Product codes

FDA product code

DRG: Transmitters And Receivers, Physiological Signal, Radiofrequency (FDA product code)

Follow DRG: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 870.2910
Review panel
Cardiovascular
Premarket route
510(k)
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
1922 in the last 12 months
Median review time
168 daysthe latest 25 decisions
Recall records
121 in the last 12 months
Adverse event reports
71last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code DRG per year, by decision date, 192 since 1977. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20177
20186
20195
20206
20212
20226
20234
20243
20253
2026 (so far)2

Latest decisions

NumberDevice and applicantDecision and review time
K261569 UbiqVue™ 2AYe Cardiac System (UX2253)Lifesignals, Inc. Jul 9, 202658 days
K252984 Perin Health System (PHD80060-2)Perin Health Devices, LLC Jan 22, 2026126 days
K252440 EasyTeleMed (2.0.2)Ingeniars S.R.L. Oct 2, 202559 days
K250259 TeleRehab Aermos Cardiopulmonary RehabilitationThe ScottCare Corporation Jun 4, 2025127 days
K243837 iBSMIorbit Digital Technologies Private Limited May 16, 2025154 days
K242018 UbiqVue™ 2A Multi-parameter System (UX2550)Lifesignals, Inc. Nov 12, 2024125 days
K241101 BioButton SystemBiointellisense, Inc. Sep 26, 2024157 days
K240251 ANNE ChestSibel Health, Inc. Jun 3, 2024125 days
K223711 ANNE OneSibel Health, Inc. Aug 10, 2023241 days
K220526 HD+12, HD+15Cardioline S.P.A May 15, 2023445 days

Recalls

Recall records for product code DRG per year, by the date the recall began. Each recalled product is one record; these 12 records belong to 11 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20172
20180
20190
20201
20211
20220
20230
20241
20250
2026 (so far)1

Latest recalls

Recall and classFirm and reasonBegan
Z-3013-2026Class II Midmark CorporationA communication anomoly may result in incorrect electrocardiogram waveforms of a patient and an inaccurate beats-per-minute calculation. Jul 20, 2026
Z-2616-2024Class II Biointellisense Inc.Software anomaly causing the skin temperature measurement accuracy to be slightly out of specification. Jul 19, 2024
Z-1162-2021Class II Nihon Kohden America IncIncorrect Date Stamp or No Data Transfer on Telemetry Receiver and Transmitter. Jan 8, 2021
Z-2537-2020Class II Vitalconnect Inc.A healthcare provider (HCP) noticed a patient had a lowered historic SpO2 reading that the software should have triggered a notification to the care team, but… May 21, 2020
Z-3132-2017Class II Boston Scientific CorporationThe device history record (DHR) was missing its test documentation for final HIPOT (high potential) electrical testing. May 15, 2017

Adverse event reports (MAUDE)

Medical device reports received per year that name product code DRG. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
2017484
2018166
201970
202082
2021107
202243
202319
202426
202553
2026 (so far)65

Last 12 months by event type

  • Deaths0
  • Injuries1
  • Malfunctions70
  • Other or not stated0

Most reported device problems, last 12 months

  1. Communication or Transmission Problem56
  2. Unable to Obtain Readings19
  3. Intermittent Communication Failure14
  4. Output Problem8
  5. Data Problem3
  6. Incorrect Measurement2

In plain English: what the newest reports describe Written by AI

  • Many reports describe the receiver intermittently showing 'Comm Loss' (communication loss), meaning the bedside transmitters briefly lose their link to the central station. In some reports all receiver cards or all telemetry units went into Comm Loss at once, and one said the outage lasted about 30 seconds and recurred roughly every 30 minutes.

    12 reports: 8030229-2026-07281, 8030229-2026-07282, 8030229-2026-06950, 8030229-2026-06946, 8030229-2026-06686, 8030229-2026-06690, 8030229-2026-06857, 8030229-2026-06606, 8030229-2026-06615, 8030229-2026-06993, 8030229-2026-06735, 8030229-2026-06755

  • Several reports describe a lit or blinking error light on the receiver, and one said it could not be seen at the central station. Power cycling and checking cables did not always clear it. In one report, replacing the receiver cards did not stop the problem, but swapping in a spare receiver did. Another report's technical support concluded the receiver itself was the cause and recommended repair.

    5 reports: 8030229-2026-07056, 8030229-2026-06993, 8030229-2026-06615, 8030229-2026-06755, 8030229-2026-06735

  • Two reports describe noise or artifacts in the signal. One says the receiver cards were showing noise, and the other says a transmitter displayed artifacts and sporadic respiration rate waveforms. Both also mention intermittent signal loss.

    2 reports: 8030229-2026-06682, 8030229-2026-06588

  • One report describes the central station showing the names of the last three patients who had used a transmitter instead of the current patient's name. The transmitter was swapping between these names.

    Report: 8030229-2026-07147

  • One report describes a patient who got an electric shock while plugging a non-manufacturer extension cord into an outlet. She had a small burn on her palm, reported an irregular heartbeat and headache, and her blood pressure reached 190/100. She believes the long-term ECG recorder she was wearing contributed to the shock.

    Report: 3011651603-2026-00003

Written by AI from the narratives of the 19 newest reports that have one, received May 3, 2026 to Oct 16, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
629in the FDA’s Global Unique Device Identification Database
Establishment listings
238registered establishments that list this product code

Watch DRG in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get DRG every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in DRG. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.