FDA product code
OVE: Intervertebral Fusion Device With Integrated Fixation, Cervical (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 888.3080
- Review panel
- Orthopedic
- Premarket route
- 510(k)
- Implant
- Yes
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 16915 in the last 12 months
- Median review time
- 76 daysthe latest 25 decisions
- Recall records
- 120 in the last 12 months
- Adverse event reports
- 155last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code OVE per year, by decision date, 169 since 2007. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 9 |
| 2018 | 7 |
| 2019 | 7 |
| 2020 | 13 |
| 2021 | 10 |
| 2022 | 10 |
| 2023 | 16 |
| 2024 | 6 |
| 2025 | 10 |
| 2026 (so far) | 13 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K263113 | ZSAlign Cervical Interbody SystemZSFab, Inc. | Oct 2, 202628 days |
| K262726 | NeckTune™ SA Cervical CageMedyssey Co., Ltd. | Sep 29, 202660 days |
| K263049 | SABER-C SystemElevation Spine | Sep 23, 202623 days |
| K261882 | Pisces-SC Standalone Cervical Interbody SystemOsseus Fusions Systems, LLC | Aug 21, 202677 days |
| K261451 | ZSAlign Cervical Interbody SystemZSFab, Inc. | Jul 31, 202691 days |
| K262094 | E3D™-C Interbody SystemEvolution Spine | Jul 22, 202630 days |
| K261279 | DeltaNEC Interbody SystemNexus Spine | Jun 22, 202666 days |
| K252982 | CONDUIT™ ZERO-P™ VA Secured Spacer SystemAvalign Technologies, Inc. | Jun 18, 2026274 days |
| K260036 | NeckTune™ 3D SA Cervical CageMedyssey Co, Ltd. | Jun 15, 2026160 days |
| K260660 | SABER-C SystemElevation Spine | Apr 27, 202656 days |
Recalls
Recall records for product code OVE per year, by the date the recall began. Each recalled product is one record; these 12 records belong to 12 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 1 |
| 2018 | 0 |
| 2019 | 2 |
| 2020 | 0 |
| 2021 | 2 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 1 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-3210-2024Class II | Stryker SpineLabeling Correction: Instructions For Use (IFU)-Updated electronic Rev 5, specific language for indications for use for the United States and Canada was inadve… | Aug 23, 2024 |
| Z-0104-2022Class II | K2M, IncUnits may arrive with missing pins and/or in a state in disassembly. Pins dislodging from the instrument during a procedure may enter the surgical cavity, requ… | Sep 8, 2021 |
| Z-0086-2022Class II | Medtronic Sofamor Danek USA IncEngraving on implants may not match what is listed on the pouch labels. | Sep 2, 2021 |
| Z-1973-2019Class II | SpineFrontier, Inc.There is a potential for the driver to bind up when attempting to advance the intervertebral body fixation screw. | Jun 13, 2019 |
| Z-1823-2020Class II | U&I CORP.The vertical dimensions of screw-hole position in some standalone cages might be out of range in engineering drawings. A possibility exists of screw pullout, l… | Jan 30, 2019 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code OVE. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 134 |
| 2018 | 210 |
| 2019 | 169 |
| 2020 | 162 |
| 2021 | 186 |
| 2022 | 200 |
| 2023 | 148 |
| 2024 | 142 |
| 2025 | 136 |
| 2026 (so far) | 147 |
Last 12 months by event type
- Deaths0
- Injuries36
- Malfunctions119
- Other or not stated0
Most reported device problems, last 12 months
- Break34
- Fracture26
- Patient Device Interaction Problem15
- Material Integrity Problem14
- Migration12
- Detachment of Device or Device Component11
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 17,973in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 794registered establishments that list this product code
Other codes under 21 CFR 888.3080
Product codes classified under the same regulation.
Watch OVE in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get OVE every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in OVE. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:OVERetrieved Nov 29, 2026, 4:34 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:OVERetrieved Nov 29, 2026, 4:34 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:OVE · count decision_dateRetrieved Nov 29, 2026, 4:34 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:OVERetrieved Nov 29, 2026, 4:34 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:OVE · count event_date_initiatedRetrieved Nov 29, 2026, 4:34 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:OVE · count res_event_numberRetrieved Nov 29, 2026, 4:34 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-3210-2024" "Z-0104-2022" "Z-0086-2022" "Z-1973-2019" "Z-1823-2020")Retrieved Nov 29, 2026, 4:34 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:OVE AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 4:34 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:OVE AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 4:34 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:OVE AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 4:34 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:OVERetrieved Nov 29, 2026, 4:34 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:OVERetrieved Nov 29, 2026, 4:34 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:888.3080Retrieved Nov 29, 2026, 12:26 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil