Product codes

FDA product code

LYJ: Stimulator, Autonomic Nerve, Implanted For Epilepsy (FDA product code)

Follow LYJ: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class III
Regulation
None listed
Review panel
Neurology
Premarket route
PMA
Implant
Yes
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Ineligible
PMA decisions
1911 original PMA, the rest supplements
Median review time
51 daysthe latest 25 decisions
Recall records
370 in the last 12 months
Adverse event reports
1,381last 12 months, 61 deaths

PMA decisions

Original PMAs and supplements for product code LYJ per year, by decision date, 191 since 1997. Review time is the days from receipt to decision.

All of them in the FDA’s PMA database

PMA decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20179
20183
20197
20206
20211
20223
20235
20243
20256
2026 (so far)3

Latest decisions

NumberDevice and applicantDecision and review time
P970003 S254 VNS Therapy® SystemLivaNova USA, Inc. Oct 2, 2026203 days
P970003 S249 VNS Therapy SystemLivaNova USA, Inc. Aug 12, 2026366 days
P970003 S253 VNS Therapy® SystemLivaNova USA, Inc. Mar 4, 202630 days
P970003 S252 VNS THERAPY SYSTEMLivaNova USA, Inc. Dec 17, 202530 days
P970003 S251 VNS Therapy Model 3100 Programmer, VNS Therapy Model 2000 Programming WandLivaNova USA, Inc. Nov 25, 202584 days
P970003 S250 VNS Therapy® SystemLivaNova USA, Inc. Sep 12, 202529 days
P970003 S248 VNS Therapy SystemLivaNova USA, Inc. Sep 5, 202530 days
P970003 S247 VNS THERAPY SYSTEMLivaNova USA, Inc. May 2, 202551 days
P970003 S246 VNS Therapy® SenTiva® Model 1000 Generator, VNS Therapy® AspireSR Model 106 Generator, VNS Therapy® Demipulse Model 103LivaNova USA, Inc. Mar 27, 202587 days
P970003 S245 VNS Therapy SystemLivaNova USA, Inc. Nov 14, 202430 days

Recalls

Recall records for product code LYJ per year, by the date the recall began. Each recalled product is one record; these 37 records belong to 28 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20173
20185
20192
20203
20211
20220
20230
20242
20252
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-1067-2025Class II LivaNova USA, Inc.Their is a potential that implanted pulse generators may stop delivering therapy due to an internal component issue. Jan 3, 2025
Z-1068-2025Class II LivaNova USA, Inc.Their is a potential that implanted pulse generators may stop delivering therapy due to an internal component issue. Jan 3, 2025
Z-1072-2025Class II LivaNova USA, Inc.Vagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanically activated component that may become stuck in a closed positi… Dec 12, 2024
Z-1073-2025Class II LivaNova USA, Inc.Vagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanically activated component that may become stuck in a closed positi… Dec 12, 2024
Z-0532-2022Class II LivaNova USA, Inc.A manufacturing issue may cause the silicone tubing wall thickness of vagus nerve stimulation leads to be less than the design specification. If the tubing bec… Dec 2, 2021

Adverse event reports (MAUDE)

Medical device reports received per year that name product code LYJ. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
20172,074
20182,317
20192,432
20201,688
20211,764
20221,584
20231,764
20241,524
20251,455
2026 (so far)1,249

Last 12 months by event type

  • Deaths61
  • Injuries590
  • Malfunctions728
  • Other or not stated2

Most reported device problems, last 12 months

  1. Adverse Event Without Identified Device or Use Problem782
  2. Fracture273
  3. High impedance177
  4. Premature Discharge of Battery46
  5. Incomplete or Inadequate Connection37
  6. Low impedance34

In plain English: what the newest reports describe Written by AI

  • Several reports describe high impedance, meaning electrical resistance in the device or lead is too high for the stimulator to work normally. In some cases this was found during checks before a battery replacement, and in one case a planned battery swap was postponed.

    7 reports: 1644487-2026-11258, 1644487-2026-11263, 1644487-2026-11265, 1644487-2026-11254, 1644487-2026-11255, 1644487-2026-11261, 1644487-2026-11276

  • Several reports describe a full revision, meaning surgery to replace the implanted parts. The manufacturer had not yet received the removed devices or leads, so it had not yet examined them.

    4 reports: 1644487-2026-11265, 1644487-2026-11254, 1644487-2026-11274, 1644487-2026-11282

  • Several reports describe changes in seizures while using the device, including three large seizures in one report and admission to intensive care with seizures in another. The reports do not state whether the device caused these changes.

    3 reports: 1644487-2026-11262, 1644487-2026-11269, 1644487-2026-11266

  • Two reports describe the stimulator or lead moving under the skin. One report describes a lead that was seen to be fractured, meaning broken, during a battery replacement, and the lead was not replaced that day.

    2 reports: 1644487-2026-11274, 1644487-2026-11261

  • In one report, a physician said the stimulation at the optimized dose caused ADHD-like behavior. Lowering the dose reduced those behaviors but also reduced seizure control.

    Report: 1644487-2026-11271

Written by AI from the narratives of the 20 newest reports that have one, received Oct 26, 2026 to Oct 30, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
30in the FDA’s Global Unique Device Identification Database
Establishment listings
1registered establishments that list this product code

Watch LYJ in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get LYJ every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in LYJ. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.