 days |
| K162152 | Erbe’s CO2 Tubing/Cap SetsErbe USA, Inc. | Aug 31, 201629 days |
| K161482 | EndoChoice Water Bottle Cap Irrigation SystemEndochoice, Inc. | Jun 27, 201627 days |
| K153588 | EndoChoice Water Bottle Cap SystemEndochoice, Inc. | Feb 24, 201670 days |
| K153362 | AquaShield system CO2, AquaShield system-PENTAXUnited States Endoscopy Group, Inc. | Dec 16, 201526 days |
| K142155 | ENDOCHOICE WATER BOTTLE CAP SYSTEMEndochoice, Inc. | Jan 20, 2015167 days |
Recalls
Recall records for product code FEQ per year, by the date the recall began. Each recalled product is one record; these 5 records belong to 5 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 1 |
| 2025 | 0 |
| 2026 (so far) | 1 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-2062-2026Class II | Erbe Medical, LLCDevices providing sterile water/CO2/air to endoscopes were manufactured with a distal irrigation segment connector configuration that may be more susceptible t… | Mar 31, 2026 |
| Z-0229-2025Class II | Olympus Corporation of the AmericasDuring product testing, Olympus identified an intermittent loss of function of the OFP-2 Flushing Pump caused by an internal component connection failure. | Aug 23, 2024 |
| Z-0113-2016Class II | Olympus Corporation of the Americassmall puncture marks in a sterile package for an accessory to an Olympus device | Sep 11, 2015 |
| Z-0329-2014Class II | Olympus America Inc.There is a possibility that the pump head was not shipped in the locked position. If the pump head is in an unlocked position this may, under certain circumsta… | May 20, 2013 |
| Z-0806-2012 | Olympus America Inc.Water may leak from the irrigation tubing used on the Olympus Flushing Pump | Dec 12, 2011 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code FEQ. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 4 |
| 2018 | 22 |
| 2019 | 6 |
| 2020 | 5 |
| 2021 | 3 |
| 2022 | 19 |
| 2023 | 11 |
| 2024 | 705 |
| 2025 | 344 |
| 2026 (so far) | 249 |
Last 12 months by event type
- Deaths0
- Injuries4
- Malfunctions278
- Other or not stated0
Most reported device problems, last 12 months
- Leak/Splash73
- Pumping Problem35
- Infusion or Flow Problem29
- Excess Flow or Over-Infusion12
- Improper Flow or Infusion12
- Break10
In plain English: what the newest reports describe Written by AI
Several reports describe the water flow or pressure behaving wrongly. Reports describe fluctuating water flow, abnormal pump head pressure, flow outside its specified range, a water jet out of control during maintenance, and backflow with a flow rate above the expected limit found during investigation.
5 reports: 9611174-2026-00207, 9611174-2026-00208, 9611174-2026-00209, 9611174-2026-00228, 9611174-2026-00234
Several reports describe water leaking or flowing when it should not. One says water leaked during inspection before use, another says the pump constantly leaked with an amber warning light showing an internal fault, and one says fluid kept coming out of the scope after the foot pedal was released, starting after new tubing was installed.
3 reports: 9611174-2026-00222, 9611174-2026-00215, 9611174-2026-00227
Several reports say the flow could not be controlled normally. Users could not adjust the pump speed or flow setting, and one had to run the pump with the foot pedal to change the flow rate.
3 reports: 9611174-2026-00209, 9611174-2026-00219, 9611174-2026-00210
Several returned pumps were found with foreign material in the water bottle, including blue spots, black residue, and a brown and black substance. The reports do not say how this material got there or what effect it had on the device.
3 reports: 9611174-2026-00232, 9611174-2026-00225, 9611174-2026-00205
One report describes a colonoscopy where the pump's clamp did not grip the tubing properly, so staff had to squeeze the assembly by hand to keep water flowing. The procedure was extended by less than thirty minutes and finished with the same device, and no patient harm was reported.
Report: 9611174-2026-00231
Written by AI from the narratives of the 18 newest reports that have one, received Sep 17, 2026 to Oct 29, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 143in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 78registered establishments that list this product code
Other codes under 21 CFR 876.1500
Product codes classified under the same regulation.
Watch FEQ in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get FEQ every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in FEQ. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:FEQRetrieved Nov 29, 2026, 5:15 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:FEQRetrieved Nov 29, 2026, 5:16 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:FEQ · count decision_dateRetrieved Nov 29, 2026, 5:16 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:FEQRetrieved Nov 29, 2026, 5:15 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:FEQ · count event_date_initiatedRetrieved Nov 29, 2026, 5:15 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:FEQ · count res_event_numberRetrieved Nov 29, 2026, 5:15 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-2062-2026" "Z-0229-2025" "Z-0113-2016" "Z-0329-2014" "Z-0806-2012")Retrieved Nov 29, 2026, 5:16 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:FEQ AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 5:16 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:FEQ AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 5:16 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:FEQ AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 5:16 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:FEQRetrieved Nov 29, 2026, 5:15 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:FEQRetrieved Nov 29, 2026, 5:15 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:876.1500Retrieved Nov 29, 2026, 8:27 AM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil