FDA product code
BZR: Mixer, Breathing Gases, Anesthesia Inhalation (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 868.5330
- Review panel
- Anesthesiology
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- Yes
- GMP exempt
- No
- Third-party review
- Eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 550 in the last 12 months
- Median review time
- 161 daysthe latest 25 decisions
- Recall records
- 30 in the last 12 months
- Adverse event reports
- 6last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code BZR per year, by decision date, 55 since 1976. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 1 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 1 |
| 2022 | 0 |
| 2023 | 4 |
| 2024 | 1 |
| 2025 | 1 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K243917 | F&P Optiflow Air/Oxygen Flow SourceFisher & Paykel Healthcare, Ltd. | Jul 14, 2025206 days |
| K241465 | Midas Flowmeter; eAVSParker Hannifin Corporation | Sep 19, 2024119 days |
| K231895 | Maxtec MaxBlend2+pMaxtec, LLC | Dec 4, 2023159 days |
| K230987 | FlowStar Analog Mixer FlowmeterBaldus Sedation GmbH & Co. KG | Sep 14, 2023161 days |
| K222794 | FlowStar Touch Digital Mixer FlowmeterBaldus Sedation GmbH & Co. KG | Jul 28, 2023315 days |
| K221494 | Quality Mix Blender, OxymixerDehas Medical Systems GmbH | Jan 23, 2023245 days |
| K202480 | Nitronox Plus 0-70, Nitronox 0-50, Nitronox Plus 50/50Parker Hannifin | Apr 22, 2021237 days |
| K173205 | Equinox AdvantageO-Two Medical Technologies, Inc. | Aug 29, 2018331 days |
| K161718 | MaxBlend 2, MaxBlend LiteMaxtec, LLC | Oct 27, 2016127 days |
| K142291 | GMX Series Medical Air/Oxygen BlenderGentec (Shanghai) Corporation | May 19, 2015274 days |
Recalls
Recall records for product code BZR per year, by the date the recall began. Each recalled product is one record; these 3 records belong to 3 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 2 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-2887-2018Class II | CAREstream Medical LLCTwo analgesic gas delivery system oxygen/nitrous oxide mixers were reported to have delivered higher levels of nitrous oxide than specification during routine… | May 15, 2018 |
| Z-2885-2018Class II | O-Two Medical Technologies, Inc.Higher levels of nitrous oxide delivered than specification | May 14, 2018 |
| Recall | Maquet IncSoftware update so US customers will be updated equally with worldwide customers. Ver. 5.0 will be upgraded to ver 7.0. Quality will improve and enhance functi… | Jan 11, 2005 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code BZR. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 22 |
| 2018 | 24 |
| 2019 | 26 |
| 2020 | 15 |
| 2021 | 33 |
| 2022 | 15 |
| 2023 | 19 |
| 2024 | 18 |
| 2025 | 13 |
| 2026 (so far) | 6 |
Last 12 months by event type
- Deaths0
- Injuries0
- Malfunctions6
- Other or not stated0
Most reported device problems, last 12 months
- Device Alarm System2
- Gas Output Problem2
- Defective Alarm1
In plain English: what the newest reports describe Written by AI
Several reports say the blender stopped delivering the set oxygen mix, so patients could receive less oxygen than intended and some became desaturated (had low blood oxygen). In some cases, loose internal parts such as a control knob or shaft extender were found.
6 reports: 21432015, 1218704-2026-00001, 1218704-2025-00004, 22001948, 1000117260-2024-00001, MW5185546
Several reports say the blender gave no alarm when oxygen flow failed. One warning from the manufacturer says the audible alarm may not sound when both air and oxygen inlet pressures are below 30 psig.
4 reports: 2020676-2024-00018, 2020676-2024-00019, MW5185546, 1000117260-2024-00001
Four reports describe blood inside the device and heavy corrosion, found by an EU distributor and during third-party servicing. No patient or user injury was reported.
4 reports: 2020676-2025-00020, 2020676-2025-00010, 2020676-2025-00006, 2020676-2024-00022
Four reports describe a resuscitation bag and mask assembled incorrectly, so they did not fit and care was delayed in the delivery room. In one, staff gave blow-by oxygen (holding tubing near the baby's face) while another device was found. No injuries were reported.
4 reports: 2028807-2025-00018, 2028807-2025-00017, 2028807-2025-00014, 2028807-2025-00010
One report describes a broken flowmeter on a helium-oxygen blender that leaked and made a squealing noise while staff tried to set it up for a patient on high flow nasal cannula. The leak was found after troubleshooting.
Report: 24130152
Written by AI from the narratives of the 18 newest reports that have one, received Oct 30, 2024 to Sep 30, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 384in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 61registered establishments that list this product code
Watch BZR in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get BZR every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in BZR. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:BZRRetrieved Nov 29, 2026, 6:38 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:BZRRetrieved Nov 29, 2026, 6:38 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:BZR · count decision_dateRetrieved Nov 29, 2026, 6:38 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:BZRRetrieved Nov 29, 2026, 6:38 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:BZR · count event_date_initiatedRetrieved Nov 29, 2026, 6:38 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:BZR · count res_event_numberRetrieved Nov 29, 2026, 6:38 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-2887-2018" "Z-2885-2018")Retrieved Nov 29, 2026, 6:38 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:BZR AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 6:38 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:BZR AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 6:38 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:BZR AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 6:38 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:BZRRetrieved Nov 29, 2026, 6:38 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:BZRRetrieved Nov 29, 2026, 6:38 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:868.5330Retrieved Nov 29, 2026, 6:38 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil