Blog

Competitors

Medical device competitive analysis with FDA data, step by step

How to size up competitors in a product code with FDA data (510(k)s, MAUDE, recalls), with a worked infusion pump example and the name-merging step.

Title card reading "Medical device competitive analysis with FDA data"

Someone asks who you're really up against in your product category, and the answer has to hold up in a strategy review, a design review or a risk file. Market reports cost thousands of dollars and rarely show their sources. The FDA's own databases are free and public, and they answer three questions directly: who is getting new devices cleared, whose devices are being reported, and who is recalling what.

The short answer: define the market as one or more FDA product codes. For each code, count 510(k) decisions by applicant over five years, MAUDE reports by manufacturer over 24 months and recalls by firm over five years. Merge the many spellings of each company's name before you rank anything. Then read what the reports and recalls say, not only how many there are.

This post walks through that method on product code FRN (infusion pumps) with real openFDA numbers, and lists the limits to carry into anything you write from it.

Step 1: Define the market as product codes

A product code is the FDA's three-letter label for a device type. Every 510(k), PMA, MAUDE report and recall carries one, so it is the one key that joins the databases. Our post on FDA product code lookup shows how to find yours from a cleared device.

FRN is "Pump, Infusion": Class II, regulated under 21 CFR 880.5725 and reviewed by the General Hospital panel. Its product code page has the classification and yearly counts.

Most markets span more than one code. Insulin pumps, for example, have their own code (LZG), and an analysis of "infusion" that leaves it out answers a different question. Write the codes you chose at the top of the analysis.

Step 2: Who is clearing new devices

Count 510(k) decisions by applicant over the last five years. In openFDA's 510(k) data (updated November 9, 2026), FRN had 30 decisions from November 1, 2021 to October 31, 2026, out of 849 since the code's first clearance. That ratio alone says something: this is a mature category where most of the history is old.

Ranked by the applicant names exactly as the FDA records them, Baxter Healthcare Corporation leads with 9, then ICU Medical with 3. But Koru Medical Systems appears under four spellings: "Repro-Medical System, Inc., Dba Koru Medical Systems" (2), "Koru Medical Systems" (1), "Koru Medical Systems, Inc." (1) and "Repro-Med Systems, Inc. Dba Koru Medical Systems" (1). Merged, Koru has 5 and is second. Deka appears under three spellings, with one decision each.

The six most recent FRN decisions show who is active now and how long their reviews took (calendar days from FDA receipt to decision):

510(k)DeviceApplicantDecisionDays
K253703RemunityPRO+ Pump for Remodulin (treprostinil) InjectionDeka R&D CorpJul 24, 2026242
K252015FreedomEdge Infusion SystemKoru Medical Systems, Inc.Jan 28, 2026215
K251636Spectrum IQ Infusion System with Dose IQ Safety SoftwareBaxter Healthcare CorporationJul 28, 202560
K251640SIGMA Spectrum Infusion Pump with Master Drug LibraryBaxter Healthcare CorporationJul 28, 202560
K242752MRidium 3870 MRI Infusion Pump SystemIradimed CorporationMay 28, 2025258
K243855BD Alaris Infusion System with Guardrails Suite MXCarefusion 303, Inc.Apr 25, 2025130

Two Baxter submissions were received the same day and decided in 60 days, which suggests a coordinated update to one pump family. A 215 to 258 day review usually means at least one hold for more information. Open each 510(k) summary to see what changed and which predicate was cited; our guide to finding a predicate device covers how.

Step 3: Whose devices are being reported

MAUDE holds medical device reports from manufacturers, importers, user facilities and voluntary reporters. For FRN, openFDA lists 107,023 reports received from November 1, 2024 to October 31, 2026 (data as of November 11, 2026). Of those, 105,124 (98.2%) are malfunctions, 1,422 injuries, 295 other and 182 deaths.

Count them by the manufacturer name on the report and the top of the list looks like this: CAREFUSION SD 46,993, CAREFUSION 303, INC. 15,852, BAXTER HEALTHCARE CORPORATION 8,788, ICU MEDICAL, INC. 7,570. Now merge.

  • Spellings. ICU Medical appears as "ICU MEDICAL, INC.", "ICU Medical, Inc.", "ICU MEDICAL, INC", "ICU MEDICAL" and "ICU Medical, Inc". Smiths Medical ASD appears three ways, one of them with a broken apostrophe ("SMITH¿S MEDICAL ASD, INC."). Fresenius Kabi USA appears three ways.
  • Sites and subsidiaries. "ICU Medical Costa Rica Ltd." and "Fresenius Vial S.A.S." file under their own names, separate from the parent's.
  • Owners. The data never says who owns whom. You add that by hand: BD completed its acquisition of CareFusion on March 17, 2015, and ICU Medical completed its acquisition of Smiths Medical on January 6, 2022.

After merging the 30 most common names, the ranking changes:

Company (names merged)MAUDE reports, 24 monthsShare
BD (CareFusion SD, CareFusion 303)62,84558.7%
ICU Medical (incl. Costa Rica site and Smiths Medical)17,02815.9%
Baxter9,0458.5%
Fresenius Kabi (incl. Fresenius Vial)7,2966.8%
Medtronic (Puerto Rico Operations, MiniMed)5,5945.2%

ICU Medical goes from fourth to second once its ten spellings and sites are counted together. An analysis that ranks raw names gets this wrong, and nothing on the page tells you.

What these numbers are not is a safety ranking. A company with more pumps in hospitals will have more reports. Some manufacturers file more thoroughly than others. One event can produce reports from a hospital and the manufacturer. The FDA's About MAUDE page says MDR data isn't meant for evaluating rates or comparing them across devices. Use report share to see where the reporting volume sits, then read the reports.

Step 4: Who is recalling, and why

Recalls are the strongest signal in the public data, because each one is a firm's own decision that something has to be corrected or removed. openFDA's recall data (updated November 19, 2026) lists 133 FRN recall records initiated from November 1, 2021 to October 31, 2026.

By firm, Baxter Healthcare Corporation has 27, Fresenius Kabi USA 21, CareFusion 303 16, Smiths Medical ASD 16 and ICU Medical 22 across two spellings (14 and 8). Merge Smiths into ICU Medical and ICU Medical has 38.

The root causes are more useful than the counts:

Root cause, as recordedRecall records
Under Investigation by firm41
Process control22
Device Design20
Software design10
Software Design Change9
Software design (manufacturing process)9
Nonconforming Material/Component6

Two things stand out. Nearly a third of the records (41 of 133) still say the cause is under investigation, so a root cause analysis of this market is incomplete by definition. And the three software labels together account for 28 records, more than device design (20). For a team designing an infusion pump, that points at software verification as the place competitors have been caught out. Our post on FDA recall classes explains how to read the class on each record.

Step 5: Read what the reports say

Counts tell you where to look. The problem codes tell you what to design against. For FRN over the same 24 months, the most reported device problems were:

  • Device Sensing Problem: 16,177
  • Break: 11,627
  • Device Alarm System: 10,146
  • Failure to Sense: 9,556
  • Corroded: 6,379
  • Failure to Infuse: 5,695
  • Contamination: 5,059

A report can carry several problem codes, so these overlap and don't add up to the report total. Pull a sample of narratives for the top three codes and the top two companies, and record what failed and how it was found. That is the part of a competitive analysis that feeds a hazard analysis and a design input list, and it is where our guide to searching MAUDE helps.

Step 6: Write it up so someone can check it

A competitive analysis built on public data is only as good as its citations. For every number, record:

  • the database and the exact query (product code, date field and range)
  • the date you retrieved it and the "data as of" date the FDA publishes
  • which names you merged, and why (spelling, site or acquisition, with the source)
  • the limits: no rates, passive surveillance, duplicates, and reports are not admissions of fault (21 CFR 803.16)

Then rerun it on a schedule. New clearances, recalls and report spikes are what change a competitive picture, and they arrive weekly in openFDA.

Doing this in DeviceVigil

DeviceVigil runs these queries for you. Search a device type, a brand, a company or a product code at devicevigil.com/search, with no account, and it lines up premarket decisions, recalls and MAUDE reports by year, compares the last 12 months with the 12 before, and ranks the most reported problems, makers, product codes and recall root causes. The report prints or saves as PDF.

In the extension, open any 510(k), De Novo or PMA on the FDA's site and the Competitors tab lists every company in its product code with its decisions, MAUDE reports, deaths and recalls, with spellings merged and the share of reports and decisions drawn as bars. Acquisitions aren’t in the FDA’s data, so those stay separate: note them in your write-up. Add the product code to the watchlist and DeviceVigil checks it daily or weekly and lists new decisions, recalls and reports.

See it beside the FDA record

DeviceVigil is a Chrome and Edge extension for regulatory, quality and post-market teams. Open a 510(k), PMA, classification or recall page and see that device type’s clearances, recalls and MAUDE reports in one panel, and get told when something changes.

Add to Chrome — free See how it works

Frequently asked questions

Can I compare competitors' failure rates with MAUDE?

No. MAUDE has no count of devices in use, events are under-reported, and one event can be reported more than once. The FDA says MDR data isn't meant for evaluating rates or comparing them across devices. Compare what is reported and how it changes over time, not rates.

Where do I find a competitor's 510(k)s?

Search the FDA's 510(k) database by applicant or by product code, or count decisions by applicant in openFDA's 510(k) data. Search by product code too: applicant names vary, and acquired companies keep filing under their old names. Our guide on how to search the FDA 510(k) database covers both.

Why does the same company appear under several names?

Each database records the name the submitter or reporter typed, and each manufacturing site can file under its own name. In FRN's MAUDE data, ICU Medical, its Costa Rica site and Smiths Medical, which it acquired in 2022, appear under ten names among the 30 most common. Merge spellings and sites always, and merge acquisitions when you can cite the deal.

How often should a competitive analysis be refreshed?

openFDA updates its device data weekly. A quarterly refresh suits a strategy review. For a product code you compete in every day, a weekly watch on new decisions, recalls and reports catches changes as they happen.

Sources

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This post describes public FDA sources and regulations; it isn’t regulatory or legal advice.