FDA product code
HNY: Cystotome (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class I
- Regulation
- 21 CFR 886.4350
- Review panel
- Ophthalmic
- Premarket route
- 510(k) exempt
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 200 in the last 12 months
- Median review time
- 38 daysthe latest 20 decisions
- Recall records
- 00 in the last 12 months
- Adverse event reports
- 7last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code HNY per year, by decision date, 20 since 1979. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K961524 | STERILAB CANNULAS, STERILAB CYSTOTMES, STERILAB NEEDLES AND STERILAB LENS MANIPULATORSSterilab, Inc. | May 16, 199624 days |
| K880345 | CYSTOTOMES/KNIFESMyocure, Inc. | May 3, 198898 days |
| K874975 | PRISMA DISPOSABLE SHARP ANGLED IRRIGAT. CYSTOTOMEAdvanced Surgical Products, Inc. | Apr 13, 1988133 days |
| K874993 | PRISMA DISPOSABLE SHARP IRRIGATING CYSTOTOMEAdvanced Surgical Products, Inc. | Apr 13, 1988133 days |
| K875001 | PRISMA DISPOSABLE BLUNT IRRIGATING CYSTOTOMEAdvanced Surgical Products, Inc. | Apr 13, 1988131 days |
| K874976 | PRISMA DISPOS. DOUBLE SHARP IRRIGATING CYSTOTOMEAdvanced Surgical Products, Inc. | Apr 13, 1988133 days |
| K874997 | PRISMA DISPOSABLE STRAIGHT IRRIGATING CYSTOTOMEAdvanced Surgical Products, Inc. | Apr 13, 1988133 days |
| K874983 | PRISMA DISPOSABLE ANGLED IRRIGATING CYSTOTOMEAdvanced Surgical Products, Inc. | Apr 13, 1988133 days |
| K873163 | VAN DELLEN IRRIGATING CYSTOTOMEVan Dellen, Ltd. | Sep 4, 198724 days |
| K871365 | CAPSULECTOMER CYSTOTOMENoetix, Inc. | Apr 16, 19879 days |
Recalls
Recall records for product code HNY per year, by the date the recall began. Each recalled product is one record.
No recalls on record for this product code.
Adverse event reports (MAUDE)
Medical device reports received per year that name product code HNY. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 3 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 4 |
| 2023 | 4 |
| 2024 | 1 |
| 2025 | 8 |
| 2026 (so far) | 7 |
Last 12 months by event type
- Deaths0
- Injuries1
- Malfunctions6
- Other or not stated0
Most reported device problems, last 12 months
- Material Twisted/Bent6
- Product Quality Problem6
- Unintended Movement1
In plain English: what the newest reports describe Written by AI
Five reports describe cannulas or cystotome tips that came loose, broke, bent or had burrs or defects. In these cases the item was removed or replaced during surgery, and one report says a cannula fell into the eye and capsule tissue was damaged. Patient outcomes were unknown or not reported.
5 reports: 2523835-2023-00397, 2523835-2025-00248, 2523835-2023-00647, 2523835-2023-00594, 2523835-2023-00168
Two reports say irrigating needles, both in a cystotome kit and sold separately, were allegedly bent, making it hard to engage the lens capsule during cataract surgery. The reports say this occasionally caused capsule tears and longer surgery, but patient outcomes were not confirmed.
2 reports: 1211998-2026-00022, 1211998-2026-00023
Two reports describe toxic anterior segment syndrome (TASS), a sterile inflammation inside the eye after surgery, in patients using an ophthalmic cannula, a thin surgical tube. One report describes steroid treatment, raised eye pressure and a later lens repositioning surgery, while the other says the condition was continuing and needed treatment.
2 reports: 2523835-2025-00744, 2523835-2025-00212
Two reports say an ophthalmic cannula became blocked or clogged before or during surgery. In both, the surgeon replaced it with another and completed the operation, with no patient impact reported.
2 reports: 2523835-2025-00019, 2523835-2024-01909
One report says the tip of a needle on the device swivelled inside the eye during surgery, even though it was securely attached to the syringe. The sharp tip punctured the capsule outside the intended area, causing a tear in the front chamber of the eye, which prolonged surgery and needed extra steps to manage. The report says the patient was affected but does not confirm permanent harm.
Report: 1211998-2026-00016
Written by AI from the narratives of the 15 newest reports that have one, received Apr 17, 2023 to May 14, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 360in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 85registered establishments that list this product code
Other codes under 21 CFR 886.4350
Product codes classified under the same regulation.
Watch HNY in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get HNY every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in HNY. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:HNYRetrieved Nov 29, 2026, 3:18 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:HNYRetrieved Nov 29, 2026, 3:18 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:HNY · count decision_dateRetrieved Nov 29, 2026, 3:18 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:HNYRetrieved Nov 29, 2026, 3:18 PM UTC.
- device/recall · search product_code:HNY · count event_date_initiatedRetrieved Nov 29, 2026, 3:18 PM UTC.
- device/recall · search product_code:HNY · count res_event_numberRetrieved Nov 29, 2026, 3:18 PM UTC.
- device/event · search device.device_report_product_code:HNY AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 3:18 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:HNY AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 3:18 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:HNY AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 3:03 AM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:HNYRetrieved Nov 29, 2026, 3:18 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:HNYRetrieved Nov 29, 2026, 3:18 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:886.4350Retrieved Nov 29, 2026, 3:03 AM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil