FTC: Light, Ultraviolet, Dermatological (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 878.4630
- Review panel
- General and Plastic Surgery
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 2744 in the last 12 months
- Median review time
- 90 daysthe latest 25 decisions
- Recall records
- 154 in the last 12 months
- Adverse event reports
- 8last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code FTC per year, by decision date, 274 since 1976. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 2 |
| 2018 | 3 |
| 2019 | 2 |
| 2020 | 5 |
| 2021 | 3 |
| 2022 | 1 |
| 2023 | 4 |
| 2024 | 1 |
| 2025 | 3 |
| 2026 (so far) | 4 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K262885 | 308nm Ultraviolet Light Therapy Device (XWYL-401A, XWYL-403A)Guangxi Hope Medical Devices Technology Co., Ltd. | Oct 14, 202658 days |
| K262285 | UV Phototherapy Device (KZX-308nm-BL65, KZX-308nm-BL55, KZX-308nm-BL50, KZX-308nm-BL45, KZX-308nm-BL35)Shijiazhuang Recovery Star Medical Device Co., Ltd. | Oct 5, 202691 days |
| K254234 | Phototherapy System (DUV-COMBO)Choyang Medics Co., Ltd. | May 1, 2026123 days |
| K253871 | Klär Lite (RCW-KL1000)Radcliffe Watts, LLC | Mar 4, 202690 days |
| K250425 | Ultraviolet Phototherapy DeviceAPK Technology Co., Ltd. | Apr 7, 202552 days |
| K244022 | 308nm UV Phototherapy System (UV-K, UV-X, UV-Y, UV-Z)Boston Aesthetics, Inc. | Mar 13, 202576 days |
| K242977 | 308nm Excimer Phototherapy Device (CN-308E, KN-4003B3, KN-4003B4, KN-5000C KN-5000D, KN-5000H, KN-5000K)Xuzhou Kernel Medical Equipment Co., Ltd. | Feb 6, 2025133 days |
| K242908 | UV Phototherapy Device (Group A: SQ308PCHFD, SQ308PCQFD, SQ308PCPFD, SQ308PCNFD, SQ308PCMFD, Group B: SQ308PHTFF, SQ308PHSFF, SQ308PHRFF, SQ308PHZFG, SQ308PPUFF, SQ308PPVFF, SQ308PPWFG)Shanghai Sigma High-Tech Co., Ltd. | Dec 19, 202486 days |
| K233811 | DT Controlled Phototherapy EquipmentDaavlin Distributing Co. | Dec 13, 202314 days |
| K223882 | Narrowband UV Phototherapy Light Lamp (Model: HB-UPLL-01, HB-UPLL-02)Shenzhen Honpal Optoelectronic Technology Co., Ltd. | Mar 27, 202390 days |
Recalls
Recall records for product code FTC per year, by the date the recall began. Each recalled product is one record; these 15 records belong to 8 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 1 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 2 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 1 |
| 2026 (so far) | 3 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-2742-2026Class II | Daavlin Distributing CompanyCalibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 ins… | Mar 5, 2026 |
| Z-2743-2026Class II | Daavlin Distributing CompanyCalibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 ins… | Mar 5, 2026 |
| Z-2744-2026Class II | Daavlin Distributing CompanyCalibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 ins… | Mar 5, 2026 |
| Z-2295-2026Class II | Daavlin Distributing CompanySome of the Calibration/Output Certificates had the outputs reversed. | Dec 10, 2025 |
| Z-1073-2022Class II | Daavlin Distributing CompanySoftware issue, resuming an interrupted treatment will result in swap of utilized calibration treatment distance 0" inches to "9. Resulting in over or under do… | Apr 26, 2022 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code FTC. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 2 |
| 2018 | 0 |
| 2019 | 14 |
| 2020 | 4 |
| 2021 | 48 |
| 2022 | 4 |
| 2023 | 5 |
| 2024 | 4 |
| 2025 | 8 |
| 2026 (so far) | 6 |
Last 12 months by event type
- Deaths0
- Injuries6
- Malfunctions2
- Other or not stated0
Most reported device problems, last 12 months
- Patient Device Interaction Problem2
- Adverse Event Without Identified Device or Use Problem1
- Missing Information1
- Off-Label Use1
- Use of Device Problem1
- Use of Incorrect Control/Treatment Settings1
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 779in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 72registered establishments that list this product code
Other codes under 21 CFR 878.4630
Product codes classified under the same regulation.
Watch FTC in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get FTC every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in FTC. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:FTCRetrieved Nov 29, 2026, 1:05 AM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:FTCRetrieved Nov 29, 2026, 4:49 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:FTC · count decision_dateRetrieved Nov 29, 2026, 1:05 AM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:FTCRetrieved Nov 29, 2026, 4:49 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:FTC · count event_date_initiatedRetrieved Nov 29, 2026, 1:05 AM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:FTC · count res_event_numberRetrieved Nov 29, 2026, 1:05 AM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-2742-2026" "Z-2743-2026" "Z-2744-2026" "Z-2295-2026" "Z-1073-2022")Retrieved Nov 29, 2026, 4:49 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:FTC AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 4:49 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:FTC AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 4:49 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:FTC AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 1:05 AM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:FTCRetrieved Nov 29, 2026, 1:05 AM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:FTCRetrieved Nov 29, 2026, 1:05 AM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:878.4630Retrieved Nov 28, 2026, 9:38 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil