FDA product code
LBI: Unit, Neonatal Phototherapy (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 880.5700
- Review panel
- General Hospital
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 711 in the last 12 months
- Median review time
- 157 daysthe latest 25 decisions
- Recall records
- 140 in the last 12 months
- Adverse event reports
- 5last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code LBI per year, by decision date, 71 since 1976. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 2 |
| 2018 | 2 |
| 2019 | 1 |
| 2020 | 1 |
| 2021 | 1 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 1 |
| 2026 (so far) | 1 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K251308 | SnugLit(TM) Wearable Phototherapy System (SNGL-01-US)Thera B Medical Products | Jan 15, 2026262 days |
| K243372 | BiliWrapGerium Medical, Ltd. | May 29, 2025211 days |
| K210289 | Infant Phototherapy EquipmentBistos Co., Ltd. | May 28, 2021115 days |
| K200031 | Neonatal Phototherapy SystemAvalon Biomedical (Shenzhen) Limited | Oct 5, 2020272 days |
| K190899 | bili-hutLittle Sparrows Technologies, Inc. | Sep 5, 2019150 days |
| K182178 | neoBLUE blanket LED Phototherapy SystemNatus Medical Incorporated | Dec 13, 2018122 days |
| K172656 | BiliLuxDraeger Medical Systems, Inc. | May 1, 2018238 days |
| K170585 | SkylifeNeolight, LLC | Oct 27, 2017241 days |
| K163526 | Bilicocoon™ Phototherapy SystemNeomedlight | Oct 6, 2017294 days |
| K160745 | neoBLUE compact LED Phototherapy SystemNatus Medical Incorporated | Dec 12, 2016269 days |
Recalls
Recall records for product code LBI per year, by the date the recall began. Each recalled product is one record; these 14 records belong to 14 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 1 |
| 2018 | 0 |
| 2019 | 1 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-0923-2020Class II | Lumitex IncThere is a potential for some systems built between 2014 and Jan 2018 to have a light output falling below the recommended minimum output. This problem has bee… | Dec 11, 2019 |
| Z-1924-2017Class II | Natus Medical IncorporatedThe second generation neoBLUE 2 lights with a black band, sold from 2003 to 2008, are recalled so it will eliminate the possibility that a neoBLUE system may h… | Apr 3, 2017 |
| Z-1503-2017Class II | Ohmeda MedicalCertain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheati… | Nov 18, 2016 |
| Z-0770-2017Class II | Natus Medical IncorporatedneoBLUE blanket LED Phototherapy System distributed since release of the product in 2011 is recalled due to customer complaints regarding discoloration/ degrad… | Oct 28, 2016 |
| Z-0769-2017Class II | Natus Medical IncorporatedThe firm sent the Technical Bulletin to consignees who received the reliability updated neoBLUE blanket LED Phototherapy Systems between March 2016 to August 2… | Sep 16, 2016 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code LBI. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 577 |
| 2018 | 37 |
| 2019 | 6 |
| 2020 | 6 |
| 2021 | 2 |
| 2022 | 8 |
| 2023 | 0 |
| 2024 | 6 |
| 2025 | 3 |
| 2026 (so far) | 5 |
Last 12 months by event type
- Deaths0
- Injuries3
- Malfunctions2
- Other or not stated0
Most reported device problems, last 12 months
- Electrical /Electronic Property Problem2
In plain English: what the newest reports describe Written by AI
The lamp or its arm came loose or fell in four reports. In one, a loosened arm socket let the lamp drop toward a baby; in others, a loose bolt or detached arm let it collapse or fall onto a patient. No patient harm was reported in any of them.
4 reports: 2510954-2025-00001, 2510954-2024-00036, 1220063-2024-00029, 1220063-2024-00010
Three reports describe skin injuries: second-degree burns in a premature infant, which a lawsuit alleges came from light sensitivity with no malfunction alleged, a minor abrasion from a baby lying on the pad without its mattress, and pressure damage behind an ear.
3 reports: 1000122810-2024-00001, 9612197-2025-00010, 9612330-2022-00044
Two reports describe a burning smell or overheating. In one, a nurse found burning metal at a fiber optic connection on the cord and the device was pulled from use. In the other, a home-use blanket showed an overheating error message, and the infant's bilirubin (the yellow jaundice level) kept falling.
2 reports: 23943066, 25259324
Two reports describe a lamp arm failing and injuring staff. One describes a faulty spring arm that could not stay in place, leaving a nurse with a facial bruise, though an X-ray showed no fracture. The other says a falling device injured staff but gives few details.
2 reports: 2510954-2025-00005, 9612330-2022-00020
One report describes a user receiving an electric shock. About 30 seconds after the AC wall plug was unplugged, the live and neutral pins could still deliver a shock, and the device was removed from service for evaluation.
Report: 9612197-2026-00018
Written by AI from the narratives of the 17 newest reports that have one, received Apr 12, 2022 to Aug 4, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 132in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 42registered establishments that list this product code
Other codes under 21 CFR 880.5700
Product codes classified under the same regulation.
Watch LBI in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get LBI every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in LBI. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:LBIRetrieved Nov 28, 2026, 10:08 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:LBIRetrieved Nov 29, 2026, 6:51 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:LBI · count decision_dateRetrieved Nov 28, 2026, 10:08 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:LBIRetrieved Nov 29, 2026, 6:51 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:LBI · count event_date_initiatedRetrieved Nov 28, 2026, 10:08 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:LBI · count res_event_numberRetrieved Nov 28, 2026, 10:08 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-0923-2020" "Z-1924-2017" "Z-1503-2017" "Z-0770-2017" "Z-0769-2017")Retrieved Nov 29, 2026, 6:51 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:LBI AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 6:51 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:LBI AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 6:51 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:LBI AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 6:51 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:LBIRetrieved Nov 28, 2026, 10:08 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:LBIRetrieved Nov 28, 2026, 10:08 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:880.5700Retrieved Nov 28, 2026, 10:08 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.