FDA product code
IRP: Massager, Powered Inflatable Tube (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 890.5650
- Review panel
- Physical Medicine
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 15812 in the last 12 months
- Median review time
- 160 daysthe latest 25 decisions
- Recall records
- 50 in the last 12 months
- Adverse event reports
- 1last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code IRP per year, by decision date, 158 since 1978. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 4 |
| 2018 | 3 |
| 2019 | 10 |
| 2020 | 9 |
| 2021 | 19 |
| 2022 | 14 |
| 2023 | 10 |
| 2024 | 12 |
| 2025 | 13 |
| 2026 (so far) | 12 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K261887 | Normatec Elite Heat (PRO26-SH, PRO26-ST, PRO26-TL, PRO26-PH, PRO26-PS, PRO26-PT)NormaTec Industries, LP | Oct 6, 2026123 days |
| K262812 | Air compression massager (ACM-C1, ACM-C2, ACM-C3, ACM-C4, ACM-C5, ACM-C6, ACM-C7, ACM-C8, ACM-C9, ACM-5C1, ACM-5C2, ACM-5C3, ACM-5C4, ACM-5C5, ACM-5C6, ACM-5C7, ACM-5C8, ACM-5C9)Sichuan Qianli-Beoka Medical Technology, Inc. | Oct 5, 202659 days |
| K260944 | ManaCold Compression SystemManamed, LLC | Aug 12, 2026142 days |
| K260529 | Air Compression Massager (HY-1165)Xiamen High Top Electronic Technology Co., Ltd. | Jul 27, 2026160 days |
| K260948 | Air Compression Massager (EP30B001 I)Wavepulse (Xiamen) Healthcare Technology Co., Ltd. | Jul 24, 2026123 days |
| K253919 | Cryo And Thermo Therapy SystemXiamen Weiyou Intelligent Technology Co., Ltd. | Jun 10, 2026184 days |
| K252396 | Cryo Sport 2.0 (AVA 501)Avanutri Equipamentos DE Saúde | May 29, 2026302 days |
| K254200 | Air Compression Leg Massager (LF-FT001, LF-FT002-1, LF-FT003, LF-FT003-1, LF-FT004-1, LF-FT005-1, LF-FT006-1, LF-FT007, LF-FT008, LF-FT009, LF-FT010, LF-FT011, LF-FT012, LF-FT013, LF-FT014, LF-FT015, LF-FT016, UM-01, UM-02, UM-03, UM-03-1, UM-04-1, UM-05-1, UM-06-1, UM-07, UM-08, UM-09, UM-10, UM-11, UM-12, UM-13, UM-14, UM-15, UM-16, AI-01, AI-02, AI-03, AI-04, AI-05, AI-06, AI01, AI02, AI03, AI-07, AI-08, AI-09, AI-10, AI-11, AI-12, AI-13, AI-14)Wenzhou Lingfeng Electronic Technology Co., Ltd. | May 5, 2026127 days |
| K253064 | Air Pressure Therapy System (VU-IPC8M, VU-IPC6M, VU-IPC4X)Xiamen Weiyou Intelligent Technology Co., Ltd. | Apr 30, 2026220 days |
| K252386 | Hot and Cold Compression System (A02-C-032)Chengdu Cryo-Push Medical Technology Co.,Ltd | Apr 3, 2026246 days |
Recalls
Recall records for product code IRP per year, by the date the recall began. Each recalled product is one record; these 5 records belong to 4 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 2 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 1 |
| 2024 | 0 |
| 2025 | 1 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-2463-2025Class II | Zhejiang Luyao Electronics Technology Co., Ltd.Regulatory documentation error and misleading color packaging claim. | Jan 17, 2025 |
| Z-1398-2023Class II | Compass Health Brands (Corporate Office)Error in the Introduction section of the user manual that indicates the device may be used for the prevention of deep vein thrombosis (DVT). Marketed without a… | Mar 7, 2023 |
| Z-0204-2018Class II | Won Industry Co.Distribution of unapproved devices with incorrect air compression level setup | Jan 13, 2017 |
| Z-0205-2018Class II | Won Industry Co.Distribution of unapproved devices with incorrect air compression level setup | Jan 13, 2017 |
| Z-3210-2011 | CoolSystems, Inc.It was discovered that some Game Ready Professional Therapy System units that were to have been quarantined pending final inspection and testing were inadverte… | Aug 23, 2011 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code IRP. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 1 |
| 2021 | 0 |
| 2022 | 1 |
| 2023 | 0 |
| 2024 | 4 |
| 2025 | 10 |
| 2026 (so far) | 1 |
Last 12 months by event type
- Deaths0
- Injuries1
- Malfunctions0
- Other or not stated0
Most reported device problems, last 12 months
- Appropriate Device Problem Term/Code Not Available1
In plain English: what the newest reports describe Written by AI
Several reports describe the battery pack of the device catching fire or smoking while it was charging, even though the device was not in use. In the reports, the battery pack was described as sparking, smoking, exploding, melting or bursting into flames, and some reports mention damage to nearby furniture or carpet. No injuries were reported in these cases.
8 reports: 3003619590-2024-00001, 3003619590-2024-00003, 3003619590-2024-00004, 3003619590-2024-00005, 3003619590-2024-00006, 3003619590-2025-00001, 3003619590-2025-00002, 3003619590-2025-00003
Several reports describe pain, nerve or swelling problems from using a compression device. One user reported intense pressure leading to a visit to the emergency room with neck spasms after a shoulder wrap was used following surgery. Another reported a painful, raised rash on the leg where the wrap was applied, and a third reported worsening leg swelling and pain after use.
3 reports: 3011270181-2024-00114, 3011270181-2024-00056, MW5095732
Several reports describe the device's air pump or wrap failing under high pressure. One user reported the wrap burst and the pump then exploded, with debris striking the face. Another reported the device exploded and caused leg pain. These reports describe the same kind of failure, but the cause is not stated.
2 reports: 3011491271-2024-00001, MW5113019
The manufacturer says the battery pack contains a lithium-ion battery from an outside supplier. In one report, the manufacturer's preliminary review found the customer had charged the device with a charger other than the one the manufacturer provided, and the manufacturer said an older battery from another supplier was discontinued in 2022.
Report: 3003619590-2025-00001
One report describes a fire at a physical therapy clinic that the manufacturer learned about from the clinic's insurer. The device was being evaluated along with several other possible causes, and the manufacturer had no further details at the time of the report.
Report: 3003619590-2024-00002
Written by AI from the narratives of the 19 newest reports that have one, received Jul 28, 2015 to Jul 27, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 1,206in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 201registered establishments that list this product code
Watch IRP in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get IRP every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in IRP. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:IRPRetrieved Nov 28, 2026, 8:44 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:IRPRetrieved Nov 29, 2026, 10:10 AM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:IRP · count decision_dateRetrieved Nov 28, 2026, 8:44 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:IRPRetrieved Nov 29, 2026, 10:10 AM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:IRP · count event_date_initiatedRetrieved Nov 28, 2026, 8:44 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:IRP · count res_event_numberRetrieved Nov 28, 2026, 8:44 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-2463-2025" "Z-1398-2023" "Z-0204-2018" "Z-0205-2018" "Z-3210-2011")Retrieved Nov 29, 2026, 10:10 AM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:IRP AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 10:10 AM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:IRP AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 10:10 AM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:IRP AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 10:10 AM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:IRPRetrieved Nov 28, 2026, 8:44 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:IRPRetrieved Nov 28, 2026, 8:44 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:890.5650Retrieved Nov 28, 2026, 8:44 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.