FDA product code
MOS: Coil, Magnetic Resonance, Specialty (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 892.1000
- Review panel
- Radiology
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 49110 in the last 12 months
- Median review time
- 31 daysthe latest 25 decisions
- Recall records
- 380 in the last 12 months
- Adverse event reports
- 8last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code MOS per year, by decision date, 491 since 1995. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 9 |
| 2018 | 10 |
| 2019 | 3 |
| 2020 | 5 |
| 2021 | 4 |
| 2022 | 11 |
| 2023 | 10 |
| 2024 | 10 |
| 2025 | 3 |
| 2026 (so far) | 10 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K262813 | InkSpace Imaging Small Body Array (SBA12PH15x)Inkspace Imaging, Inc. | Oct 5, 202659 days |
| K262330 | dS Breast 7ch 1.5TdS Breast 7ch 3.0TdS Breast 16ch 1.5TdS Breast 16ch 3.0TdS Breast Coil 7ch 1.5TdS Sentinelle Breast 16ch 1.5TdS Sentinelle Breast 16ch 3.0TPhilips Medical Systems Nederland B.V. | Jul 30, 202621 days |
| K262294 | dS FootAnkle 16Ch 1.5T; dS FootAnkle 16Ch 3.0T; dS HiRes Hand/Wrist 16Ch 1.5T; dS HiRes Hand/Wrist 16CH 3.0T; dS Small Extremity 16Ch 1.5T; dS Small Extremity 16Ch 3.0TPhilips Medical Systems Nederland B.V. | Jul 24, 202617 days |
| K260947 | 0.55T 12Ch Breast CoilNeocoil, LLC | Apr 23, 202631 days |
| K252587 | HD 8Ch Wrist ArrayShenzhen RF Tech Co., Ltd. | Apr 14, 2026242 days |
| K252876 | GEM Flex Coil 16-L Array, 3.0T Receive Only; GEM Flex Coil 16-M Array, 3.0T Receive Only; GEM Flex Coil 16-S Array, 3.0T Receive OnlyShenzhen RF Tech Co., Ltd. | Apr 1, 2026203 days |
| K260519 | Smart Fit TorsoCardiac 1.5TPhilips Healthcare (Suzhou) Co., Ltd. | Mar 20, 202631 days |
| K252179 | GEM Flex Coil 16-L Array, 1.5T Receive Only; GEM Flex Coil 16-M Array, 1.5T Receive Only; GEM Flex Coil 16-S Array, 1.5T Receive OnlyShenzhen RF Tech Co., Ltd. | Feb 10, 2026214 days |
| K253738 | 3.0T AIR 32CH HNAGe Healthcare Coils (Usa Instruments, Inc.) | Jan 23, 202660 days |
| K254190 | dS Base 1.5T; dS Base 3.0T; dS Head 1.5T; dS Head 3.0T; dS HeadNeck 1.5T; dS HeadNeck 3.0TPhilips Medical Systems Nederland B.V. | Jan 14, 202622 days |
Recalls
Recall records for product code MOS per year, by the date the recall began. Each recalled product is one record; these 38 records belong to 17 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 8 |
| 2020 | 1 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 16 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-0536-2025Class II | Philips North America LlcPotential safety issue where a patient may be harmed while preparing for or during a scan. | Nov 4, 2024 |
| Z-0537-2025Class II | Philips North America LlcPotential safety issue where a patient may be harmed while preparing for or during a scan. | Nov 4, 2024 |
| Z-0538-2025Class II | Philips North America LlcPotential safety issue where a patient may be harmed while preparing for or during a scan. | Nov 4, 2024 |
| Z-0539-2025Class II | Philips North America LlcPotential safety issue where a patient may be harmed while preparing for or during a scan. | Nov 4, 2024 |
| Z-0540-2025Class II | Philips North America LlcPotential safety issue where a patient may be harmed while preparing for or during a scan. | Nov 4, 2024 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code MOS. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 9 |
| 2018 | 7 |
| 2019 | 8 |
| 2020 | 2 |
| 2021 | 3 |
| 2022 | 5 |
| 2023 | 4 |
| 2024 | 8 |
| 2025 | 2 |
| 2026 (so far) | 8 |
Last 12 months by event type
- Deaths0
- Injuries8
- Malfunctions0
- Other or not stated0
Most reported device problems, last 12 months
- Adverse Event Without Identified Device or Use Problem5
- Insufficient Device Problem Information5
- Improper or Incorrect Procedure or Method2
In plain English: what the newest reports describe Written by AI
Several reports describe burns, redness or warmth on the skin during MRI scans of other body parts, such as a foot, pelvis, shoulder, chest or wrist. In one case the burn was blamed on the coil's cord, and in another on the coil's hard housing touching the skin after the patient may have shifted position.
7 reports: MW5185981, MW5151781, MW5115653, 3007350713-2023-06191, 1056069-2024-00002, MW5160126, 3003768277-2024-007125
Several reports describe burns, blisters or redness on the chest or breast area during breast biopsies done with a breast coil used alongside an MRI scanner in 2025 and 2026. Several of these are part of a cluster the manufacturer was told about in March 2026.
6 reports: 3004929307-2026-00001, 3004929307-2026-00002, 3004929307-2026-00003, 3004929307-2026-00005, 1056069-2026-000001, 1056069-2026-000002
Several reports describe first and second degree burns (skin reddening and blisters) during prostate, pelvis, hip or thigh scans with the torso coil. The manufacturer checked these against its earlier burn reports and concluded they pose a potential risk of temporary, reversible harm.
4 reports: 1056069-2025-000001, 1056069-2024-000005, 1056069-2024-00004, 1056069-2024-00003
For the cluster of breast biopsy burns, the manufacturer's early checks and MRI testing of the coil found no faults. The cause of the burns is still unconfirmed, and detailed laboratory testing was still pending in the earlier reports.
3 reports: 3004929307-2026-00001, 3004929307-2026-00002, 3004929307-2026-00003
One report says a burn from an MRI scan of the hand got worse over time and that the patient later lost the tip of a finger. The patient links this to a known issue with the torso coil, but the report does not confirm the cause.
Report: 1051786-2025-000005
Written by AI from the narratives of the 20 newest reports that have one, received Mar 10, 2023 to Aug 25, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 1,965in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 820registered establishments that list this product code
Other codes under 21 CFR 892.1000
Product codes classified under the same regulation.
Watch MOS in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get MOS every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in MOS. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:MOSRetrieved Nov 29, 2026, 12:28 AM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:MOSRetrieved Nov 29, 2026, 4:49 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:MOS · count decision_dateRetrieved Nov 29, 2026, 12:28 AM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:MOSRetrieved Nov 29, 2026, 4:49 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:MOS · count event_date_initiatedRetrieved Nov 29, 2026, 12:28 AM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:MOS · count res_event_numberRetrieved Nov 29, 2026, 12:28 AM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-0536-2025" "Z-0537-2025" "Z-0538-2025" "Z-0539-2025" "Z-0540-2025")Retrieved Nov 29, 2026, 4:49 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:MOS AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 4:49 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:MOS AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 4:49 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:MOS AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 12:28 AM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:MOSRetrieved Nov 29, 2026, 12:28 AM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:MOSRetrieved Nov 29, 2026, 12:28 AM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:892.1000Retrieved Nov 29, 2026, 10:07 AM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.