The FDA cleared 47 510(k)s, granted no De Novo requests and approved no original PMAs. 15 device recalls were posted, 1 of them Class I. Every 510(k), De Novo and PMA decision dated January 27 to February 2, 2025, and every device recall the FDA posted those days.
- 510(k) clearances
- 47median review 121 days
- De Novo grants
- 0
- Original PMA approvals
- 0plus 23 supplements
- Recalls posted
- 1548 products, 1 Class I
Recalls posted
Device recalls the FDA posted January 27 to February 2, 2025, Class I first. Class I is the most serious: a reasonable chance the product causes serious harm or death. The class comes from the FDA’s weekly enforcement report, so the newest recalls can show “Class pending” for a few days.
| Recall and class | Firm and reason | Product code |
|---|---|---|
| Z-0991-2025Class I | MEDLINE INDUSTRIES, LP - NorthfieldThe catheter subassembly in the Integrated Arterial Catheter was manufactured with excess material on the catheter hub. | LRO LRP2 products |
| Z-1072-2025Class II | LivaNova USA, Inc.Vagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanically activated component that may become stuck in a closed position, which may result in patients returning to baseline seizu… | LYJ2 products |
| Z-1050-2025Class II | CareFusion 303, Inc.Due to an increase in complaints of door latch assembly failures that may cause the doors to be left in an open state potentially causing inventory discrepancies, a delay in access to medication and/or diversion of medi… | BRY1 product |
| Z-1051-2025Class II | Karl Storz EndoscopyScope IFUs contain a reprocessing modalities that have not been reviewed and/or approved for safety and efficacy by the FDA. | FGB14 products |
| Z-1065-2025Class II | Alphatec Spine, Inc.Complaints have been received regarding post-operative implant collapse. | MAX1 product |
| Z-1066-2025Class II | Draeger Medical Systems, Inc.Edifier speakers (shipped in or after April 2023) may not annunciate when configured in analog mode which can lead to delayed treatment to the patient. | MHX1 product |
| Z-1067-2025Class II | LivaNova USA, Inc.Their is a potential that implanted pulse generators may stop delivering therapy due to an internal component issue. | LYJ2 products |
| Z-1069-2025Class II | Murata Vios, Inc.During the set-up and workflow to begin Vios monitoring, it has been noticed that patient vital signs and ECG waveforms can become distorted and will flicker or flash very fast. This issue only occurs on the Central Sta… | DRT1 product |
| Z-0922-2025Class II | SIEMENS MEDICAL SOLUTIONS USA, INCA potential issue with ARTIS One systems was identified. In very rare situations, the first x-ray release following patient registration may be performed with incorrect copper prefiltration. If the described issue occur… | OWB1 product |
| Z-1024-2025Class II | Boston Scientific CorporationReported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one. | KNS14 products |
| Z-1048-2025Class II | MEDLINE INDUSTRIES, LP - NorthfieldCertain lots Sterile Infusion Catheters which were mislabeled as having a heparin coating were either distributed or manufactured into medical convenience kits and distributed. | DYB LRO2 products |
| Z-1038-2025Class II | Zimmer, Inc.It was identified internally that two commingle events occurred where the metal support post within the polyethylene articular surface was assembled incorrectly for three units within the scope. | JWH2 products |
| Z-1040-2025Class II | MEDLINE INDUSTRIES, LP - NorthfieldCertain Medline kits were manufactured and distributed with Procedure Masks with SO SOFT Earloops, and SOFT TOUCH II Surgical Masks which were subsequently recalled for not meeting the bioburden testing requirements of… | LRO LRP OXQ3 products |
| Z-1022-2025Class II | Medtronic Navigation, Inc.Due to an increase in complaint that the percutaneous reference pin would not fit into the patient reference frame or percutaneous pin adapter. | OLO1 product |
| Z-1023-2025Class II | Accuray IncorporatedSnap rings may become partially or completed detached from the shafts within the internal mechanism. Uncontrolled rotation may occur on the left and right roll axes when the rings are removed or damaged. | IYE1 product |
De Novo grants
New device types the FDA classified through the De Novo pathway. Each one creates a product code and becomes a possible predicate for later 510(k)s.
No De Novo decisions this week.
Original PMA approvals
Original premarket approvals for Class III devices. The FDA also decided 23 PMA supplements this week, which change devices already approved.
No original PMAs were approved this week.
510(k) clearances
47 510(k) clearances by review panel. The median took 121 calendar days from receipt to decision; the fastest, K250014, 25 days; the longest, K231297, 634 days.
Calendar days include the time a submission waits on the company, so they run longer than the FDA’s 90-day review goal. How to search the 510(k) database
- Orthopedic9
- General, Plastic Surgery6
- Cardiovascular5
- Gastroenterology, Urology4
- Physical Medicine4
- Microbiology3
- Radiology3
- Anesthesiology2
- Clinical Chemistry2
- Dental2
- General Hospital2
- Neurology2
- Ear, Nose, Throat1
- Hematology1
- Obstetrics/Gynecology1
Every clearance
| Number | Device and applicant | Product code | Decision and review time |
|---|---|---|---|
| K242717 | Laboratorios Biogalenic Sterile Water for Inhalation, USPLaboratorios Biogalenic S.A. DE C.V. | BTTAnesthesiology | Jan 30, 2025142 days |
| K244027 | SANSA HSATHuxley Medical | MNRAnesthesiology | Jan 28, 202529 days |
| K233956 | Aulisa Infant Oximeter Module (2nd Gen.) (GA-OM0018)Taiwan Aulisa Medical Devices Technologies, Inc. | DQACardiovascular | Jan 30, 2025412 days |
| K241334 | Semi-Floating Temporary Pacing Electrode Catheter, Bipolar, Stainless Steel Electrodes, 115 cm, 4Fr. (006221P); Semi-Floating Temporary Pacing Electrode Catheter, Bipolar, Stainless Steel Electrodes, 115 cm, 5Fr. (006225P); Semi-Floating Temporary Pacing Electrode Catheter, Bipolar, Stainless Steel Electrodes, 125 cm, 6Fr. (006241P); Semi-Floating Temporary Pacing Electrode Catheter, Bipolar, Stainless Steel Electrodes, 125 cm, 7Fr. (006242P); NBIH™ Temporary Pacing Electrode Catheter,C.R. Bard, Inc. | LDFCardiovascular | Jan 31, 2025266 days |
| K242568 | Wrist Blood Pressure Monitor (OHMS11, OHMS12)Shenzhen Zhongkemingwang Telecommunications Software Corp. | DXNCardiovascular | Jan 30, 2025154 days |
| K242757 | Rotarex Atherectomy SystemBard Peripheral Vascular, Inc. | MCWCardiovascular | Jan 30, 2025140 days |
| K242966 | Gentuity® HF-OCT Imaging System with Vis-Rx Prime® Micro-Imaging CatheterGentuity, LLC | DQOCardiovascular | Jan 31, 2025127 days |
| K242135 | Synthgene Home Test HCG Test Strip; Synthgene Home Test HCG Test Cassette; Synthgene Home Test HCG Test MidstreamNanjing Synthgene Medical Technology Co., Ltd. | LCXClinical Chemistry | Jan 31, 2025193 days |
| K243060 | TeleRPM Gen2 Blood Glucose Monitoring SystemGuangdong Transtek Medical Electronics Co., Ltd. | NBWClinical Chemistry | Jan 30, 2025125 days |
| K231297 | Firstar Dental UnitFirstar Dental Company | EIADental | Jan 27, 2025634 days |
| K242217 | Surcam Dental Implant SystemSurcam Medical Devices and Developments, Ltd. | DZEDental | Jan 29, 2025184 days |
| K243150 | Nuance Audio Hearing Aid SoftwareLuxottica Group S.P.A. | SCREar, Nose, Throat | Jan 31, 2025123 days |
| K243025 | Ureteral Access SheathSeplou (Zhuhai) Co., Ltd. | FEDGastroenterology, Urology | Jan 30, 2025125 days |
| K243117 | Digital Video ProcessorHunan Vathin Medical Instrument Co., Ltd. | FGBGastroenterology, Urology | Jan 29, 2025121 days |
| K243439 | Medical Monitor (27HS714S)Lg Electronics.Inc | GCJGastroenterology, Urology | Jan 28, 202583 days |
| K243820 | Hoover Negative Pressure Access Sheath (Hoover) (Hoover Model #'s IVX-NS-1040, IVX-NS-1050, IVX-NS-1140, IVX-NS-1150, IVX-NS-1240, IVX-NS-1250Dornier Medtech America | FEDGastroenterology, Urology | Jan 31, 202550 days |
| K243441 | Nitrile Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate, Simulated Gastric Acid (Light blue, Dark blue)Basic Medical Technology, Inc. | LZAGeneral Hospital | Jan 31, 202586 days |
| K243501 | 3M™ Attest™ Super Rapid Steam Extended Cycle Clear Challenge Pack 1492PCDE; 3M™ Attest™ Auto-reader 490; 3M™ Attest™ Auto-reader 490H; 3M™ Attest™ Mini Auto-reader 490M3M Company | FRCGeneral Hospital | Jan 28, 202577 days |
| K241270 | BTL-754BTL Industries, Inc. | GEXGeneral, Plastic Surgery | Jan 30, 2025269 days |
| K242941 | CO2 Laser Therapy Systems (SHE-LSP003-1)Beijing Sano Laser S&T Development Co.,Ltd | GEXGeneral, Plastic Surgery | Jan 27, 2025124 days |
| K243042 | Co-Ablation System with Sterile Co-Ablation ProbeHygea Medical Technology Co., Ltd. | GEHGeneral, Plastic Surgery | Jan 29, 2025124 days |
| K243213 | Self-Forming Magnet (FLEX SFM)GI Windows, Inc. | SAHGeneral, Plastic Surgery | Jan 29, 2025118 days |
| K243423 | LED Light Therapy Mask (Model(s): RB-008, RB-030, RB-081, RB-008G, RB-008GB, RB-008J, RB-008JB)Shenzhen Rainbow Technology Co., Ltd. | OHSGeneral, Plastic Surgery | Jan 28, 202585 days |
| K250014 | Symani Surgical System (820-03830)Medical Microinstruments, Inc. | SAQGeneral, Plastic Surgery | Jan 27, 202525 days |
| K243374 | HemosIL CL HIT-IgG(PF4-H)Instrumentation Laboratory (IL) Co. | LCOHematology | Jan 28, 202590 days |
| K241915 | CareSuperb™ COVID-19 Antigen Home TestAccess Bio, Inc. | QYTMicrobiology | Jan 29, 2025212 days |
| K243202 | MTS Ceftobiprole 0.002-32 µg/mLLiofilchem s.r.l. | JWYMicrobiology | Jan 30, 2025121 days |
| K243410 | simpli-COLLECT STI TestAbbott Molecular | QYAMicrobiology | Jan 30, 202590 days |
| K243387 | GYDER® Hip SystemGyder Surgical Pty, Ltd. | OLONeurology | Jan 29, 202590 days |
| K243698 | Alignment System Cranial, with Alignment Software Cranial with LITT; Cirq Alignment Software Cranial Biopsy; Cirq Alignment Software Cranial sEEG; VarioGuide Alignment Software Cranial; Cirq Alignment Software Cranial LITTBrainlab AG | HAWNeurology | Jan 28, 202560 days |
| K241348 | SwimCount® Harvester (1 mL, 3 mL)Motilitycount Aps | MQKObstetrics/Gynecology | Jan 31, 2025263 days |
| K241292 | MyShoulder Planner (5.3SSWPL)Medacta International S.A. | KWSOrthopedic | Jan 30, 2025267 days |
| K242442 | Quick-Start Orthopaedic Fixation Screw and Reverse Thread ScrewInnovate Orthopaedics, Ltd. | HWCOrthopedic | Jan 31, 2025168 days |
| K243665 | Crius PEEK Suture Anchor System (Thread-Fix Anchor, Thread-Fix Ultra Anchor, Helicoil Anchor, Pile Anchor)ZheJiang Decans Medical Devices Co., Ltd. | MBIOrthopedic | Jan 31, 202565 days |
| K243673 | Xenco Medical CancelleX Cervical Interbody SystemXenco Medical, LLC | ODPOrthopedic | Jan 29, 202563 days |
| K243712 | WasherCap™ Mini Fixation System (Model 45)Abanza Tecnomed S.L | MBIOrthopedic | Jan 30, 202559 days |
| K243726 | Osprey Suture AnchorResponsive Arthroscopy, Inc. | MBIOrthopedic | Jan 30, 202558 days |
| K243740 | Versalock Upper Limb Plating SystemGM Dos Reis Industria e Comercio Ltda. | HRSOrthopedic | Jan 29, 202556 days |
| K243742 | Arthrex DynaNite Nitinol StaplesArthrex, Inc. | JDROrthopedic | Jan 31, 202558 days |
| K244025 | Argo Knotless GENESYS AnchorConmed Corporation | MBIOrthopedic | Jan 28, 202532 days |
| K241193 | AGILIS - Multi-function Physical Therapy Table (Model: Plus, Pro and Basic.)Marzzan Lucas Jorge | JFBPhysical Medicine | Jan 31, 2025277 days |
| K242982 | Mobility Scooter (MS160A)Suzhou Master Machinery Manufacturing Co.,Ltd | INIPhysical Medicine | Jan 31, 2025127 days |
| K243442 | iBOT® PMDMobius Mobility | IMKPhysical Medicine | Jan 30, 202585 days |
| K243572 | Game Ready system (550500-04); Game Ready Control Unit (550550-04); Game Ready Full Leg Boot Wrap - Medium (590923-04); Game Ready Full Leg Boot Wrap - Large (590922-04); Game Ready C-T Spine Wrap (590502-04); Game Ready Half Leg Boot (590906-04); Game Ready Back Wrap (590500-04); Game Ready Hand/Wrist Wrap (590811-04); Game Ready Flexed Elbow Wrap (590903-04); Game Ready Straight Elbow Wrap (590200-04); Game Ready Shoulder Wrap, Medium, Right (590422-04); Game Ready Shoulder Wrap, LaAvanos Medical, Inc. | IRPPhysical Medicine | Jan 31, 202574 days |
| K241266 | Imaging system of positron emission and X-ray computed tomography (DigitMI 930)Raysolution Healthcare Co., Ltd. | KPSRadiology | Jan 27, 2025266 days |
| K241694 | ZeTTA PACSTae Young Soft Co., Ltd. | LLZRadiology | Jan 31, 2025233 days |
| K243229 | Bunkerhill AVCBunkerHill Health | JAKRadiology | Jan 27, 2025111 days |
Only see the weeks that matter to you
DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and every day it checks openFDA and tells you about new clearances, recalls and MAUDE reports for them. It also puts all of this beside the FDA’s own record pages.
Only your product codes, every Monday
One short email after each openFDA update with just the decisions and recalls in up to three product codes you choose. Free.
Earlier weeks
Each week keeps its own page.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Decisions arrive in weekly updates; check the FDA record before you rely on a figure.
- device/510k · search decision_date:[20250127 TO 20250202]Retrieved Nov 29, 2026, 10:43 PM UTC. FDA data updated Nov 16, 2026.
- device/pma · search decision_date:[20250127 TO 20250202]Retrieved Nov 29, 2026, 10:43 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search event_date_posted:[20250127 TO 20250202]Retrieved Nov 29, 2026, 10:43 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-1072-2025" "Z-1073-2025" "Z-1050-2025" "Z-1051-2025" "Z-1052-2025" "Z-1053-2025" "Z-1054-2025" "Z-1055-2025" "Z-1056-2025" "Z-1057-2025" "Z-1058-2025" "Z-1059-2025" "Z-1060-2025" "Z-1061-2025" "Z-1062-2025" "Z-1063-2025" "Z-1064-2025" "Z-1065-2025" "Z-1066-2025" "Z-1067-2025" "Z-1068-2025" "Z-1069-2025" "Z-0922-2025" "Z-0991-2025" "Z-0992-2025" "Z-1024-2025" "Z-1025-2025" "Z-1026-2025" "Z-1027-2025" "Z-1028-2025" "Z-1029-2025" "Z-1030-2025" "Z-1031-2025" "Z-1032-2025" "Z-1033-2025" "Z-1034-2025" "Z-1035-2025" "Z-1036-2025" "Z-1037-2025" "Z-1048-2025" "Z-1049-2025" "Z-1038-2025" "Z-1039-2025" "Z-1040-2025" "Z-1041-2025" "Z-1042-2025" "Z-1022-2025" "Z-1023-2025")Retrieved Nov 29, 2026, 10:43 PM UTC. FDA data updated Nov 18, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.