Weekly archive
Week of September 23, 2024
The FDA cleared 54 510(k)s, granted 1 De Novo request and approved 1 original PMA. 27 device recalls were posted, 2 of them Class I. Every 510(k), De Novo and PMA decision dated September 23 to 29, 2024, and every device recall the FDA posted those days.
- 510(k) clearances
- 54median review 119 days
- De Novo grants
- 1
- Original PMA approvals
- 1plus 57 supplements
- Recalls posted
- 27104 products, 2 Class I
Recalls posted
Device recalls the FDA posted September 23 to 29, 2024, Class I first. Class I is the most serious: a reasonable chance the product causes serious harm or death. The class comes from the FDA’s weekly enforcement report, so the newest recalls can show “Class pending” for a few days.
| Recall and class | Firm and reason | Product code |
|---|---|---|
| Z-3207-2024Class I | Percussionaire CorporationUpdated labeling to ensure that blue cap on in-line valve is removed, prior to delivery of (Intrapulmonary percussive ventilation) IPV therapy. | BZO2 products |
| Z-3153-2024Class I | Fresenius Kabi USA, LLCThe software has anomalies that have the potential to cause alarms, nonfunctioning pump, or unresponsive screen while continuing therapy. These could cause serious patient harm or death. | FRN1 product |
| Z-3303-2024Class II | Healgen Scientific LlcThe reason for this recall is Healgen Scientific Ltd. has determined that their Urinary Reagent Strip products have shown reduced sensitivity, or potential false negative results for the detection of urinary protein by… | JIR1 product |
| Z-3304-2024Class II | Siemens Healthcare Diagnostics, Inc.Positive bias in patient results at and below the cut-off of 60 U/mL with the lots listed in the table below when compared to alternate lots on the Atellica IM analyzer. | JZO2 products |
| Z-3306-2024Class II | FENG CHUN YUAN MEDICAL EQUIPMENT (SHENZHEN)CO.,LTDThe "DVOT brand" rapid test kits have not obtained the necessary authorization to be sold in the United States. | QYT3 products |
| Z-3309-2024Class II | Tornier, IncDevices from one lot of Perform" Reversed Inserts may contain a locking ring that was assembled upside down which would prevent the humeral insert from securely seating in the stem or spacer. | PAO1 product |
| Z-3264-2024Class II | TMJ Solutions IncPatient-fitted temporomandibular (TMJ) prothesis contained incorrect components, SterileEO | LZD1 product |
| Z-3265-2024Class II | Boston Scientific CorporationWallFlexTM Esophageal Stent System and AgileTM Esophageal Over the Wire (OTW) Stent System due to the potential for delivery catheter tip detachment. | ESW31 products |
| Z-3296-2024Class II | CooperSurgical, Inc.Low concentration of sodium pyruvate in the wash could compromise gamete development | MQL3 products |
| Z-3299-2024Class II | ARROW INTERNATIONAL Inc.Reports have been received regarding open seals on the packaging. | DQX1 product |
| Z-3300-2024Class II | Dr Franz Koehler Chemie GmbhPotential for particles to be present in solution. | KDL1 product |
| Z-3301-2024Class II | Cardinal Health 200, LLCThe firm received reports of breakage in the Anti-Reflux Valve (ARV) due to improper use. As a result, changes to labeling were made. | FEG KNT2 products |
| Z-3158-2024Class II | FUJIFILM Healthcare Americas CorporationWhen the key switch is turned on, the exposure conditions are set to the default values (85kV, 1.6 mAs, approx. 0.13mGy@ SID 1m) intended for adult chest imaging. In addition, if exposure conditions are set from the con… | IZL1 product |
| Z-3248-2024Class II | Hologic IncIdentified an issue with Selenia Dimensions 1.12.0 and 3Dimensions 2.3.0 system software when used in combination with C-View software in which the quality of the final C-View synthesized 2D image is impacted and may ap… | OTE1 product |
| Z-3249-2024Class II | Philips North AmericaDuring a production process, Philips became aware of one IntelliVue power supply with a broken ground bolt upon disconnection of the ground cable from the equipotential ground connector. Loss of electrical grounding may… | MHX1 product |
| Z-3250-2024Class II | Cook Biotech, Inc.Products expire prior to the expiration date printed on the product labeling | FTM12 products |
| Z-3262-2024Class II | Fisher & Paykel Healthcare, Ltd.Due to a software issue, affected devices that are set up with High Pressure Oxygen (HPO), if the flow alignment alarm occurs, the device will deliver room air only. If this happens, a patient my experience oxygen desat… | QAV1 product |
| Z-3212-2024Class II | Centinel Spine, Inc.During the assembly of Prodisc C, a blue discoloration was observed on a Prodisc C inlay. The discoloration is considered a foreign contaminate that may not have been fully evaluated in the manufacturing validations. | MJO2 products |
| Z-3214-2024Class II | GE HealthcareGE HealthCare has become aware of an issue where image storage is not possible when Centricity PACS reaches the maximum number of unique image identifiers (Ckeys). If this occurs, it can result in acquisition of a parti… | LLZ1 product |
| Z-3215-2024Class II | Siemens Medical Solutions USA, Inc.If ultrasound systems with software, are changed from factory default to : 1) Milliliters per second (ml/sec, mL/sec) or 2) Milliliters per minute (ml/min, mL/min); then systems will perform incorrect unit conversion of… | IYN3 products |
| Z-3218-2024Class II | Karl Storz EndoscopyVarious medical device product IFUs contain reprocessing modalities that have not been reviewed and/or approved for safety and efficacy by the FDA. | FGA8 products |
| Z-3226-2024Class II | American Contract Systems, Inc.ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requ… | FDE LRO MLS19 products |
| Z-3245-2024Class II | Waldemar Link GmbH & Co. KG (Corp. Hq.)Some lots of the LINK Universal Handle, with quick coupling, Stainless Steel, straight may suffer an immediate material failure intraoperatively when the instrument is used for the first time. The instrument can break i… | LXH1 product |
| Z-3246-2024Class II | Becton Dickinson & Co.Escherichia coli (E. coli) misidentification for clinical samples that were tested on BD Phoenix" M50 instrument and BD Phoenix" 100, may lead to misdiagnosis and inappropriate treatment of infections caused by E. coli, | LON2 products |
| Z-3209-2024Class II | Microcare MedicalContaminated equipment and packaging caused Klebsiella oxytoca, Enterobacter asburiae, Enterobacter bugandensis, pathogenic and opportunistic bacteria, to occur in ultrasonic detergent, which caused product bloating/lea… | JCB1 product |
| Z-3210-2024Class II | Stryker SpineLabeling Correction: Instructions For Use (IFU)-Updated electronic Rev 5, specific language for indications for use for the United States and Canada was inadvertently removed. | OVE1 product |
| Z-3263-2024Class III | T.G. Eakin LimitedOn lot of cartons of Eakin Cohesive small seals, Model Number 839002 may contain blistered seals labelled as 839005. | EXE1 product |
Original PMA approvals
Original premarket approvals for Class III devices. The FDA also decided 57 PMA supplements this week, which change devices already approved.
510(k) clearances
54 510(k) clearances by review panel. The median took 119 calendar days from receipt to decision; the fastest, K242632, 23 days; the longest, K233446, 344 days.
Calendar days include the time a submission waits on the company, so they run longer than the FDA’s 90-day review goal. How to search the 510(k) database
- General, Plastic Surgery11
- Cardiovascular8
- Orthopedic7
- Radiology5
- Anesthesiology4
- General Hospital4
- Neurology4
- Obstetrics/Gynecology4
- Microbiology2
- Clinical Toxicology1
- Dental1
- Ear, Nose, Throat1
- Gastroenterology, Urology1
- Ophthalmic1
Every clearance
| Number | Device and applicant | Product code | Decision and review time |
|---|---|---|---|
| K234134 | AirFit F30i Mask System; AirFit F30i NM Mask System; Arcadia Mask SystemResmed Pty Ltd (Registration Number: 3004604967) | BZDAnesthesiology | Sep 24, 2024270 days |
| K240163 | Hudson RCI® Comfort Flo Nasal Cannula; Hudson RCI® Comfort Flo Plus CannulaMedline Industries, LP | BTTAnesthesiology | Sep 27, 2024249 days |
| K241216 | myAirResmed Corp | BZDAnesthesiology | Sep 26, 2024148 days |
| K241939 | EasyCare Tx 2Resmed Corp | BZDAnesthesiology | Sep 27, 202487 days |
| K233446 | AMC Health CareConsoleAmc Health | MWICardiovascular | Sep 27, 2024344 days |
| K240013 | EchoGo Heart Failure (2.0)Ultromics Limited | QUOCardiovascular | Sep 23, 2024265 days |
| K240236 | Happy Ring Health Monitoring SystemHappy Health, Inc. | MWICardiovascular | Sep 24, 2024239 days |
| K241101 | BioButton SystemBiointellisense, Inc. | DRGCardiovascular | Sep 26, 2024157 days |
| K241414 | CrossWise RF Transseptal Access System (Models: CW-1085S, CW-1085A, CW-1012W, CW-1012C); CrossWise RF Adapter Cable (Model CW-1002)Circa Scientific, Inc. | DXFCardiovascular | Sep 26, 2024132 days |
| K242075 | Indigo® Aspiration System – Lightning Bolt Aspiration TubingPenumbra, Inc. | QEWCardiovascular | Sep 25, 202471 days |
| K242213 | Shockwave Intravascular Lithotripsy (IVL) System with the Javelin Peripheral Intravascular Lithotripsy (IVL) CatheterShockwave Medical, Inc. | PPNCardiovascular | Sep 27, 202460 days |
| K242557 | ClotTriever XL Catheter (41-102)Inari Medical, Inc. | QEWCardiovascular | Sep 27, 202430 days |
| K242540 | AllTest Multi-Drug Urine Test Panel; AllTest Multi-Drug Rapid Urine Test PanelHangzhou AllTest Biotech Co., Ltd. | NFTClinical Toxicology | Sep 27, 202432 days |
| K241147 | Any-ComMediclus Co., Ltd. | EBFDental | Sep 26, 2024154 days |
| K242070 | Ion Endoluminal System (IF 1000)Intuitive Surgical, Inc. | EOQEar, Nose, Throat | Sep 24, 202470 days |
| K241874 | Aspiration Catheter (2189)Hobbs Medical, Inc. | OCXGastroenterology, Urology | Sep 24, 202488 days |
| K240220 | Aulisa Temperature Module (TM0002)Taiwan Aulisa Medical Devices Technologies, Inc. | FLLGeneral Hospital | Sep 27, 2024245 days |
| K241909 | Halyard Purple Nitrile* Powder-Free Exam Gloves, Low Dermatitis Potential, Tested for Use with Chemotherapy Drugs, Fentanyl Citrate, Simulated Gastric Acid and Fentanyl in Simulated Gastric AcidO&M Halyard, Inc. | LZAGeneral Hospital | Sep 27, 202488 days |
| K241916 | TearRepair Liquid Skin ProtectantOptmed, Inc. | KMFGeneral Hospital | Sep 27, 202488 days |
| K242632 | Verifine® Pen NeedlesPromisemed Hangzhou Meditech Co., Ltd. | FMIGeneral Hospital | Sep 26, 202423 days |
| K234149 | MOLLI 2 SystemMolli Surgical, Inc. | NEUGeneral, Plastic Surgery | Sep 26, 2024272 days |
| K240042 | MOLLI 2 SystemMolli Surgical, Inc. | NEUGeneral, Plastic Surgery | Sep 25, 2024264 days |
| K240776 | Celsio Flexible Cryocatheter SystemEndocision Technologies, Inc. | GEHGeneral, Plastic Surgery | Sep 24, 2024187 days |
| K241535 | Vybe RF IIShenB Co., Ltd. | GEIGeneral, Plastic Surgery | Sep 25, 2024118 days |
| K241827 | Microwave Ablation Generator (KY-2000A, KY-2100A)Canyon Medical, Inc. | NEYGeneral, Plastic Surgery | Sep 26, 202494 days |
| K241834 | IPL Hair Removal Device (BE932C, BE932D, BE932E)Shenzhen Yang WO Electronic Co., Ltd. | OHTGeneral, Plastic Surgery | Sep 23, 202490 days |
| K241860 | Diode Laser Hair Removal SystemBeijing Winkonlaser Technology Limited | GEXGeneral, Plastic Surgery | Sep 25, 202490 days |
| K241933 | HIGHERDOSE Red and Infrared Light Mask (MK66-L)Shenzhen Kaiyan Medical Equipment Co., Ltd. | OHSGeneral, Plastic Surgery | Sep 27, 202487 days |
| K242204 | Disposable Neutral Electrodes (GP201B-A, GP202B-A, GP202B-P, GP202B-I, GP202M-A, GP202M-P, GP202M-I, GP202B-AC, GP202B-PC, GP202B-IC, GP202M-AC, GP202M-PC, GP202M-IC, GP203M-A, GP203B-A, GP203M-AC, GP203B-AC, GP203M-P, GP203B-P, GP203M-PC, GP203B-PC, GP203M-I, GP203B-I, GP203M-IC, GP203B-IC)Dongguan Quanding Medical Supplies Co., Ltd. | GEIGeneral, Plastic Surgery | Sep 25, 202461 days |
| K242206 | ACTIVODaeju Meditech Engineering Co., Ltd. | GEXGeneral, Plastic Surgery | Sep 27, 202460 days |
| K242571 | Secret Line up (FC1938E80 FC1950E90 FC1960E100 FC1970E120 FC1990E150 FC19100E150 FC19100E280 FC19100E420 FC2038E80 FC2050E90 FC2060E100 FC2070E120 FC2090E150 FC20100E170 FC2138G80 FC2150G90 FC2160G100 FC2170G120 FC2190G150 FC21100G170 FCL18100C160 FCL1938E80 FCL1950E90 FCL1960E100 FCL1970E120 FCL1990E150 FCL19100E160 FCL2138G80 FCL2150G90 FCL2160G100 FCL2170G120 FCL2190G150 FCL21100G160 FCRL1870C162 FCRL18100C185 FCRL1970E162 FCRL19100E185 FCHL1960E90 FCHL1970E120 FCHL19100E160 FCW1938Hyundae Meditech Co., Ltd. | NEWGeneral, Plastic Surgery | Sep 27, 202429 days |
| K240041 | IVD CAPSULE PSP; abioSCOPEAbionic SA | SCXMicrobiology | Sep 25, 2024264 days |
| K240854 | Accelerate Arc SystemAccelerate Diagnostics, Inc. | QNJMicrobiology | Sep 26, 2024182 days |
| K240993 | encevis (2.1)Ait Austrian Institute of Technology GmbH | OMBNeurology | Sep 27, 2024169 days |
| K241027 | Permatage Settable Bone PuttyOrthocon, Inc. | GXPNeurology | Sep 26, 2024164 days |
| K241513 | SourcererBrain Electrophysiology Laboratory Company, LLC | OLXNeurology | Sep 27, 2024121 days |
| K242582 | Optima Coil System (OptiBlock Line Extension)Balt USA, LLC | HCGNeurology | Sep 27, 202428 days |
| K240434 | RELIEEV Suction Curette (Flexible 3.0/ Standard 3.6)Li Medical Corporation , Ltd. | HHKObstetrics/Gynecology | Sep 25, 2024224 days |
| K240745 | ID Free Personal LubricantWestridge Laboratories, Inc. | NUCObstetrics/Gynecology | Sep 27, 2024192 days |
| K241341 | cryo-GO Vitrification DeviceFujifilm Irvine Scientific | MQKObstetrics/Gynecology | Sep 26, 2024136 days |
| K242125 | Elvie Stride 2Chiaro Technology, Ltd. | HGXObstetrics/Gynecology | Sep 27, 202470 days |
| K241447 | iCare ST500 (TA04)Icare Finland OY | HKYOphthalmic | Sep 27, 2024128 days |
| K233807 | LumiVy™ Lumbar IBF SystemVy Spine, LLC | MAXOrthopedic | Sep 24, 2024300 days |
| K241570 | Unity Total Knee SystemCorin U.S.A. Limited | JWHOrthopedic | Sep 23, 2024115 days |
| K241651 | Affixus Tibial and Antegrade Femoral Nailing SystemZimmer, Inc. | HSBOrthopedic | Sep 26, 2024111 days |
| K241815 | Protean Fragment Plating SystemSkeletal Dynamics, Inc. | HRSOrthopedic | Sep 26, 202494 days |
| K241895 | Cannulated PsiFGuardSpineguard | SCYOrthopedic | Sep 26, 202490 days |
| K242273 | Cove Putty, OsteoCove PuttySeaSpine Orthopedics Corporation | MQVOrthopedic | Sep 25, 202455 days |
| K242299 | NovaBone Putty - Synthetic Bioactive Bone GraftNovabone Products, LLC | MQVOrthopedic | Sep 25, 202454 days |
| K240243 | EConsole2DRTECH Corporation | LLZRadiology | Sep 23, 2024237 days |
| K240369 | CAC (gated) AlgorithmBunkerHill Health | JAKRadiology | Sep 27, 2024233 days |
| K241480 | JBS-LVOJLK, Inc. | QASRadiology | Sep 27, 2024126 days |
| K241536 | trophon Wireless Ultrasound Probe HolderNanosonics Limited | OUJRadiology | Sep 27, 2024120 days |
| K242292 | uAI Easy Triage ICHShanghai United Imaging Intelligence Co., Ltd. | QASRadiology | Sep 24, 202453 days |
Only see the weeks that matter to you
DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and every day it checks openFDA and tells you about new clearances, recalls and MAUDE reports for them. It also puts all of this beside the FDA’s own record pages.
Only your product codes, every Monday
One short email after each openFDA update with just the decisions and recalls in up to three product codes you choose. Free.
Earlier weeks
Each week keeps its own page.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Decisions arrive in weekly updates; check the FDA record before you rely on a figure.
- device/510k · search decision_date:[20240923 TO 20240929]Retrieved Nov 29, 2026, 8:50 AM UTC. FDA data updated Nov 16, 2026.
- device/pma · search decision_date:[20240923 TO 20240929]Retrieved Nov 29, 2026, 8:50 AM UTC. FDA data updated Nov 16, 2026.
- device/recall · search event_date_posted:[20240923 TO 20240929]Retrieved Nov 29, 2026, 8:50 AM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-3207-2024" "Z-3208-2024" "Z-3303-2024" "Z-3304-2024" "Z-3305-2024" "Z-3306-2024" "Z-3307-2024" "Z-3308-2024" "Z-3309-2024" "Z-3153-2024" "Z-3264-2024" "Z-3265-2024" "Z-3266-2024" "Z-3267-2024" "Z-3268-2024" "Z-3269-2024" "Z-3270-2024" "Z-3271-2024" "Z-3272-2024" "Z-3273-2024" "Z-3274-2024" "Z-3275-2024" "Z-3276-2024" "Z-3277-2024" "Z-3278-2024" "Z-3279-2024" "Z-3280-2024" "Z-3281-2024" "Z-3282-2024" "Z-3283-2024" "Z-3284-2024" "Z-3285-2024" "Z-3286-2024" "Z-3287-2024" "Z-3288-2024" "Z-3289-2024" "Z-3290-2024" "Z-3291-2024" "Z-3292-2024" "Z-3293-2024" "Z-3294-2024" "Z-3295-2024" "Z-3296-2024" "Z-3297-2024" "Z-3298-2024" "Z-3299-2024" "Z-3300-2024" "Z-3301-2024" "Z-3302-2024" "Z-3158-2024")Retrieved Nov 29, 2026, 8:50 AM UTC. FDA data updated Nov 18, 2026.
- device/enforcement · search recall_number:("Z-3248-2024" "Z-3249-2024" "Z-3250-2024" "Z-3251-2024" "Z-3252-2024" "Z-3253-2024" "Z-3254-2024" "Z-3255-2024" "Z-3256-2024" "Z-3257-2024" "Z-3258-2024" "Z-3259-2024" "Z-3260-2024" "Z-3261-2024" "Z-3262-2024" "Z-3263-2024" "Z-3212-2024" "Z-3213-2024" "Z-3214-2024" "Z-3215-2024" "Z-3216-2024" "Z-3217-2024" "Z-3218-2024" "Z-3219-2024" "Z-3220-2024" "Z-3221-2024" "Z-3222-2024" "Z-3223-2024" "Z-3224-2024" "Z-3225-2024" "Z-3226-2024" "Z-3227-2024" "Z-3228-2024" "Z-3229-2024" "Z-3230-2024" "Z-3231-2024" "Z-3232-2024" "Z-3233-2024" "Z-3234-2024" "Z-3235-2024" "Z-3236-2024" "Z-3237-2024" "Z-3238-2024" "Z-3239-2024" "Z-3240-2024" "Z-3241-2024" "Z-3242-2024" "Z-3243-2024" "Z-3244-2024" "Z-3245-2024")Retrieved Nov 29, 2026, 8:50 AM UTC. FDA data updated Nov 18, 2026.
- device/enforcement · search recall_number:("Z-3246-2024" "Z-3247-2024" "Z-3209-2024" "Z-3210-2024")Retrieved Nov 29, 2026, 8:50 AM UTC. FDA data updated Nov 18, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.