FDA product code
HNR: Forceps, Ophthalmic (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class I
- Regulation
- 21 CFR 886.4350
- Review panel
- Ophthalmic
- Premarket route
- 510(k) exempt
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 440 in the last 12 months
- Median review time
- 35 daysthe latest 25 decisions
- Recall records
- 1613 in the last 12 months
- Adverse event reports
- 114last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code HNR per year, by decision date, 44 since 1976. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K022835 | STEPHENS DISPOSABLE FORCEPSStephens Instruments | Nov 1, 200266 days |
| K895434 | ALCON FOLDING FORCEPSAlcon Laboratories | Dec 4, 198987 days |
| K891487 | SOFT LENS FOLDING BLOCKWestern Medical Products, Inc. | Aug 15, 1989151 days |
| K893199 | PHACO-INJECTOR(TM) INTRAOCULAR LENS IMPLANT INSTRAllergan Medical Optics | Jun 16, 198951 days |
| K885230 | MCDONALD FOLDING FORCEPSAlcon Laboratories | May 17, 1989148 days |
| K881049 | BRADBURY INTRAOCULAR RETINAL KIT CAT #200Microline Pentax, Inc. | Sep 12, 1988185 days |
| K883197 | IOL IMPLANTATION FORCEPSD & D Surgical | Aug 22, 198825 days |
| K880274 | ADVANCED SURGICAL PRODUCTS SCLERAL PLUG FORCEPSAdvanced Surgical Products, Inc. | Feb 5, 198815 days |
| K864266 | OPHTHALMIC KNIFE & FORCEPSWestern Medical Products, Inc. | Nov 28, 198630 days |
| K864183 | OPHTHALMIC SURGICAL INSTRUMENTSTechno Ophthalmics Intl. | Nov 28, 198632 days |
Recalls
Recall records for product code HNR per year, by the date the recall began. Each recalled product is one record; these 16 records belong to 5 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 1 |
| 2024 | 0 |
| 2025 | 1 |
| 2026 (so far) | 12 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-2069-2026Class II | Katalyst Surgical, LLCField Safety Corrective Action for IFU in DEX Forceps and Scissors. | Mar 25, 2026 |
| Z-2070-2026Class II | Katalyst Surgical, LLCField Safety Corrective Action for IFU in DEX Forceps and Scissors. | Mar 25, 2026 |
| Z-2071-2026Class II | Katalyst Surgical, LLCField Safety Corrective Action for IFU in DEX Forceps and Scissors. | Mar 25, 2026 |
| Z-2072-2026Class II | Katalyst Surgical, LLCField Safety Corrective Action for IFU in DEX Forceps and Scissors. | Mar 25, 2026 |
| Z-2073-2026Class II | Katalyst Surgical, LLCField Safety Corrective Action for IFU in DEX Forceps and Scissors. | Mar 25, 2026 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code HNR. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 18 |
| 2018 | 39 |
| 2019 | 82 |
| 2020 | 135 |
| 2021 | 63 |
| 2022 | 58 |
| 2023 | 98 |
| 2024 | 81 |
| 2025 | 115 |
| 2026 (so far) | 111 |
Last 12 months by event type
- Deaths0
- Injuries16
- Malfunctions97
- Other or not stated1
Most reported device problems, last 12 months
- Break32
- Activation Problem30
- Sharp Edges9
- Structural Problem9
- Difficult to Open or Close6
- Adverse Event Without Identified Device or Use Problem5
In plain English: what the newest reports describe Written by AI
Reports describe sharp or wide capsulorhexis forceps tearing or splitting the corneal incision during cataract surgery. In these cases surgeons placed corneal sutures to close the wound, and one report says prolonged surgery and possible reduced wound sealing or astigmatism could follow.
5 reports: 3002808441-2026-00007, 3002808441-2026-00006, 1211998-2026-00029, 1211998-2026-00031, 3002808441-2026-00005
In reports of vitrectomy and lens exchange surgery, the jaws of two or three forceps from the same lot stuck together in the closed position, even when the surgeon was not squeezing the handle. Surgeons switched to new forceps to finish the procedure, and no patient harm was reported.
3 reports: 3003398873-2026-00199, 3003398873-2026-00196, MW6001808
In several reports, the tip of an ophthalmic forceps broke or the instrument came apart inside the eye. Broken pieces were retrieved from the eye, and no patient harm was reported.
3 reports: 3003398873-2026-00198, 3007589150-2026-00029, 3003398873-2026-00186
One report describes a forceps that failed to spring back during a retinal procedure, so it could not grasp and release tissue. The surgeon stopped using it and replaced it with a different instrument, and the patient's impact was not reported.
Report: 3003398873-2026-00201
One report says four forceps came back rusted after autoclave sterilization. The report gives no further detail.
Report: 1417592-2026-47379
Written by AI from the narratives of the 15 newest reports that have one, received Sep 2, 2026 to Oct 14, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 9,424in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 287registered establishments that list this product code
Other codes under 21 CFR 886.4350
Product codes classified under the same regulation.
Watch HNR in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get HNR every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in HNR. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:HNRRetrieved Nov 29, 2026, 5:18 AM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:HNRRetrieved Nov 29, 2026, 5:18 AM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:HNR · count decision_dateRetrieved Nov 28, 2026, 5:12 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:HNRRetrieved Nov 29, 2026, 5:18 AM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:HNR · count event_date_initiatedRetrieved Nov 29, 2026, 5:18 AM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:HNR · count res_event_numberRetrieved Nov 28, 2026, 5:12 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-2069-2026" "Z-2070-2026" "Z-2071-2026" "Z-2072-2026" "Z-2073-2026")Retrieved Nov 29, 2026, 5:18 AM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:HNR AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 5:18 AM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:HNR AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 5:18 AM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:HNR AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 5:18 AM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:HNRRetrieved Nov 28, 2026, 5:12 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:HNRRetrieved Nov 28, 2026, 5:12 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:886.4350Retrieved Nov 29, 2026, 3:03 AM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.