FDA product code
PZP: Fluid Jet Removal System (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 876.4350
- Review panel
- Gastroenterology and Urology
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 92 in the last 12 months
- Median review time
- 162 daysthe latest 9 decisions
- Recall records
- 20 in the last 12 months
- Adverse event reports
- 231last 12 months, 10 deaths
510(k) and De Novo decisions
Decisions for product code PZP per year, by decision date, 9 since 2017. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 1 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 2 |
| 2022 | 0 |
| 2023 | 1 |
| 2024 | 2 |
| 2025 | 1 |
| 2026 (so far) | 2 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K260723 | HYDROS Robotic System (HY1000); HYDROS Handpiece (HH1000); HYDROS TRUS Probe (HU1000)Procept Biorobotics | Jul 9, 2026126 days |
| K253211 | AQUABEAM Robotic SystemProcept Biorobotics | Jun 23, 2026267 days |
| K251082 | HYDROS Robotic System (HY1000); HYDROS Handpiece (HH1000); HYDROS TRUS Probe (HU1000)Procept Biorobotics | Oct 10, 2025184 days |
| K241952 | AQUABEAM Robotic System (AB2000)Procept Biorobotics | Sep 30, 202489 days |
| K240200 | HYDROS Robotic System; HYDROS TRUS Probe; HYDROS HandpieceProcept Biorobotics | Aug 20, 2024208 days |
| K231024 | AquaBeam Robotic SystemProcept Biorobotics, Corporation | Aug 30, 2023141 days |
| K212835 | AquaBeam Robotic SystemProcept Biorobotics, Corporation | Oct 6, 202129 days |
| K202961 | AquaBeam Robotic SystemProcept Biorobotics, Corporation | Mar 11, 2021162 days |
| DEN170024 | AQUABEAM SystemProcept Biorobotics, Corporation | Dec 21, 2017248 days |
Recalls
Recall records for product code PZP per year, by the date the recall began. Each recalled product is one record; these 2 records belong to 2 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 1 |
| 2020 | 0 |
| 2021 | 1 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-2330-2021Class II | PROCEPT BIOROBOTICS CORPORATIONScope tube tip may detach from the telescoping tube, which is attached to a single-use handpiece, a component of an aquablation robotic system. Both tip detach… | Jul 9, 2021 |
| Z-1110-2019Class II | PROCEPT BIOROBOTICS CORPORATIONCertain lots may develop an internal leak due to a micro-crack in a specific weld joint within the device. When the Robot detects this failure, it immediately… | Mar 6, 2019 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code PZP. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 0 |
| 2018 | 14 |
| 2019 | 70 |
| 2020 | 85 |
| 2021 | 99 |
| 2022 | 174 |
| 2023 | 254 |
| 2024 | 309 |
| 2025 | 291 |
| 2026 (so far) | 211 |
Last 12 months by event type
- Deaths10
- Injuries144
- Malfunctions74
- Other or not stated3
Most reported device problems, last 12 months
- Adverse Event Without Identified Device or Use Problem147
- Visual Prompts will not Clear39
- Connection Problem12
- Activation Failure4
- Poor Quality Image3
- Computer Software Problem2
In plain English: what the newest reports describe Written by AI
Reports describe bleeding after the procedure. The catheter urine turned dark, red, or grade 3 or 4 (the catheter color scale used to judge bleeding), and patients were taken back to the operating room for cautery (heat sealing of bleeding vessels) and clot removal. The manufacturer did not report a device malfunction.
5 reports: 3012977056-2026-00216, 3012977056-2026-00214, 3012977056-2026-00208, 3012977056-2026-00205, 3012977056-2026-00200
Reports describe patients whose heart or breathing stopped around the procedure. In one case the patient died, and in another the patient died after cardiac arrest and a pulmonary embolism (a blood clot in the lungs). The treating surgeons did not attribute these events to the device, and one report says the cause of death is unknown.
4 reports: 3012977056-2026-00202, 3012977056-2026-00193, 3012977056-2026-00204, 3012977056-2026-00191
Reports describe the bladder being punctured or torn during the procedure, with the waterjet or cautery tool cited in one case. Patients needed surgical repair, one by laparoscopy and another by open surgery through the abdomen.
3 reports: 3012977056-2026-00213, 3012977056-2026-00199, 3012977056-2026-00188
Reports describe the waterjet or resection tool injuring the ureters, the tubes that carry urine from the kidneys to the bladder. One patient needed a precautionary kidney drainage tube, and another needed a ureteral stent (a small tube placed to keep the ureter open). One patient's status is unknown.
2 reports: 3012977056-2026-00203, 3012977056-2026-00187
Reports describe the system's foot pedal, which controls the waterjet, failing to engage or showing error codes during setup or treatment. Troubleshooting, including a handpiece swap and a reboot, did not fix it, so the surgeon stopped the procedure. In one report the surgeon switched to a different surgery, TURP (transurethral resection of the prostate).
2 reports: 3012977056-2026-00211, 3012977056-2026-00206
Written by AI from the narratives of the 16 newest reports that have one, received Sep 4, 2026 to Oct 20, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 15in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 21registered establishments that list this product code
Watch PZP in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get PZP every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in PZP. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:PZPRetrieved Nov 29, 2026, 1:31 AM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:PZPRetrieved Nov 29, 2026, 6:33 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:PZP · count decision_dateRetrieved Nov 29, 2026, 1:31 AM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:PZPRetrieved Nov 29, 2026, 6:33 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:PZP · count event_date_initiatedRetrieved Nov 29, 2026, 1:31 AM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:PZP · count res_event_numberRetrieved Nov 29, 2026, 1:31 AM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-2330-2021" "Z-1110-2019")Retrieved Nov 29, 2026, 6:33 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:PZP AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 6:33 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:PZP AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 6:33 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:PZP AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 1:31 AM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:PZPRetrieved Nov 29, 2026, 1:31 AM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:PZPRetrieved Nov 29, 2026, 1:31 AM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:876.4350Retrieved Nov 29, 2026, 1:31 AM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil