FDA product code
JAQ: System, Applicator, Radionuclide, Remote-Controlled (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 892.5700
- Review panel
- Radiology
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 2083 in the last 12 months
- Median review time
- 107 daysthe latest 25 decisions
- Recall records
- 523 in the last 12 months
- Adverse event reports
- 5last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code JAQ per year, by decision date, 208 since 1978. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 7 |
| 2018 | 7 |
| 2019 | 2 |
| 2020 | 4 |
| 2021 | 0 |
| 2022 | 1 |
| 2023 | 2 |
| 2024 | 10 |
| 2025 | 6 |
| 2026 (so far) | 2 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K262383 | LumenCare AzureNucletron B.V. | Jul 24, 202611 days |
| K252717 | MAASTRO Rectal Applicator SetVarian Medical Systems | May 20, 2026265 days |
| K251037 | Rectal ApplicatorNucletron B.V. | Dec 19, 2025260 days |
| K252884 | Guided Aarhus Applicator SetVarian Medical Systems | Nov 19, 202570 days |
| K250299 | CT/ MR Tandem and Ovoid SetVarian Medical Systems, Inc. | May 9, 202598 days |
| K250289 | Intracavitary/Interstitial SystemVarian Medical Systems | Apr 25, 202584 days |
| K243939 | Titanium Flexible Geometry FSD Applicator Set (GM11013400)Varian Medical Systems, Inc. | Mar 3, 202573 days |
| K241764 | Mould Probe MR SafeVarian Medical Systems, Inc. | Jan 10, 2025204 days |
| K242961 | Intraluminal Applicator SetVarian Medical Systems | Nov 26, 202462 days |
| K243045 | “BRAXX” Esophageal Brachytherapy Applicator (Model: BKMI2001, BKMI2002)Braxx Biotech Co., Ltd. | Nov 19, 202453 days |
Recalls
Recall records for product code JAQ per year, by the date the recall began. Each recalled product is one record; these 52 records belong to 27 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 1 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 3 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-2710-2026Class II | Elekta, Inc.Due to safety concerns regarding the detachment of the tip of the intrauterine tube (IU tube) that may remain within a patient. | Apr 3, 2026 |
| Z-2711-2026Class II | Elekta, Inc.Due to safety concerns regarding the detachment of the tip of the intrauterine tube (IU tube) that may remain within a patient. | Apr 3, 2026 |
| Z-2712-2026Class II | Elekta, Inc.Due to safety concerns regarding the detachment of the tip of the intrauterine tube (IU tube) that may remain within a patient. | Apr 3, 2026 |
| Z-0772-2020Class II | Nucletron BVA component failure may lead to treatment interruption or incorrect source positioning. | Nov 18, 2019 |
| Z-1829-2015Class II | Varian Medical Systems IncVarian has discovered that there is a discrepancy in the absolute dose rate given with the Leipzig-style Surface Applicator's Instructions for Use (IFU): Dose… | May 18, 2015 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code JAQ. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 5 |
| 2018 | 2 |
| 2019 | 8 |
| 2020 | 8 |
| 2021 | 5 |
| 2022 | 5 |
| 2023 | 3 |
| 2024 | 1 |
| 2025 | 3 |
| 2026 (so far) | 5 |
Last 12 months by event type
- Deaths0
- Injuries3
- Malfunctions2
- Other or not stated0
Most reported device problems, last 12 months
- Fracture1
- Intermittent Communication Failure1
In plain English: what the newest reports describe Written by AI
During removal of applicator parts, pieces broke or came loose, and in some cases a piece stayed in the patient. One report describes a plastic needle piece left inside, and another describes a tip that was missing after removal. Doctors or surgeons removed the retained pieces.
7 reports: 9611894-2026-00002, 9611894-2025-00001, 9611894-2025-00002, MW5182495, 1721504-2021-00042, 1721504-2021-00028, 15885418
Two reports describe a part breaking outside the patient. In one, the tandem stem of the applicator broke off after removal. In the other, the metal tip of a quality-check measuring tool broke off, and a CT scan showed it was not inside the patient, though care was delayed.
2 reports: 17623019, MW5116903
The software step that rebuilds the catheter's shape may produce an incorrect result. The manufacturer found a possible bending problem in version 4.0 and newer that may be hard for users to notice. A customer separately reported a possible catheter reconstruction issue.
2 reports: 9611894-2022-00001, 9611894-2021-00002
One report describes an obstruction or friction in the applicator tube that stopped the full radiation dose from being delivered. The patient received a partial dose, needed another procedure and an extra night in hospital. The cause is unknown and could be a defect or debris.
Report: MW5111998
One report says a transfer tube channel on the afterloader measured 119.9 cm in 2021 but 122.9 cm in 2022. The reporter said the longer tube may have caused underexposure, so patients may have received doses different from those prescribed.
Report: MW5111948
Written by AI from the narratives of the 20 newest reports that have one, received May 6, 2021 to May 11, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 4,244in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 182registered establishments that list this product code
Watch JAQ in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get JAQ every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in JAQ. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:JAQRetrieved Nov 29, 2026, 7:07 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:JAQRetrieved Nov 29, 2026, 7:07 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:JAQ · count decision_dateRetrieved Nov 29, 2026, 7:07 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:JAQRetrieved Nov 29, 2026, 7:07 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:JAQ · count event_date_initiatedRetrieved Nov 29, 2026, 7:07 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:JAQ · count res_event_numberRetrieved Nov 29, 2026, 7:07 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-2710-2026" "Z-2711-2026" "Z-2712-2026" "Z-0772-2020" "Z-1829-2015")Retrieved Nov 29, 2026, 7:07 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:JAQ AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 7:07 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:JAQ AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 7:07 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:JAQ AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 7:07 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:JAQRetrieved Nov 29, 2026, 7:07 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:JAQRetrieved Nov 29, 2026, 7:07 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:892.5700Retrieved Nov 29, 2026, 7:07 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil