FDA product code
QNI: Resorbable Implant For Anterior Cruciate Ligament (Acl) Repair (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 888.3044
- Review panel
- Orthopedic
- Premarket route
- 510(k)
- Implant
- Yes
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Ineligible
- 510(k) and De Novo decisions
- 31 in the last 12 months
- Median review time
- 195 daysthe latest 3 decisions
- Recall records
- 21 in the last 12 months
- Adverse event reports
- 13last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code QNI per year, by decision date, 3 since 2020. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 1 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 1 |
| 2026 (so far) | 1 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K251214 | BEAR® (Bridge-Enhanced ACL Restoration) ImplantMiach Orthopaedics, Inc. | Jan 13, 2026270 days |
| K243578 | BEAR® (Bridge-Enhanced ACL Restoration) ImplantMiach Orthopaedics, Inc. | Mar 6, 2025107 days |
| DEN200035 | BEAR (Bridge-Enhanced ACL Repair) ImplantMiach Orthopaedics, Inc. | Dec 16, 2020195 days |
Recalls
Recall records for product code QNI per year, by the date the recall began. Each recalled product is one record; these 2 records belong to 2 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 1 |
| 2024 | 0 |
| 2025 | 1 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-1220-2026Class II | Miach OrthopaedicsIncorrect expiration date on label that extends beyond the product's documented/approved shelf life. With an incorrect expiration date, the UDI will also be in… | Dec 31, 2025 |
| Z-0479-2026Class II | Miach OrthopaedicsThe field action was initiated in February of 2023 and was limited to the affected lot from sales force trunk stock only. | Feb 2, 2023 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code QNI. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 2 |
| 2025 | 29 |
| 2026 (so far) | 12 |
Last 12 months by event type
- Deaths0
- Injuries5
- Malfunctions8
- Other or not stated0
Most reported device problems, last 12 months
- Insufficient Device Problem Information5
- Material Too Soft/Flexible3
- Device Handling Problem2
- Adverse Event Without Identified Device or Use Problem1
- No Apparent Adverse Event1
- Separation Problem1
In plain English: what the newest reports describe Written by AI
Several reports describe the implant hydrating (soaking up blood, which it must do before placement) far too quickly, often in seconds, and in some cases falling apart or turning to mush in the surgeon's hands before it could be placed in the knee. In these cases the surgeon discarded the implant and used a second one from the same lot, and the procedure was completed.
8 reports: 3020652-2025-00008, 3020652-2025-00009, 3020652-2025-00010, 3020652-2025-00011, 3020652-2025-00012, 3020652-2025-00014, 3020652-2025-00026, 3020652-2025-00027
Some reports describe the implant looking unusual when opened. One report says the top and bottom were hard to tell apart and the surfaces looked rough, and a sales representative reported an implant that fell apart after hydration and was discarded without being used on a patient.
3 reports: 3020652-2025-00010, 3020652-2025-00015, 3020652-2026-00015
In some reports the implant did not soak up blood at all: it stayed firm, would not hydrate, and in one case it was removed from the knee still firm. In one report a surgeon tried to force blood into it, but it still did not hydrate properly.
2 reports: 3020652-2025-00031, 3020652-2025-00007
One patient reports that they were not told about or asked to consent to the implanted devices, and that after surgery for a Baker's cyst removal they had extreme pain, swelling, a reaction that has not healed, and bone thinning (osteopenia) in the thigh bone. An MRI described the ACL as having a torn area and degeneration, and the knee is unstable.
2 reports: MW5188938, 3020652-2026-00005
One report describes a small piece of the implant's base that stayed firm and did not soften, breaking off while the knee was being closed. The surgeon was not concerned, and the procedure was completed as normal.
Report: 3020652-2025-00007
Written by AI from the narratives of the 20 newest reports that have one, received Sep 4, 2025 to Sep 3, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 1in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 5registered establishments that list this product code
Watch QNI in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get QNI every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in QNI. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:QNIRetrieved Nov 29, 2026, 1:48 AM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:QNIRetrieved Nov 29, 2026, 5:16 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:QNI · count decision_dateRetrieved Nov 29, 2026, 1:48 AM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:QNIRetrieved Nov 29, 2026, 5:16 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:QNI · count event_date_initiatedRetrieved Nov 29, 2026, 1:48 AM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:QNI · count res_event_numberRetrieved Nov 29, 2026, 1:47 AM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-1220-2026" "Z-0479-2026")Retrieved Nov 29, 2026, 5:16 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:QNI AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 5:16 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:QNI AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 5:16 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:QNI AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 1:48 AM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:QNIRetrieved Nov 29, 2026, 1:48 AM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:QNIRetrieved Nov 29, 2026, 1:48 AM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:888.3044Retrieved Nov 29, 2026, 1:48 AM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil