FDA product code
EBW: Controller, Foot, Handpiece And Cord (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class I
- Regulation
- 21 CFR 872.4200
- Review panel
- Dental
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 634 in the last 12 months
- Median review time
- 269 daysthe latest 25 decisions
- Recall records
- 20 in the last 12 months
- Adverse event reports
- 10last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code EBW per year, by decision date, 63 since 1977. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 4 |
| 2018 | 1 |
| 2019 | 2 |
| 2020 | 1 |
| 2021 | 2 |
| 2022 | 6 |
| 2023 | 0 |
| 2024 | 4 |
| 2025 | 3 |
| 2026 (so far) | 4 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K260464 | Dental Electric Motor System (iM2X)DTL Dental Equipment Manufacture Co., Ltd. | Nov 6, 2026267 days |
| K260027 | Firstar Dental Electric MotorFDC Technology Industry Co. | Oct 1, 2026269 days |
| K251407 | Motor Handpiece and Control Unit (SDI10)Saeshin Precision Co., Ltd. | Jan 30, 2026269 days |
| K251701 | Star E900 Electric SystemDentalez, Inc., Stardental Division | Jan 12, 2026224 days |
| K251811 | Motor and Apex Module (MaAM)Dentsply Sirona, Inc. | Aug 25, 202574 days |
| K231562 | ELEC ENGINE (Model: ISE-170L)Micro-Nx Co., Ltd. | Jul 18, 2025779 days |
| K242646 | Dental Implant UnitGuilin Aesthedent Medical Instruments Co., Ltd. | Apr 11, 2025220 days |
| K233288 | NLZ Built-In Motor SystemNakanishi, Inc. | Jun 24, 2024269 days |
| K233117 | Surgic Pro2 (Surgic Pro2 OPT), Surgic Pro2 (Surgic Pro2 NON-OPT)Nakanishi, Inc. | Jun 20, 2024267 days |
| K230895 | CEFLA Dental Micromotors: i-MMr, i-MMr L, i-MMr L FLUO, i-MMs, i-MMs FLUO; i-XR3, i-XR3 L, i-XR3 L FLUO, i-XS4,Cefla S.C. | Apr 1, 2024367 days |
Recalls
Recall records for product code EBW per year, by the date the recall began. Each recalled product is one record; these 2 records belong to 2 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 1 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-1351-2018Class II | W & H DentalWerk Burmoos GMBH - Site 1During continuous postmarket surveillance activities, we have noticed a potential malfunction of individual foot controls of the type S-N1/S-N2. In case the er… | Dec 21, 2017 |
| Z-0273-2014Class II | Nobel Biocare Usa LlcThe recall was initiated because there is a potential for malfunction of the OsseoCare Pro when used with the OseeoCare Pro iPad application software release 1… | Oct 24, 2013 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code EBW. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 13 |
| 2018 | 46 |
| 2019 | 55 |
| 2020 | 71 |
| 2021 | 57 |
| 2022 | 31 |
| 2023 | 17 |
| 2024 | 14 |
| 2025 | 13 |
| 2026 (so far) | 9 |
Last 12 months by event type
- Deaths0
- Injuries0
- Malfunctions10
- Other or not stated0
Most reported device problems, last 12 months
- Mechanical Problem8
- Device Handling Problem3
- Activation, Positioning or Separation Problem1
- Overheating of Device1
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 547in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 148registered establishments that list this product code
Other codes under 21 CFR 872.4200
Product codes classified under the same regulation.
Watch EBW in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get EBW every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in EBW. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:EBWRetrieved Nov 29, 2026, 11:18 AM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:EBWRetrieved Nov 29, 2026, 11:18 AM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:EBW · count decision_dateRetrieved Nov 29, 2026, 11:18 AM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:EBWRetrieved Nov 29, 2026, 11:18 AM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:EBW · count event_date_initiatedRetrieved Nov 29, 2026, 11:18 AM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:EBW · count res_event_numberRetrieved Nov 29, 2026, 11:18 AM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-1351-2018" "Z-0273-2014")Retrieved Nov 29, 2026, 11:18 AM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:EBW AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 11:18 AM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:EBW AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 11:18 AM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:EBW AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 11:18 AM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:EBWRetrieved Nov 29, 2026, 11:18 AM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:EBWRetrieved Nov 29, 2026, 11:18 AM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:872.4200Retrieved Nov 28, 2026, 8:13 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.