FDA product code
LYJ: Stimulator, Autonomic Nerve, Implanted For Epilepsy (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class III
- Regulation
- None listed
- Review panel
- Neurology
- Premarket route
- PMA
- Implant
- Yes
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Ineligible
- PMA decisions
- 1911 original PMA, the rest supplements
- Median review time
- 51 daysthe latest 25 decisions
- Recall records
- 370 in the last 12 months
- Adverse event reports
- 1,381last 12 months, 61 deaths
PMA decisions
Original PMAs and supplements for product code LYJ per year, by decision date, 191 since 1997. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 9 |
| 2018 | 3 |
| 2019 | 7 |
| 2020 | 6 |
| 2021 | 1 |
| 2022 | 3 |
| 2023 | 5 |
| 2024 | 3 |
| 2025 | 6 |
| 2026 (so far) | 3 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| P970003 S254 | VNS Therapy® SystemLivaNova USA, Inc. | Oct 2, 2026203 days |
| P970003 S249 | VNS Therapy SystemLivaNova USA, Inc. | Aug 12, 2026366 days |
| P970003 S253 | VNS Therapy® SystemLivaNova USA, Inc. | Mar 4, 202630 days |
| P970003 S252 | VNS THERAPY SYSTEMLivaNova USA, Inc. | Dec 17, 202530 days |
| P970003 S251 | VNS Therapy Model 3100 Programmer, VNS Therapy Model 2000 Programming WandLivaNova USA, Inc. | Nov 25, 202584 days |
| P970003 S250 | VNS Therapy® SystemLivaNova USA, Inc. | Sep 12, 202529 days |
| P970003 S248 | VNS Therapy SystemLivaNova USA, Inc. | Sep 5, 202530 days |
| P970003 S247 | VNS THERAPY SYSTEMLivaNova USA, Inc. | May 2, 202551 days |
| P970003 S246 | VNS Therapy® SenTiva® Model 1000 Generator, VNS Therapy® AspireSR Model 106 Generator, VNS Therapy® Demipulse Model 103LivaNova USA, Inc. | Mar 27, 202587 days |
| P970003 S245 | VNS Therapy SystemLivaNova USA, Inc. | Nov 14, 202430 days |
Recalls
Recall records for product code LYJ per year, by the date the recall began. Each recalled product is one record; these 37 records belong to 28 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 3 |
| 2018 | 5 |
| 2019 | 2 |
| 2020 | 3 |
| 2021 | 1 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 2 |
| 2025 | 2 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-1067-2025Class II | LivaNova USA, Inc.Their is a potential that implanted pulse generators may stop delivering therapy due to an internal component issue. | Jan 3, 2025 |
| Z-1068-2025Class II | LivaNova USA, Inc.Their is a potential that implanted pulse generators may stop delivering therapy due to an internal component issue. | Jan 3, 2025 |
| Z-1072-2025Class II | LivaNova USA, Inc.Vagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanically activated component that may become stuck in a closed positi… | Dec 12, 2024 |
| Z-1073-2025Class II | LivaNova USA, Inc.Vagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanically activated component that may become stuck in a closed positi… | Dec 12, 2024 |
| Z-0532-2022Class II | LivaNova USA, Inc.A manufacturing issue may cause the silicone tubing wall thickness of vagus nerve stimulation leads to be less than the design specification. If the tubing bec… | Dec 2, 2021 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code LYJ. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 2,074 |
| 2018 | 2,317 |
| 2019 | 2,432 |
| 2020 | 1,688 |
| 2021 | 1,764 |
| 2022 | 1,584 |
| 2023 | 1,764 |
| 2024 | 1,524 |
| 2025 | 1,455 |
| 2026 (so far) | 1,249 |
Last 12 months by event type
- Deaths61
- Injuries590
- Malfunctions728
- Other or not stated2
Most reported device problems, last 12 months
- Adverse Event Without Identified Device or Use Problem782
- Fracture273
- High impedance177
- Premature Discharge of Battery46
- Incomplete or Inadequate Connection37
- Low impedance34
In plain English: what the newest reports describe Written by AI
Several reports describe high impedance, meaning electrical resistance in the device or lead is too high for the stimulator to work normally. In some cases this was found during checks before a battery replacement, and in one case a planned battery swap was postponed.
7 reports: 1644487-2026-11258, 1644487-2026-11263, 1644487-2026-11265, 1644487-2026-11254, 1644487-2026-11255, 1644487-2026-11261, 1644487-2026-11276
Several reports describe a full revision, meaning surgery to replace the implanted parts. The manufacturer had not yet received the removed devices or leads, so it had not yet examined them.
4 reports: 1644487-2026-11265, 1644487-2026-11254, 1644487-2026-11274, 1644487-2026-11282
Several reports describe changes in seizures while using the device, including three large seizures in one report and admission to intensive care with seizures in another. The reports do not state whether the device caused these changes.
3 reports: 1644487-2026-11262, 1644487-2026-11269, 1644487-2026-11266
Two reports describe the stimulator or lead moving under the skin. One report describes a lead that was seen to be fractured, meaning broken, during a battery replacement, and the lead was not replaced that day.
2 reports: 1644487-2026-11274, 1644487-2026-11261
In one report, a physician said the stimulation at the optimized dose caused ADHD-like behavior. Lowering the dose reduced those behaviors but also reduced seizure control.
Report: 1644487-2026-11271
Written by AI from the narratives of the 20 newest reports that have one, received Oct 26, 2026 to Oct 30, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 30in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 1registered establishments that list this product code
Watch LYJ in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get LYJ every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in LYJ. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:LYJRetrieved Nov 29, 2026, 10:53 PM UTC. FDA data updated Nov 16, 2026.
- device/pma · search product_code:LYJRetrieved Nov 29, 2026, 10:53 PM UTC. FDA data updated Nov 16, 2026.
- device/pma · search product_code:LYJ · count decision_dateRetrieved Nov 29, 2026, 10:53 PM UTC. FDA data updated Nov 16, 2026.
- device/pma · search product_code:LYJ · count pma_numberRetrieved Nov 29, 2026, 10:53 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:LYJRetrieved Nov 29, 2026, 10:53 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:LYJ · count event_date_initiatedRetrieved Nov 29, 2026, 10:53 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:LYJ · count res_event_numberRetrieved Nov 29, 2026, 10:53 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-1067-2025" "Z-1068-2025" "Z-1072-2025" "Z-1073-2025" "Z-0532-2022")Retrieved Nov 29, 2026, 10:53 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:LYJ AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 10:53 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:LYJ AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 10:53 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:LYJ AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 10:53 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:LYJRetrieved Nov 29, 2026, 10:53 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:LYJRetrieved Nov 29, 2026, 10:53 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil