FDA product code
CAC: Apparatus, Autotransfusion (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 868.5830
- Review panel
- Cardiovascular
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- Yes
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 1950 in the last 12 months
- Median review time
- 126 daysthe latest 25 decisions
- Recall records
- 890 in the last 12 months
510(k) and De Novo decisions
Decisions for product code CAC per year, by decision date, 195 since 1977. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 1 |
| 2018 | 2 |
| 2019 | 1 |
| 2020 | 1 |
| 2021 | 4 |
| 2022 | 1 |
| 2023 | 2 |
| 2024 | 3 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K241236 | XTRA Collection sets; XTRA Sequestration set XSorin Group Italia S.R.L. | Aug 16, 2024106 days |
| K240426 | Aventus Clot Management SystemInquis Medical | May 14, 202491 days |
| K240584 | XTRA Autotransfusion System (with XTRA Bowl sets)Sorin Group Italia S.R.L. | Apr 29, 202459 days |
| K231782 | FlowSaver Blood Return System (80-101)Inari Medical, Inc. | Jul 14, 202328 days |
| K221483 | FlowSaver Blood Return SystemInari Medical, Inc. | Feb 17, 2023270 days |
| K221722 | Haemonetics Cell Saver Elite/Elite+ Autotransfusion System (CSE-E-US/CSE-EW-US)Haemonetics Corporation | Nov 15, 2022154 days |
| K201305 | Auto transfusion (ATS) Chest Drains: Ocean Water Seal Chest Drain, Oasis Dry Suction Water Seal Chest Drain, Express Dry Seal Chest DrainAtrium Medical Corporation | Sep 3, 2021476 days |
| K210862 | Hemafuse SystemSisu Global Health | Jul 29, 2021128 days |
| K210176 | FlowSaver Blood FilterInari Medical | Jul 22, 2021181 days |
| K210089 | CATSmart, Automated Blood Processing Autotransfusion SystemFresenius Kabi AG | Feb 11, 202129 days |
Recalls
Recall records for product code CAC per year, by the date the recall began. Each recalled product is one record; these 89 records belong to 29 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 2 |
| 2018 | 4 |
| 2019 | 24 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 7 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-0709-2024Class I | Maquet Cardiovascular, LLCExpress chest drains are pre-packaged with sterile water syringes which have bene recalled. If the device is not kept in an upright position, the patient may b… | Dec 13, 2023 |
| Z-0710-2024Class I | Maquet Cardiovascular, LLCExpress chest drains are pre-packaged with sterile water syringes which have bene recalled. If the device is not kept in an upright position, the patient may b… | Dec 13, 2023 |
| Z-0621-2024Class II | Atrium Medical CorporationA sterilization nonconformance prematurely aged the product by 48.3 days, resulting in the expiration date on the label not being accurate. This nonconforming… | Nov 1, 2023 |
| Z-0316-2024Class II | Atrium Medical CorporationInstructions for Use (IFU) for the Atrium Ocean, Oasis, and Express chest drains do not provide sufficient precaution instruction for proper set up of catheter… | Sep 18, 2023 |
| Z-0317-2024Class II | Atrium Medical CorporationInstructions for Use (IFU) for the Atrium Ocean, Oasis, and Express chest drains do not provide sufficient precaution instruction for proper set up of catheter… | Sep 18, 2023 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code CAC. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
This section didn’t load from openFDA. It’s retried within the hour.
Last 12 months by event type
This section didn’t load from openFDA. It’s retried within the hour.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 377in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 148registered establishments that list this product code
Other codes under 21 CFR 868.5830
Product codes classified under the same regulation.
Watch CAC in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get CAC every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in CAC. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:CACRetrieved Nov 29, 2026, 4:59 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:CACRetrieved Nov 24, 2026, 5:13 PM UTC. FDA data updated Nov 9, 2026.
- device/510k · search product_code:CAC · count decision_dateRetrieved Nov 24, 2026, 5:13 PM UTC. FDA data updated Nov 9, 2026.
- device/recall · search product_code:CACRetrieved Nov 29, 2026, 4:59 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:CAC · count event_date_initiatedRetrieved Nov 29, 2026, 4:59 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:CAC · count res_event_numberRetrieved Nov 29, 2026, 4:59 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-0709-2024" "Z-0710-2024" "Z-0621-2024" "Z-0316-2024" "Z-0317-2024")Retrieved Nov 24, 2026, 5:13 PM UTC. FDA data updated Nov 11, 2026.
- device/udi · search product_codes.code:CACRetrieved Nov 29, 2026, 4:59 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:CACRetrieved Nov 29, 2026, 4:59 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:868.5830Retrieved Nov 29, 2026, 2:23 AM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil