Weekly archive
Week of October 28, 2024
The FDA cleared 63 510(k)s, granted 1 De Novo request and approved no original PMAs. 27 device recalls were posted, 3 of them Class I. Every 510(k), De Novo and PMA decision dated October 28 to November 3, 2024, and every device recall the FDA posted those days.
- 510(k) clearances
- 63median review 112 days
- De Novo grants
- 1
- Original PMA approvals
- 0plus 31 supplements
- Recalls posted
- 2760 products, 3 Class I
Recalls posted
Device recalls the FDA posted October 28 to November 3, 2024, Class I first. Class I is the most serious: a reasonable chance the product causes serious harm or death. The class comes from the FDA’s weekly enforcement report, so the newest recalls can show “Class pending” for a few days.
| Recall and class | Firm and reason | Product code |
|---|---|---|
| Z-0140-2025Class I | DATEX--OHMEDA, INC.Prior to current established manufacturing practices, higher levels of airborne formaldehyde exposure from newly manufactured incubators could be seen during clinical use that would rapidly decrease over a week. Affecte… | FMZ2 products |
| Z-0222-2025Class I | Smiths Medical ASD Inc.Damage to the battery pack may have caused a short to a capacitor within the battery pack. While the battery encasement is designed to be flame retardant, a short to the capacitor could lead to melting of the battery pa… | FRN1 product |
| Z-0208-2025Class I | Percussionaire CorporationAs a result of product complaints, breathing circuits are experiencing discolored fluid which are formed when therapy was being given with specific medication through the nebulizer cup. Discoloration is caused by a reac… | NHJ1 product |
| Z-0283-2025Class II | West Pharmaceutical Services AZ, Inc.Due to packaging damage from heat exposure during shipping/transportation that may result in a breach in the sterile barrier. | LHI1 product |
| Z-0284-2025Class II | Canon Medical System, USA, INC.When PET-CT system is executing reconstruction, if PET acquisition for another patient is performed (or PET reconstruction for another patient is performed from raw data processing), PET reconstruction control processin… | KPS1 product |
| Z-0285-2025Class II | Zimmer, Inc.A Gen 5 tail piece was assembled to a Gen 4 Finished Device, resulting in a 0.178" difference in the end of the endoscope slot to the end of the tail piece. This allows the scope to move further into the plastic sheath… | GEI1 product |
| Z-0286-2025Class II | Siemens Healthcare Diagnostics IncSiemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in one lot of epoc BGEM BUN Test Cards approximately 2 months prior to expiration. The observed maximum negative bias from internal… | CHL1 product |
| Z-0287-2025Class II | American Surgical Company, LLCProduct defect where the X-ray detectable barium strip (aka radio-opaque marker) becomes detached from the sponge (pattie). May delay surgical process as the surgeon may require extra time to remove the markers from the… | HBA1 product |
| Z-0279-2025Class II | Accriva Diagnostics, Inc.Values used to calibrate whole blood oximeter instruments were observed to be higher than normal, which may result in incorrect patient results where oxygen levels are higher than actual results, which may lead to patie… | DQA1 product |
| Z-0280-2025Class II | Pro-Dex IncThe packaging of a sterile battery, intended to power a surgical screwdriver, may contain Tyvek shavings inside the sterile package. | MOQ1 product |
| Z-0281-2025Class II | Smith & Nephew IncComplaint received indicating that a package contained a LEGION NARROW PS OXIN SZ 5N RIGHT implant instead of a GENESIS II POSTERIOR STABILIZED FEMORAL SIZE 5 LEFT as described on the product label. | JWH2 products |
| Z-0259-2025Class II | Princeton Biomeditech CorpThis recall was initiated due to customer complaints of false negative of hCG test results when running patient urine samples. Use of the above listed product lot may result in false negative results. An end user report… | JHI1 product |
| Z-0261-2025Class II | Ambu Inc.Ambu has received complaints on Ambu VivaSight 2 DLTs regarding hyper angulation of the distal end. The hyper angulation can lead to an increased risk of complications during intubation and potential airway injury. | CBI1 product |
| Z-0262-2025Class II | Integra LifeSciences Corp.Out of specification endotoxin levels. | GXQ1 product |
| Z-0263-2025Class II | Westmed, Inc.Due to connector within the circuit flex extender not meeting specifications resulting in a low connection force and as a result, the connector may leak or disconnect and may cause life threating conditions or death. | BZA1 product |
| Z-0264-2025Class II | Olympus Corporation of the AmericasThere was an increase in complaints indicating an occurrence of system error E1226 "Could not write to Audio Buffer" during procedures using the ESG-410 (SEP). | GEI1 product |
| Z-0265-2025Class II | Beckman Coulter, Inc.IFUs for the device incorrectly states the biotin interference was remediated. The issue could result in potentially false-low results. | JNL1 product |
| Z-0266-2025Class II | Remel, IncA color change of the solution within the dropper can occur from orange-yellow to light yellow resulting in out-of-specification fluorescent stain results. | JFO1 product |
| Z-0267-2025Class II | Nihon Kohden America IncDue to oximeters not having FDA market approval or clearance to distribute in the U.S. | DQA2 products |
| Z-0269-2025Class II | King Systems Corp. dba Ambu, Inc.Exceed the limitations to the 510 (k) exemption in 21 CFR 868.9(a) because it introduces a new indication for pediatric use in an Oropharyngeal Airway. | CAE8 products |
| Z-0277-2025Class II | Smith & Nephew IncComplaint received indicating that a package contained a LGN OX CONSTRAINED FEM 3 LT implant instead of a LGN OX CONSTRAINED FEM 4 LT as described on the product label. | JWH2 products |
| Z-0233-2025Class II | Abbott Molecular, Inc.Abbott has identified an increase of incidences regarding Error Code (EC) 9198 (Positive control is non-reactive) while using the Alinity m HR HPV AMP Kit and Alinity m STI AMP Kit. Certain invalidated positive assay co… | MAQ QEP2 products |
| Z-0235-2025Class II | Integra LifeSciences Corp.Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods | HBA22 products |
| Z-0257-2025Class II | Covidien LPSpecific lots of the Chameleon PTA Balloon Catheter failed to meet in-house balloon pressure testing. | LIT1 product |
| Z-0288-2025Class III | Iotamotion IncIncorrect GTIN number. | QQH1 product |
| Z-0260-2025Class III | Trackx Technology IncThe patient label stickers may have the wrong expiration date recorded. The contract manufacturer entered the manufacturing date (2024-04-01) as the expiration date. The actual expiration date of the product is 2027-04-… | OWB1 product |
| Z-0258-2025Class III | DRG International, Inc.The High and Low Control acceptance ranges in the Certificate of Analysis (CofA) did not match the acceptance range information on the High and Low Control Component labels. Due to the defect, the device failed to provi… | JLX1 product |
De Novo grants
New device types the FDA classified through the De Novo pathway. Each one creates a product code and becomes a possible predicate for later 510(k)s.
Original PMA approvals
Original premarket approvals for Class III devices. The FDA also decided 31 PMA supplements this week, which change devices already approved.
No original PMAs were approved this week.
510(k) clearances
63 510(k) clearances by review panel. The median took 112 calendar days from receipt to decision; the fastest, K243381, 2 days; the longest, K240177, 281 days.
Calendar days include the time a submission waits on the company, so they run longer than the FDA’s 90-day review goal. How to search the 510(k) database
- Orthopedic15
- General, Plastic Surgery8
- Cardiovascular7
- Gastroenterology, Urology7
- Radiology5
- Dental4
- General Hospital3
- Anesthesiology2
- Clinical Chemistry2
- Neurology2
- Physical Medicine2
- Clinical Toxicology1
- Ear, Nose, Throat1
- Microbiology1
- Ophthalmic1
- Panel not stated1
- Pathology1
Every clearance
| Number | Device and applicant | Product code | Decision and review time |
|---|---|---|---|
| K240375 | V80 Anesthetic Vaporizer (V80)Shenzhen Mindray Bio-Medical Electronics Co., Ltd. | CADAnesthesiology | Oct 31, 2024267 days |
| K241366 | QualityFlow O2 Series (QualityFLOW O2, QualityFLOW O2 MTV)Dehas Medical Systems GmbH | CBPAnesthesiology | Oct 29, 2024168 days |
| K240177 | Zio AT® device (A100A1001)iRhythm Technologies, Inc. | DSICardiovascular | Oct 30, 2024281 days |
| K240366 | EPXT™ Unidirectional Steerable Diagnostic Catheter; Dynamic Tip™ Unidirectional Steerable Diagnostic Catheter; Dynamic XT™ Unidirectional Steerable Diagnostic CatheterBoston Scientific Corporation | DRFCardiovascular | Nov 1, 2024269 days |
| K240653 | SmartCardia 7L Platform (MCT)Smartcardia SA | QYXCardiovascular | Oct 31, 2024238 days |
| K241828 | Affera Integrated Mapping System; Impedance Localization Patch KitMedtronic, Inc. | DQKCardiovascular | Nov 1, 2024130 days |
| K242076 | VersaCross™ RF WireBaylis Medical Company | DXFCardiovascular | Nov 1, 2024108 days |
| K242623 | Disposable Blood Pressure cuff; Reusable Blood Pressure cuffShenzhen Medke Technology Co., Ltd. | DXQCardiovascular | Oct 31, 202458 days |
| K243080 | Riptide Aspiration SystemMicro Therapeutics, Inc. d/b/a ev3 Neurovascular | NRYCardiovascular | Oct 29, 202429 days |
| K240455 | BD Vacutainer® Citrate Blood Collection TubesBecton, Dickinson and Company | GIMClinical Chemistry | Oct 31, 2024259 days |
| K242320 | BD Vacutainer® One Use HolderBecton, Dickinson and Company | JKAClinical Chemistry | Nov 1, 202488 days |
| K240670 | DRI Ecstasy Plus AssayMicrogenics Corporation | DKZClinical Toxicology | Oct 30, 2024233 days |
| K240383 | Highness Implant SystemHighness Co., Ltd. | DZEDental | Oct 31, 2024266 days |
| K240560 | ZENEX Implant System_LongIzenimplant Co., Ltd. | DZEDental | Oct 30, 2024244 days |
| K242591 | XERO-alphaDRTECH Corporation | EHDDental | Nov 1, 202463 days |
| K242675 | FineEtch (FineEtch-10 1.2ml, FineEtch-10 5ml, FineEtch-37 1.2ml FineEtch-37 5ml, FineEtch 37% 1.2ml 20ea pack)Spident Co., Ltd. | KLEDental | Nov 1, 202456 days |
| K241612 | Video Bronchoscope System; Single-use Flexible Bronchoscope (BC-S1E00-L, BC-S1H00-L, BC-S1J00-L, BC-S1E00, BC-S1H00, BC-S1J00.); Digital Video Monitor (DVM-D1, DVM-D2)Hunan Vathin Medical Instrument Co., Ltd. | EOQEar, Nose, Throat | Oct 29, 2024146 days |
| K240334 | LAGIS Endoscopic Instruments - Grasper (DI-5670-33D, DI-5670-45D, DI-5716-33D, DI-5716-45D, DI-5719-33D, DI-5731-33D, DI-5731-45D, DI-5901-33D, DI-5901-45D)Lagis Enterprise Co., Ltd. | GCJGastroenterology, Urology | Oct 30, 2024268 days |
| K240514 | AMT Low-Profile Suprapubic Catheter & Drainage SetApplied Medical Technology, Inc. | KOBGastroenterology, Urology | Nov 1, 2024253 days |
| K240847 | MedicCO2LON (MedicCO2LON)Ide Vision, Ltd. | FCXGastroenterology, Urology | Oct 30, 2024217 days |
| K241533 | NeurAxis IB-Stim (01-1020)Neuraxis | QHHGastroenterology, Urology | Oct 30, 2024153 days |
| K242269 | SC+ Hemodialysis Device (SC-14269); Dialysate Cartridge (SC-14656); Blood Tube Set (SC-14651)Quanta Dialysis Technologies, Ltd. | ONWGastroenterology, Urology | Nov 1, 202492 days |
| K243091 | Over-tube (TR-1108A)Fujifilm Corporation | FEDGastroenterology, Urology | Nov 1, 202432 days |
| K243095 | Mammotome AutoCore™ Single Insertion Core Biopsy SystemDevicor Medical Products, Inc. | KNWGastroenterology, Urology | Oct 29, 202429 days |
| K240333 | Wearable Digital Thermometer (T31)Guangdong Genial Technology Co., Ltd. | FLLGeneral Hospital | Oct 30, 2024268 days |
| K240881 | Signia™ Circular Adapter (Standard Length) (SIGCIRSTND); Signia™ Circular Adapter XL Length (SIGCIRXL)Covidien (Part of Medtronic) | GAGGeneral Hospital | Nov 1, 2024214 days |
| K242328 | PowerPort™ ClearVUE™ Slim Implantable Ports and PowerPort™ ClearVUE™ Slim Implantable PortsBard Access Systems, Inc. | LJTGeneral Hospital | Oct 31, 202486 days |
| K241630 | ECHELON 4000 60mm Compact Stapler (EC3D60C); ECHELON 4000 60mm Standard Stapler (EC3D60S); ECHELON 4000 60mm Long Stapler (EC3D60L); ECHELON 3D 60mm White Reload (ER60W); ECHELON 3D 60mm Blue Reload (ER60B); ECHELON 3D 60mm Green Reload (ER60G); ECHELON 3D 60mm Black Reload (ER60T)Ethicon Endo-Surgery, LLC | GDWGeneral, Plastic Surgery | Oct 28, 2024144 days |
| K241646 | ORTIV (ORTIV-BL, ORTIV-WH)Mvitro Co., Ltd. | GEXGeneral, Plastic Surgery | Nov 1, 2024147 days |
| K241718 | Micro-current Facial Beauty Device (AM-810B, AM-810W, AM-812B, AM-812W)Shenzhen Aozemei Technology Co., Ltd. | OHSGeneral, Plastic Surgery | Oct 28, 2024136 days |
| K241902 | Edison SystemHistoSonics, Inc. | QGMGeneral, Plastic Surgery | Oct 30, 2024121 days |
| K242622 | Sterile Lancets for Single UseNingbo Caremed Medical Products Co., Ltd. | QRKGeneral, Plastic Surgery | Oct 28, 202455 days |
| K242627 | Safety LancetNingbo Medsun Medical Co., Ltd. | FMKGeneral, Plastic Surgery | Oct 30, 202457 days |
| K243126 | TRACHEOBRONXANE™ DUMON®Novatech SA | NWAGeneral, Plastic Surgery | Oct 29, 202429 days |
| K243217 | Legend X Desktop SystemPollogen, Ltd. | GEIGeneral, Plastic Surgery | Nov 1, 202429 days |
| K242256 | QIAstat-Dx Meningitis/Encephalitis (ME) PanelQIAGEN GmbH | PLOMicrobiology | Oct 29, 202490 days |
| K242271 | Caduceus STaiwan Main Orthopaedic Biotechnology Co., Ltd. | SBFNeurology | Oct 30, 202490 days |
| K243157 | AtriCure cryoICE BOX (ACM)AtriCure, Inc. | GXHNeurology | Oct 29, 202429 days |
| K242393 | Acuity 181™ (tisilfocon A) and Acuity 181™ (tisilfocon A) with Tangible™ Hydra-PEG® Rigid Gas Permeable Contact LensesAcuity Polymers, Inc. | HQDOphthalmic | Nov 1, 202480 days |
| K240175 | Tigon Medical Fractures, Instability, and Reconstruction (FIRE)Tigon Medical | HWCOrthopedic | Oct 29, 2024280 days |
| K240388 | Osteotomy Truss System (OTS)4Web Medical, Inc. | HRSOrthopedic | Nov 1, 2024267 days |
| K241077 | DeGen Medical Patient Specific Implant (PSI) SystemDegen Medical | OVDOrthopedic | Oct 29, 2024193 days |
| K241992 | Catalyft™ LS Expandable Interbody SystemMedtronic Sofamor Danek USA, Inc. | OVDOrthopedic | Oct 28, 2024112 days |
| K241999 | Prophecy Surgical Planning SystemWright Medical Technology, Inc. (Stryker) | HSNOrthopedic | Nov 1, 2024115 days |
| K242311 | EXPERT - Flexible Joint Fixation SystemGM Dos Reis Industria e Comercio Ltda. | HTNOrthopedic | Nov 1, 202488 days |
| K242350 | Spinal Alignment Solutions Pelvic Incidence (PI) Rod SystemSpinal Alignment Solutions, Inc. | NKBOrthopedic | Oct 28, 202481 days |
| K242399 | Bonebridge Osteosynthesis Plating SystemBonebridge AG | HRSOrthopedic | Oct 31, 202479 days |
| K242452 | Monkey Bars Pin to Bar External Fixation SystemParagon 28, Inc. | KTTOrthopedic | Nov 1, 202474 days |
| K242665 | DePuy ATTUNE™ Total Knee SystemDepuy Orthopedics, Inc. | OIYOrthopedic | Nov 1, 202458 days |
| K242711 | JOURNEY II Unicompartmental Knee System (JOURNEY II UK)Smith & Nephew, Inc. | HSXOrthopedic | Oct 30, 202451 days |
| K243137 | Curiteva Porous PEEK Laminoplasty SystemCuriteva, Inc. | NQWOrthopedic | Oct 28, 202428 days |
| K243203 | OC JuggerKnotless Soft Anchor; OC JuggerLoop Soft AnchorRiverpoint Medical | MBIOrthopedic | Oct 29, 202428 days |
| K243247 | Persona the Personalized Knee SystemZimmer, Inc. | MBHOrthopedic | Oct 29, 202418 days |
| K243381 | Total Wrist Arthroplasty System (TWA)Skeletal Dynamics, Inc. | JWJOrthopedic | Nov 1, 20242 days |
| K243181 | Foundation Dermal Regeneration Scaffold Plus (DRS+) SoloBionova Medical, Inc. | FROPanel not stated | Oct 31, 202431 days |
| K240303 | MetaLite DX Digital Pathology SoftwareJellox Biotech, Inc. | QKQPathology | Oct 28, 2024269 days |
| K241189 | Electric Wheelchair (ME23-EW16-BC200-01, ME23-EW16-BC200-02)Jiaxing Xiaoxiang Medical Equipment Technology Co., Ltd. | ITIPhysical Medicine | Oct 29, 2024183 days |
| K242620 | Breo Laser Hair Growth Comb (Scalp 3L)Shenzhen Breo Technology Co., Ltd. | OAPPhysical Medicine | Nov 1, 202459 days |
| K241068 | uDR 780iShanghai United Imaging Healthcare Co., Ltd. | KPRRadiology | Nov 1, 2024196 days |
| K241108 | RemedyLogic AI MRI Lumbar Spine ReaderRemedy Logic, Inc. | QIHRadiology | Oct 30, 2024191 days |
| K241223 | eRAD PACSErad, Inc. | LLZRadiology | Oct 31, 2024183 days |
| K241965 | Catheter Surface Flap Applicator Set (GM11011830)Varian Medical Systems | JAQRadiology | Oct 29, 2024118 days |
| K243186 | Symbia Pro.specta Q3 (11364751); Symbia Pro.specta X3 (11364752); Symbia Pro.specta X7 (11364753); Symbia Pro.specta VA30A FamilySiemens Medical Solutions USA, Inc. | KPSRadiology | Oct 30, 202430 days |
Only see the weeks that matter to you
DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and every day it checks openFDA and tells you about new clearances, recalls and MAUDE reports for them. It also puts all of this beside the FDA’s own record pages.
Only your product codes, every Monday
One short email after each openFDA update with just the decisions and recalls in up to three product codes you choose. Free.
Earlier weeks
Each week keeps its own page.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Decisions arrive in weekly updates; check the FDA record before you rely on a figure.
- device/510k · search decision_date:[20241028 TO 20241103]Retrieved Nov 29, 2026, 11:17 PM UTC. FDA data updated Nov 16, 2026.
- device/pma · search decision_date:[20241028 TO 20241103]Retrieved Nov 29, 2026, 11:17 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search event_date_posted:[20241028 TO 20241103]Retrieved Nov 29, 2026, 11:17 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-0283-2025" "Z-0284-2025" "Z-0285-2025" "Z-0286-2025" "Z-0287-2025" "Z-0288-2025" "Z-0140-2025" "Z-0141-2025" "Z-0222-2025" "Z-0279-2025" "Z-0280-2025" "Z-0281-2025" "Z-0282-2025" "Z-0208-2025" "Z-0259-2025" "Z-0260-2025" "Z-0261-2025" "Z-0262-2025" "Z-0263-2025" "Z-0264-2025" "Z-0265-2025" "Z-0266-2025" "Z-0267-2025" "Z-0268-2025" "Z-0269-2025" "Z-0270-2025" "Z-0271-2025" "Z-0272-2025" "Z-0273-2025" "Z-0274-2025" "Z-0275-2025" "Z-0276-2025" "Z-0277-2025" "Z-0278-2025" "Z-0233-2025" "Z-0234-2025" "Z-0235-2025" "Z-0236-2025" "Z-0237-2025" "Z-0238-2025" "Z-0239-2025" "Z-0240-2025" "Z-0241-2025" "Z-0242-2025" "Z-0243-2025" "Z-0244-2025" "Z-0245-2025" "Z-0246-2025" "Z-0247-2025" "Z-0248-2025")Retrieved Nov 29, 2026, 11:17 PM UTC. FDA data updated Nov 18, 2026.
- device/enforcement · search recall_number:("Z-0249-2025" "Z-0250-2025" "Z-0251-2025" "Z-0252-2025" "Z-0253-2025" "Z-0254-2025" "Z-0255-2025" "Z-0256-2025" "Z-0257-2025" "Z-0258-2025")Retrieved Nov 29, 2026, 11:17 PM UTC. FDA data updated Nov 18, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil