Product codes

FDA product code

LCJ: Tissue Expander And Accessories (FDA product code)

Follow LCJ: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Unclassified
Regulation
None listed
Review panel
General and Plastic Surgery
Premarket route
510(k)
Implant
Yes
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Ineligible
510(k) and De Novo decisions
551 in the last 12 months
Median review time
153 daysthe latest 25 decisions
Recall records
2110 in the last 12 months
Adverse event reports
347last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code LCJ per year, by decision date, 55 since 1977. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20170
20183
20190
20200
20210
20220
20233
20241
20252
2026 (so far)1

Latest decisions

NumberDevice and applicantDecision and review time
K254126 Natrelle 133S Tissue ExpandersAbbVie Jan 16, 202628 days
K242963 MENTOR™ PliaForm™ Breast Tissue Expander w/ Suture TabsMentor Worldwide, LLC Apr 24, 2025211 days
K243836 Mentor™ CPX ™ 4 PLUS Enhance Breast Tissue ExpanderMentor Worldwide, LLC Jan 12, 202530 days
K241918 MENTOR™ CPX™4 PLUS Enhance Breast Tissue ExpanderMentor Worldwide, LLC Aug 2, 202432 days
K211676 Motiva Flora SmoothSilk Tissue ExpanderMotiva USA, LLC Oct 13, 2023864 days
K214124 AlloX2 Pro Tissue ExpandersSientra, Inc. Jun 8, 2023525 days
K221127 Sientra, inc. PortfinderSientra, Inc. May 10, 2023387 days
K180826 Natrelle 133 Plus MICROCELL Tissue ExpanderAllergan Dec 21, 2018266 days
K182335 CPX 4 Breast Tissue Expander with Smooth SurfaceMentor Worldwide, LLC Sep 25, 201828 days
K182054 Natrelle 133S Tissue ExpanderAllergan Aug 29, 201829 days

Recalls

Recall records for product code LCJ per year, by the date the recall began. Each recalled product is one record; these 21 records belong to 7 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20170
20180
20194
20200
20210
20220
20231
20240
20250
2026 (so far)10

Latest recalls

Recall and classFirm and reasonBegan
Z-1681-2026Class II Mentor Texas, LP.Infusion sets may contain dull/blunt needle tip, which may be difficult to advance, and/or break. Feb 18, 2026
Z-1682-2026Class II Mentor Texas, LP.Infusion sets may contain dull/blunt needle tip, which may be difficult to advance, and/or break. Feb 18, 2026
Z-1683-2026Class II Mentor Texas, LP.Infusion sets may contain dull/blunt needle tip, which may be difficult to advance, and/or break. Feb 18, 2026
Z-1684-2026Class II Mentor Texas, LP.Infusion sets may contain dull/blunt needle tip, which may be difficult to advance, and/or break. Feb 18, 2026
Z-1685-2026Class II Mentor Texas, LP.Infusion sets may contain dull/blunt needle tip, which may be difficult to advance, and/or break. Feb 18, 2026

Adverse event reports (MAUDE)

Medical device reports received per year that name product code LCJ. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
2017111
2018518
2019554
2020639
2021440
2022499
2023473
2024488
2025500
2026 (so far)313

Last 12 months by event type

  • Deaths0
  • Injuries274
  • Malfunctions72
  • Other or not stated1

Most reported device problems, last 12 months

  1. Material Rupture137
  2. Adverse Event Without Identified Device or Use Problem71
  3. Fluid/Blood Leak37
  4. Break31
  5. Device Appears to Trigger Rejection19
  6. Migration12

In plain English: what the newest reports describe Written by AI

  • Multiple reports describe the expander shell splitting, tearing or rupturing, sometimes along the seams or at a suture tab. Saline can then leak out and the expander deflates. Several of these devices were removed or exchanged.

    7 reports: 9617229-2026-20243, 9617229-2026-20245, 9617229-2026-20221, 1645337-2026-11628, 9617229-2026-19894, 9617229-2026-19895, 1645337-2026-11493

  • Multiple reports describe infection and fluid buildup (seroma) around the expander, with fever, cellulitis (skin infection) and pus. Some patients received antibiotics and were hospitalized, and at least two devices were removed.

    5 reports: 9617229-2026-20273, 9617229-2026-19975, 9617229-2026-19650, 1645337-2026-11746, 9617229-2026-19390

  • Two reports describe reduced blood flow to the nipple or superficial skin death around the areola after endoscopic breast surgery. A published article says the areola skin healed with conservative treatment.

    2 reports: 9617229-2026-20269, 9617229-2026-20273

  • One report describes bilateral breast pain after implantation. The patient was given expanders that were too large, and the pressure against the ribs and chest was felt as pain.

    Report: 1645337-2026-11487

  • One report says the outer plastic shell of the expander was nearly impossible to separate from the inner sterile shell. The device remained implanted.

    Report: 9617229-2026-19305

Written by AI from the narratives of the 17 newest reports that have one, received Oct 6, 2026 to Oct 27, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
2,298in the FDA’s Global Unique Device Identification Database
Establishment listings
53registered establishments that list this product code

Watch LCJ in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get LCJ every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in LCJ. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.