FDA device news · Product code LNS
FDA 510(k) clearance
Break Wave (K252913): Sonomotion, Inc. 510(k) clearance
San Mateo, CA
- Decision
- Substantially Equivalent
- Decision date
- Jan 12, 2026
- Received
- Sep 12, 2025
- 510(k) type
- Traditional
- Product code
- LNSLithotriptor, Extracorporeal Shock-Wave, Urological
- Review panel
- Gastroenterology, Urology
- Device class
- Class II
- Regulation
- 21 CFR 876.5990
- Third-party review
- No
- Summary or statement
- Summary
- Review time
- 122 daysreceipt to decision, calendar days
- Median in LNS
- 178 daysthe latest 24 other decisions
- LNS decisions
- 2in the last 12 months
- Applicant’s 510(k)s
- 4in openFDA, this one included
In plain English Written by AI
What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.
- What it is
- The Break Wave is an extracorporeal shock wave lithotripter, a portable device that uses focused acoustic (sound) pulses, applied through the skin, to break kidney and ureter stones into small fragments that can pass on their own. It includes a diagnostic ultrasound imaging unit that guides the operator to the stone and lets them watch the fragmentation in real time.
- What it’s cleared for
- The Break Wave is cleared to fragment urinary stones in the kidney (renal pelvis and renal calyces) and the ureter (upper and lower ureter). The summary says the indication is the same as the predicate device.
- Compared with (the predicate)
- K072538, Piezolith 3000 Triple - focus extracorporeal lithotripter
- How it differs
- The summary says the Break Wave uses a lower amplitude, longer duration pulse and a higher pulse rate, and has a smaller probe head with a fixed focus lens. It can be hand-held or held on a mechanical arm, couples directly to the skin without a water-filled membrane, and does not need an integrated table or water circulation system.
- Testing the summary lists
- Electrical safety and electromagnetic compatibility (IEC 60601-1, IEC 60601-1-2)
- Lithotripsy device requirements (IEC 61846, IEC 60601-2-36)
- Reprocessing (ANSI AAMI ST98:2022)
- Biocompatibility (ISO 10993-1)
- Usability (IEC 62366)
- Software verification and validation (IEC 62304)
Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.
How the review time compares
Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.
K252913 took 122 calendar days from receipt to decision. Of the latest 24 other decisions in product code LNS, 15 took longer, 9 were faster; their median was 178 days.
Latest decisions in LNS
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K261086 | Break WaveSonomotion, Inc. · Special | Apr 29, 202628 days |
| K252913 | Break WaveSonomotion, Inc. · Traditional | Jan 12, 2026122 days |
| K242922 | Extracorporeal Shock Wave Lithotripter (U200)Shenzhen Wikkon Precision Technologies Co., Ltd. · Traditional | Jun 18, 2025267 days |
| K221903 | Delta III ProDornier Medtech America, Inc. · Traditional | Feb 22, 2023237 days |
| K213772 | LM-9300 Plus LithotripterLite-Med, Inc. · Traditional | Jan 19, 2023413 days |
| K201001 | Modulith SLX-F2Karl Storz Endoscopy America · Traditional | Mar 30, 2021348 days |
| K201074 | Delta III LithotripterDornier Medtech America, Inc. · Traditional | Sep 3, 2020134 days |
| K172084 | Delta III LithotripterDornier Medtech America, Inc. · Special | Aug 8, 201729 days |
| K170122 | Delta III LithotripterDornier Medtech America · Traditional | Jun 28, 2017166 days |
| K151298 | Gemini XXP-HPDornier Medtech America · Traditional | Jul 30, 201576 days |
Other 510(k)s from Sonomotion, Inc.
The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.
Watch LNS in DeviceVigil
DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.
Get LNS every Monday
One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in LNS. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/510k · search k_number:K252913Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:LNSRetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/classification · search product_code:LNSRetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:LNS AND decision_date:[20251129 TO 20261129]Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search applicant:"Sonomotion, Inc."Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.