The FDA cleared 50 510(k)s, granted 2 De Novo requests and approved no original PMAs. 21 device recalls were posted, 4 of them Class I. Every 510(k), De Novo and PMA decision dated November 11 to 17, 2024, and every device recall the FDA posted those days.

510(k) clearances
50median review 109 days
De Novo grants
2
Original PMA approvals
0plus 69 supplements
Recalls posted
21137 products, 4 Class I

Recalls posted

Device recalls the FDA posted November 11 to 17, 2024, Class I first. Class I is the most serious: a reasonable chance the product causes serious harm or death. The class comes from the FDA’s weekly enforcement report, so the newest recalls can show “Class pending” for a few days.

Recall and classFirm and reasonProduct code
Z-0290-2025Class I Philips Respironics, Inc.Potential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements. CBK7 products
Z-0304-2025Class I Draeger Medical, Inc.The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use. BSZ2 products
Z-0327-2025Class I Baxter Healthcare CorporationBaxter is aware of several recalls by other manufacturers related to the potential risk of exposure to non-dioxin-like (NDL) polychlorinated biphenyl acids (PCBAs) and NDL polychlorinated biphenyls (PCBs) when using cer… KDI1 product
Z-0297-2025Class I Boston Scientific CorporationBoston Scientific is updating the instructions for use of their POLARx and POLARx FIT Cryoablation Balloon Catheters related to Atrio-esophageal Fistula Risk. OAE7 products
Z-0466-2025Class II Smiths Medical ASD, Inc.The adapter input plug can become damaged or broken. If the input plug is damaged, the metal contacts to the body of the AC adapter may be exposed or one or more of the AC mains prongs may separate from the input plug. MEA1 product
Z-0467-2025Class II Bard Peripheral Vascular IncDisposable biopsy instruments were potentially manufactured missing the backend notch retention feature on the needle that results in inadequate securing of the needle inside the device, which could lead to needle detac… KNW1 product
Z-0468-2025Class II Hobbs Medical, Inc.Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that blocked the ID of the nee… FBK9 products
Z-0477-2025Class II Biomet, Inc.Firm received complaints reporting that guides were warped when pulled out of packaging. This issue may result in extension of surgery to find another product, or may lead to adverse local tissue reaction leading to sur… KDD1 product
Z-0478-2025Class II Caire, Inc.Potential failure of affected units to recharge the battery of the oxygen concentrators. CAW4 products
Z-0383-2025Class II AVID Medical, Inc.Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or… LRO OES OEZ73 products
Z-0456-2025Class II Getinge Usa Sales IncError code "50037" occurs on the IR-Hand Control when a MEERA-table is controlled, and the table stops the movement. This error may occur sporadically when the MEERA-table is controlled via an IR-Hand Control and may re… FQO8 products
Z-0464-2025Class II Siemens Healthcare Diagnostics, Inc.The IMMULITE 2000 PSA assay for use on the IMMULITE 2000 and IMMULITE 2000 XPi analyzers did not meet the High-dose Hook Effect claim as stated in the Instructions for Use (IFU). MTF1 product
Z-0465-2025Class II MEDLINE INDUSTRIES, LP - NorthfieldTip protector may fall off within the packaging, resulting in compromised sterile barrier system and loss of sterility. LRW1 product
Z-0365-2025Class II Stryker CorporationThere is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in localized burns to… GEI6 products
Z-0371-2025Class II Bard Peripheral Vascular IncThe Venclose" RF Ablation Catheters may have been manufactured with internal wiring error. The red and yellow signal wires were soldered on swapped ports of the catheter PCB. This error in wire configuration would lead… GEI1 product
Z-0372-2025Class II Medtronic NeuromodulationThe Wireless Rechargers in the kits may be unable to enter a recharging session upon first use. LGW1 product
Z-0352-2025Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Potential for the Mains Control Unit board fuse may blow out during system start-up or during a procedure. OWB2 products
Z-0354-2025Class II MEDLINE INDUSTRIES, LP - NorthfieldSpecimen container included in kits is labeled as sterile, but has been identified to be non-sterile. Long-term consequences may include unnecessary treatment and/or prolonged hospitalization. LRO8 products
Z-0362-2025Class II Topcon Medical Systems, Inc.The overlay of visual field test locations on a probability map, as presented in IMAGEnet 6 in the so-called "Hood Report", is not appropriate as it can imply a structure/function correlation for which clinical evidence… NFJ1 product
Z-0363-2025Class II Karl Storz EndoscopyDue to improper labeling of products. An intended use on the label has not been reviewed and approved by FDA for distribution in the U.S. NVT1 product
Z-0364-2025Class II Olympus Corporation of the AmericasExposure of the UroPass product to Ultraviolet ( UV ) Radiation can cause brittleness of the device dilator tip, which may lead to broken dilator tips in the package or in patients during surgical procedures. KNY1 product

De Novo grants

New device types the FDA classified through the De Novo pathway. Each one creates a product code and becomes a possible predicate for later 510(k)s.

NumberDevice and applicantProduct codeDecision and review time
DEN240023 SmartFlow Neuro CannulaClearPoint Neuro, Inc. SDG Nov 13, 2024175 days
DEN230085 VERIFY RESI-TEST SLIDE-THRU Cleaning Process Protein (CPP) IndicatorSteris SDC Nov 12, 2024330 days

Original PMA approvals

Original premarket approvals for Class III devices. The FDA also decided 69 PMA supplements this week, which change devices already approved.

No original PMAs were approved this week.

510(k) clearances

50 510(k) clearances by review panel. The median took 109 calendar days from receipt to decision; the fastest, K243306, 25 days; the longest, K211498, 1,279 days.

Calendar days include the time a submission waits on the company, so they run longer than the FDA’s 90-day review goal. How to search the 510(k) database

  • Radiology9
  • Cardiovascular8
  • General, Plastic Surgery7
  • Dental6
  • Orthopedic6
  • Anesthesiology4
  • Physical Medicine3
  • Panel not stated2
  • Gastroenterology, Urology1
  • General Hospital1
  • Microbiology1
  • Neurology1
  • Obstetrics/Gynecology1

Every clearance

NumberDevice and applicantProduct codeDecision and review time
K230915 AirLife™ Open Et Oxygen Mask, AirLife™ Open Et+ Oxygen MaskVyaire Medical, Inc. CCKAnesthesiology Nov 12, 2024592 days
K241135 RESPOND(R) VentilatorCorvent Medical, Inc. CBKAnesthesiology Nov 13, 2024203 days
K241339 Besmed Bacterial Filter and HMEFBesmed Health Business Corp CAHAnesthesiology Nov 14, 2024185 days
K242775 InPen System App (MMT-8060 (iOS), MMT-8061 (Android))Medtronic Minimed NDCAnesthesiology Nov 12, 202460 days
K211498 Nonin OTC Pulse Oximeter Model 3250Nonin Medical, Inc. OLKCardiovascular Nov 13, 20241,279 days
K240860 EchoGo Amyloidosis (1.0)Ultromics Limited SDJCardiovascular Nov 15, 2024232 days
K241070 Gelsoft™ Plus Vascular ProsthesesVascutek, Ltd. DSYCardiovascular Nov 15, 2024210 days
K241224 Medline ReNewal Reprocessed Biosense Webster Webster CS Catheter with EZ Steer Technology and Medline ReNewal Reprocessed Biosense Webster Webster CS Catheter with EZ Steer Technology and Auto ID Technology (BD710DF282CT, BD710DF282RTS, BD710FJ282RTS, BD710FJ282CT)Surgical Instrument Service and Savings Inc.(Dba Medline Ren NLHCardiovascular Nov 15, 2024197 days
K241397 HinscopeHinlab Sas MSXCardiovascular Nov 14, 2024182 days
K242018 UbiqVue™ 2A Multi-parameter System (UX2550)Lifesignals, Inc. DRGCardiovascular Nov 12, 2024125 days
K242419 018 Hydrophilic Coated OTW PTA Balloon Dilatation CatheterCreagh Medical Ltd. Dba Surmodics, Inc. LITCardiovascular Nov 15, 202492 days
K242420 pNOVUS 21 MicrocatheterPhenox, Ltd. DQOCardiovascular Nov 12, 202489 days
K233081 NUVENTUS NV.C™ Dental Implant SystemInfinitum Eta, Ltd. DZEDental Nov 15, 2024416 days
K240337 Dental Handpiece (Model: J6-126, J6-136, SH11W2, SH11W3, W401LW, W15LW, W142LW, W11L)Guangdong Jinme Medical Technology Co., Ltd. EFBDental Nov 13, 2024282 days
K241074 High-speed Air Turbine Handpieces (SCHG01, SCHG03, SCHG18, SCHF19, SCHF22), Dental Low Speed Handpieces including Straight and Geared Angle Handpieces and Air Motors (Straight Handpieces: SCHD05-S, SCHD16-S, SCHD17-S, SCHD24-S, SCHD27-S Geared Angle Handpieces: SCHD05-C, SCHD06-C, SCHD16-C, SCHD17-C, SCHD24-C, SCHD27-C, SCHD29-S, SCHB4, SCHB5, SCHB20, SCHB64 Air Motors: SCHD16-M, SCHD17-M, SCHD24-M, SCHD27-M)Foshan Soco Precision Instrument Co., Ltd. EFBDental Nov 12, 2024207 days
K241575 Straumann® Anatomic Healing Abutments XC (AHA)Straumann USA, LLC NHADental Nov 11, 2024161 days
K242360 Speedex Light Body; Speedex Medium; Speedex Putty; Speedex Putty Soft; Speedex Universal ActivatorDent4you AG ELWDental Nov 15, 202498 days
K242407 Dental Glass Ceramics Blocks (HT, LT, MT)Hunan Vsmile Biotechnology Co., Ltd. EIHDental Nov 13, 202491 days
K242173 Retrace Ureteral Access Sheath (ASXL10, ASXL12, ACXL10, ACXL12, AXXL10, AXXL12, ALXL10, ALXL12)Coloplast Corp. FEDGastroenterology, Urology Nov 12, 2024111 days
K243273 CeQur Simplicity™ On-Demand Insulin Delivery SystemCequr SA OPPGeneral Hospital Nov 13, 202428 days
K241138 ArtiSeal Vessel Sealing System-ArtiSeal Instruments; ArtiSeal Vessel Sealing System-ArtiSeal GeneratorLivsmed, Inc. GEIGeneral, Plastic Surgery Nov 15, 2024205 days
K241930 VeincareWontech Co., Ltd. GEXGeneral, Plastic Surgery Nov 14, 2024136 days
K242440 Intense Pulsed Light Treatment System (LK-PT)Nanjing Bestview Laser S&T Co., Ltd. ONFGeneral, Plastic Surgery Nov 14, 202490 days
K242502 Aurelia Surgical Mask ASTM Level-3 (2130)Supermax Healthcare Canada – Supermax Medical FXXGeneral, Plastic Surgery Nov 12, 202482 days
K242598 DEFINE System (AG612444A)InMode , Ltd. PBXGeneral, Plastic Surgery Nov 14, 202476 days
K242960 AdVance™ XP Male Sling System (720163-03)Boston Scientific Corporation OTMGeneral, Plastic Surgery Nov 15, 202451 days
K243306 Lancet (IA, IB, IC, ID, IE, IK, IL, IM, IIA, IIB, III, V, VI, VII, VIII, IX); Lancing device (HH-X-T, HH-XV-T, HH-XVI-T, HH-XVII-T, HH-XVIII-T, HH-XIX, HH-XXI-T, HH-XXII-T, HH-XXIII-T, HH-XXIV-T, HH-XXV-T, HH-XXVI-T, HH-XXVIII-T, HH-XXIX-T, HH-XXX-T, HH-XIII-T, HH-XXVII-T)Tianjin Huahong Technology Co., Ltd. QRLGeneral, Plastic Surgery Nov 15, 202425 days
K242465 Panther Fusion SARS-CoV-2/Flu A/B/RSV assayHologic, Inc. QOFMicrobiology Nov 15, 202487 days
K243285 TMINI Miniature Robotic SystemTHINK Surgical, Inc. OLONeurology Nov 15, 202428 days
K240523 VitaVitro® Single Lumen Oocyte Retrieval Needle (ORNS-17G, ORNS-18G), VitaVitro® Double Lumen Oocyte Retrieval Needle (ORND-17G)Shenzhen Vitavitro Biotech Co., Ltd. MQEObstetrics/Gynecology Nov 15, 2024266 days
K240528 b-ONE® Total Hip SystemB-One Ortho, Corp. LZOOrthopedic Nov 15, 2024266 days
K242079 Arthrex Elbow Fracture Plating SystemArthrex, Inc. HRSOrthopedic Nov 14, 2024121 days
K242260 aprevo® Cervical ACDF; aprevo® Cervical ACDF-X; aprevo® Cervical ACDF-X NO CAMCarlsmed, Inc. ODPOrthopedic Nov 15, 2024107 days
K242436 Proximal Humerus Fixation SystemSkeletal Dynamics, Inc. HRSOrthopedic Nov 14, 202490 days
K242764 MediRod Pedicle Screw SystemCmf Medicon Surgical, Inc. NKBOrthopedic Nov 13, 202461 days
K242814 Tandry Q-Locking Plate SystemMicroware Precision Co., Ltd. HRSOrthopedic Nov 14, 202457 days
K242272 CORIOGRAPH Pre-Op Planning and Modeling ServicesBlue Belt Technologies, Inc. Panel not stated Nov 15, 2024106 days
K242856 LUOFUCON® Silicone Ag+ Foam Dressing (Prescription use)/ LUOFUCON® Silicone Ag+ Antibacterial Foam Dressing (OTC use) (Bordered, Bordered Lite, Non-Bordered, and Transfer)Huizhou Foryou Medical Devices Co., Ltd. FROPanel not stated Nov 13, 202454 days
K242430 Power-Flex (PFX1214, PFX1517, PFX1820)Soul Mobility ITIPhysical Medicine Nov 12, 202489 days
K242791 Electric wheelchair (KR-8807)Zhejiang Kairui Medical Device Co., Ltd. ITIPhysical Medicine Nov 12, 202457 days
K242940 Xrecovery (XR-001)Shenzhen Xinrun Electric Appliances Co, Ltd. IRPPhysical Medicine Nov 14, 202450 days
K240771 PRORAD X-Ray Flat Panel Detector with DROCPrognosys Medical Systems Private Limited MQBRadiology Nov 12, 2024236 days
K241125 VIVIX-S 1751SVieworks Co., Ltd. MQBRadiology Nov 15, 2024206 days
K241439 VUNO Med-Chest X-ray Triage/VUNO Med-CXR Link TriageVuno, Inc. QFMRadiology Nov 15, 2024178 days
K241981 Change Healthcare Stratus Imaging PACSChange Healthcare Canada Company LLZRadiology Nov 12, 2024130 days
K242062 1CMR ProMycardium AI Limited LLZRadiology Nov 15, 2024123 days
K242130 Koios DSKoios Medical, Inc. POKRadiology Nov 15, 2024116 days
K242342 Fetal EchoScanBrightheart POKRadiology Nov 14, 202499 days
K242800 The 5000 Compact Series Ultrasound SystemsPhilips Ultrasound, LLC IYNRadiology Nov 15, 202459 days
K243299 ACUSON Redwood Diagnostic Ultrasound SystemSiemens Medical Solutions USA, Inc. IYNRadiology Nov 15, 202428 days

Only see the weeks that matter to you

DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and every day it checks openFDA and tells you about new clearances, recalls and MAUDE reports for them. It also puts all of this beside the FDA’s own record pages.

Add to Chrome — free Add to Edge Or follow this page by RSS

Only your product codes, every Monday

One short email after each openFDA update with just the decisions and recalls in up to three product codes you choose. Free.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Decisions arrive in weekly updates; check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.