FDA product code
ELC: Scaler, Ultrasonic (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 872.4850
- Review panel
- Dental
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 1952 in the last 12 months
- Median review time
- 200 daysthe latest 25 decisions
- Recall records
- 100 in the last 12 months
- Adverse event reports
- 240last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code ELC per year, by decision date, 195 since 1976. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 2 |
| 2018 | 1 |
| 2019 | 4 |
| 2020 | 1 |
| 2021 | 3 |
| 2022 | 2 |
| 2023 | 5 |
| 2024 | 4 |
| 2025 | 2 |
| 2026 (so far) | 2 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K260773 | Varios Combi Pro2Nakanishi, Inc. | Mar 10, 20261 day |
| K253254 | GBT Machine Airflow Prophylaxis MasterE.M.S Electro Medical Systems S.A | Jan 28, 2026121 days |
| K251689 | VARIOS 170 LUX SCALER SYSTEM (VA170LUXS10, VA170LUXS11); VARIOS 170 SCALER SYSTEM (VA170S11)Nakanishi, Inc. | Sep 19, 2025109 days |
| K242317 | Integrated Endo System (Meet Endo-II)Denjoy Dental Co., Ltd. | Feb 13, 2025192 days |
| K233844 | OdneCleanOdne AG | Aug 8, 2024248 days |
| K240707 | Ultrasonic Scaler (WD-JY-B002,WD-JY-W001,WD-JY-R003,WD-JY-G004,WD-JY-B003,WD-JY-C006,WD-JY-M0014, WD-JY-N0015,WD-JY-N0016,WD-JY-N0017,WD-JY-E007, WD-JY-F008,WD-JY-A005,WD-JY-H009,WD-JY-I0011, WD-JY-K0012,WD-JY-K0013,WD-JY-K0014, WD-JY-L0013,WD-JY-L0014,WD-JY-L0015, WD-JY-L0016,WD-JY-L0017)Shenzhen Micro Electric Intelligence Co., Ltd. | Jul 2, 2024109 days |
| K240801 | BioSonic US200 Ultrasonic Scaler (60034537)Dent4you AG | Jul 1, 202498 days |
| K233922 | SP NEWTRON CAN-A; MODULE NEWTRON CAN-A EMB; SUPPORT XINETIC; NWT CAN SLIM LED CORD G3 SHORT; NWT CAN SLIM LED CORD G3 LONGSatelec - Acteon Group | Mar 11, 202489 days |
| K223340 | EDS Calculus Softener TURBO CREssential Dental Systems, Inc. | Dec 11, 2023405 days |
| K230641 | Ultrasonic Scaler TipsGuilin Refine Medical Instrument Co., Ltd. | Nov 21, 2023258 days |
Recalls
Recall records for product code ELC per year, by the date the recall began. Each recalled product is one record; these 10 records belong to 5 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 1 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 2 |
| 2023 | 0 |
| 2024 | 5 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-2177-2024Class II | Dental EZ Group Star Dental DivisionTip holder portion of the scaler tip assembly may have an oversized condition that will not allow the scaler tip to stay seated in the holder may cause injury | May 30, 2024 |
| Z-2178-2024Class II | Dental EZ Group Star Dental DivisionTip holder portion of the scaler tip assembly may have an oversized condition that will not allow the scaler tip to stay seated in the holder may cause injury | May 30, 2024 |
| Z-1686-2024Class II | Electro Medical Systems SAReceived complaints that bottles assembled with one batch of bottle collars can be disassembled in two parts when used under pressure (during treatment or stor… | Mar 13, 2024 |
| Z-1687-2024Class II | Electro Medical Systems SAReceived complaints that bottles assembled with one batch of bottle collars can be disassembled in two parts when used under pressure (during treatment or stor… | Mar 13, 2024 |
| Z-1688-2024Class II | Electro Medical Systems SAReceived complaints that bottles assembled with one batch of bottle collars can be disassembled in two parts when used under pressure (during treatment or stor… | Mar 13, 2024 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code ELC. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 356 |
| 2018 | 289 |
| 2019 | 302 |
| 2020 | 164 |
| 2021 | 166 |
| 2022 | 165 |
| 2023 | 210 |
| 2024 | 219 |
| 2025 | 289 |
| 2026 (so far) | 211 |
Last 12 months by event type
- Deaths0
- Injuries23
- Malfunctions217
- Other or not stated0
Most reported device problems, last 12 months
- No Flow89
- Overheating of Device89
- Restricted Flow rate45
- Break39
- Power Problem10
- Vibration10
In plain English: what the newest reports describe Written by AI
Scalers and their inserts or tips overheated in six reports, and in three of them no water came out at the same time. In one report the manufacturer's fix of changing the o-rings did not stop the heat. None of these reports describes an injury.
6 reports: 2424472-2026-05886, 2424472-2026-05871, 2424472-2026-05842, 2424472-2026-05872, 2424472-2026-05835, 2424472-2026-05809
Patients complained that the ultrasonic scaler, a dental tool that vibrates to remove tartar, felt too strong and painful, according to three reports summarized by the manufacturer.
3 reports: 3010364969-2026-00005, 3010364969-2026-00006, 3010364969-2026-00008
Water stopped reaching the handpiece or insert in three reports. In one, the water stopped after a single spit. In another, a patient felt a painful "zing" and the office stopped using the unit.
3 reports: 2424472-2026-05806, 1416605-2026-00016, 2424472-2026-05836
Two reports describe the same dental visit, where a patient felt pain while a dental handpiece was used to clean a tooth. The dentist found subcutaneous emphysema, which is air trapped under the skin, and prescribed antibiotics. The reports do not say which of two handpieces was involved.
2 reports: 9611253-2026-00049, 9611253-2026-00050
In one report, a scaler got too hot and caused a minor burn to a doctor's finger. The manufacturer's summary is the only description of this event.
Report: 3010364969-2026-00010
Written by AI from the narratives of the 17 newest reports that have one, received Jul 15, 2026 to Oct 27, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 3,779in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 268registered establishments that list this product code
Watch ELC in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get ELC every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in ELC. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:ELCRetrieved Nov 29, 2026, 6:00 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:ELCRetrieved Nov 29, 2026, 6:01 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:ELC · count decision_dateRetrieved Nov 29, 2026, 6:01 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:ELCRetrieved Nov 29, 2026, 6:01 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:ELC · count event_date_initiatedRetrieved Nov 29, 2026, 6:00 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:ELC · count res_event_numberRetrieved Nov 29, 2026, 6:00 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-2177-2024" "Z-2178-2024" "Z-1686-2024" "Z-1687-2024" "Z-1688-2024")Retrieved Nov 29, 2026, 6:01 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:ELC AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 6:00 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:ELC AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 6:01 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:ELC AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 6:01 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:ELCRetrieved Nov 29, 2026, 6:00 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:ELCRetrieved Nov 29, 2026, 6:00 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:872.4850Retrieved Nov 29, 2026, 6:01 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil