Product codes

FDA product code

IRT: Pad, Heating, Powered (FDA product code)

Follow IRT: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 890.5740
Review panel
Physical Medicine
Premarket route
510(k) exempt
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
420 in the last 12 months
Median review time
188 daysthe latest 25 decisions
Recall records
241 in the last 12 months
Adverse event reports
114last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code IRT per year, by decision date, 42 since 1977. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20170
20180
20190
20200
20210
20221
20230
20240
20251
2026 (so far)0

Latest decisions

NumberDevice and applicantDecision and review time
K250047 bite away twomibeTec GmbH Sep 12, 2025245 days
K220514 bite away neomibeTec GmbH Jun 23, 2022121 days
K160943 Bite AwayRiemser Pharma Aug 26, 2016143 days
K142228 Heated Eye PadDigital Heat Corporation Feb 9, 2015180 days
K133981 AVACEN 100Avacen, Inc. Aug 15, 2014232 days
K111273 FIR HEAT THERAPHY SYSYEMSShenzhen Anpan Technology Co., Ltd. Nov 8, 2011188 days
K072534 BIO-MAT 2000Richway Intl., Inc. Feb 8, 2008154 days
K963887 PT-PAC PORTABLE HEATING PADAnthony A. Chiurco Mar 31, 1997186 days
K963824 LI'S ITCH STOPPERHontech Foundation For Medical Technology Dec 12, 199680 days
K962195 HEALTH O METER ELECTRIC HEATING PADS MODELS EHP2, EHP3Sunbeam Health Division Nov 27, 1996173 days

Recalls

Recall records for product code IRT per year, by the date the recall began. Each recalled product is one record; these 24 records belong to 7 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20170
20180
20190
20200
20210
20223
20230
20241
20250
2026 (so far)1

Latest recalls

Recall and classFirm and reasonBegan
Z-1588-2026Class I Navajo Manufacturing CompanyIf heating pad is folded while in use, the power density increases in the folded area causing excessive temperatures, which can also be caused by placing pad u… Feb 18, 2026
Z-2586-2024Class II EYE COMFORT CARE LLCDue to a change in the safety clips for the heated eye masks, the wires to the safety clips may detach as a result of tugging and pulling and may cause a short… Feb 8, 2024
Z-0236-2023Class I Whele LLC DBA PerchThe products may have the potential to burn customers. Firm has received complaints of burns, mild shocks, and rashes/irritation. Oct 24, 2022
Z-0237-2023Class I Whele LLC DBA PerchThe products may have the potential to burn customers. Firm has received complaints of burns, mild shocks, and rashes/irritation. Oct 24, 2022
Z-0238-2023Class I Whele LLC DBA PerchThe products may have the potential to burn customers. Firm has received complaints of burns, mild shocks, and rashes/irritation. Oct 24, 2022

Adverse event reports (MAUDE)

Medical device reports received per year that name product code IRT. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
2017220
2018325
2019205
2020117
202197
2022134
2023321
202477
202571
2026 (so far)112

Last 12 months by event type

  • Deaths0
  • Injuries80
  • Malfunctions24
  • Other or not stated10

Most reported device problems, last 12 months

  1. Labelling, Instructions for Use or Training Problem23
  2. Overheating of Device18
  3. Improper or Incorrect Procedure or Method10
  4. Temperature Problem9
  5. Insufficient Device Problem Information6
  6. Fire5

In plain English: what the newest reports describe Written by AI

  • Several reports describe skin burns after people lay, leaned or sat on the pad, including burns to the back, leg, shoulder, hip and breast. One user fell asleep on it and reported burns and blisters over much of the body. Some reports are attorney letters describing claimed injuries.

    11 reports: 3010805625-2026-00023, 3007790958-2026-00022, 3010805625-2026-00020, 3007790958-2026-00018, 3007790958-2026-00017, 3010805625-2026-00016, 3010805625-2026-00021, 3010805625-2026-00015, 1222304-2026-00009, 1222304-2026-00007, 3030539818-2026-00016

  • Reports describe the device smoking, sparking or catching fire, sometimes at the plug or cord, and scorching bed sheets, carpet or a pillowcase. One report gives a plug temperature over 250°F, measured with a heat gun, and a skin burn. Another report describes a fire but no injury, and the manufacturer is still investigating it.

    6 reports: 3011164092-2026-00102, 3011164092-2025-00059, 3030539818-2026-00017, 1000143485-2026-29201, 3007790958-2026-00018, 3011077581-2026-00001

  • Reports say the pad itself was burned or burnt out. A dialysis patient, who uses the pad during peritoneal dialysis (a blood-cleaning treatment done in the abdomen), needed a replacement.

    3 reports: MW6002936, 1222304-2026-00009, 3030539818-2026-00016

  • One report says a dialysis patient found the pad was not warming as intended, and that patients sometimes experience cramping when using it. The patient was sent a replacement.

    Report: MW6000729

  • One report describes a separate heated medical unit overheating at its power port during sleep, causing smoke and burns close to a 73-year-old user's head. The reporter says the manufacturer refused a full refund and asked for the burned unit back for an internal review.

    Report: MW5189524

Written by AI from the narratives of the 20 newest reports that have one, received Jun 6, 2026 to Oct 22, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
1,093in the FDA’s Global Unique Device Identification Database
Establishment listings
306registered establishments that list this product code

Product codes classified under the same regulation.

Watch IRT in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get IRT every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in IRT. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.