FDA product code
ITX: Transducer, Ultrasonic, Diagnostic (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 892.1570
- Review panel
- Radiology
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 4156 in the last 12 months
- Median review time
- 139 daysthe latest 25 decisions
- Recall records
- 1470 in the last 12 months
- Adverse event reports
- 649last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code ITX per year, by decision date, 415 since 1978. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 6 |
| 2018 | 2 |
| 2019 | 1 |
| 2020 | 3 |
| 2021 | 3 |
| 2022 | 2 |
| 2023 | 6 |
| 2024 | 3 |
| 2025 | 4 |
| 2026 (so far) | 6 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K260114 | Align® Cover and Pad Guidance ToolCIVCO Medical Instruments Co., Inc. | Sep 24, 2026253 days |
| K261170 | BA-004CCFujifilm Corporation | Aug 12, 2026125 days |
| K254143 | EM-TECH Passive Ultrasound Probe StrapEM-TECH Solutions, Inc. | Aug 10, 2026231 days |
| K260161 | SmartGuide Disposable Guides and GridsUC-CARE , Ltd. | Jul 30, 2026191 days |
| K262474 | ND Needle Guidance SystemAccurate Access, LLC | Jul 27, 202610 days |
| K261014 | Accuro® 3S Needle Guide KitRivanna Medical, Inc. | Jul 24, 2026119 days |
| K250883 | ULTRASONIC PROBE UM-3R (UM-3R); ULTRASONIC PROBE UM-G20-29R (UM-G20-29R)Olympus Medical Systems Corporation | Sep 18, 2025178 days |
| K250762 | ULTRASONIC PROBE UM-S20-17S (UM-S20-17S); ULTRASONIC PROBE UM-S20-20R (UM-S20-20R)Olympus Medical Systems Corporation | Jul 11, 2025120 days |
| K250524 | Mendaera Guidance SystemMendaera, Inc. | Jul 2, 2025131 days |
| K250469 | Accuro® 3S Needle Guide KitRivanna Medical, Inc. | Jun 13, 2025115 days |
Recalls
Recall records for product code ITX per year, by the date the recall began. Each recalled product is one record; these 147 records belong to 59 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 1 |
| 2018 | 18 |
| 2019 | 4 |
| 2020 | 4 |
| 2021 | 35 |
| 2022 | 6 |
| 2023 | 4 |
| 2024 | 1 |
| 2025 | 38 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-1006-2026Class II | Sheathing Technologies IncVaginal/rectal ultrasound probe covers with a viral barrier may have product defects, which may result in contamination during use, posing an infection risk. | Nov 17, 2025 |
| Z-1618-2026Class III | Philips Ultrasound, IncTo provide clarification and labelling to define the useful life of ultrasound transducers in the field. | Sep 5, 2025 |
| Z-1619-2026Class III | Philips Ultrasound, IncTo provide clarification and labelling to define the useful life of ultrasound transducers in the field. | Sep 5, 2025 |
| Z-1620-2026Class III | Philips Ultrasound, IncTo provide clarification and labelling to define the useful life of ultrasound transducers in the field. | Sep 5, 2025 |
| Z-1621-2026Class III | Philips Ultrasound, IncTo provide clarification and labelling to define the useful life of ultrasound transducers in the field. | Sep 5, 2025 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code ITX. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 412 |
| 2018 | 130 |
| 2019 | 119 |
| 2020 | 116 |
| 2021 | 146 |
| 2022 | 87 |
| 2023 | 138 |
| 2024 | 988 |
| 2025 | 575 |
| 2026 (so far) | 599 |
Last 12 months by event type
- Deaths0
- Injuries12
- Malfunctions635
- Other or not stated2
Most reported device problems, last 12 months
- Poor Quality Image156
- Break72
- Electrical /Electronic Property Problem65
- Material Deformation54
- No Apparent Adverse Event38
- Material Twisted/Bent33
In plain English: what the newest reports describe Written by AI
Several reports describe kinks or bends in the probe cable or insertion tube, some found during reprocessing (cleaning and preparing between uses) and some in returned devices. One report also noted the probe shaking violently, and another noted oil leaking from the insertion tube.
7 reports: 3002808148-2026-36958, 3002808148-2026-36961, 3002808148-2026-36965, 3002808148-2026-35254, 3002808148-2026-35313, 3002808148-2026-35265, 3002808148-2026-37251
Tips and tubes were cracked, fractured, elongated or delaminated (layers separated), with wires exposed in one report. Other damage included bumps, scratches and folds on the probe.
5 reports: 3002808148-2026-37518, 3002808148-2026-37241, 3002808148-2026-37244, 3002808148-2026-35312, 3002808148-2026-35449
During breast reconstruction surgery, Doppler probes (devices that use sound to check blood flow) that made sound when tested gave no sound once placed on the artery. One report says two more probes from the same lot also failed, and another says a second probe was used and worked. No patient harm was reported.
3 reports: 1216677-2026-00199, 1216677-2026-00205, 1216677-2026-00177
One report describes a structural break about three inches from the tip of a heart ultrasound probe used through the esophagus during a heart valve procedure, with exposed metal and blood on the probe. Tests afterward showed no esophageal or stomach injury, but the patient needed an additional procedure because of the device failure.
Report: MW6004271
One report describes endoscopes used to examine airways where patient specimens tested positive for mold and fungus. Scope cultures were negative and no patient infections were reported, but reprocessor culture results were still pending.
Report: 3002808148-2026-35429
Written by AI from the narratives of the 19 newest reports that have one, received Oct 8, 2026 to Oct 29, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 8,904in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 3,411registered establishments that list this product code
Other codes under 21 CFR 892.1570
Product codes classified under the same regulation.
Watch ITX in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get ITX every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in ITX. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:ITXRetrieved Nov 29, 2026, 6:00 AM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:ITXRetrieved Nov 29, 2026, 6:38 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:ITX · count decision_dateRetrieved Nov 29, 2026, 6:00 AM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:ITXRetrieved Nov 29, 2026, 6:38 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:ITX · count event_date_initiatedRetrieved Nov 29, 2026, 6:00 AM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:ITX · count res_event_numberRetrieved Nov 29, 2026, 6:00 AM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-1006-2026" "Z-1618-2026" "Z-1619-2026" "Z-1620-2026" "Z-1621-2026")Retrieved Nov 29, 2026, 6:38 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:ITX AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 6:38 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:ITX AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 6:38 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:ITX AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 6:00 AM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:ITXRetrieved Nov 29, 2026, 6:00 AM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:ITXRetrieved Nov 29, 2026, 6:00 AM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:892.1570Retrieved Nov 28, 2026, 11:19 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil