Weekly archive
Week of September 30, 2024
The FDA cleared 61 510(k)s, granted 1 De Novo request and approved 1 original PMA. 25 device recalls were posted, 2 of them Class I. Every 510(k), De Novo and PMA decision dated September 30 to October 6, 2024, and every device recall the FDA posted those days.
- 510(k) clearances
- 61median review 99 days
- De Novo grants
- 1
- Original PMA approvals
- 1plus 52 supplements
- Recalls posted
- 2569 products, 2 Class I
Recalls posted
Device recalls the FDA posted September 30 to October 6, 2024, Class I first. Class I is the most serious: a reasonable chance the product causes serious harm or death. The class comes from the FDA’s weekly enforcement report, so the newest recalls can show “Class pending” for a few days.
| Recall and class | Firm and reason | Product code |
|---|---|---|
| Z-0001-2025Class I | Medtronic MiniMed, Inc.Insulin pumps that have been dropped, bumped, or experienced physical impact may have damage to internal electrical components, which may cause reduced pump battery life. Since the defect reduces the battery life overal… | OYC OZO OZP3 products |
| Z-0004-2025Class I | Mercury Enterprises, Inc. dba Mercury MedicalManual resuscitator circuit flow controller may come apart due to a missing manufacturing step. | BTL1 product |
| Z-0020-2025Class II | Philips North America LlcDevices with Option CP2 and initially programmed to software version P.01.01 were configured incorrectly at the factory. Therefore, Enhanced ECG Capabilities provided by Option CP2 were not enabled in these devices. Wit… | MHX1 product |
| Z-0009-2025Class II | Becton, Dickinson and Company, BD Bio SciencesSample preparation system with software may cause: 1) Multidispense feature enabled with 2-tube assay: Specimen volume dispensed into tube1 at least 3% higher than volume in tube2, or 2) Absolute counts calculated using… | PER1 product |
| Z-0019-2025Class II | Outset Medical, Inc.Blood leak sensor, a component of hemodialysis device, may show decreased sensitivity, at the lowest level, specified as 0.35 mL/min blood leak rate, which could delay triggering the ALARM_DIALYZER_BLOOD_LEAK alarm, if… | KDI1 product |
| Z-0006-2025Class II | Ossur H / FPressure sensitive adhesive hooks attached to the sides of the anterior cervical orthosis collar panel can occasionally exhibit poor adhesion, which may lead to them gradually peeling off with repeated doffing, possibly… | IQK1 product |
| Z-0007-2025Class II | Philips North AmericaThe 866173 lntelliVue G7m Anesthesia Gas Modules could have a component issue that may trigger an interruption of the gas measurement. This issue concerns the G7m's sample gas pump which could stop working, producing a… | DSI1 product |
| Z-0008-2025Class II | Young Dental Manufacturing I, LLCThe bulk product used in the recalling firm's repackaged finished product was recalled due to bacterial contamination. | FLG1 product |
| Z-0224-2024Class II | Philips North AmericaPhilips has become aware of an issue with the Philips Integris-Allura, Allura Xper and Azurion systems that have been delivered to Puerto Rico without a warning label and the Source-to-Skin Distance Spacer (Tube-spacer)… | OWB6 products |
| Z-0533-2024Class II | MEDLINE INDUSTRIES, LP - NorthfieldUndeclared latex | FAQ3 products |
| Z-0622-2024Class II | DeRoyal Industries IncKit were manufactured and distributed with Edwards Life Sciences TruWave Disposable pressure transducers which were subsequently recalled by Edwards Life Sciences. | LRO1 product |
| Z-1157-2024Class II | MEDLINE INDUSTRIES, LP - NorthfieldPackaging material was updated for the Bulb Irrigation Syringe (DYND20125), and additional testing done at Medline Corporate indicated the potential for the sterile barrier to be breached during transportation. This onl… | KYZ1 product |
| Z-1276-2024Class II | ConvaTec, IncThe kit contains a flange that has a 45mm coupling ring, however, the pouch within the kit is for a 57mm coupling ring. | FON1 product |
| Z-1286-2023Class II | Randox LaboratoriesThe Antibody (R2 reagent) is showing a positive bias compared to previous batches on patient samples. Quality Controls do not show the same bias and will be within range | DCK1 product |
| Z-1304-2023Class II | Instrumentation LaboratoryRemoves the test definition for HemosIL Liquid Anti-Xa (PN 0020302600 and 0020302601) from the ACL Elite/Elite Pro instruments | KFF4 products |
| Z-1647-2023Class II | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Potential Loss of X-ray-When the full disk capacity is reached, X-Ray functionality will cease to be available there will be a sudden interruption of the procedure | OWB1 product |
| Z-2069-2023Class II | TEI Biosciences, Inc.Possible out of specification endotoxin test results due to issues with in-process and finished goods endotoxin testing. | FTM KGN17 products |
| Z-2415-2023Class II | Carefusion 2200 IncSpecific lots of the product have the potential to exhibit cracking or breaking of the rotation knob due to the incorrect use of an adhesive in the manufacturing process. | GCJ15 products |
| Z-2588-2023Class II | BioFire Diagnostics, LLCThe firm has identified an increased risk of obtaining false positive Cryptosporidium results from customers using the BIOFIRE FILMARRAY Gastrointestinal (GI) Panel | PCH1 product |
| Z-2616-2023Class II | ETHICON, LLCInternal testing on returned product from this lot confirmed that some PDS" II (polydioxanone) Sutures from this lot did not meet tensile strength requirement. | NEW2 products |
| Z-2620-2023Class II | Copan Diagnostics IncDue to plastic defects, the pipette aspiration capability potentially doesn't meet the expected tolerance | JRC1 product |
| Z-3310-2024Class II | CooperVision, Inc.Manufactured with a misaligned axis resulting in lenses with the incorrect power, may experience poor visual acuity | LPM2 products |
| Z-3312-2024Class II | Philips North America LlcIf the DIVA touch display generates touch inputs without user interaction, due to possibility of the erroneous acknowledgement of a patient alarm, there is a potential for a delay in the detection of a change in a patie… | MHX1 product |
| Z-3313-2024Class II | Alcon Research LLCMaterial with a quality issue was used in contact lens production, so contact lenses may not meet quality and/or performance standards for the entirety of their labeled shelf life, and wearing these lenses may lead to t… | LPL1 product |
| Z-0595-2024Class III | Maquet Medical Systems USACARDIOHELP-i System was not properly tested to measure leakage current | DTQ1 product |
Original PMA approvals
Original premarket approvals for Class III devices. The FDA also decided 52 PMA supplements this week, which change devices already approved.
510(k) clearances
61 510(k) clearances by review panel. The median took 99 calendar days from receipt to decision; the fastest, K242721, 24 days; the longest, K210933, 1,281 days.
Calendar days include the time a submission waits on the company, so they run longer than the FDA’s 90-day review goal. How to search the 510(k) database
- Orthopedic11
- Radiology10
- Cardiovascular8
- Gastroenterology, Urology7
- Dental6
- General, Plastic Surgery5
- Neurology4
- Microbiology3
- Clinical Toxicology2
- General Hospital2
- Physical Medicine2
- Clinical Chemistry1
Every clearance
| Number | Device and applicant | Product code | Decision and review time |
|---|---|---|---|
| K240529 | 8F Modified Sheath SystemRoute 92 Medical, Inc. | QJPCardiovascular | Oct 2, 2024222 days |
| K240808 | Pulse Oximeter (WS20A)Hunan Accurate Bio-Medical Technology Co., Ltd. | DQACardiovascular | Oct 3, 2024192 days |
| K240917 | Esperance 3+ Aspiration Catheter SystemWallaby Medical | NRYCardiovascular | Oct 1, 2024181 days |
| K241746 | UMED REPROCESSED AIRCAST VENAFLOW FOOT CUFF; UMED REPROCESSED KENDALL SCD EXPRESS SLEEVE; UMED REPROCESSED AIRCAST ELITE CALF CUFF 3040 3050; UMED REPROCESSED CTC VP 501 CALF CUFF; UMED REPROCESSED CTC VP 520 FOOT CUFFUnited Medical Resources, LLC | JOWCardiovascular | Oct 4, 2024108 days |
| K242106 | AuST Steerable SheathCenterPoint Systems, LLC | DYBCardiovascular | Oct 4, 202478 days |
| K242276 | CrossFAST™ Integrated Microcatheter Guide Extension System (CFM55)Vantis Vascular, Inc. | DQYCardiovascular | Sep 30, 202460 days |
| K242507 | OBSIDIO™ Conformable EmbolicBoston Scientific | KRDCardiovascular | Oct 3, 202442 days |
| K242721 | Upper Arm Electronic Blood Pressure Monitor (FC-BP103, FC-BP106, FC-BP113, FC-BP120, FC-BP121, FC-BP130, FC-BP131)Shenzhen Finicare Co., Ltd. | DXNCardiovascular | Oct 4, 202424 days |
| K210933 | YSI 2900C Biochemistry AnalyzerYsi, Inc. | CGAClinical Chemistry | Sep 30, 20241,281 days |
| K241869 | BioSieve™ Fentanyl FIA Home Test Kit; BioSieve™ Fentanyl FIA Pro Test Kit; BioSieve™ Toxismart ReaderVivachek Biotech (Hangzhou) Co., Ltd. | NGLClinical Toxicology | Oct 4, 202499 days |
| K242498 | Pocguide Multi-Drug Test Panel, Pocguide Multi-Drug Test Panel OTCAicheck Biotech, Inc. | NFTClinical Toxicology | Oct 1, 202440 days |
| K233497 | Rest Assure SystemSomnomed, Inc. | PLCDental | Oct 3, 2024338 days |
| K240153 | Ceramic Etchant (HF-5 and HF-9)Liaoning Upcera Co., Ltd. | KLEDental | Oct 2, 2024257 days |
| K240586 | Additive Manufacturing Zirconia Customized RestorationHangzhou Thales Medtech Co., Ltd. | ELLDental | Oct 3, 2024216 days |
| K241151 | LUXEN 5GDentalmax Co., Ltd. | EIHDental | Sep 30, 2024158 days |
| K241977 | Injectable Root Canal Bioceramic Sealer (i-MTA SP)Longly Biotechnology (Wuhan) Co., Ltd. | KIFDental | Oct 3, 202490 days |
| K242277 | Crown HTSprintray, Inc. | EBFDental | Oct 1, 202461 days |
| K240074 | Single use flexible ureteroscope (UR-D1, UR-D2, UR-F1, UR-F2)Shenzhen Besdata Technology Co., Ltd. | FGBGastroenterology, Urology | Oct 4, 2024268 days |
| K240675 | Endoscopic Injection NeedlesPromisemed Hangzhou Meditech Co., Ltd. | FBKGastroenterology, Urology | Oct 4, 2024207 days |
| K241052 | Bladeless Trocar; Bladed Trocar; Hasson TrocarAdvamedix, GmbH | GCJGastroenterology, Urology | Oct 4, 2024170 days |
| K241285 | Disposable Endoscope Valves Set (AF series)Alton (Shanghai) Medical Instruments Co., Ltd. | ODCGastroenterology, Urology | Oct 1, 2024147 days |
| K241434 | Video Ureteroscope SystemHunan Vathin Medical Instrument Co., Ltd. | FGBGastroenterology, Urology | Oct 4, 2024136 days |
| K241796 | TroCare TroKit Laparoscope Lens Wiper (CP000626)TroCare, LLC | GCJGastroenterology, Urology | Oct 4, 2024105 days |
| K242325 | GripTract-GI Endoscopic Tissue Manipulator Lower GI ModelsActuated Medical, Inc. | FDFGastroenterology, Urology | Oct 4, 202459 days |
| K241836 | Kem Medical Lead-free Chemical Indicators for Steam SterilizationKem Medical Products Corporation | JOJGeneral Hospital | Sep 30, 202497 days |
| K242660 | Curlin 8000 Ambulatory Infusion System; Enterprise Solution (ES) Software; Curlin 380-series Administration SetsZevex, Inc. | FRNGeneral Hospital | Oct 3, 202429 days |
| K240376 | Electroporation System (N3000)Surgnova Healthcare Technologies (Zhejiang) Co., Ltd. | OABGeneral, Plastic Surgery | Oct 4, 2024240 days |
| K241907 | Anovo Surgical System (model 6N)Momentis Surgical , Ltd. | QNMGeneral, Plastic Surgery | Oct 2, 202493 days |
| K241952 | AQUABEAM Robotic System (AB2000)Procept Biorobotics | PZPGeneral, Plastic Surgery | Sep 30, 202489 days |
| K242302 | RejuvaKnee™ Collagen Meniscus ImplantCollagen Matrix, Inc. | OLCGeneral, Plastic Surgery | Oct 2, 202458 days |
| K242679 | Disposable Powered Articulating Endoscopic Linear Cutter Stapler & Reloads (30mm series, 45mm series, 60mm series)Fengh Medical Co., Ltd. | GDWGeneral, Plastic Surgery | Oct 4, 202428 days |
| K241317 | Wondfo 2019-nCoV Antigen Test (Lateral Flow Method)Guangzhou Wondfo Biotech Co., Ltd. | QYTMicrobiology | Sep 30, 2024143 days |
| K241967 | The Sensititre 20 - 24 hour Haemophilus influenzae/Streptococcus pneumoniae (HP) MIC or Breakpoint Susceptibility System with ceftriaxone in the dilution range of 0.015 - 2 ug/ml.Thermo Fisher Scientific | JWYMicrobiology | Oct 1, 202490 days |
| K242613 | NxTAG® Respiratory Pathogen PanelLuminex Molecular Diagnostics, Inc. | OCCMicrobiology | Oct 2, 202429 days |
| K241298 | QUADSENSEEventum Orthopaedics, Ltd. | ONNNeurology | Oct 4, 2024149 days |
| K242154 | Numen™ Coil Embolization System; NumenFR™ Detachment SystemMicroPort NeuroTech (Shanghai) Co., Ltd. | HCGNeurology | Sep 30, 202469 days |
| K242351 | Zeta Cranial Navigation System (ZNS131-US)Zeta Surgical, Inc. | HAWNeurology | Oct 4, 202457 days |
| K242616 | Lantern® HipOrthalign, Inc. | ONNNeurology | Oct 3, 202430 days |
| K240393 | Exactech® TRULIANT® Knee SystemExactech, Inc. | MBHOrthopedic | Oct 3, 2024238 days |
| K240889 | ARTFX Lumbar PEEK CagesArtfx Medical | MAXOrthopedic | Oct 2, 2024184 days |
| K241097 | Arthrex Virtual Implant Positioning (VIP) System SoftwareArthrex, Inc. | QHEOrthopedic | Oct 2, 2024163 days |
| K241472 | Icona Hip StemCorin U.S.A. Limited | MEHOrthopedic | Oct 3, 2024132 days |
| K241487 | Interwedge® Standalone LateralFoundation Surgical Group, Inc. | OVDOrthopedic | Oct 2, 2024131 days |
| K241657 | MSFX MIKRON SPINAL FIXATION SYSTEMMikron Makina Sanayi VE Ticaret Ltd. Sti. | NKBOrthopedic | Sep 30, 2024112 days |
| K242343 | BPS Wrist Fracture SystemOrthonovis, Inc. | HRSOrthopedic | Oct 4, 202458 days |
| K242364 | IdentiTi™ II Interbody SystemAlphatec Spine, Inc. | MAXOrthopedic | Oct 4, 202456 days |
| K242377 | REMEDY Stemmed Knee Spacer System with All-Poly Tibial ComponentOsteoremedies, LLC | JWHOrthopedic | Oct 2, 202454 days |
| K242445 | Pangea Platform; Pangea Femur Plating System; Pangea Humerus Plating System; Pangea Tibia Plating SystemStryker GmbH | HRSOrthopedic | Oct 6, 202451 days |
| K242671 | Treace Medical Concepts (TMC) Screw Fixation SystemTreace Medical Concepts, Inc. | HWCOrthopedic | Oct 4, 202428 days |
| K240331 | Electric wheelchair (YHWL002, YHWL003)Yongkang Youha Electric Appliance Co., Ltd. | ITIPhysical Medicine | Oct 1, 2024239 days |
| K241854 | Power WheelchairAnhui Jbh Medical Apparatus Co., Ltd. | ITIPhysical Medicine | Oct 4, 202499 days |
| K240553 | LVivo Software ApplicationDia Imaging Analysis, Ltd. | QIHRadiology | Oct 4, 2024219 days |
| K240943 | LungVisionBodyvision Medical , Ltd. | QIHRadiology | Oct 1, 2024179 days |
| K241245 | EchoSolv ASEcho IQ, Ltd. | POKRadiology | Oct 4, 2024154 days |
| K241331 | MuscleViewSpringbok, Inc. | LNHRadiology | Oct 1, 2024144 days |
| K241561 | MammoScreen BDTherapixel | QIHRadiology | Oct 2, 2024124 days |
| K241925 | VitruvianScan (v1.0)Perspectum, Ltd. | LNHRadiology | Oct 2, 202493 days |
| K242005 | Versana Premier; Versana Premier Lotus; LOGIQ FGE Medical Systems Ultrasound and Primary Care Diagnostics | IYNRadiology | Oct 2, 202485 days |
| K242651 | GM85Samsung Electronics Co., Ltd. | IZLRadiology | Oct 1, 202427 days |
| K242652 | Lunit INSIGHT DBT v1.1Lunit, Inc. | QDQRadiology | Oct 4, 202430 days |
| K242678 | Definium Pace Select ETGe Hualun Medical Systems Co. , Ltd. | KPRRadiology | Oct 1, 202425 days |
Only see the weeks that matter to you
DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and every day it checks openFDA and tells you about new clearances, recalls and MAUDE reports for them. It also puts all of this beside the FDA’s own record pages.
Only your product codes, every Monday
One short email after each openFDA update with just the decisions and recalls in up to three product codes you choose. Free.
Earlier weeks
Each week keeps its own page.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Decisions arrive in weekly updates; check the FDA record before you rely on a figure.
- device/510k · search decision_date:[20240930 TO 20241006]Retrieved Nov 29, 2026, 5:05 AM UTC. FDA data updated Nov 16, 2026.
- device/pma · search decision_date:[20240930 TO 20241006]Retrieved Nov 29, 2026, 5:05 AM UTC. FDA data updated Nov 16, 2026.
- device/recall · search event_date_posted:[20240930 TO 20241006]Retrieved Nov 29, 2026, 5:05 AM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-0020-2025" "Z-0001-2025" "Z-0002-2025" "Z-0003-2025" "Z-0009-2025" "Z-0019-2025" "Z-0004-2025" "Z-0006-2025" "Z-0007-2025" "Z-0008-2025" "Z-0224-2024" "Z-0226-2024" "Z-0227-2024" "Z-0228-2024" "Z-0229-2024" "Z-0230-2024" "Z-0533-2024" "Z-0534-2024" "Z-0535-2024" "Z-0595-2024" "Z-0622-2024" "Z-1157-2024" "Z-1276-2024" "Z-1286-2023" "Z-1304-2023" "Z-1305-2023" "Z-1306-2023" "Z-1307-2023" "Z-1647-2023" "Z-2069-2023" "Z-2070-2023" "Z-2071-2023" "Z-2072-2023" "Z-2073-2023" "Z-2074-2023" "Z-2075-2023" "Z-2076-2023" "Z-2077-2023" "Z-2078-2023" "Z-2079-2023" "Z-2080-2023" "Z-2081-2023" "Z-2082-2023" "Z-2083-2023" "Z-2084-2023" "Z-2085-2023" "Z-2415-2023" "Z-2416-2023" "Z-2417-2023" "Z-2418-2023")Retrieved Nov 29, 2026, 5:05 AM UTC. FDA data updated Nov 18, 2026.
- device/enforcement · search recall_number:("Z-2419-2023" "Z-2420-2023" "Z-2421-2023" "Z-2422-2023" "Z-2423-2023" "Z-2424-2023" "Z-2425-2023" "Z-2426-2023" "Z-2427-2023" "Z-2428-2023" "Z-2429-2023" "Z-2588-2023" "Z-2616-2023" "Z-2617-2023" "Z-2620-2023" "Z-3310-2024" "Z-3311-2024" "Z-3312-2024" "Z-3313-2024")Retrieved Nov 29, 2026, 5:05 AM UTC. FDA data updated Nov 18, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil