Weekly archive
Week of January 12, 2026
The FDA cleared 77 510(k)s, granted 2 De Novo requests and approved no original PMAs. 31 device recalls were posted, 2 of them Class I. Every 510(k), De Novo and PMA decision dated January 12 to 18, 2026, and every device recall the FDA posted those days.
Week beforeWeek afterRSS feedThis week as lead lists
- 510(k) clearances
- 77median review 173 days
- De Novo grants
- 2
- Original PMA approvals
- 0plus 43 supplements
- Recalls posted
- 3191 products, 2 Class I
Recalls posted
Device recalls the FDA posted January 12 to 18, 2026, Class I first. Class I is the most serious: a reasonable chance the product causes serious harm or death. The class comes from the FDA’s weekly enforcement report, so the newest recalls can show “Class pending” for a few days.
| Recall and class | Firm and reason | Product code |
|---|---|---|
| Z-1011-2026Class I | Medline Industries, LPMedline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its specified weight limit, encountering an ob… | LLI6 products |
| Z-1020-2026Class I | Abbott Diabetes Care, Inc.Due to a manufacturing process issue that may cause carbon build up in the sensor that may result in incorrect low glucose results | QLG4 products |
| Z-1102-2026Class II | RAYSEARCH LABORATORIES ABPotential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override… | MUJ9 products |
| Z-1111-2026Class II | ICU Medical, Inc.Needle free access device intended as accessory to vascular access used in hemodialysis or accessory to intravascular administration sets has a silicone seal which may bulge at the top, separate from the body, tear; may… | FPA1 product |
| Z-1112-2026Class II | GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGINGThere is a potential intermittent issue on certain Omni Legend systems that can result in a streaking artifact in the PET clinical scan images. This streaking artifact is most easily identified in transaxial slices on t… | KPS1 product |
| Z-1113-2026Class II | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Philips has identified two (2) software issues affecting device systems that may result in loss of imaging (X-ray) functionality and/or loss of motorized movement, and/or incorrect image content and/or loss of data. Iss… | OWB1 product |
| Z-1118-2026Class II | ICU Medical, Inc.Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at th… | FRN3 products |
| Z-1121-2026Class II | Baxter Healthcare CorporationCertain pumps have potentially been released from service with defective grease applied to the cam and motor gears. Due to its low viscosity, the defective grease may lead to the device having insufficient lubrication o… | FRN1 product |
| Z-1122-2026Class II | Mazor Robotics LtdSoftware errors that can result in incorrect surgical instrument positioning during spinal surgery. | OLO1 product |
| Z-1123-2026Class II | Merit Medical Systems, Inc.Inflation device handle may detach from the syringe during procedure. | OEQ OEZ PPD7 products |
| Z-1130-2026Class II | LEASEIR TECHNOLOGIES SLUThe console label for affected devices is missing the "DANGER" symbol. | GEX1 product |
| Z-1093-2026Class II | Hologic, IncFungal contamination of affected lot with Parengyodontium album. | LEA1 product |
| Z-1094-2026Class II | ICU Medical, Inc.ICU Medical received one lot of defective primary speakers from our supplier that may fail to produce an audible sound. If the primary speaker in your Plum Duo Infusion Pump fails, the pump may not produce audible alarm… | FRN1 product |
| Z-1095-2026Class II | Instrumentation LaboratoryPotential for microbial contamination. | GFO1 product |
| Z-1096-2026Class II | Intuitive Surgical, Inc.Access Port System tray may develop cracks potentially resulting in a sterility breach. | NAY1 product |
| Z-1098-2026Class II | Medtronic IncStents may be impacted by a Nitinol anomaly which causes a reduced stent diameter when deployed and corresponding reduced radial force. Issue may result in stent migration. | QAN1 product |
| Z-1099-2026Class II | Thoratec LLCAcute circulatory support system pump may not fit into the motor and reports received of the pump not being securely locked into the motor using the screw-in locking feature. Failure to properly lock the pump into motor… | QNR1 product |
| Z-1100-2026Class II | ICU Medical Inc.Affected lots of syringes may have a crack on the syringe collar, which may lead to delayed or repeated blood draw/fluid delivery. This issue may also lead to blood leakage during sampling of arterial blood or leakage d… | CAZ JKA2 products |
| Z-1079-2026Class II | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Nine (9) identified software issues which may result in the following: loss of X-ray after Pedal Tap, loss of X-ray after Phase Fault, missing live X-ray on the FlexVision monitor, system continues to restart, AMD tripl… | OWB1 product |
| Z-1080-2026Class II | Heraeus Medical GmbH (Dental Division)increase of complaints regarding the breakage of the ampoules within their device systems. An extensive investigation revealed that multiple factors have an impact on the ampoule breakage. An unfortunate combination of… | LOD4 products |
| Z-1084-2026Class II | Diagnostica Stago, Inc.Potential for defect for Asserachrom HPIA lot 271288, which can affect every well in the impacted strip and produce negative results. | LCO1 product |
| Z-1085-2026Class II | Medline Industries, LPMedline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigat… | FDE LRO OEZ6 products |
| Z-1091-2026Class II | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Six (6) identified software issues which may result in the following: system keeps restarting, AMC triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after the sta… | OWB1 product |
| Z-1092-2026Class II | C-RAD POSITIONING ABDue to issues with the system's stereotactic radiosurgery (SRS) treatment cannot be guaranteed for all couch angles and system setups. n certain cases, the system may indicate the patient as correctly positioned at the… | IYE1 product |
| Z-1017-2026Class II | Canon Medical System, USA, INC.There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed ven… | LNH2 products |
| Z-1024-2026Class II | REPLIGEN CORPORATIONPotential Elevated Laser Output A disconnected cable or a poorly crimped cable under mechanical bending stress can cause loss of electrical contact. At one connection on the board, the failure can drive the laser output… | 1 product |
| Z-1062-2026Class II | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documen… | OWB16 products |
| Z-1078-2026Class II | Elekta, Inc.Using oncology information system that manages workflow may result in overtreatment if:Plan delivered with Particle Therapy IHE-RO TDW-II interface,With one field per fraction or last field partial treatment,Partial tre… | IYE1 product |
| Z-1048-2026Class II | Fresenius Kabi USA, LLCEmphasizing instructions for LVP duration programming located in the IFU. | FRN1 product |
| Z-1049-2026Class II | Medline Industries, LPMedline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positio… | CAI FTN LRO12 products |
| Z-1061-2026Class II | HANGZHOU BEVER MEDICAL DEVICES CO., LTD.Labeling error, Incorrect expiration date | GBM1 product |
De Novo grants
New device types the FDA classified through the De Novo pathway. Each one creates a product code and becomes a possible predicate for later 510(k)s.
Original PMA approvals
Original premarket approvals for Class III devices. The FDA also decided 43 PMA supplements this week, which change devices already approved.
No original PMAs were approved this week.
510(k) clearances
77 510(k) clearances by review panel. The median took 173 calendar days from receipt to decision; the fastest, K254190, 22 days; the longest, K250135, 364 days.
Calendar days include the time a submission waits on the company, so they run longer than the FDA’s 90-day review goal. How to search the 510(k) database
- Cardiovascular12
- Orthopedic12
- Radiology10
- General, Plastic Surgery6
- Dental5
- Gastroenterology, Urology5
- Neurology4
- Panel not stated4
- Physical Medicine4
- Anesthesiology3
- General Hospital3
- Clinical Chemistry2
- Microbiology2
- Clinical Toxicology1
- Ear, Nose, Throat1
- Hematology1
- Immunology1
- Ophthalmic1
Every clearance
| Number | Device and applicant | Product code | Decision and review time |
|---|---|---|---|
| K251419 | Laboratorios Biogalenic Sterile Water for Inhalation, USPLaboratorios Biogalenic S.A. DE C.V. | BTTAnesthesiology | Jan 14, 2026252 days |
| K251448 | FL-10000U Respiratory HumidifierFlexicare Medical Limited. | BTTAnesthesiology | Jan 16, 2026252 days |
| K251847 | Sleepnet Arie Full Face Vented Mask; Sleepnet Arie Nasal Vented MaskSleepnet Corporation | BZDAnesthesiology | Jan 15, 2026213 days |
| K250135 | WAVE Clinical Platform (2.0.000)Baxter Healthcare Corp/ Excel Medical | MWICardiovascular | Jan 16, 2026364 days |
| K251307 | Clinical Automatic Blood Pressure Monitor (DBP-20,DBP-20i)Shenzhen Hingmed Medical Instrument Co., Ltd. | DXNCardiovascular | Jan 12, 2026259 days |
| K251357 | Esperance pHLO Aspiration SystemPhenox, Ltd. | NRYCardiovascular | Jan 15, 2026260 days |
| K251367 | OptoMonitor 3Opsens, Inc. | DXOCardiovascular | Jan 16, 2026260 days |
| K251744 | Affinity Pixie™ Arterial Filter with Balance™ Biosurface; Affinity® Pediatric Arterial Blood FilterMedtronic | DTMCardiovascular | Jan 16, 2026224 days |
| K251831 | Bio-Medicus Life Support Catheter and IntroducerMedtronic, Inc. | QHWCardiovascular | Jan 15, 2026216 days |
| K251970 | Sprinter Legend Rapid Exchange Balloon Dilatation CatheterMedtronic, Ireland | LOXCardiovascular | Jan 17, 2026205 days |
| K252297 | Toro 88 Superbore CatheterToro Neurovascular, Inc. | QJPCardiovascular | Jan 16, 2026177 days |
| K253511 | Concerto Versa™ Detachable CoilMedtronic, Inc. | KRDCardiovascular | Jan 14, 202670 days |
| K253925 | Aventus Thrombectomy SystemInquis Medical | QEZCardiovascular | Jan 15, 202638 days |
| K254085 | CARTO™ 3 EP Navigation System V9.0 with PIU PlusBiosense Webster, Inc. | DQKCardiovascular | Jan 18, 202631 days |
| K254114 | Baylis Connector CableBaylis Medical Technologies, Inc. | DSACardiovascular | Jan 18, 202630 days |
| K253470 | MiniMed 780G Insulin Pump (MMT-1884); MiniMed 780G Insulin Pump Kit (MMT-1894)Medtronic Minimed, Inc. | QFGClinical Chemistry | Jan 15, 202697 days |
| K253585 | SmartGuard technology; Predictive Low Glucose technologyMedtronic Minimed, Inc. | QJIClinical Chemistry | Jan 14, 202658 days |
| K253082 | LZI Buprenorphine II Enzyme ImmunoassayLin-Zhi International, Inc. | DJGClinical Toxicology | Jan 16, 2026115 days |
| K251388 | Dental air-motor handpiece (TM-203B2, TM-203M4, TM-205B2, TM-205M4, TM-IS, TM-ES, TM-SG20, TM-X25, TM-BB, TM-IC, TM-RA41)Foshan Topmed Dental Co., Ltd. | EFBDental | Jan 12, 2026252 days |
| K251389 | High speed air turbine handpieceFoshan Topmed Dental Co., Ltd. | EFBDental | Jan 12, 2026252 days |
| K251701 | Star E900 Electric SystemDentalez, Inc., Stardental Division | EBWDental | Jan 12, 2026224 days |
| K253167 | Any-PasteMediclus Co., Ltd. | KIFDental | Jan 12, 2026108 days |
| K253248 | Bright Cavity LinerDmp Dental Industry S.A. | EJKDental | Jan 13, 2026106 days |
| K251790 | SeptAlignSpirair, Inc. | NHBEar, Nose, Throat | Jan 16, 2026219 days |
| K251341 | VITAL Tubing Sets for HemodialysisVital Healthcare Sdn. Bhd. | FJKGastroenterology, Urology | Jan 16, 2026261 days |
| K251739 | da Vinci Surgical System (IS5000)Intuitive Surgical, Inc. | NAYGastroenterology, Urology | Jan 16, 2026224 days |
| K252913 | Break WaveSonomotion, Inc. | LNSGastroenterology, Urology | Jan 12, 2026122 days |
| K253556 | da Vinci SP Surgical System (SP1098); da Vinci E-200 GeneratorIntuitive Surgical, Inc. | NAYGastroenterology, Urology | Jan 16, 202663 days |
| K254014 | Connected OR Hub with Device and Voice Control; SDC4K Information Management System with Device and Voice ControlStryker Endoscopy | GCJGastroenterology, Urology | Jan 13, 202629 days |
| K251308 | SnugLit(TM) Wearable Phototherapy System (SNGL-01-US)Thera B Medical Products | LBIGeneral Hospital | Jan 15, 2026262 days |
| K252313 | Cardinal Health Nitrile Examination Gloves Extended CuffCardinalhealth | LZCGeneral Hospital | Jan 14, 2026173 days |
| K252549 | Non-sterile Powder Free Nitrile Examination Glove, Dual Color (White-Green), Tested for Use with Chemotherapy Drugs, Fentanyl Citrate, Xylazine HCl, and Simulated Gastric Acid; Non-sterile Powder Free Nitrile Examination Glove, Dual Color (White-Blue), Tested for Use with Chemotherapy Drugs, Fentanyl Citrate, Xylazine HCl, and Simulated Gastric AcidKossan International Sdn Bhd | LZAGeneral Hospital | Jan 13, 2026153 days |
| K251254 | Ignite RF SystemInMode , Ltd. | GEIGeneral, Plastic Surgery | Jan 15, 2026267 days |
| K251524 | Dr. Yglo Skin Tag RemoverTheotclab Healthcare B.V. | GEHGeneral, Plastic Surgery | Jan 12, 2026241 days |
| K251988 | Boston iFace (Boston iFace)Boston Aesthetics, Inc. | GEIGeneral, Plastic Surgery | Jan 12, 2026199 days |
| K251996 | ENTire IRE SystemEntire Medical , Ltd. | GEIGeneral, Plastic Surgery | Jan 14, 2026201 days |
| K252830 | Disposable Surgical Face Mask (3P00B, C2900)Beatles Medical Supplies (Xiantao) Co., Ltd. | FXXGeneral, Plastic Surgery | Jan 12, 2026129 days |
| K252960 | SOLIUS PRO UVB Light PanelSolius Labs, Inc. | SGZGeneral, Plastic Surgery | Jan 16, 2026122 days |
| K251249 | Tru Hematology TestTruvian Sciences, Inc. | GKZHematology | Jan 16, 2026268 days |
| K252775 | NOVEOS Specific IgE (sIgE): Capture Reagent F013, Peanut (Arachis hypogaea); Capture Reagent F076, Bos d 4 a-lactalbumin, Milk; Capture Reagent F077, Bos d 5 ß-lactoglobulin, Milk; Capture Reagent F232, Gal d 2 Ovalbumin, EggHycor Biomedical | DHBImmunology | Jan 16, 2026136 days |
| K251721 | Xpert GI PanelCepheid | PCHMicrobiology | Jan 16, 2026226 days |
| K252283 | Nano-Check Influenza A+B TestNano-Ditech Corporation | PSZMicrobiology | Jan 15, 2026177 days |
| K252767 | actiTENS miniSublimed | GZJNeurology | Jan 16, 2026140 days |
| K252958 | METICULY Patient-specific titanium mesh implantMeticuly Co., Ltd. | GXNNeurology | Jan 14, 2026120 days |
| K252962 | DYNAMIS™ Navigation SystemPromethean Restorative, LLC | OLONeurology | Jan 12, 2026118 days |
| K253363 | NeuroSpan BridgeAuxilium Biotechnologies, Inc. | JXINeurology | Jan 17, 2026109 days |
| K251353 | Eyer 2Phelcom Technologies | HKIOphthalmic | Jan 16, 2026261 days |
| K251214 | BEAR® (Bridge-Enhanced ACL Restoration) ImplantMiach Orthopaedics, Inc. | QNIOrthopedic | Jan 13, 2026270 days |
| K251558 | Precision AI Surgical Planning System (PAI-SPS)Precision AI Pty, Ltd. | QHEOrthopedic | Jan 12, 2026236 days |
| K251961 | SDS Growing RodBAAT Medical Products B.V. | PGMOrthopedic | Jan 15, 2026203 days |
| K252205 | Curiteva Porous PEEK Cervical Interbody System; Curiteva Porous PEEK Lumbar Interbody System; Curiteva Porous PEEK Laminoplasty System; Curiteva Porous PEEK Standalone ALIF SystemCuriteva, Inc. | ODPOrthopedic | Jan 16, 2026186 days |
| K252219 | Cervical Interbody and VBR Fusion SystemSync Surgical | ODPOrthopedic | Jan 14, 2026183 days |
| K252461 | Swedge™ Pedicle Screw Fixation System Bezier RodSpinal Resources, Inc. | NKBOrthopedic | Jan 13, 2026161 days |
| K252516 | N22 EZ GlenosphereShoulder Innovations, Inc. | PHXOrthopedic | Jan 15, 2026157 days |
| K252567 | AltiVate Reverse® ADLC GlenosphereEncore Medical L.P. | PHXOrthopedic | Jan 15, 2026154 days |
| K252807 | Arthrex Anatomic Lapidus Plates and I-Beam Lapidus PlatesArthrex, Inc. | HRSOrthopedic | Jan 12, 2026131 days |
| K253171 | Libertas Taper Short (TS) Uncemented Femoral Stem (Libertas TS Stem)Maxx Orthopedics, Inc. | LZOOrthopedic | Jan 15, 2026111 days |
| K254002 | TriMed Volar Bearing Plates (VBEAL-13-7S, VBEAR-13-7S, VBEAL-14-7S, VBEAR-14-7S, VBEAL-16-7S, VBEAR-16-7S)TriMed, Inc. | HRSOrthopedic | Jan 15, 202631 days |
| K254061 | Curiteva Porous PEEK Cervical Interbody Fusion SystemCuriteva, Inc. | ODPOrthopedic | Jan 15, 202629 days |
| K250890 | SBC1 CreamSioxmed, LLC | FROPanel not stated | Jan 12, 2026293 days |
| K251714 | Ion-CSurGenTec, LLC | MRWPanel not stated | Jan 16, 2026227 days |
| K251900 | MY01 Continuous Compartmental Pressure MonitorMY01, Inc. | LXCPanel not stated | Jan 16, 2026210 days |
| K254126 | Natrelle 133S Tissue ExpandersAbbVie | LCJPanel not stated | Jan 16, 202628 days |
| K251458 | Electric Wheelchair (DH01168)Foshan Dahao Medical Technology Co., Ltd. | ITIPhysical Medicine | Jan 15, 2026248 days |
| K251886 | EmpowerRide NAVIGATOREmpower Ride, LLC | IORPhysical Medicine | Jan 12, 2026206 days |
| K253276 | Electric Wheelchair (Robooter E80) (BBR-E80-01, BBR-E80-02)Shanghai Bangbang Robotics Co., Ltd. | ITIPhysical Medicine | Jan 15, 2026108 days |
| K253408 | CoolToneZimmer Medizinsysteme GmbH | NGXPhysical Medicine | Jan 16, 2026108 days |
| K251934 | qXR-DetectQure.Ai Technologies | MYNRadiology | Jan 16, 2026206 days |
| K252188 | EMLA (Elekta Evo, VersaHD, Elekta Harmony Pro, Elekta Infintiy, Elekta Harmony, Elekta Synergy)Elekta Solutions AB | IYERadiology | Jan 15, 2026185 days |
| K252579 | Orthoscan TAU MVP Mini C-Arm SystemZiehm-Orthoscan, Inc. | OXORadiology | Jan 14, 2026152 days |
| K252634 | Imagine® Enterprise SuiteSoftlink International Private Limited | QIHRadiology | Jan 16, 2026149 days |
| K252934 | DiagnocatDGNCT, LLC | MYNRadiology | Jan 15, 2026122 days |
| K253023 | BIOGRAPH OneSiemens Healthineers AG | OUORadiology | Jan 15, 2026118 days |
| K254001 | VERITON CT 300 Series Digital SPECT/CT System (VERITON CT 316/364); VERITON CT 400 Series Digital SPECT/CT System (VERITON CT 416/464)Spectrum Dynamics Medical, Ltd. | KPSRadiology | Jan 13, 202629 days |
| K254010 | ExacTrac Dynamic (2.0.2); ExacTrac Dynamic SurfaceBrainlab SE | IYERadiology | Jan 14, 202630 days |
| K254186 | Azurion R3.1Philips Medical Systems B.V. | OWBRadiology | Jan 16, 202624 days |
| K254190 | dS Base 1.5T; dS Base 3.0T; dS Head 1.5T; dS Head 3.0T; dS HeadNeck 1.5T; dS HeadNeck 3.0TPhilips Medical Systems Nederland B.V. | MOSRadiology | Jan 14, 202622 days |
Only see the weeks that matter to you
DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and every day it checks openFDA and tells you about new clearances, recalls and MAUDE reports for them. It also puts all of this beside the FDA’s own record pages.
Only your product codes, every Monday
One short email after each openFDA update with just the decisions and recalls in up to three product codes you choose. Free.
Earlier weeks
Each week keeps its own page.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Decisions arrive in weekly updates; check the FDA record before you rely on a figure.
- device/510k · search decision_date:[20260112 TO 20260118]Retrieved Nov 29, 2026, 7:58 AM UTC. FDA data updated Nov 16, 2026.
- device/pma · search decision_date:[20260112 TO 20260118]Retrieved Nov 29, 2026, 7:58 AM UTC. FDA data updated Nov 16, 2026.
- device/recall · search event_date_posted:[20260112 TO 20260118]Retrieved Nov 29, 2026, 9:40 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-1011-2026" "Z-1012-2026" "Z-1013-2026" "Z-1014-2026" "Z-1015-2026" "Z-1016-2026" "Z-1102-2026" "Z-1103-2026" "Z-1104-2026" "Z-1105-2026" "Z-1106-2026" "Z-1107-2026" "Z-1108-2026" "Z-1109-2026" "Z-1110-2026" "Z-1111-2026" "Z-1112-2026" "Z-1113-2026" "Z-1118-2026" "Z-1119-2026" "Z-1120-2026" "Z-1121-2026" "Z-1122-2026" "Z-1123-2026" "Z-1124-2026" "Z-1125-2026" "Z-1126-2026" "Z-1127-2026" "Z-1128-2026" "Z-1129-2026" "Z-1130-2026" "Z-1093-2026" "Z-1094-2026" "Z-1095-2026" "Z-1096-2026" "Z-1098-2026" "Z-1099-2026" "Z-1100-2026" "Z-1101-2026" "Z-1020-2026" "Z-1021-2026" "Z-1022-2026" "Z-1023-2026" "Z-1079-2026" "Z-1080-2026" "Z-1081-2026" "Z-1082-2026" "Z-1083-2026" "Z-1084-2026" "Z-1085-2026")Retrieved Nov 29, 2026, 9:40 PM UTC. FDA data updated Nov 18, 2026.
- device/enforcement · search recall_number:("Z-1086-2026" "Z-1087-2026" "Z-1088-2026" "Z-1089-2026" "Z-1090-2026" "Z-1091-2026" "Z-1092-2026" "Z-1017-2026" "Z-1018-2026" "Z-1024-2026" "Z-1062-2026" "Z-1063-2026" "Z-1064-2026" "Z-1065-2026" "Z-1066-2026" "Z-1067-2026" "Z-1068-2026" "Z-1069-2026" "Z-1070-2026" "Z-1071-2026" "Z-1072-2026" "Z-1073-2026" "Z-1074-2026" "Z-1075-2026" "Z-1076-2026" "Z-1077-2026" "Z-1078-2026" "Z-1048-2026" "Z-1049-2026" "Z-1050-2026" "Z-1051-2026" "Z-1052-2026" "Z-1053-2026" "Z-1054-2026" "Z-1055-2026" "Z-1056-2026" "Z-1057-2026" "Z-1058-2026" "Z-1059-2026" "Z-1060-2026" "Z-1061-2026")Retrieved Nov 29, 2026, 7:58 AM UTC. FDA data updated Nov 18, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.