FDA product code
LDF: Electrode, Pacemaker, Temporary (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 870.3680
- Review panel
- Cardiovascular
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- Yes
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 1301 in the last 12 months
- Median review time
- 130 daysthe latest 25 decisions
- Recall records
- 151 in the last 12 months
- Adverse event reports
- 147last 12 months, 1 death
510(k) and De Novo decisions
Decisions for product code LDF per year, by decision date, 130 since 1978. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 1 |
| 2018 | 1 |
| 2019 | 1 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 1 |
| 2025 | 5 |
| 2026 (so far) | 1 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K252042 | Sync AV II Temporary Pacing CatheterSwift Sync, Inc. | Jul 30, 2026395 days |
| K242863 | Bioptimal Bipolar Pacing CatheterBioptimal International Pte. , Ltd. | Jun 15, 2025268 days |
| K251186 | Bard® Temporary Pacing Electrode Catheter Needle / Cannula (Introducer)C.R. Bard, Inc. | May 15, 202529 days |
| K241334 | Semi-Floating Temporary Pacing Electrode Catheter, Bipolar, Stainless Steel Electrodes, 115 cm, 4Fr. (006221P); Semi-Floating Temporary Pacing Electrode Catheter, Bipolar, Stainless Steel Electrodes, 115 cm, 5Fr. (006225P); Semi-Floating Temporary Pacing Electrode Catheter, Bipolar, Stainless Steel Electrodes, 125 cm, 6Fr. (006241P); Semi-Floating Temporary Pacing Electrode Catheter, Bipolar, Stainless Steel Electrodes, 125 cm, 7Fr. (006242P); NBIH™ Temporary Pacing Electrode Catheter,C.R. Bard, Inc. | Jan 31, 2025266 days |
| K241199 | Rotatable Connector (5944RL)Shenzhen Launch Electrical Co., Ltd. | Jan 23, 2025268 days |
| K242705 | Streamline Unipolar Pediatric Temporary Pacing Lead (6491) Streamline Unipolar Temporary Atrial Pacing Lead (6492) Streamline Unipolar Temporary Myocardial Pacing Wire (6494) Streamline Bipolar Temporary Myocardial Pacing Lead (6495) Streamline Unipolar Temporary Myocardial Pacing Lead (6500)Medtronic, Inc. | Jan 17, 2025130 days |
| K232261 | TME Temporary Myocardial ElectrodeOsypka AG | Apr 26, 2024270 days |
| K190716 | Streamline 6491 Unipolar Pediatric Temporary Pacing Lead, Streamline 6492 Unipolar Temporary Atrial Pacing Lead, Streamline 6494 Unipolar Temporary Myocardial Pacing Wire, Streamline 6495 Unipolar Temporary Myocardial Pacing LeadMedtronic | Aug 8, 2019142 days |
| K173923 | Temporary Cardiac Pacing WireETHICON, Inc. | Aug 2, 2018219 days |
| K171253 | Streamline 6491 Unipolar Pediatric Temporary Pacing Lead, Streamline 6492 Unipolar Temporary Pacing Lead, Streamline 6495 Bipolar Temporary Myocardial Pacing Lead, Streamline 6500 Unipolar Temporary Myocardial Pacing LeadMedtronic | May 25, 201727 days |
Recalls
Recall records for product code LDF per year, by the date the recall began. Each recalled product is one record; these 15 records belong to 11 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 1 |
| 2019 | 1 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 3 |
| 2025 | 0 |
| 2026 (so far) | 1 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-2383-2026Class I | ARROW INTERNATIONAL, LLCLidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality is… | Apr 21, 2026 |
| Z-2991-2024Class II | MEDLINE INDUSTRIES, LP - NorthfieldMedical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have be… | Apr 8, 2024 |
| Z-2992-2024Class II | MEDLINE INDUSTRIES, LP - NorthfieldMedical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have be… | Apr 8, 2024 |
| Z-2993-2024Class II | MEDLINE INDUSTRIES, LP - NorthfieldMedical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have be… | Apr 8, 2024 |
| Z-2098-2019Class II | Alto Development CorpUpdated instructions for use: Ultra-Flex MYO/WIRE Monopolar pacing wire looped electrode (M-21 Series) to be consistent between electrode IFU and instructions… | Jun 21, 2019 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code LDF. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 79 |
| 2018 | 91 |
| 2019 | 130 |
| 2020 | 83 |
| 2021 | 125 |
| 2022 | 117 |
| 2023 | 212 |
| 2024 | 190 |
| 2025 | 139 |
| 2026 (so far) | 137 |
Last 12 months by event type
- Deaths1
- Injuries23
- Malfunctions121
- Other or not stated2
Most reported device problems, last 12 months
- Pacing Problem35
- Break13
- No Pacing12
- Material Rupture9
- Adverse Event Without Identified Device or Use Problem7
- Inflation Problem7
In plain English: what the newest reports describe Written by AI
In five reports the device did not pace the heart. In most cases a new catheter or lead was used, and one report describes a stopped heart rhythm that needed external pacing and sedation.
5 reports: 2015691-2026-19853, 2015691-2026-18065, 2015691-2026-18066, 2024168-2026-03976, 1018233-2026-05540
The balloon at the tip of the pacing catheter burst or leaked during pre-use inflation, insertion, or passage through a sleeve. In each case a new catheter was used to finish the procedure, and no patient harm was reported.
4 reports: 2024168-2026-03617, 2024168-2026-03923, 2024168-2026-03844, 2024168-2026-03324
Two reports describe a balloon that ruptured inside the patient after a fluid infusion was attached to its inflation port, which is a misuse. The balloon fragment was left in the patient, and tests were under way to locate it.
2 reports: 1018233-2026-06461, 1018233-2026-06441
Two reports describe the electrode coming loose. One says yellow pins detached from the wires more than once during procedures. In a returned catheter the manufacturer found no defect, but its instructions warn that pulling it through the sheath can dislodge the electrode.
2 reports: 1018233-2026-05294, 2015691-2026-17616
Two reports describe fluid around the heart after a pacing lead was removed. One patient needed the fluid drained and intensive care observation, and a study report describes a heart wall perforation and one death in its pacing group without stating the cause.
2 reports: 2024168-2026-03152, 2024168-2026-03199
Written by AI from the narratives of the 18 newest reports that have one, received Aug 7, 2026 to Oct 30, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 502in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 118registered establishments that list this product code
Other codes under 21 CFR 870.3680
Product codes classified under the same regulation.
Watch LDF in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get LDF every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in LDF. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:LDFRetrieved Nov 29, 2026, 4:40 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:LDFRetrieved Nov 29, 2026, 4:40 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:LDF · count decision_dateRetrieved Nov 29, 2026, 4:40 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:LDFRetrieved Nov 29, 2026, 4:40 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:LDF · count event_date_initiatedRetrieved Nov 29, 2026, 4:40 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:LDF · count res_event_numberRetrieved Nov 29, 2026, 4:40 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-2383-2026" "Z-2991-2024" "Z-2992-2024" "Z-2993-2024" "Z-2098-2019")Retrieved Nov 29, 2026, 4:40 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:LDF AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 4:40 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:LDF AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 4:40 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:LDF AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 4:18 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:LDFRetrieved Nov 29, 2026, 4:40 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:LDFRetrieved Nov 29, 2026, 4:40 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:870.3680Retrieved Nov 28, 2026, 10:12 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil