Product codes

FDA product code

DWE: Tubing, Pump, Cardiopulmonary Bypass (FDA product code)

Follow DWE: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 870.4390
Review panel
Cardiovascular
Premarket route
510(k)
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
210 in the last 12 months
Median review time
85 daysthe latest 21 decisions
Recall records
480 in the last 12 months
Adverse event reports
28last 12 months, 2 deaths

510(k) and De Novo decisions

Decisions for product code DWE per year, by decision date, 21 since 1980. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20170
20180
20190
20200
20210
20220
20230
20241
20250
2026 (so far)0

Latest decisions

NumberDevice and applicantDecision and review time
K233702 SMARxT Tubing and ConnectorsLivaNova USA, Inc. Aug 8, 2024265 days
K150542 Terumo Pump TubingTerumo Cardiovascular Systems Corporation May 1, 201559 days
K142607 AngioVac CircuitAngiodynamics Dec 11, 201487 days
K092486 VORTEX MEDICAL ANGIOVAC CARDIOPULMONARY BYPASS CIRCUITVortex Medical, Inc. Aug 28, 200915 days
K080592 HLM TUBING SET WITH BIOLINE COATINGMaquet Cardiopulmonary, AG Sep 11, 2008192 days
K053025 JOSTRA HLM TUBING SETMaquet Cardiopulmonary, AG Nov 10, 200514 days
K022857 PUMP TUBING WITH X-COATINGTerumo Cardiovascular Systems Corporation Sep 19, 200222 days
K013578 PUMP TUBINGTerumo Cardiovascular Systems Corp. Jan 22, 200285 days
K993189 CARDIOPULMONARY BYPASS PUMP TUBING - 1/4 INCH TUBING, CARDIOPULMONARY BYPASS PUMP TUBING - 3/8 INCH TUBINGTerumo Corp. Mar 1, 2000160 days
K993062 CHEST DRAINAGE TUBINGOlson Medical Sales, Inc. Dec 10, 199988 days

Recalls

Recall records for product code DWE per year, by the date the recall began. Each recalled product is one record; these 48 records belong to 23 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
201728
20180
20191
20200
20210
20220
20230
20240
20250
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-1389-2020Class III LivaNova USA Inc.Perfusion tubing pack was packaged with the incorrect cuvette. Jul 3, 2019
Z-0664-2018Class III LivaNova USAThe product is labeled with the incorrect expiration date. Dec 19, 2017
Z-1709-2018Class II Datascope CorporationThe Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer. Seal separation from the housing may allow air to… Oct 19, 2017
Z-1717-2018Class II Datascope CorporationThe Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer. Seal separation from the housing may allow air to… Oct 19, 2017
Z-1718-2018Class II Datascope CorporationThe Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer. Seal separation from the housing may allow air to… Oct 19, 2017

Adverse event reports (MAUDE)

Medical device reports received per year that name product code DWE. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
2017131
2018134
201982
202045
202141
2022123
2023286
2024221
202582
2026 (so far)24

Last 12 months by event type

  • Deaths2
  • Injuries2
  • Malfunctions24
  • Other or not stated0

Most reported device problems, last 12 months

  1. Delivered as Unsterile Product9
  2. Component Misassembled5
  3. Fluid/Blood Leak3
  4. Decrease in Pressure2
  5. Detachment of Device or Device Component2
  6. Gas/Air Leak2

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
3,187in the FDA’s Global Unique Device Identification Database
Establishment listings
32registered establishments that list this product code

Watch DWE in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get DWE every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in DWE. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.