FDA product code
JHI: Visual, Pregnancy Hcg, Prescription Use (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 862.1155
- Review panel
- Clinical Chemistry
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 3761 in the last 12 months
- Median review time
- 113 daysthe latest 25 decisions
- Recall records
- 474 in the last 12 months
510(k) and De Novo decisions
Decisions for product code JHI per year, by decision date, 376 since 1976. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 1 |
| 2018 | 1 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 1 |
| 2023 | 0 |
| 2024 | 1 |
| 2025 | 0 |
| 2026 (so far) | 1 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K262299 | Instant-View® hCG Combo II Cassette TestAlfa Scientific Designs, Inc. | Sep 22, 202677 days |
| K241919 | Innovita HCG Pregnancy Rapid Combo TestInnovita (Tangshan) Biological Technology Co., Ltd. | Aug 2, 202432 days |
| K203272 | Alltest Pregnancy Rapid Combo Test CassetteHangzhou AllTest Biotech Co., Ltd. | Jan 31, 2022451 days |
| K172322 | Atellica IM Total hCG (ThCG)Siemens Healthcare Diagnostics, Inc. | Mar 29, 2018240 days |
| K172257 | TrueDX hCG Early Result Pregnancy Test (Midstream Format), TrueDX hCG Early Result Pregnancy Test (Cassette Format), VeriClear Early Result Pregnancy Test (Midstream Format), VeriClear Early Result Pregnancy Test (Cassette Format)True Diagnostics, Inc. | Dec 22, 2017149 days |
| K152768 | Assure Tech hCG Pregnancy Serum/Urine Combo Test Cassette, Assure Tech hCG Pregnancy Serum/Urine Combo Test StripAssure Tech. Co., Ltd. | Feb 24, 2016153 days |
| K130456 | WUNDER PREGNANCY TESTJames Nguyen | Apr 8, 2014410 days |
| K132834 | FASTEP S10 HCG SERUM/URINE COMBO TESTPolymed Therapeutics, Inc. | Jan 9, 2014121 days |
| K131166 | SOFIA(R) HCG FIAQuidel Corp. | Aug 2, 2013100 days |
| K123844 | CHEMTRUE HCG PREGNANCY URINE DIPSTICK, HCG PREGNANCY URINE CASSETTE TEST, AND HCG COMBO SERUM/URINE CASSETTE TESTChemtron Biotech, Inc. | May 30, 2013168 days |
Recalls
Recall records for product code JHI per year, by the date the recall began. Each recalled product is one record; these 47 records belong to 28 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 1 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 2 |
| 2023 | 6 |
| 2024 | 1 |
| 2025 | 0 |
| 2026 (so far) | 4 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-3271-2026Class II | Alere San Diego, Inc.hCG rapid tests may detect low levels of hCG below the stated assay sensitivity, which may result in false positive result. | Aug 20, 2026 |
| Z-3272-2026Class II | Alere San Diego, Inc.hCG rapid tests may detect low levels of hCG below the stated assay sensitivity, which may result in false positive result. | Aug 20, 2026 |
| Z-3273-2026Class II | Alere San Diego, Inc.hCG rapid tests may detect low levels of hCG below the stated assay sensitivity, which may result in false positive result. | Aug 20, 2026 |
| Z-3274-2026Class II | Alere San Diego, Inc.hCG rapid tests may detect low levels of hCG below the stated assay sensitivity, which may result in false positive result. | Aug 20, 2026 |
| Z-0259-2025Class II | Princeton Biomeditech CorpThis recall was initiated due to customer complaints of false negative of hCG test results when running patient urine samples. Use of the above listed product… | Aug 23, 2024 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code JHI. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
This section didn’t load from openFDA. It’s retried within the hour.
Last 12 months by event type
This section didn’t load from openFDA. It’s retried within the hour.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 231in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 215registered establishments that list this product code
Other codes under 21 CFR 862.1155
Product codes classified under the same regulation.
Watch JHI in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get JHI every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in JHI. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:JHIRetrieved Nov 28, 2026, 3:24 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:JHIRetrieved Nov 28, 2026, 3:24 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:JHI · count decision_dateRetrieved Nov 28, 2026, 3:24 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:JHIRetrieved Nov 28, 2026, 3:24 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:JHI · count event_date_initiatedRetrieved Nov 28, 2026, 3:24 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:JHI · count res_event_numberRetrieved Nov 28, 2026, 3:24 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-3271-2026" "Z-3272-2026" "Z-3273-2026" "Z-3274-2026" "Z-0259-2025")Retrieved Nov 28, 2026, 3:24 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:JHIRetrieved Nov 28, 2026, 3:24 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:JHIRetrieved Nov 28, 2026, 3:24 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:862.1155Retrieved Nov 29, 2026, 5:00 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.