DXO: Transducer, Pressure, Catheter Tip (FDA product code)

Follow DXO: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 870.2870
Review panel
Cardiovascular
Premarket route
510(k)
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
641 in the last 12 months
Median review time
112 daysthe latest 25 decisions
Recall records
100 in the last 12 months
Adverse event reports
246last 12 months, 3 deaths

510(k) and De Novo decisions

Decisions for product code DXO per year, by decision date, 64 since 1979. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20172
20181
20194
20202
20210
20221
20230
20241
20251
2026 (so far)1

Latest decisions

NumberDevice and applicantDecision and review time
K251367 OptoMonitor 3Opsens, Inc. Jan 16, 2026260 days
K241418 OptoMonitor 3Opsens, Inc. Feb 12, 2025268 days
K233904 ACIST RXi System (016616); ACIST Navvus II Catheter (016675_Acist Medical Systems, Inc. Jul 17, 2024219 days
K222216 TruWave Disposable Pressure TransducerEdwards Lifesciences, LLC Dec 22, 2022150 days
K202943 OptoMonitor 3Opsens, Inc. Nov 24, 202055 days
K193620 OptoMonitor 3Opsens, Inc. Jun 18, 2020175 days
K193279 Comet II Pressure GuidewireBoston Scientific Corporation Dec 23, 201926 days
K192340 OptoMonitorOpsens, Inc. Dec 12, 2019106 days
K190852 Zurich Pressure Guidewire System Model 100Zurich Medical, Inc. Aug 14, 2019134 days
K183413 TruWave Disposable Pressure TransducerEdwards Lifeciences, LLC May 1, 2019142 days

Recalls

Recall records for product code DXO per year, by the date the recall began. Each recalled product is one record; these 10 records belong to 3 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20170
20180
20190
20200
20218
20220
20230
20240
20250
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-1959-2021Class II Edwards Lifesciences, LLCThere is a potential that the pressure tubing may detach from the blood sampling system May 26, 2021
Z-1960-2021Class II Edwards Lifesciences, LLCThere is a potential that the pressure tubing may detach from the blood sampling system May 26, 2021
Z-1961-2021Class II Edwards Lifesciences, LLCThere is a potential that the pressure tubing may detach from the blood sampling system May 26, 2021
Z-1962-2021Class II Edwards Lifesciences, LLCThere is a potential that the pressure tubing may detach from the blood sampling system May 26, 2021
Z-1963-2021Class II Edwards Lifesciences, LLCThere is a potential that the pressure tubing may detach from the blood sampling system May 26, 2021

Adverse event reports (MAUDE)

Medical device reports received per year that name product code DXO. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
2017186
2018179
2019153
2020158
2021116
2022144
2023194
2024197
2025183
2026 (so far)218

Last 12 months by event type

  • Deaths3
  • Injuries27
  • Malfunctions214
  • Other or not stated2

Most reported device problems, last 12 months

  1. Detachment of Device or Device Component39
  2. Break31
  3. Communication or Transmission Problem24
  4. Difficult to Remove22
  5. Contamination20
  6. Material Separation20

In plain English: what the newest reports describe Written by AI

  • Several reports describe the wireless guidewire (used to measure pressure in heart arteries) bending, kinking, deforming or having difficulty advancing or torquing. Users switched to another guidewire to finish the procedure in most of these cases.

    6 reports: 2024168-2026-04089, 2024168-2026-03901, 2024168-2026-03777, 2024168-2026-03589, 2024168-2026-03989, 2024168-2026-04055

  • Several reports describe the tip of a pressure guidewire breaking off or tearing inside the patient. In one report the detached tip could not be retrieved with a snare and the patient needed cardiac surgery to remove it.

    3 reports: 2124215-2026-51993, 2124215-2026-50860, 2024168-2026-04014

  • Several reports describe disposable pressure transducers (devices that convert blood pressure into readings) leaking from the drip chamber, housing, or tubing joint. In one case the tubing had fully detached from its bond. Reports say no patient injury was alleged.

    2 reports: 2015691-2026-19694, 2015691-2026-19202

  • Evaluation of pressure tubing in two transducer reports found dark particles inside the line, and one report describes a burn mark on the tubing. The source of these particles was not identified.

    2 reports: 2015691-2026-19694, 2015691-2026-19526

  • One report describes a stopcock (a valve on the line) on a pressure transducer shearing off, leaving sharp edges and a broken line. Another report describes an inaccurate arterial pressure reading with no alarm, which was resolved by replacing the unit.

    2 reports: 50802623, 2015691-2026-18401

Written by AI from the narratives of the 20 newest reports that have one, received Sep 4, 2026 to Oct 29, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
791in the FDA’s Global Unique Device Identification Database
Establishment listings
116registered establishments that list this product code

Watch DXO in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get DXO every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in DXO. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.