FDA product code
DPS: Electrocardiograph (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 870.2340
- Review panel
- Cardiovascular
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 5796 in the last 12 months
- Median review time
- 238 daysthe latest 25 decisions
- Recall records
- 984 in the last 12 months
- Adverse event reports
- 122last 12 months, 2 deaths
510(k) and De Novo decisions
Decisions for product code DPS per year, by decision date, 579 since 1976. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 9 |
| 2018 | 14 |
| 2019 | 3 |
| 2020 | 6 |
| 2021 | 5 |
| 2022 | 6 |
| 2023 | 10 |
| 2024 | 8 |
| 2025 | 6 |
| 2026 (so far) | 5 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K260640 | Kinetic Software LibraryMonebo Technologies, Inc. | Oct 23, 2026238 days |
| K260005 | MyoVista wavECG (DG-200-B15-003)HeartSciences, Inc. | Oct 5, 2026276 days |
| K260250 | QCA 10 Electrocardiograph (QCARD-10); QCA 10W Electrocardiograph (QCARD-10W)QT Medical, Inc. | Sep 29, 2026245 days |
| K252463 | SE-1202 Series Electrocardiograph (Model: SE-1202, SE-1202E)Edan Instruments, Inc. | Jul 2, 2026330 days |
| K251670 | MAC 7 Resting ECG Analysis SystemGe Medical Systems Information Technologies, Inc. | Jan 26, 2026241 days |
| K252361 | AccurECG Analysis System (v2.0)Accurkardia, Inc. | Dec 22, 2025146 days |
| K252474 | Withings BeamO (SCT02)Withings | Nov 10, 202596 days |
| K250569 | Cardiologs Holter PlatformPhilips France Commercial | Aug 6, 2025161 days |
| K243252 | ZBPro DiagnosticZbeats, Inc. | Jul 10, 2025268 days |
| K243305 | Masimo W1Masimo Corporation | Apr 3, 2025164 days |
Recalls
Recall records for product code DPS per year, by the date the recall began. Each recalled product is one record; these 98 records belong to 34 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 9 |
| 2020 | 13 |
| 2021 | 5 |
| 2022 | 2 |
| 2023 | 3 |
| 2024 | 6 |
| 2025 | 4 |
| 2026 (so far) | 4 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-3012-2026Class II | Midmark CorporationA communication anomoly may result in incorrect electrocardiogram waveforms of a patient and an inaccurate beats-per-minute calculation. | Jul 20, 2026 |
| Z-2955-2026Class II | Philips North America LlcDevice will intermittently provide inaccurate data related to native heart rate. | Jul 1, 2026 |
| Z-2956-2026Class II | Philips North America LlcDevice will intermittently provide inaccurate data related to native heart rate. | Jul 1, 2026 |
| Z-2876-2026Class II | GE Medical Systems Information Technologies IncSoftware issue issue with the MAC VU360 v3.0 system. After selecting the order refresh icon, an open preview or review screen might fail to close automatically… | Jun 29, 2026 |
| Z-0165-2026Class II | Philips North AmericaFailure of Environmental Stress Testing from a pinched power module wire within the housing of the Cardiac Workstation device which resulted in a short. | Sep 12, 2025 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code DPS. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 15 |
| 2018 | 19 |
| 2019 | 27 |
| 2020 | 59 |
| 2021 | 24 |
| 2022 | 59 |
| 2023 | 54 |
| 2024 | 64 |
| 2025 | 57 |
| 2026 (so far) | 121 |
Last 12 months by event type
- Deaths2
- Injuries7
- Malfunctions112
- Other or not stated1
Most reported device problems, last 12 months
- Incorrect Measurement43
- High Readings35
- Incorrect, Inadequate or Imprecise Result or Readings17
- Patient Data Problem17
- Imprecise or Erratic Results7
- Human-Device Interface Problem2
In plain English: what the newest reports describe Written by AI
Several reports describe the ECG waveform on the display flickering, showing incorrectly, or not staying stable. In one report, the flickering appeared during function checks before use. The manufacturer planned to replace the systems in that case.
4 reports: MW6004352, 1218950-2026-100716, 1218950-2026-100687, 1218950-2026-100505
Three reports describe patient information being wrong on the device. One showed the wrong age for a patient after use, one recorded an exam under the wrong patient, and one reported duplicate and mismatched records of unknown cause. None reported patient injury.
3 reports: 3006087789-2026-00027, 3006087789-2026-00025, 3006087789-2026-00022
Two reports describe the heart rate reading being too high or doubled. The manufacturer found the errors happened when the ECG software add-on (plug-in) was opened more than once at the same time, or when several acquisition windows and workflows ran together. Customers were sent a firmware update, and one customer was also given a plug-in update. The physicians in one report gave no medication and no harm occurred.
2 reports: 1523530-2026-00012, 1523530-2026-00013
Two reports describe heart rate or readings that are inaccurate or change erratically without explanation. The reports do not state a cause. No harm was reported.
2 reports: 9610816-2026-101917, 9610816-2026-102178
One report describes a patient death under review. The monitor flagged a 10-second pause that was processed automatically rather than checked by a technician. The manufacturer's metrics found six pauses, the longest 2.25 seconds, and 26% noise. The report does not say whether the device caused the death.
Report: 1000341190-2026-00002
Written by AI from the narratives of the 18 newest reports that have one, received Jul 20, 2026 to Oct 28, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 3,379in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 496registered establishments that list this product code
Other codes under 21 CFR 870.2340
Product codes classified under the same regulation.
Watch DPS in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get DPS every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in DPS. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:DPSRetrieved Nov 29, 2026, 1:01 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:DPSRetrieved Nov 29, 2026, 1:01 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:DPS · count decision_dateRetrieved Nov 29, 2026, 1:01 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:DPSRetrieved Nov 29, 2026, 1:01 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:DPS · count event_date_initiatedRetrieved Nov 29, 2026, 1:01 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:DPS · count res_event_numberRetrieved Nov 29, 2026, 1:01 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-3012-2026" "Z-2955-2026" "Z-2956-2026" "Z-2876-2026" "Z-0165-2026")Retrieved Nov 29, 2026, 1:01 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:DPS AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 1:01 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:DPS AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 1:01 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:DPS AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 1:01 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:DPSRetrieved Nov 29, 2026, 1:01 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:DPSRetrieved Nov 29, 2026, 1:01 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:870.2340Retrieved Nov 28, 2026, 11:06 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil