Product codes

FDA product code

MCW: Catheter, Peripheral, Atherectomy (FDA product code)

Follow MCW: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 870.4875
Review panel
Cardiovascular
Premarket route
510(k)
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
1461 in the last 12 months
Median review time
73 daysthe latest 25 decisions
Recall records
340 in the last 12 months
Adverse event reports
861last 12 months, 31 deaths

510(k) and De Novo decisions

Decisions for product code MCW per year, by decision date, 146 since 1990. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20177
20188
20193
20202
20216
20226
20234
20243
20253
2026 (so far)1

Latest decisions

NumberDevice and applicantDecision and review time
K260244 Auryon Atherectomy SystemEximo Medical Apr 10, 202673 days
K250723 FreedomFlow™ Orbital Circumferential Atherectomy System (H6005/ 4Fr 3-Sphere Configuration)Cardio Flow Inc., Apr 25, 202546 days
K250385 Turbo-Elite Laser Atherectomy CatheterSpectranetics Mar 13, 202530 days
K242757 Rotarex Atherectomy SystemBard Peripheral Vascular, Inc. Jan 30, 2025140 days
K242947 FreedomFlow™ Orbital Circumferential Atherectomy SystemCardio Flow Inc., Nov 5, 202441 days
K241553 Auryon Atherectomy Catheter 1.7mm; Auryon Atherectomy Catheter 1.7mm, Hydrophilic CoatingEximo Medical, Ltd. Jun 27, 202427 days
K233483 FreedomFlow Orbital Circumferential Atherectomy System (H6004/5Fr 3-Sphere Configuration)Cardio Flow Inc., Feb 15, 2024111 days
K233668 Auryon Atherectomy Catheter 1.5 mm XL, Hydrophilic Coating (EXM-4010-H000), Auryon Atherectomy Catheter 0.9 mm XL, Hydrophilic Coating (EXM-4011-H000), Auryon Atherectomy Catheter 1.5 mm XL (EXM -4010-0000), Auryon Atherectomy Catheter 0.9 mm XL (EXM-4011-0000)Eximo Medical, Ltd. Dec 15, 202330 days
K231538 FreedomFlow Orbital Circumferential Atherectomy SystemCardio Flow Inc., Sep 27, 2023120 days
K230709 Auryon Atherectomy SystemEximo Medical, Ltd. Jun 9, 202387 days

Recalls

Recall records for product code MCW per year, by the date the recall began. Each recalled product is one record; these 34 records belong to 29 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20176
20181
20190
20200
20214
20222
20231
20242
20251
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-1244-2025Class I Bard Peripheral Vascular IncAtherectomy Catheter eIFU updated to clarify and emphasize procedural steps intended to reduce the likelihood of catheter breakage events. Catheter has an oute… Feb 5, 2025
Z-0680-2025Class II Spectranetics CorporationDue to complaints received for the marker band detaching from the distal tip of the catheter and remaining in the patient as laser catheter was retracted. Nov 25, 2024
Z-0770-2025Class II Angiodynamics, Inc.Potential for procedural delays or interruptions during use of the Auryon Atherectomy System due to the Auryon Atherectomy System not advancing to the Activati… Nov 25, 2024
Z-0509-2024Class II Spectranetics CorporationIncorrect product labeling. Exterior product box label does not match internal pouch label. Nov 15, 2023
Z-0694-2022Class I ev3 Inc.The device has similarities in design to another device that the firm recently issued a field correction on due to instances of tip detachment and embolization. Feb 4, 2022

Adverse event reports (MAUDE)

Medical device reports received per year that name product code MCW. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
2017415
2018537
2019618
2020631
2021734
2022566
2023574
2024677
2025689
2026 (so far)796

Last 12 months by event type

  • Deaths31
  • Injuries292
  • Malfunctions538
  • Other or not stated0

Most reported device problems, last 12 months

  1. Break220
  2. Adverse Event Without Identified Device or Use Problem197
  3. Detachment of Device or Device Component128
  4. Mechanical Jam105
  5. Entrapment of Device84
  6. Material Deformation76

In plain English: what the newest reports describe Written by AI

  • In several reports, the catheter became stuck or jammed, often on the guidewire, and could not be freed. In some cases the catheter and guidewire were pulled out together, and in others the guidewire had to be cut so the two could be separated.

    6 reports: 2124215-2026-57717, 2124215-2026-56636, 2124215-2026-56656, 2124215-2026-56239, 3008439199-2026-00395, 3008439199-2026-00404

  • In multiple reports, the cutting or rotating part of the catheter, such as the helix or spiral inside the device, fractured or detached from the catheter shaft. The rotation stopped, and the procedure was finished with another device.

    4 reports: 3008439199-2026-00398, 3008439199-2026-00395, 3008439199-2026-00405, 3008439199-2026-00404

  • In several reports, the catheter was damaged in the middle of its shaft, causing fluid leakage or an abnormal noise and loss of suction. The device was removed and the procedure was completed with another device.

    3 reports: 3008439199-2026-00399, 3008439199-2026-00400, 3008439199-2026-00402

  • In a few reports, the catheter clogged during suction and could not be cleared by flushing or reverse-flow, so it was replaced.

    Report: 3008439199-2026-00396

  • In one report, a vessel ruptured in a narrow artery during the procedure, causing contrast leakage. It was treated with prolonged balloon inflation and a covered stent, and the patient was reported stable.

    Report: 3008439199-2026-00393

Written by AI from the narratives of the 20 newest reports that have one, received Oct 23, 2026 to Oct 30, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
347in the FDA’s Global Unique Device Identification Database
Establishment listings
230registered establishments that list this product code

Product codes classified under the same regulation.

Watch MCW in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get MCW every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in MCW. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.