FDA product code
MEA: Pump, Infusion, Pca (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 880.5725
- Review panel
- General Hospital
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Ineligible
- 510(k) and De Novo decisions
- 310 in the last 12 months
- Median review time
- 92 daysthe latest 25 decisions
- Recall records
- 460 in the last 12 months
- Adverse event reports
- 2,687last 12 months, 7 deaths
510(k) and De Novo decisions
Decisions for product code MEA per year, by decision date, 31 since 1991. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 2 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K162165 | ambIT PCA*PIB, ambIT PIB, ambIT PIB*PCA, ambIT PIEB, and ambIT Programmable Intermittent PumpSummit Medical Products, Inc. | Aug 29, 2017391 days |
| K170982 | CADD®- Solis Ambulatory Infusion Pump with Wireless Communication; CADD®-Solis Infusion Pump, Model 2110; PharmGuard® Administrator Medication Safety Software; CADD® Administration Sets; Wireless Communication Module (CM)Smiths Medical Ads, Inc. | Aug 24, 2017143 days |
| K143612 | LifeCare PCA Infusion SystemHospira, Inc. | Apr 8, 2016476 days |
| K130394 | CADD -SOLIS AMBULATORY INFUSION PUMP, VERSION 3.0Smiths Medical Asd, Inc. | Dec 9, 2013297 days |
| K092313 | PERFUSOR SPACE INFUSION SYRINGE PUMP SYSTEMB.Braun Medical, Inc. | Nov 4, 200997 days |
| K072144 | CADD -SOLIS AMBULATORY INFUSION PUMP AND CADD -SOLIS MEDICATION SAFETY SOFTWARE - ADMINISTRATORSmiths Medical MD, Inc. | Mar 7, 2008218 days |
| K051253 | MODIFICATION TO: MULTIRATE INFUSOR SV )WITH AND WITHOUT PCM), MULTIRATE INFUSOR LV (WITH AND WITHOUT PCM)Baxter Healthcare Corp | Jul 15, 200560 days |
| K043299 | MEDLEY PCA MODULE, MODEL 8120Alaris Medical Systems, Inc. | Dec 10, 200410 days |
| K043256 | HOSPIRA LIFECARE INFUSION SYSTEM, MODEL 12384-04-07; LIFECARE PCA INFUSION SYSTEM STERILE EMPY VIALS, MODEL 6021-04-03Hospira, Inc. | Dec 6, 200412 days |
| K042800 | LIFECARE PCA INFUSION SYSTEM WITH HOSPIRA MEDNET SOFTWAREHospira, Inc. | Oct 18, 200410 days |
Recalls
Recall records for product code MEA per year, by the date the recall began. Each recalled product is one record; these 46 records belong to 27 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 2 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 1 |
| 2023 | 0 |
| 2024 | 3 |
| 2025 | 3 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-1662-2025Class I | Smiths Medical ASD, Inc.Pumps may experience Wireless Connection Modules intermittent connection alarms, which will interrupt an active infusion. Interruption or delay of therapy can… | Apr 10, 2025 |
| Z-1667-2025Class I | Smiths Medical ASD, Inc.Affected pumps may trigger an erroneous (false) Upstream Occlusion Alarm under certain conditions, which will interrupt an active infusion. Interruption or del… | Apr 10, 2025 |
| Z-1681-2025Class I | Smiths Medical ASD, Inc.There is a potential for thermal damage in CADD-Solis and CADD- Solis VIP infusion pumps. | Apr 10, 2025 |
| Z-0466-2025Class II | Smiths Medical ASD, Inc.The adapter input plug can become damaged or broken. If the input plug is damaged, the metal contacts to the body of the AC adapter may be exposed or one or mo… | Jun 26, 2024 |
| Z-2419-2024Class I | Smiths Medical ASD Inc.CADD-Solis Ambulatory Infusion Pumps, with software versions before v4.3, may have the following issues:1. Upstream Occlusion, 2. Stop and Power Keys Unrespons… | Feb 27, 2024 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code MEA. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 44 |
| 2018 | 256 |
| 2019 | 132 |
| 2020 | 2,331 |
| 2021 | 3,776 |
| 2022 | 6,443 |
| 2023 | 3,984 |
| 2024 | 4,560 |
| 2025 | 3,330 |
| 2026 (so far) | 2,465 |
Last 12 months by event type
- Deaths7
- Injuries9
- Malfunctions2,670
- Other or not stated1
Most reported device problems, last 12 months
- Failure to Infuse616
- Device Sensing Problem332
- Crack296
- Break177
- Failure to Sense131
- Device Alarm System116
In plain English: what the newest reports describe Written by AI
Other reports describe error codes, a system error, a failed accuracy test, a failed air-in-line sensor test, or a faulty logic board. Reports state no patient involvement, and only the logic board report gives a cause.
6 reports: 2016493-2026-75901, 2016493-2026-75961, 2016493-2026-75936, 2016493-2026-75809, 3012307300-2026-09758, 2016493-2026-75076
Pumps showed 'cassette not attached' or 'cassette not connected correctly' errors during setup or testing, which stop the pump from operating. One report says the pump became loose when locked. Patient harm was unknown or not reported.
5 reports: 3012307300-2026-09127, 3012307300-2026-09705, 3012307300-2026-09717, 3012542015-2026-00077, 3012307300-2026-09670
The downstream occlusion sensor, which detects blockages in the infusion line, was found out of calibration, defective, or with a detached seal. One report says the PCA button was not working and the sensor was found in the investigation. No patient harm was reported.
4 reports: 3012542015-2026-00079, 3012307300-2026-09703, 3012307300-2026-09718, 3012307300-2026-09722
The pump showed error 41622 (ERROR_MOTOR_TIMEOUT), a high-priority alarm that stops the infusion. Reports say it could interrupt therapy if it recurs. Patient outcome was unknown or not reported.
2 reports: 3012307300-2026-09697, 3012307300-2026-09714
Testing found that the binary switches failed, and in one report the flange detect switch was visibly distorted. No patient involvement was reported.
2 reports: 2016493-2026-75346, 2016493-2026-75394
Written by AI from the narratives of the 20 newest reports that have one, received Oct 26, 2026 to Oct 30, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 184in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 60registered establishments that list this product code
Other codes under 21 CFR 880.5725
Product codes classified under the same regulation.
Watch MEA in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get MEA every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in MEA. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:MEARetrieved Nov 29, 2026, 3:19 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:MEARetrieved Nov 29, 2026, 3:19 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:MEA · count decision_dateRetrieved Nov 29, 2026, 3:19 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:MEARetrieved Nov 29, 2026, 3:19 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:MEA · count event_date_initiatedRetrieved Nov 29, 2026, 3:19 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:MEA · count res_event_numberRetrieved Nov 29, 2026, 3:19 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-1662-2025" "Z-1667-2025" "Z-1681-2025" "Z-0466-2025" "Z-2419-2024")Retrieved Nov 29, 2026, 3:19 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:MEA AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 3:19 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:MEA AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 3:19 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:MEA AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 2:50 AM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:MEARetrieved Nov 29, 2026, 3:19 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:MEARetrieved Nov 29, 2026, 3:19 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:880.5725Retrieved Nov 29, 2026, 3:13 PM UTC. FDA data updated Nov 16, 2026.