FDA product code
QFG: Alternate Controller Enabled Insulin Infusion Pump (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 880.5730
- Review panel
- Clinical Chemistry
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Ineligible
- 510(k) and De Novo decisions
- 339 in the last 12 months
- Median review time
- 89 daysthe latest 25 decisions
- Recall records
- 2412 in the last 12 months
- Adverse event reports
- 383,925last 12 months, 195 deaths
510(k) and De Novo decisions
Decisions for product code QFG per year, by decision date, 33 since 2019. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 2 |
| 2020 | 1 |
| 2021 | 0 |
| 2022 | 2 |
| 2023 | 9 |
| 2024 | 5 |
| 2025 | 5 |
| 2026 (so far) | 9 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K262992 | MiniMed Fit pumpMedtronic MiniMed | Nov 13, 202678 days |
| K261816 | Tandem Mobi insulin pump with interoperable technologyTandem Diabetes Care, Inc. | Nov 12, 2026164 days |
| K263756 | twiist systemDeka Research and Development Corp. | Nov 12, 202627 days |
| K263422 | Mint ACE PumpBeta Bionics, Inc. | Oct 21, 202626 days |
| K260001 | Mint ACE PumpBeta Bionics, Inc. | Sep 11, 2026252 days |
| K261672 | twiist Automated Insulin Delivery (AID) SystemDeka Research and Development | Jun 17, 202627 days |
| K253976 | iLet ACE PumpBeta Bionics, Inc. | Apr 30, 2026139 days |
| K253743 | MiniMed Flex pumpMedtronic Minimed, Inc. | Mar 12, 2026108 days |
| K253470 | MiniMed 780G Insulin Pump (MMT-1884); MiniMed 780G Insulin Pump Kit (MMT-1894)Medtronic Minimed, Inc. | Jan 15, 202697 days |
| K253074 | Tandem Mobi insulin pump with interoperable technologyTandem Diabetes Care, Inc. | Nov 6, 202544 days |
Recalls
Recall records for product code QFG per year, by the date the recall began. Each recalled product is one record; these 24 records belong to 22 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 1 |
| 2023 | 2 |
| 2024 | 1 |
| 2025 | 8 |
| 2026 (so far) | 12 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-0316-2027Class I | Tandem Diabetes Care, Inc.Tandem Diabetes Care, Inc. ( Tandem ) has identified that certain Tandem Mobi insulin pumps may generate false-positive occlusion alarms when no actual insulin… | Sep 23, 2026 |
| Z-0088-2027Class II | Medtronic MiniMed, Inc.Some MiniMed Flex insulin pumps were released with missing (i.e., devices with no result or abort test results are presumed to have failed) or failed verificat… | Aug 20, 2026 |
| Z-3330-2026Class II | Tandem Diabetes Care, Inc.A beta (early-release) version of the iOS application, version 2.13 (TestFlight) had a software issue preventing successful pairing of the t:slim application w… | Jul 16, 2026 |
| Z-2980-2026Class II | Medtronic MiniMed, Inc.Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress, and result… | Jul 13, 2026 |
| Z-2982-2026Class II | Medtronic MiniMed, Inc.Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress, and result… | Jul 13, 2026 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code QFG. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 30,522 |
| 2021 | 74,701 |
| 2022 | 129,119 |
| 2023 | 188,014 |
| 2024 | 273,190 |
| 2025 | 328,240 |
| 2026 (so far) | 346,598 |
Last 12 months by event type
- Deaths195
- Injuries50,955
- Malfunctions332,636
- Other or not stated139
Most reported device problems, last 12 months
- Pumping Stopped82,409
- Unintended Movement37,879
- Loss of or Failure to Bond35,691
- Insufficient Device Problem Information31,748
- Obstruction of Flow26,457
- Failure to Charge19,119
In plain English: what the newest reports describe Written by AI
Pumps repeatedly showed occlusion alarms, which means the insulin flow was thought to be blocked somewhere in the tubing, infusion set or cartridge. Some customers changed the infusion set or cartridge and resumed insulin, and many reports say blood glucose was not affected.
14 reports: 3013756811-2026-197917, 3013756811-2026-197913, 3013756811-2026-197915, 3013756811-2026-197909, 3013756811-2026-197912, 3013756811-2026-197908, 3013756811-2026-197905, 3013756811-2026-197903, 3013756811-2026-197899, 3013756811-2026-197895, 3013756811-2026-197893, 3013756811-2026-197890, 3013756811-2026-197901, 3013756811-2026-197914
In two reports, the disposable pod's cannula, the thin tube that sits under the skin, was found blocked after removal while blood glucose rose to 300 mg/dL and above 360 mg/dL. One report also describes bleeding at the site, and each patient was given a new pod.
2 reports: 3014585508-2026-79826, 3014585508-2026-79827
One report describes a customer with occlusion alarms whose blood glucose reached 426 mg/dL and who was admitted to intensive care with ketone levels described as dangerous. The customer had kept one cartridge for over 48 hours, and the manufacturer's support advised changing it every 48 hours.
Report: 3013756811-2026-197917
One report says the pump battery could not be charged, which caused the pump to shut down. No harm to the customer was reported, and the customer switched to another method of insulin therapy.
Report: 3013756811-2026-197923
One report says the pump showed a malfunction alarm, and the customer went back to manual insulin injections. No effect on blood glucose was reported.
Report: 3013756811-2026-197884
Written by AI from the narratives of the 19 newest reports that have one, received Oct 30, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 124in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 77registered establishments that list this product code
Watch QFG in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get QFG every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in QFG. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:QFGRetrieved Nov 29, 2026, 3:20 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:QFGRetrieved Nov 29, 2026, 3:20 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:QFG · count decision_dateRetrieved Nov 29, 2026, 3:20 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:QFGRetrieved Nov 29, 2026, 3:20 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:QFG · count event_date_initiatedRetrieved Nov 29, 2026, 3:20 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:QFG · count res_event_numberRetrieved Nov 29, 2026, 3:20 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-0316-2027" "Z-0088-2027" "Z-3330-2026" "Z-2980-2026" "Z-2982-2026")Retrieved Nov 29, 2026, 3:20 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:QFG AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 3:20 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:QFG AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 3:20 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:QFG AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 3:20 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:QFGRetrieved Nov 29, 2026, 3:20 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:QFGRetrieved Nov 29, 2026, 3:20 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:880.5730Retrieved Nov 29, 2026, 3:20 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.